510(k) Data Aggregation

    K Number
    K082571

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-09-30

    (25 days)

    Product Code
    Regulation Number
    884.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

    Device Description

    The Avaulta™ Support System includes a sterile, single use, permanent implant that provides long term reinforcement to support structures in the correction of anterior or posterior vaginal wall prolapse. The central soft knit section provides compliant organ support while the strong knit arms provide improved strength for tension free fixation of the implant.

    The Avaulta™ Plus Biosynthetic Support System and Avaulta™ Solo support system both utilize a nonabsorbable monofilament, polypropylene mesh to provide long-term reinforcement for support structures. The Avaulta™ Plus Biosynthetic Support System adds a porous, acellular, ultrathin sheet of crosslinked collagen attached to the polypropylene mesh which serves to establish a protective barrier between mucosal tissue and the polypropylene mesh and contains apertures uniformly sized to allow for ingrowth of host tissue and capillary vessels.

    AI/ML Overview

    The provided text is a 510(k) summary for the Avaulta™ Solo Support System and Avaulta™ Plus Biosynthetic Support System (K082571). It describes a surgical mesh device for vaginal wall prolapse.

    However, the document does not contain any information regarding acceptance criteria or the study that proves the device meets specific performance criteria through an AI/algorithm-based evaluation. The entire submission is focused on establishing substantial equivalence to a predicate device based on similar intended use, general design, fundamental scientific technology, and performance data from physical testing in accordance with guidance for surgical mesh.

    Therefore, I cannot provide the requested information. The document details the device's description, intended use, and substantial equivalence to a predicate device, but it lacks any mention of an AI or algorithm-based system, its performance metrics, or related studies.

    To illustrate what a comprehensive response would look like if the information were present, I'll use placeholders for each section, emphasizing that this information is missing from the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    Performance Metric (Hypothetical)Acceptance Criteria (Hypothetical)Reported Device Performance (Hypothetical)
    [Specific AI Performance Metric 1][Target Value or Range][Achieved Value]
    [Specific AI Performance Metric 2][Target Value or Range][Achieved Value]
    .........

    2. Sample Size for Test Set and Data Provenance

    • Test Set Sample Size: Not provided (no AI/algorithm study described).
    • Data Provenance: Not provided (no AI/algorithm study described).

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    • Number of Experts: Not provided (no AI/algorithm study described).
    • Qualifications of Experts: Not provided (no AI/algorithm study described).

    4. Adjudication Method for Test Set

    • Adjudication Method: Not provided (no AI/algorithm study described).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, an MRMC study is not mentioned as part of this 510(k) submission.
    • Effect Size of Human Readers with vs. without AI assistance: Not applicable, as no AI-assisted study is described.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? No, a standalone algorithm performance study is not mentioned. The submission focuses on the performance of the physical surgical mesh itself.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Not provided (no AI/algorithm study described). For the physical device, "substantial equivalence" to a predicate device is the primary "ground truth" for regulatory approval, supported by physical testing data, not clinical ground truth in the AI sense.

    8. Sample Size for Training Set

    • Training Set Sample Size: Not applicable, as no AI/algorithm is described.

    9. How Ground Truth for Training Set Was Established

    • Ground Truth Establishment (Training Set): Not applicable, as no AI/algorithm is described.

    Conclusion based on provided text:

    The provided 510(k) summary for the Avaulta™ Solo Support System and Avaulta™ Plus Biosynthetic Support System does not describe an AI/algorithm-based device nor any studies related to its performance against acceptance criteria typically associated with AI systems. The document focuses on the mechanical and biological properties of a surgical mesh and its substantial equivalence to a predicate device, as per the requirements for surgical mesh products. Therefore, all the requested information pertaining to AI/algorithm performance studies is absent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063712

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-03-12

    (88 days)

    Product Code
    Regulation Number
    884.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

    Device Description

    The Avaulta™ Support System includes a sterile, single use, permanent implant that provides long term reinforcement to support structures in the correction of anterior or posterior vaginal wall prolapse. The central soft knit section provides compliant organ support while the strong knit arms provide improved strength for tension free fixation of the implant.

    The Avaulta™ Plus Biosynthetic Support System and Avaulta™ Solo support system both utilize a nonabsorbable monofilament, polypropylene mesh to provide long-term reinforcement for support structures. The Avaulta™ Plus Biosynthetic Support System adds a porous, acellular, ultrathin sheet of crosslinked collagen attached to the polypropylene mesh which serves to establish a protective barrier between mucosal tissue and the polypropylene mesh and contains apertures uniformly sized to allow for ingrowth of host tissue and capillary vessels.

    AI/ML Overview

    This 510(k) premarket notification describes a surgical mesh device, the Avaulta™ Solo Support System and Avaulta™ Plus Biosynthetic Support System. The notification focuses on establishing substantial equivalence to predicate devices and provides limited information regarding detailed acceptance criteria and a specific study proving the device meets those criteria.

    Based on the provided document, here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of specific acceptance criteria (e.g., tensile strength, pore size, biocompatibility metrics with defined thresholds) and corresponding reported performance metrics for the Avaulta™ Support System.

    Instead, the document states:

    • "The appropriate testing to determine substantial equivalence was completed."
    • "This includes testing in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999)."

    This implies that the acceptance criteria were based on the guidance document for surgical mesh and that the device performance met the requirements outlined in that guidance, thereby demonstrating substantial equivalence to the predicate devices. The specific numerical values for these criteria are not provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The 510(k) summary does not detail specific sample sizes for particular tests or the provenance of any data (e.g., country of origin, retrospective/prospective). It generally refers to "appropriate testing."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and therefore not provided in the context of this device. The Avaulta™ Support System is a physical surgical mesh, not an AI or diagnostic device that requires expert-established ground truth for its performance evaluation in the way a medical imaging AI would. Its performance is evaluated through material testing and clinical equivalence to predicate devices.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided. Adjudication methods like 2+1 or 3+1 are used in studies involving human readers and AI for diagnostic tasks, not for evaluating a physical medical device like surgical mesh.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    This information is not applicable and therefore not provided. An MRMC study is relevant for evaluating the impact of AI on human reader performance in diagnostic tasks. The Avaulta™ Support System is a surgical implant, not a diagnostic tool assisted by AI, so such a study would not be performed for this device.

    6. If a Standalone Study (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This information is not applicable and therefore not provided. This concept is relevant to AI algorithms. The Avaulta™ Support System is a physical medical device.

    7. The Type of Ground Truth Used

    The concept of "ground truth" in the way it's used for AI or diagnostic studies (e.g., pathology, outcomes data) is not directly applicable to the primary performance evaluation documented here for a surgical mesh.

    Instead, the "truth" for this device's performance evaluation would be established through:

    • Material properties testing: Adherence to established mechanical, physical, and chemical standards (e.g., tensile strength, burst strength, pore size, biocompatibility using ISO standards).
    • Clinical equivalence: Demonstrating that the device performs as intended for tissue reinforcement in vaginal wall prolapse, comparable to the predicate devices, which have a known safety and effectiveness profile. This "ground truth" would be established through clinical experience with the predicate devices and the understanding that the new device's materials and design are substantially equivalent.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. "Training set" is a concept used in machine learning for AI model development. This device is a physical product, not an AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided. As explained above, the concept of a training set and its ground truth does not apply to the development and evaluation of this surgical mesh.

    Summary of Device Performance and Study Information (Based on the document):

    The document functions as a 510(k) summary, aiming to demonstrate substantial equivalence to legally marketed predicate devices. The "study" mentioned is the "appropriate testing to determine substantial equivalence" which "includes testing in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999)." This implies that standard material science, biocompatibility, and potentially preclinical testing (as dictated by the guidance) were conducted, and the results demonstrated that the Avaulta™ Support System met the requirements of the guidance and was comparable to the predicate devices. Specific quantitative data, sample sizes for these tests, or detailed expert reviews (beyond regulatory review) are not typically presented in this condensed form of a 510(k) summary.