510(k) Data Aggregation

    K Number
    K112998

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-03-15

    (160 days)

    Product Code
    Regulation Number
    870.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.

    Device Description

    The AutoPulse Resuscitation System Model 100 is an automated, portable, batterv powered device that compresses the chest of an adult human as an adjunct to manual CPR. The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. The AutoPulse Platform contains the mechanical drive mechanism, control system, software and electronics necessary to generate and control the force required to perform mechanical chest compressions. User controls and indicators are contained in the User Control Panel. The AutoPulse Platform and LifeBand are unchanged. A new battery and battery charger are the subject devices of this 510(k).

    AI/ML Overview

    The provided documentation is a 510(k) summary for the AutoPulse® Resuscitation System Model 100, focusing on changes related to its battery and charger. It is not a study reporting on the clinical performance of the device in relation to patient outcomes or a diagnostic algorithm's accuracy.

    Therefore, many of the requested fields are not applicable to this specific document as it details a submission for a modification to an already cleared device, primarily concerning electrical components, rather than a clinical effectiveness study.

    Based on the provided text, here is the information that can be extracted:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Bench Testing)Reported Device Performance
    System level compatibility with new Li-Ion battery meets operating parameters in a safe and intended manner.Data showed that the new battery chemistry did not alter the functioning of the AutoPulse in any way and that it operates identically to the NiMH Battery powered AutoPulse.
    New multi-chemistry charger capable of safely and consistently charging NiMH battery, testing the battery, and correctly identifying end of life.Bench testing showed the new charger was capable of safely and consistently charging the NiMH battery, testing the battery, and correctly identifying when the end of life has been reached.
    New multi-chemistry charger capable of safely and consistently charging, testing, maintaining, and identifying end of life conditions per Li-Ion battery specifications.The new charger successfully demonstrated its ability to charge, test, maintain, and identify the end of life conditions per the Li-Ion battery specifications.
    Li-Ion battery meets performance safety specifications (including charging/discharging characteristics, ability to power AutoPulse for specified runtime throughout life, environmental testing, electrical safety, electromagnetic compatibility).Extensive bench testing was conducted to verify ability of the Li-Ion battery to meet the performance safety specifications. (Specific quantitative results not provided in this summary, but the general statement indicates criteria were met).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in terms of clinical patient data. The testing was "extensive bench testing" on the device components (new battery and charger).
    • Data Provenance: Not applicable as this was bench testing, not human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable. Ground truth for electrical and mechanical performance is established through engineering specifications and standards, not expert clinical consensus.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method for the test set

    • Adjudication Method: Not applicable. This was bench testing against engineering specifications and international standards, not a study requiring adjudication of clinical findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This document is about a mechanical resuscitation device's electrical component update, not an AI-assisted diagnostic tool.
    • Effect size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Standalone Study: The "standalone" performance here refers to the device's functional integrity with the new battery and charger. The summary states that "the system level compatibility was verified to ensure that the AutoPulse powered with the new Li-lon battery meets the operating parameters in a safe and intended manner" and that "the new battery chemistry did not alter functioning of the AutoPulse in any way and that it operates identically to the NiMH Battery powered AutoPulse." This effectively describes the standalone performance of the modified device meeting its operational specifications.

    7. The type of ground truth used

    • Ground Truth: Engineering specifications for power output, charging characteristics, battery life, environmental resilience, electrical safety, and electromagnetic compatibility established by relevant international standards (e.g., IEC 60601-1, IEC 60068, CISPR11, IEC 62133).

    8. The sample size for the training set

    • Sample Size: Not applicable. This is not a machine learning or AI-based device requiring a training set.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072527

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-03-19

    (194 days)

    Product Code
    Regulation Number
    870.5200
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K063602

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-12-21

    (17 days)

    Product Code
    Regulation Number
    870.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.

    Device Description

    AutoPulse Resuscitation System Model 100 ("Device") is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. The Device consists of a single use chest compression assembly (CCA) that includes a patient liner, and a reusable platform that contains a user control panel, a drive mechanism, a control system, and a power system (rechargeable battery).

    AI/ML Overview

    The provided text describes the AutoPulse™ Resuscitation System Model 100, an automatic mechanical chest compressor. However, the document does not contain specific acceptance criteria or study details in the format requested. It states that "Appropriate product testing was conducted to evaluate conformance to product specification and substantial equivalence to the predicate device" and that "Testing of the device modifications demonstrates the device meets and improves upon the predicate device." This indicates that internal testing was performed, but no details about the methodology, results, or specific criteria are provided.

    Therefore, many of the requested fields cannot be populated based on the given information.

    Here's an analysis of what can be inferred or what is explicitly missing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in this document"meets and improves upon the predicate device"

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The document only mentions "product testing was conducted."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable; the testing described is for device conformance and substantial equivalence, not a ground truth establishment by experts for diagnostic or interpretative performance.

    4. Adjudication method for the test set

    • Not applicable; the testing described is for device conformance and substantial equivalence.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a mechanical chest compressor, not an AI diagnostic imaging device that would involve human reader interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • The device operates as a standalone mechanical chest compressor, but this question typically refers to the "standalone performance" of an AI algorithm's diagnostic accuracy. The document does not describe such a study for this device.

    7. The type of ground truth used

    • The document implies that the "ground truth" for its performance would be its ability to perform mechanical chest compressions according to its specifications and "substantially equivalent" to (or improved upon) its predicate device. No external "ground truth" like pathology or outcomes data is mentioned in relation to specific performance metrics for the device itself in this summary.

    8. The sample size for the training set

    • Not applicable. This device is a mechanical system, not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable.

    Conclusion:

    The provided 510(k) summary focuses on general device description, intended use, predicate device comparison, and regulatory status. It states that testing was conducted to demonstrate conformance and substantial equivalence but does not provide details about the specific acceptance criteria, study design, sample sizes, or performance metrics that would answer the posed questions. The questions are largely tailored to AI/diagnostic devices, which this medical device (a mechanical chest compressor) is not.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040453

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-03-11

    (17 days)

    Product Code
    Regulation Number
    870.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AutoPulse™ Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.

    Device Description

    AutoPulse Resuscitation System Model 100 ("Device") is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. The Device consists of a single use chest compression assembly (CCA) that includes a patient liner, and a reusable platform that contains a user control panel, a drive mechanism, a control system, and a power system (rechargeable battery).

    AI/ML Overview

    This device description is for the AutoPulse Resuscitation System Model 100, an automated chest compression device.

    Based on the provided text, there is no specific acceptance criteria or detailed study results that prove the device meets acceptance criteria mentioned.

    The document states:

    • "Appropriate product testing was conducted to evaluate conformance to product specification and substantial equivalence to predicate devices." (Section 1)

    However, it does not provide any data, metrics, or specific acceptance criteria that were used in this testing. It only concludes that the device is "substantially equivalent" to predicate products based on indications for use, basic overall function, and materials.

    Therefore, I cannot populate the requested table or provide details for most of the other questions.

    Here's an assessment based on the available information:

    1. Table of acceptance criteria and the reported device performance:
      • Acceptance Criteria: Not specified in the provided text.
      • Reported Device Performance: Not specified in the provided text.
    Acceptance CriteriaReported Device Performance
    Not providedNot provided
    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not provided.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as no test set details or ground truth establishment are described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is an automated mechanical device, not an AI-based diagnostic or assistive system for human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document describes the device itself as an "automated, portable, battery powered device that compresses the chest." While its performance would inherently be standalone in its function, no specific "standalone performance study" details (like accuracy or efficacy metrics) are provided. It operates as an adjunct to manual CPR, implying it's often used with human oversight, but its core function is automated.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not provided.
    7. The sample size for the training set: Not applicable, as this is a mechanical device, not a machine learning model with a training set.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K022345

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-08-15

    (27 days)

    Product Code
    Regulation Number
    870.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.

    Device Description

    AutoPulse Resuscitation System Model 100 ("Device") is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. The Device consists of a single use chest compression assembly (CCA) that includes a patient liner, and a reusable platform that contains a user control panel, a drive mechanism, a control system, and a power system (rechargeable battery).

    AI/ML Overview

    The provided document describes a Special 510(k) for a device modification, focusing on substantial equivalence to predicate devices rather than providing detailed acceptance criteria and study proving its performance. The document explicitly states: "Appropriate product testing was conducted to evaluate conformance to product specification and substantial equivalence to predicate devices." However, it does not disclose the specific acceptance criteria or the full details of the studies conducted.

    Therefore, many of the requested details cannot be extracted from this document.

    Here's a summary of what can be inferred or directly stated, along with what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    This information is not provided in the document. The submission focuses on substantial equivalence based on overall function and materials, rather than specific performance metrics against defined acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified.
    • Data provenance: Not specified (e.g., country of origin, retrospective/prospective). The document only mentions "Appropriate product testing was conducted."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided. The nature of this device (a resuscitation system) means "ground truth" would likely relate to physiological outcomes or device performance in simulated/clinical settings, not expert interpretation of data like images.

    4. Adjudication method for the test set:

    Not applicable and not provided. As no expert-based ground truth interpretation is mentioned, an adjudication method for such a ground truth would not be relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This device is an automated resuscitation system, not an AI-assisted diagnostic or interpretative device that augments human readers. Therefore, an MRMC study with human readers assisting AI or vice-versa is not applicable and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is an "automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR." This implies its primary function is standalone mechanical operation, though as an "adjunct to manual CPR," it inherently involves human interaction. The document doesn't detail standalone performance tests, focusing broadly on "conformance to product specification and substantial equivalence."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The document does not explicitly state the type of ground truth. For a resuscitation device, ground truth would typically involve:
    * Physiological measurements during CPR (e.g., compression depth, rate, effectiveness in maintaining blood flow).
    * Biomechanical performance (e.g., force applied, consistency).
    * Safety parameters.
    The document broadly states "conformance to product specification," which would imply testing against internal specifications for these types of parameters.

    8. The sample size for the training set:

    Not applicable and not provided. This device is a mechanical resuscitation system, not a machine learning or AI-driven system that would require a "training set" in the conventional sense of machine learning.

    9. How the ground truth for the training set was established:

    Not applicable and not provided, as there is no mention of a training set for an AI/ML model for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1