Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K033597

    Validate with FDA (Live)

    Date Cleared
    2003-11-25

    (11 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATR5000, LC5000 and Implant System are intended for the preparation of intra-oral bone for implantology procedures. These ATR Surgical Micromotors are intended to prepare Intraoral bone for implantology procedures.

    Device Description

    The ATR Surgical Micromotors ATR5000, LC5000 and Implant System, are microprocessor driven surgical micromotors used in implantology procedures. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System consists of a microprocessor controlled unit, foot pedal, electric micromotor, support rods and sterile, disposable irrigation tubes. The units also house the peristaltic pump. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System provide electronic control of velocity and torque. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System can be programmed and retain programs in memory. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System are fully operational from foot pedal.

    AI/ML Overview

    This document describes acceptance criteria and testing for the ATR Surgical Micromotors (ATR5000, LC5000, and Implant System), which are microprocessor driven surgical micromotors used in implantology procedures.

    Here's the breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance is demonstrated by its compliance with a series of electrical, safety, and electromagnetic compatibility (EMC) standards. The acceptance criteria essentially are compliance with these listed standards.

    Acceptance Criteria (Standard)Reported Device Performance
    UL 2601-1 Standard for Medical Electrical Equipment, General requirements for Safety (IEC 60601-1 with U.S. deviations)Complies with the standard
    UL 601-1 Medical Electrical Equipment, General Requirement for SafetyComplies with the standard
    IEC 60601-1 Medical Electrical Equipment Part 2, "General Safety Norms"Complies with the standard
    IEC 60601-1-2 Medical Electrical Equipment Part 1 EMCComplies with the standard
    EMC Conducted RF emissions IEC 55011Complies with the standard
    Degrees of protection provided by enclosures (IP Code) IEC 60529Complies with the standard
    IEC 61000-4-2 EMC Electrostatic Discharge ImmunityComplies with the standard
    IEC 61000-4-3 Radiated RF immunityComplies with the standard
    IEC 61000-4-4 Fast Transient ImmunityComplies with the standard
    IEC 61000-4-5 Pulse immunityComplies with the standard
    IEC 61000-4-6 Conducted RF immunityComplies with the standard
    IEC 61000-4-11 Supply Voltage Hole ImmunityComplies with the standard
    Transformer safety IEC 61558-1Complies with the standard
    IEC 61558-2-6 Transformer safety-Particular prescriptionsComplies with the standard
    ISO 11498:1997 Dental handpieces Dental low-voltage electrical motorsComplies with the standard
    Withstanding voltage testPerformed in conformance with UL 2601-1 requirements
    Leak current testPerformed in conformance with UL 2601-1 requirements

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify a numerical "sample size" in terms of number of devices tested. It refers to "All evaluation of the ATR 5000, LC5000 and Implant System" being performed. This implies at least one unit of each model (ATR5000, LC5000, and Implant System) was tested.
    • Data Provenance: The tests were "performed by certified body Laboratory," but the specific country of origin for the data is not explicitly stated. The submitting company (ATR srl) is based in Italy, suggesting the testing might have been done in Italy or a European certified lab. The predicate device was also evaluated "with the same regulation standard," indicating historical consistency in testing approach.
    • Retrospective or Prospective: Not applicable in the context of device performance testing against engineering standards. This is a prospective testing of the manufactured device models.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of device submission. The "ground truth" here is compliance with established engineering and safety standards, not a medicaldiagnosis or interpretation by experts. The "certified body Laboratory" acts as the independent verifier of compliance.

    4. Adjudication method for the test set

    This information is not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies where conflicting expert interpretations need resolution. For compliance testing against engineering standards, the results are objectively measured against the standard's specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. MRMC studies and the concept of "human readers" improving with "AI assistance" are relevant to diagnostic imaging devices or other AI-enabled tools for interpretation. This submission is for surgical micromotors, which are hand-held mechanical and electronic devices, not AI-driven diagnostic tools.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This device is a surgical micromotor, not a software algorithm. Its performance is evaluated through physical and electrical tests, not algorithmic performance.

    7. The type of ground truth used

    The "ground truth" in this context is the specific technical requirements and limits defined by the listed international and national standards (e.g., UL 2601-1, IEC 60601-1, IEC 60601-1-2, IEC 61000 series, ISO 11498). The device is deemed safe and effective if its measured performance falls within the acceptable parameters set by these standards.

    8. The sample size for the training set

    This information is not applicable. Surgical micromotors are hardware devices that do not use machine learning algorithms, and therefore, there is no "training set."

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as above; there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1