510(k) Data Aggregation

    K Number
    K221821

    Validate with FDA (Live)

    Date Cleared
    2022-09-23

    (92 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

    Device Description

    The ATEC IOM Accessory Instruments are surgical instruments that provide electrical stimulation to the body to locate and identify nerves in either open, minimally invasive, or percutaneous procedures. These surgical instruments are compatible with common FDA cleared neuromonitoring platforms as they are connected via a compatible clip or probe depending on the system. The neuromonitoring capability provides the surgeon with spinal nerve location, proximity, and integrity information. This information assists the surgeon during targeting, bone preparation, and placement of orthopedic implants such as intervertebral fusion devices (e.g., interbodies) and bone screws (e.g., pedicle screws). All sterile instruments are single use only and all reusable instruments are offered non-sterile to be steam sterilized by the end user.

    The purpose of this submission is to gain clearance for new dilators made of conductive sliver ink, aluminum alloy, and dielectric coating.

    AI/ML Overview

    The provided document is a 510(k) summary for the ATEC IOM Accessory Instruments. It primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and performance testing. It does not describe an acceptance criteria and study proving device meets acceptance criteria in the context of clinical performance or accuracy of an AI/ML device.

    Therefore, I cannot extract the specifically requested information regarding acceptance criteria and study details as they pertain to such a device type because this information is not present in the provided document.

    The document discusses:

    • Manufacturing and material equivalence: The new dilators are made of conductive silver ink, aluminum alloy, and dielectric coating and are compared to existing predicate devices.
    • Performance testing (non-clinical): This includes electrical safety testing (IEC 60601-1), functional performance testing (insulation effectiveness and electrical resistance), and reprocessing validation. All samples passed acceptance criteria for these tests.
    • Biocompatibility testing: Evaluated per ISO 10993-1 for cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity.

    None of these tests involve "device performance" in terms of clinical accuracy, AI/ML output, or human-in-the-loop effectiveness.

    However, I can extract the information that is present about the non-clinical performance testing:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaReported Device Performance
    Electrical Safety Testing and/or EvaluationIn accordance with IEC 60601-1: 2005(R)2012All samples passed acceptance criteria
    Functional Performance Testing and VerificationNot explicitly stated, but includes:All samples passed acceptance criteria
    - Insulation Effectiveness- Meets functional requirements (implied by passing)
    - Electrical Resistance- Meets functional requirements (implied by passing)
    ReprocessingCleaning validation or adoption study based on AAMIAll samples passed acceptance criteria
    TIR30:2011
    Steam sterilization validation or adoption performed
    per ANSI/AAMI/ISO 17665-1:2006/(R)2013

    2. Sample size used for the test set and the data provenance: Not applicable. The tests mentioned are laboratory-based electrical, functional, and reprocessing tests, not clinical studies with patients or data sets with provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These are engineering and sterilization tests, not expert-adjudicated clinical assessments.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. This refers to clinical or image-based adjudication, not engineering tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an accessory instrument, not an AI/ML diagnostic or assistance tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is an accessory instrument, not an AI/ML algorithm.

    7. The type of ground truth used: Not applicable in the context of clinical accuracy or AI performance. For the engineering tests, "ground truth" would be the specifications and requirements of the applicable standards (e.g., IEC 60601-1, AAMI TIR30, ANSI/AAMI/ISO 17665-1).

    8. The sample size for the training set: Not applicable. This device is not an AI/ML product.

    9. How the ground truth for the training set was established: Not applicable. This device is not an AI/ML product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191723

    Validate with FDA (Live)

    Date Cleared
    2019-10-18

    (113 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

    Device Description

    The ATEC IOM Accessory Instruments are surgical instruments that provide electrical stimulation to the body to locate and identify nerves in either open, minimally invasive, or percutaneous procedures. These surgical instruments are compatible with common FDA cleared neuromonitoring platforms as they are connected via a compatible clip or probe depending on the system. The neuromonitoring capability provides the surgeon with spinal nerve location, proximity, and integrity information. This information assists the surgeon during targeting, bone preparation, and placement of orthopedic implants such as intervertebral fusion devices (e.g., interbodies) and bone screws (e.g., pedicle screws). All sterile instruments are single use only and all reusable instruments are offered non-sterile to be steam sterilized by the end user.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the ATEC IOM Accessory Instruments. This document focuses on demonstrating substantial equivalence to predicate devices based on design, technology, and performance data, not clinical performance data. Therefore, the information typically required to describe acceptance criteria and a study proving a device meets these criteria in the context of clinical performance (e.g., sample size for test sets, expert adjudication, MRMC studies, ground truth establishment) is largely absent.

    However, the document does contain information about performance testing for non-clinical aspects. Below is an interpretation of the available information in the context of your request.

    Please note: The document explicitly states "Clinical Information: Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data." This means that the study you are asking about, which would typically involve human subject data and clinical endpoints, was not conducted or submitted for this 510(k) clearance. The performance data presented refers to engineering and safety testing.


    Acceptance Criteria and Device Performance (Non-Clinical)

    The document primarily focuses on demonstrating substantial equivalence through non-clinical performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    Test CategoryTest Method SummaryAcceptance CriteriaReported Device Performance
    Electrical Safety TestingEvaluation and testing in accordance with IEC 60601-1: 2005."All samples passed acceptance criteria" (Implicitly, compliance with IEC 60601-1)All samples passed acceptance criteria
    Reliability TestingTesting to verify no adverse effect on safety or effectiveness based on intended environment and storage conditions."All samples passed acceptance criteria" (Implicitly, device remains safe and effective after reliability tests)All samples passed acceptance criteria
    Functional Performance TestingInsulation Effectiveness; Electrical Resistance"All samples passed acceptance criteria" (Implicitly, functional specifications for insulation and resistance met)All samples passed acceptance criteria
    Biocompatibility TestingPer ISO 10993-1, including Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Hemocompatibility."All samples passed acceptance criteria" (Implicitly, compliance with ISO 10993-1 for these endpoints)Biocompatibility testing was conducted and data was leveraged (e.g., existing data, end-specific testing, Master File, predicate and reference devices, well known materials) to support biocompatibility. The conclusion states the device materials are "manufactured from the same materials as other legally US-marketed devices" and implicitly passed the evaluation.

    Information Regarding a Clinical Study (Not Applicable as per the Document)

    As explicitly stated in the document, no clinical performance data was provided or relied upon for this 510(k) clearance. Therefore, the following points cannot be addressed from the given text as they pertain to clinical studies.

    1. Sample size used for the test set and the data provenance: Not applicable for a clinical study. For the non-clinical tests, the sample sizes are not specified beyond "All samples passed," meaning at least one sample was tested.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an instrument, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm-based device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for a clinical ground truth. For non-clinical tests, the ground truth would be defined by the testing standards (e.g., IEC 60601-1, ISO 10993-1).
    7. The sample size for the training set: Not applicable for a clinical study. This device does not use a "training set" in the sense of machine learning.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of Device and Clearance Basis:

    The ATEC IOM Accessory Instruments are surgical tools designed to assist in locating nerves during spinal surgical procedures. Their clearance (K191723) was based on substantial equivalence to existing legally marketed predicate devices, specifically K182617 (ATEC IOM Accessory Instruments) and K132138 (Rhythmlink Disposable Concentric Stimulating Probe).

    The substantial equivalence was demonstrated through:

    • Comparison of indications for use, design, function, and technology.
    • Compliance with relevant electrical safety standards (IEC 60601-1).
    • Reliability testing.
    • Functional performance testing (insulation effectiveness, electrical resistance).
    • Biocompatibility testing per ISO 10993-1, largely leveraging existing data and material characterization.

    The crucial point is that this clearance did not involve a clinical performance study with human subjects, as the FDA determined it was not necessary to establish substantial equivalence for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182617

    Validate with FDA (Live)

    Date Cleared
    2019-05-09

    (227 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

    Device Description

    The ATEC IOM Accessory Instruments are surgical instruments that provide electrical stimulation to the body to locate and identify nerves in either open, minimally invasive, or percutaneous procedures. These surgical instruments are compatible with common FDA cleared neuromonitoring platforms as they are connected via a compatible clip or probe depending on the system. The neuromonitoring capability provides the surgeon with spinal nerve location, proximity, and integrity information. This information assists the surgeon during targeting, bone preparation, and placement of orthopedic implants such as intervertebral fusion devices (e.g., interbodies) and bone screws (e.g., pedicle screws).

    AI/ML Overview

    The document provided is a 510(k) premarket notification for the ATEC IOM Accessory Instruments. It primarily focuses on demonstrating substantial equivalence to predicate devices based on design, functionality, and performance testing. However, it explicitly states:

    "Clinical Information: Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data."

    This means the submission does not contain information about a study proving the device meets acceptance criteria related to clinical performance or human-in-the-loop studies involving AI. The acceptance criteria and performance data provided are related to functional, electrical safety, biocompatibility, and reprocessing aspects of the surgical instruments themselves, not their performance in a clinical diagnostic or prognostic context with AI assistance.

    Therefore, I cannot provide the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, or ground truth establishment relevant to an AI/clinical performance study.

    The document covers the following types of acceptance criteria and proven performance for the ATEC IOM Accessory Instruments (which are surgical tools, not an AI diagnostic device):

    1. Table of Acceptance Criteria and Reported Device Performance (Summary of Engineering/Safety Tests):

    TestAcceptance Criteria (Implied by Test Method)Reported Device Performance
    Functional performance testing and verification analysisInsulation Effectiveness, Electrical Resistance, Current Density (as per standard engineering principles for electrical stimulation instruments)"All functional performance testing passed. Substantial equivalence has been shown via analysis and testing."
    Electrical safety testing and/or evaluationCompliance with IEC 60601-1: 2005 for medical electrical equipment."Subject devices passed electrical safety testing and/or evaluation, demonstrating that the devices meet the requirements."
    Biocompatibility testingCompliance with ISO 10993 for Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Hemocompatibility. Risk analysis performed considering body contact and duration."A risk analysis was performed taking into account nature of body contact and duration to categorization the use of existing data, end-specific testing, and endpoint assessment to cover the identified test methods. Additionally, data was leveraged by other means (e.g., authorized use of Master File, predicate and reference devices, well known and characterized materials) to support the biocompatibility of the subject devices."
    Reprocessing (Cleaning and Sterilization)Cleaning validation study based on AAMI TIR30:2011 acceptance criteria. Steam sterilization validation performed per ANSI/AAMI/ISO 17665-1:2006/(R)2013 to assure an SAL of 10^-6."The subject devices can be adequately cleaned and steam sterilized prior to use."

    The document does NOT provide information for the following points, as they are not relevant to a device submission for surgical instruments focused on substantial equivalence based on material and functional safety rather than AI performance:

    1. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    7. The sample size for the training set
    8. How the ground truth for the training set was established