    K Number
    K181435

    Validate with FDA (Live)

    Date Cleared
    2018-08-21

    (81 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATEC Cervical Spacer System is intended for spinal fusion procedures at one or two levels from C2 – T1 in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The ATEC Cervical Spacer System is intended for use with supplemental fixation systems and or allograft comprised of cancellous and/or corticocancellous bone graft. Patients should have had six weeks of non-operative treatment.

    Device Description

    The ATEC Cervical Spacer System is an intervertebral body fusion system. The implants consist of various lengths, widths, heights and degrees of lordosis to accommodate individual patient anatomy. These implants are manufactured from PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, tantalum per ASTM F560, titanium alloy (Ti-6Al-4V ELI) per ASTM F136, and commercially pure titanium (CPTi Grade 2) per ASTM F67. The device includes rows of teeth on the surface of each end of the device which serve to grip the adjacent vertebrae to resist migration and expulsion of the device. Additionally, the commercially pure titanium implants are offered with a microstructure due to the layering of material that forms the porous geometry. This porous geometry extends to the superior and inferior surfaces of the device for implant fixation. All interbodies feature an internal graft aperture for placement of graft material to promote fusion through the cage.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "ATEC Cervical Spacer System." It is not a study demonstrating the device meets acceptance criteria related to an AI/ML algorithm's performance. Instead, it's a submission to the FDA to demonstrate "substantial equivalence" to legally marketed predicate devices, which allows the device to be marketed without requiring a full Pre-Market Approval (PMA) application.

    Therefore, the requested information about acceptance criteria, study design for AI/ML performance, sample sizes, ground truth establishment, expert qualifications, and MRMC studies is not applicable to this document.

    The document states:

    • "Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data. No clinical studies were conducted."

    The "performance data" section refers to nonclinical testing to demonstrate the physical and mechanical properties of the intervertebral fusion device itself, not the performance of an AI/ML algorithm.

    Here's what the document does describe regarding the device itself (not an AI algorithm):

    1. Acceptance Criteria and Reported Device Performance (Nonclinical Testing):

    Acceptance Criterion (Type of Test)Reported Device Performance
    Static and dynamic axial compressionSupports substantial equivalence to predicate devices
    Static and dynamic torsionSupports substantial equivalence to predicate devices
    SubsidenceSupports substantial equivalence to predicate devices
    ExpulsionSupports substantial equivalence to predicate devices
    Particulate and gravimetric analysisSupports substantial equivalence to predicate devices
    Bacterial endotoxin testing (BET)Supports substantial equivalence to predicate devices

    2. Sample size used for the test set and the data provenance: Not applicable. This document refers to nonclinical, bench-top mechanical testing, not a clinical test set for an AI algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.

    4. Adjudication method: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No.

    7. The type of ground truth used: Not applicable. For nonclinical tests, the "ground truth" is defined by the physical or mechanical properties measured against the established standards (e.g., ASTM standards).

    8. The sample size for the training set: Not applicable. There is no AI algorithm.

    9. How the ground truth for the training set was established: Not applicable. There is no AI algorithm.