    Intended Use

    The ASSURE® Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

    The PROVIDENCE™ Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

    The VIP® Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

    The XTEND® Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

    The UNIFY® Dynamic Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis.

    The CITADEL® Anterior Lumbar Plate System is intended for use by an anterior or anterolateral approach in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudarthrosis, spondylolisthesis, scoliosis, kyphosis, kyphosis, lordosis, spinal stenosis, or failed previous spine surgery.

    The TRUSS® Thoracolumbar Plate System is intended for use in the treatment of thoracolumbar (T1-L5) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, spinal stenosis, or failed previous spine surgery.

    The PLYMOUTH® Thoracolumbar Plate System is intended for use in the treatment of thoracolumbar (T1-L5) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or failed previous spine surgery.

    The SP-Fix® Spinous Process Fixation Plate is a posterior non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), and/or tumor. The SP-Fix® Spinous Process Fixation Plate is intended for use with allograft or autograft bone and is not intended for standalone use.

    The RELIEVE® Laminoplasty Fixation System is intended for use in the lower cervical and upper thoracic spine (C3-T3) in laminoplasty procedures. The RELIEVE® Laminoplasty Fixation System is used to hold the bone allograft material in place in order to prevent the allograft from expulsion, or impinging the spinal cord.

    Device Description

    These plate and screw systems are used to provide structural stability in skeletally mature individuals following discectomy, corpectomy, vertebrectomy, or laminectomy and may be inserted using an anterior, posterior, anterolateral, or lateral approach. The devices are available in various lengths and widths to accommodate varying patient anatomy. The plates are secured through the plate's screw holes into the vertebral bodies or a ratchet design which automatically locks during compression on the spinous process. Some devices can be used in conjunction with autograft and/or allogenic bone graft to be packed inside the device. These devices are manufactured from titanium alloy, or radiolucent PEEK polymer with titanium alloy or tantalum markers.

    AI/ML Overview

    The provided text is a 510(k) summary for the MRI compatibility update and addition of sterile components for a range of spinal plate systems. It does not present acceptance criteria or detailed study results in the typical format of a clinical or performance study summary. Instead, it refers to standard ASTM tests conducted to determine MRI compatibility and confirms that no further performance or biocompatibility testing was required for this specific submission because the core device performance and technological characteristics remain unchanged from previously cleared versions.

    Here's a breakdown of the information that can be extracted, and where the requested information is absent:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not provide a table with specific acceptance criteria or quantitative performance results. It states that MRI testing was performed according to specific ASTM standards (F2052, F2119, F2182, F2213). For such MRI compatibility tests, the "acceptance criteria" usually refer to the device meeting the requirements outlined within those ASTM standards to be labeled as MR Conditional at a certain field strength. The document implies compliance without listing the specific numerical outcomes or acceptance thresholds used.

    2. Sample Size for the Test Set and Data Provenance:

    • Sample Size: The document states "MRI testing was performed on the worst case subject devices." It does not specify the exact number of devices tested for each ASTM standard.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The testing was likely conducted in a controlled laboratory setting by Globus Medical or a contracted test facility. The nature of MRI compatibility testing for medical devices is typically in-vitro (lab-based) rather than using patient data.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications:

    • Not applicable. This submission is for MRI compatibility and sterile components, not for a diagnostic or AI device that requires expert-established ground truth from clinical cases.

    4. Adjudication Method:

    • Not applicable. This is not a study involving human readers or expert consensus for ground truth.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    • No. This type of study is not relevant to a submission for MRI compatibility of spinal implants.

    6. Standalone Performance Study:

    • Yes (for specific performance aspects implied by conformity to ASTM standards). The MRI compatibility tests are "standalone" in the sense that they assess the device's characteristics in a controlled environment as per the ASTM standards. However, it's not a standalone clinical performance study in the context of an AI algorithm performing a task. The document states "No further device performance testing was required for this submission. The performance testing remains the same for the subject and predicate devices." This implies that the fundamental mechanical/structural performance of the spinal implants was established in prior submissions (e.g., K040721 for ASSURE® and other predicates listed in {12}).

    7. Type of Ground Truth Used:

    • Not applicable in the conventional sense of a clinical ground truth. For MRI compatibility testing, the "ground truth" is adherence to the physical and safety parameters defined by the ASTM standards (e.g., a certain displacement force is considered safe, a certain artifact size is acceptable).

    8. Sample Size for the Training Set:

    • Not applicable. This submission does not involve a "training set" for an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. This submission does not involve a "training set" or a ground truth established for it.

    In summary, the document details a regulatory submission to update labeling for MRI compatibility and add sterile components. It relies on established ASTM standards for MRI safety assessment, rather than a clinical performance study with patient data and expert ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040721

    Validate with FDA (Live)

    Date Cleared
    2004-06-17

    (90 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ASSURE™ Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

    Device Description

    The ASSURE™ Anterior Cervical Plate System consists of plates used with either standard screws or rigid screws and set screws. The plate attaches to the anterior portion of the vertebral body of the cervical spine (levels C2-C7). The implants are composed of titanium alloy.

    AI/ML Overview

    The acceptance criteria and study proving device performance are detailed below:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    In accordance with ASTM F1717 for static and dynamic mechanical testingMeets static and dynamic mechanical testing requirements of ASTM F1717

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify a distinct "test set" in the traditional sense of a clinical or image-based study. The testing performed was mechanical testing, specifically static and dynamic tests, to assess the physical properties of the device. Therefore, the concept of data provenance (e.g., country of origin, retrospective/prospective) is not applicable in this context. The sample size for the mechanical testing is not explicitly stated, but typically involves multiple samples for each test condition as per ASTM F1717.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This question is not applicable. The "ground truth" for mechanical testing is established by the test method (ASTM F1717) and the physical properties observed. There were no human experts establishing ground truth in the context of these mechanical performance tests.

    4. Adjudication Method for the Test Set

    This question is not applicable. Mechanical testing results are objective measurements and do not require expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study conducted was mechanical testing, not a clinical study involving human readers or cases.

    6. Standalone Performance Study

    Yes, a standalone study was done, but it was a standalone mechanical performance study, not an algorithm-only without human-in-the-loop performance study (which would be relevant for AI/software devices). The device, the ASSURE™ Anterior Cervical Plate System, underwent static and dynamic mechanical testing in accordance with ASTM F1717. This evaluates the intrinsic physical performance of the device itself.

    7. Type of Ground Truth Used

    The ground truth used was conformance to a recognized standard, specifically ASTM F1717. This standard defines the methodologies and expected performance criteria for static and dynamic mechanical testing of spinal intervertebral body fixation orthoses.

    8. Sample Size for the Training Set

    This question is not applicable. There was no "training set" as this was mechanical testing of a physical device, not a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there was no "training set." The performance was evaluated against a standardized mechanical testing protocol (ASTM F1717).