Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K112648

    Validate with FDA (Live)

    Date Cleared
    2012-03-13

    (183 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Asfora Bullet Cage® is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) and instability in the lumbar spine at one or two contiguous levels from L2 to S1. DDD for lumbar systems is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The ABC cage devices are used with autogenous bone graft. Patients should be skeletally mature and have had at least six (6) months of non-operative treatment prior to implant. When implanted via a posterior (PLIF, TLIF) approach, this device should be used with supplemental fixation.

    Device Description

    The Asfora Bullet Cage (ABC) is a threaded titanium alloy interbody fusion device (cage) with a threaded titanium end cap. The ABC is a threaded, self-tapping, bullet-shaped, hollow-body, device designed to immobilize adjacent vertebrae and promote arthrodesis (fusion) across the disc space. The Asfora Bullet Cage is available in ten sizes: 5 diameters (10mm, 12mm, 14mm, 16mm, and 18mm) and 2 lengths (21mm or 25mm).

    AI/ML Overview

    The provided text is related to a 510(k) submission for a medical device called the "Asfora Bullet Cage™ System, Intervertebral Body Fusion Device." This submission is a request for substantial equivalence to previously cleared devices.

    Key takeaway: This document is a 510(k) summary and clearance letter, which focuses on demonstrating substantial equivalence to predicate devices, not on a new, formal study with specific acceptance criteria and performance data for the device itself. Therefore, many of the requested points below about specific performance metrics and study design are not applicable to this type of regulatory submission. The document explicitly states: "No significant changes to the design of the device were made. Risk assessment was performed which concluded no additional testing was required."

    Here's a breakdown based on the provided text, indicating where information is not available:


    1. Table of Acceptance Criteria and Reported Device Performance

    Criterion TypeAcceptance Criteria (Not explicitly stated for new device performance)Reported Device Performance (Not explicitly stated for new device performance)
    Substantial Equivalence (Primary Goal of 510(k) Submission)The device should be "substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976... or to devices that have been reclassified..." This implies that the device's technological characteristics, intended use, and safety and effectiveness profile should be similar to authorized predicate devices, posing no new questions of safety or effectiveness.The FDA reviewed the 510(k) submission and determined the device is substantially equivalent to the predicate devices outlined in the submission (K090048, Asfora Bullet Cage System and P950019/S9 Ray TFC Threaded Fusion Cage). The conclusion from Medical Designs, LLC was "based upon the devices’ similarities in principles of operation, technology, materials and indications for use." The FDA's clearance letter confirms this determination.
    Functional and Safety TestingImplicitly, the device should meet the functional and safety standards expected for such an intervertebral body fusion device. While specific criteria aren't provided for this submission for new testing, the previous predicate devices would have met these.The document explicitly states: "No significant changes to the design of the device were made. Risk assessment was performed which concluded no additional testing was required." This indicates that the device's performance is assumed to meet existing standards based on its similarity to the predicate devices and the lack of significant design changes.

    2. Sample size used for the test set and the data provenance

    • Not Applicable: This 510(k) submission is for substantial equivalence without new performance testing. There is no "test set" of patients or data from a clinical study for this specific submission to demonstrate novel device performance. The clearance is based on comparison to predicate devices, which would have undergone their own testing previously.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable: As there was no specific "test set" or clinical study for this submission, there was no ground truth to be established by experts for new performance data. The regulatory decision was made by FDA reviewers (e.g., Mark N. Melkerson, Director Division of Surgical, Orthopedic and Restorative Devices).

    4. Adjudication method for the test set

    • Not Applicable: No clinical or performance study with a test set requiring adjudication was conducted for this 510(k) submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This device is an implantable medical device (intervertebral body fusion device), not an imaging or AI-assisted diagnostic tool. Therefore, MRMC studies or AI assistance are not relevant to its evaluation, and no such study was conducted or reported here.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable: This is an implantable medical device, not an algorithm. Standalone algorithm performance is not relevant to this submission.

    7. The type of ground truth used

    • Not Applicable (in the context of new device performance): The "ground truth" for this 510(k) submission is the pre-existing FDA clearance and established safety and effectiveness of the identified predicate devices. The new device demonstrates "substantial equivalence" to these known safe and effective devices.

    8. The sample size for the training set

    • Not Applicable: This is an implantable medical device. The concept of a "training set" is usually associated with machine learning algorithms, which is not applicable here.

    9. How the ground truth for the training set was established

    • Not Applicable: As there is no training set for a machine learning algorithm, there is no ground truth to be established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090048

    Validate with FDA (Live)

    Date Cleared
    2009-08-14

    (219 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Asfora Bullet Cage™ (ABC) is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) and instability in the lumbar spine at one or two levels from L2 to S1. The DDD patients may also have up to Grade I spondylolisthesis at the involved level. The ABC cage devices are used with autogenous bone graft and are implanted via an open posterior approach. Patients should have had at least six (6) months of non-operative treatment prior to implant.

    Device Description

    The Asfora Bullet Cage™ is comprised of two hollow, threaded, self-tapping titanium fusion devices (cages) with closed tapered medial ends. The distal end has a small orifice that can be closed with a threaded cap. There are six equidistant slit apertures along the longitudinal axis of the device for placement of autogenous bone graft material. The cage has large cutting threads, angled toward the proximal end. The Asfora Bullet Cage™ is available in ten sizes: 5 diameters (10mm, 12mm, 14mm, 16mm, and 18mm) and 2 lengths (21mm or 25mm). This device, with associated instrumentation is intended for use in posterior lumbar interbody fusion to immobilize adjacent vertebrae and promote arthrodesis (fusion) across the disc space.

    AI/ML Overview

    The provided document is a 510(k) summary for the Asfora Bullet Cage™. It details device description, indications for use, and a statement of substantial equivalence to predicate devices. However, the document does not contain information about:

    • Specific acceptance criteria for device performance (e.g., success rates, mechanical thresholds in a clinical context).
    • Any studies involving patient data or device performance against such criteria. The "testing" mentioned in the summary refers to technical characteristics comparison, not clinical or performance outcome studies that would generate the requested metrics.
    • Sample sizes for test or training sets, data provenance, expert ground truth establishment, or adjudication methods for an AI/device performance study.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance.
    • Specific types of ground truth used (e.g., pathology, outcomes data).

    Therefore, based only on the provided text, it's not possible to answer most of the prompt's questions as they pertain to clinical performance studies or AI/device performance metrics against acceptance criteria. The document describes a device comparison/equivalence filing, not a study demonstrating performance against specific clinical acceptance criteria.

    The "study" that proves the device meets acceptance criteria, in this context, is the assessment of substantial equivalence to predicate devices. This is based on comparing technical characteristics and intended uses, implying that if the device is substantially equivalent to legally marketed and safe predicate devices, it also meets the necessary safety and effectiveness "acceptance criteria" for market clearance via the 510(k) pathway.

    Here's an attempt to answer the questions based only on the available text, highlighting where information is absent:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial Equivalence to Predicate Devices (INTER FIX™ Threaded Fusion Device and Ray TFC™ Device) in terms of technical characteristics and intended uses."Testing and comparison of technological characteristics and intended uses found that all components of the Asfora Bullet Cage™ are equivalent to the predicates."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    No information provided in the document regarding a test set or data provenance for performance evaluation. The "testing" refers to technical comparisons, not clinical performance data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable. No ground truth establishment for a test set is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is a medical device clearance document, not an AI performance study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This is a medical device clearance document, not an AI performance study.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    Not applicable. The "ground truth" for this 510(k) submission is the established safety and effectiveness of the predicate devices based on their prior marketing clearance.

    8. The sample size for the training set
    Not applicable. No training set is described for device performance evaluation.

    9. How the ground truth for the training set was established
    Not applicable. No training set is described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1