Search Results
Found 1 results
510(k) Data Aggregation
The ASAHI PTCA Guide Wire, Fielder FC J is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The Asahi PTCA Guide Wires are not to be used in the cerebral blood vessel.
The Asahi PTCA Guide Wire, Fielder FC J is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 180 cm and 300 cm length. The extension wire is connected to the end of the guide wire outside the body for 180cm wire. The wire is constructed from a stainless steel core wire. The core wire and coil are soldered. The distal tip of the guide wire has a radiopaque tip to achieve visibility, and is made soft and easily bendable with the vessel curve, and is provided in a pre-shaped J configuration. There is polyurethane coating covered with hydrophilic coating applied to the distal section of the guide wire and dimethylpolysiloxane (silicone oil - not contacting blood) is applied to gape space in spring-coil in the tip part. The proximal section of this guide wire is coated with PTFE.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Asahi PTCA Guide Wire, Fielder FC J:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Implicit Acceptance Criteria (based on predicate device equivalence and intended use): | Reported Performance: |
| - Device is substantially equivalent to predicate devices (Asahi PTCA Guide Wire, Fielder - K052022; Asahi PTCA Guide Wire, Fielder FC- K063819) regarding material, dimensional specifications, and design. | "The ASAHI PTCA Guide Wire. Fielder FC J is of the same materials as the predicate devices. The dimensional specifications and design of the device ensures compatibility for the intended use." |
| - Performs as intended for facilitating the placement of balloon dilatation catheters during PTCA and PTA. | "This 510(k) notice includes mechanical and functional bench testing that demonstrates that the ASAHI PTCA Guide Wire, Fielder FC J performs as intended.""Bench testing demonstrates that the device functions as intended." |
| - Not to be used in cerebral blood vessels. | This is an indication for use, implying that its performance outside this scope is not claimed or tested for. |
| - Meets safety standards (implied by 510(k) clearance). | FDA's clearance of the 510(k) submission implies that the device meets safety requirements for its intended use. |
Explanation: The document does not explicitly list quantitative acceptance criteria or specific performance metrics with target values (e.g., minimum tensile strength of X N, maximum friction of Y mN). Instead, it relies on the concept of substantial equivalence to predicate devices. The "acceptance criteria" are therefore implicitly tied to demonstrating that the new device is as safe and effective as the previously cleared predicate devices for the same indication. The reported performance confirms this equivalence and that the device functions as intended through bench testing.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified. The document only mentions "mechanical and functional bench testing." It does not provide details on the number of wires tested or the specific tests conducted.
- Data Provenance: Not specified. The testing described is "bench testing," which implies laboratory testing. There is no mention of clinical data or data from a specific country or population. It's a pre-market submission for a medical device, so the data would have been generated by the manufacturer (Asahi Intecc Co., Ltd., Japan). The study is not a retrospective or prospective clinical study but rather bench-top engineering testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This information is not applicable to this submission. The device is a physical medical instrument (guide wire), and the testing described is mechanical and functional bench testing. "Ground truth" in this context would typically refer to objective measurements of the device's physical properties or performance in a controlled environment, not expert interpretation of data.
4. Adjudication Method for the Test Set
- Not applicable. As the testing is mechanical and functional bench testing, there would be no need for expert adjudication in the way it's used for AI/diagnostic studies (e.g., resolving discrepancies in expert interpretations). The results would be objective measurements against predefined engineering specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study was not done. The submission describes mechanical and functional bench testing, not a clinical trial involving human readers or cases.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not applicable. This device is a physical guide wire, not an algorithm or software. Therefore, the concept of "standalone performance" for an algorithm doesn't apply.
7. The Type of Ground Truth Used
- For the bench testing, the "ground truth" would be objective engineering measurements and specifications. For example, measurements of tensile strength, torque transmission, coating integrity, lubricity, radiopacity, and dimensions would be compared against predefined internal specifications or industry standards. There is no mention of expert consensus, pathology, or outcomes data.
8. The Sample Size for the Training Set
- Not applicable. This device is a physical medical device, not an AI model or algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no training set for an AI model, this question is not relevant to this medical device submission.
Ask a specific question about this device
Page 1 of 1