Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K230949

    Validate with FDA (Live)

    Date Cleared
    2023-12-15

    (255 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARTIS is a family of dedicated angiography systems developed for single plane and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

    Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angtography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography.

    Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions.

    The ARTIS systems can also support the acquisition of position triggered imaging for spatial data synthesis.

    The ARTIS systems include also the software option DynaCT with following indications for use:

    DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.

    DynaCT is intended for imaging both hard soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

    Device Description

    The ARTIS pheno (VE30A) is a multi-axis system, specifically designed to meet the growing demands of high-end imaging for interventional radiology, interventional cardiology, and minimally invasive and hybrid surgery procedures. The stand allows positioning in angular, orbital, lateral, longitudinal, and vertical directions, leveraging the flexible isocenter. The ARTIS pheno (VE30A) is equipped with a robotic multi-axis floor stand, C-arm, flat panel detector, x-ray tube, collimator, high voltage generator, patient table, and image post-processing. The ARTIS pheno is partially coated with an optional anti-microbial coating. syngo Application Software is optional, and available for the support of dedicated clinical workflows.

    The ARTIS pheno (VE30A) covers the complete range of angiographic applications, cardiac angiography, neuro-angiography, general angiography, surgery and surgical angiography, multipurpose angiography, rotational angiography, radiographic/fluoroscopic procedures.

    The following components are configured to create the floor configuration:

    • (1) Floor stand with C-arm, X-ray tube assembly and FD
    • (2) Patient table
    • (3) Display ceiling suspension with displays
    • (4) Footswitch for releasing radiation
    • (5) Control console for controlling the stand, patient table, collimator, and imaging system

    lmages and operating elements are displayed on screens. Different display variants are used to visualize image and information content. Panoramic display configurations or large displays can be used, configurable to visualize multiple images and information content in various layouts.

    Post-processing can be done in the exam room or in the control room that offers monitors as well, with a footswitch location in the exam room or the control room. The ARTIS pheno (VE30A) is capable of 2D and 3D imaging.

    Other systems and software synqo Application Software, syngo X Workplace, Sensis Vibe, and or third-party systems may also be integrated into the ARTIS pheno (VE30A) screen configuration. Different screen configurations and layouts are possible in the examination room and the control room.

    AI/ML Overview

    The provided text describes modifications to an existing device, ARTIS pheno (VE21) to create ARTIS pheno (VE30A). The study is primarily a non-clinical performance evaluation focused on verifying that the changes introduced in VE30A do not negatively impact the safety and effectiveness compared to the predicate device.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list a table of "acceptance criteria" alongside specific "reported device performance" values in a quantitative manner for most of the listed modifications. Instead, it generally states that "All testing results passed" and "All test results met all acceptance criteria" for various tests performed on the modifications.

    However, based on the descriptions, we can infer some general acceptance criteria and the reported performance as "met" or "passed".

    Acceptance Criteria (Inferred from descriptions)Reported Device Performance
    New System Software (VE30A)
    Conformance to "Guidance for the content of Premarket submission for software in Medical Devices".All System Software modifications conform to "Guidance for the content of Premarket submission for software in Medical Devices".
    No new risks or issues regarding safety or effectiveness (implied for software/hardware changes).These software and hardware changes do not raise any new risks or any issues regarding the safety or effectiveness of the device.
    Software functional, verification, and system validation testing with passing results.Software functional, verification, and System validation testing with passing results.
    Optional Second WorkplaceFunctional as intended, without introducing new risks (implied by general statements).
    Updated Memory Data Storage to 600KMemory data storage updated to 600K (implied functional).
    Updated NOMSIE DSA (IQ) for Perivision organ programNOMSIE algorithm, overlay, configuration, vessel presentation DSA, and roadmap testing conducted and passed. (Implies image quality is acceptable or improved).
    Updated Roadmap Phase 3Configuration, DSA Roadmap, Subtracted Fluoro and transfer images testing conducted and passed. (Implies functionality and image quality are maintained or improved).
    Added new elements for Organ Program "Structure Scout": Calcium, Gadolinium, and BismuthCNR image Quality exposure control, fluoro auto exposes values cardiology, DR, CNR, DSA, and neuro testing conducted for Structure Scout new elements for the Organ program. All testing results passed. (Implies accurate detection and display of these elements, and image quality is maintained).
    Updated Usability Features (F.1 - F.12)All test results met all acceptance criteria. Usability features were validated to ensure safe and effective operation. Human Factor Usability Validation showed human factors addressed according to operator's manual. Customer employees are trained in the use of this equipment. (Implies improvements are functional and intuitive).
    New PC for Imaging SystemFunctionality maintained, without introducing new risks (implied by general statements).
    NFJ-0157 and NFJ #0163 (Integration with other systems)Functionality and safe integration confirmed (implied by general statements).
    Compliance with electrical safety, performance, and EMC standardsCertified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with a list of specified standards (e.g., AAMI ANSI ES60601-1:2005, IEC 60601-1-2:2020, etc.).
    Software specifications met acceptance criteria.The testing results support that all the software specifications have met the acceptance criteria.
    Risk control implemented to mitigate identified hazards.Risk analysis completed, and risk control implemented.
    Cybersecurity requirements met.Conforms to cybersecurity requirements, considering IEC 80001-1:2010.

    2. Sample size used for the test set and the data provenance

    The document does not specify a quantitative sample size for any test set (e.g., number of images, number of patients, number of cases). The testing appears to be primarily focused on system-level verification and validation of the software and hardware modifications rather than a clinical study involving patient data.

    • Test Set: Not explicitly stated as a distinct "test set" in the context of clinical data. The testing involved "software functional, verification, and System validation testing," "NOMSIE algorithm, overlay, configuration, vessel presentation DSA, and roadmap testing," and "CNR image Quality exposure control, fluoro auto exposes values cardiology, DR, CNR, DSA, and neuro testing." These are likely performed on test harnesses, simulated data, and potentially a limited number of test cases or phantoms.
    • Data Provenance: Not applicable in the context of clinical data. The testing is non-clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. Since the study is a non-clinical performance evaluation, it did not involve clinical experts establishing ground truth in the traditional sense for a diagnostic AI device. The "ground truth" for the non-clinical tests would be the expected functional and performance outcomes defined by engineering specifications.

    4. Adjudication method for the test set

    Not applicable. There is no indication of multiple readers or an adjudication method for establishing ground truth from human experts, as this was not a clinical study involving diagnostic interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an image-intensified fluoroscopic x-ray system, and the changes are related to system software, hardware, and workflow features, not a diagnostic AI component intended for human reader assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device itself is an imaging system. The "algorithms" mentioned (e.g., NOMSIE, Structure Scout) are integrated parts of the overall imaging functionality. While these algorithms perform specific tasks (like image quality enhancement, reconstruction, element identification), their "standalone" performance isn't described in isolation from the system's output. The performance evaluation is inherently within the context of the ARTIS pheno system. It's not an AI model providing a separate diagnostic output.

    7. The type of ground truth used

    The ground truth used for these non-clinical tests would be based on:

    • Engineering specifications and requirements for software functionality.
    • Expected image quality parameters (e.g., CNR for Structure Scout) as measured by phantoms or controlled test conditions.
    • Defined operational workflows for usability features.
    • Compliance with recognized standards for electrical safety, performance, and EMC.

    It is not based on expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    This information is not applicable and not provided. The document describes modifications to an existing x-ray system, implying traditional software and hardware development and testing, not the training of a machine learning model from a distinct "training set" of data.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided as no machine learning training set is mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220409

    Validate with FDA (Live)

    Date Cleared
    2022-06-29

    (135 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

    Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general anglography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography.

    Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions.

    ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.

    The ARTIS family include also the software option DynaCT with following indications for use:

    DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.

    DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

    Device Description

    The ARTIS pheno (VE21) system is a multi-axis system, specifically designed to meet the growing demands of high-end imaging for interventional radiology, interventional cardiology, minimally invasive and hybrid surgery procedures. The stand allows positioning in angular, orbital, lateral, longitudinal, and vertical directions, leveraging the flexible isocenter. The ARTIS pheno (VE21) system is equipped with a robotic multi-axis floor stand, C-arm, flat panel detector, x-ray tube, collimator, high voltage generator, patient table, and image post processing. The ARTIS pheno is partially coated with an optional anti-microbial coating. syngo Application Software is optional, available for the support of dedicated clinical workflows.

    The ARTIS pheno (VE21) covers the complete range of angiographic applications, cardiac angiography, neuro-angiography, general angiography, surgery and surgical angiography, multipurpose angiography, rotational angiography, and radiographic/fluoroscopic procedures.

    The following components are configured to create the floor configuration:

    • (1) Floor stand with C-arm, X-ray tube assembly and FD
    • (2) Patient table
    • (3) Display ceiling suspension with displays
    • (4) Footswitch for releasing radiation

    (5) Control console for controlling the stand, patient table, collimator, and imaging system

    Images and operating elements are displayed on screens. Different display variants are used to visualize image and information content. Panoramic display configurations or large displays can be used, configurable to visualize multiple images and information content in various layouts.

    Post processing can be done in the exam room or in the control room that offers monitors as well, with a footswitch location in the exam room or the control room. The ARTIS pheno (VE21) System is capable of 2D and 3D imaging.

    Other systems and software syngo Application Software, syngo X Workplace, Sensis, and or third-party systems may also be integrated into the ARTIS pheno (VE21) screen configuration. Different screen configurations and layouts are possible in the examination room and in the control room.

    AI/ML Overview

    This document, K220409, concerns the Siemens Medical Solutions USA, Inc. ARTIS pheno (VE21) System, an image-intensified fluoroscopic X-ray system. The submission seeks to demonstrate substantial equivalence to the previously cleared ARTIS pheno (VE20) system. The modifications in the VE21 system include updated system software (from VE20 to VE21) to support NOMSIE DSA (IQ) customization of contrast in DSA images and an optional new generator (Polydoros ACX).

    Here's a breakdown of the requested information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that a summary of comparison of technological characteristics was performed. However, it does not provide specific quantitative acceptance criteria or detailed reported device performance in a table format. Instead, it offers a qualitative statement about the testing results.

    Acceptance Criteria CategoryReported Device Performance (Summary)
    System Software ModificationsAll test results met all acceptance criteria.
    Safety & EffectivenessTest results demonstrate that the ARTIS pheno (VE21) System acceptance criteria are adequate for the intended use of the device. Results of all conducted testing and clinical assessment were found acceptable and do not raise any new issues of safety or effectiveness.
    Electrical SafetyComplies with AAMI ANSI ES60601-1:2005/(R)2012, IEC 60601-1-2:2014, IEC 60601-1-3:2013, IEC 60601-1-6:2010/A1:2013, IEC 60825-1:2007, TR 60878:2015, IEC 60601-2-28:2017, IEC 60601-2-43:2017, IEC 60601-2-54:2009/A1:2015.
    EMCComplies with IEC 60601-1-2:2014.
    Software ValidationPerformance data demonstrates continued conformance with special controls for medical devices containing software. Risk analysis was completed, and risk control implemented to mitigate identified hazards. All software specifications have met the acceptance criteria.
    UsabilityHuman Factor Usability Validation showed that human factors are addressed in the system test according to the operator's manual.
    CybersecurityConforms to cybersecurity requirements, considering IEC 80001-1:2010.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample size (e.g., number of images, or patient cases) used for the test set for image quality or clinical performance evaluations related to the software updates.

    Regarding data provenance, the document mentions "non-clinical tests" and "bench test summaries." The nature of these tests suggests they are primarily technical and engineering validations rather than studies involving patient data from a specific country of origin or a retrospective/prospective design. The modifications are described as "software and hardware changes [that] do not raise any new risks or any issues regarding the safety nor effectiveness of the device."


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not mention the use of experts or the establishment of ground truth for a test set in the context of the software or hardware modifications. The testing described focuses on technical compliance, verification, and validation, rather than clinical efficacy studies involving expert review of patient data.


    4. Adjudication Method for the Test Set

    Since there is no mention of a clinical test set requiring expert ground truth establishment, there is no adjudication method described in the document.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned in the provided document. The submission focuses on demonstrating substantial equivalence through technical and functional testing of modifications rather than comparative clinical performance with human readers.


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The document does not explicitly describe a standalone performance study in the context of an algorithm's diagnostic accuracy without human intervention. The "NOMSIE DSA (IQ) Customization of Contrast in DSA images" is a software modification related to image quality customization, and its performance is assessed through verification and validation testing, not typically as an AI algorithm in a standalone diagnostic capacity as might be reported for a disease detection or quantification algorithm.


    7. Type of Ground Truth Used

    The "ground truth" for the testing performed appears to be based on:

    • Engineering Specifications and Standard Compliance: For electrical safety, EMC, software functionality, and other technical aspects, ground truth is conformity to established industry standards (e.g., IEC 60601 series, IEC 62304) and internal product specifications.
    • Verification and Validation Testing Results: The software specifications are considered met based on the outcomes of verification and validation tests.
    • Risk Analysis: Ground truth for safety is established by mitigating identified hazards through risk control measures.

    There's no indication of ground truth based on expert consensus, pathology, or outcomes data in the context of clinical diagnostic accuracy for an algorithm.


    8. Sample Size for the Training Set

    The document does not mention a training set sample size. This is consistent with the nature of the submission, which focuses on modifications to an existing device (ARTIS pheno VE20 to VE21) and the associated technical validation, rather than the development and training of a novel machine learning algorithm requiring a distinct training dataset.


    9. How the Ground Truth for the Training Set was Established

    Since no training set is mentioned for a new machine learning algorithm, there is no information provided on how ground truth for a training set would have been established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201156

    Validate with FDA (Live)

    Date Cleared
    2020-06-29

    (60 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

    Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography.

    Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions.

    ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.

    The ARTIS family include also the software option DynaCT with following indications for use:

    DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.

    DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

    Device Description

    The ARTIS pheno (VE2) system is a multi-axis system, specifically designed to meet the growing demands of high-end imaging for interventional radiology, interventional cardiology, minimally invasive and hybrid surgery procedures. The stand allows positioning in angular, orbital, lateral, longitudinal and vertical directions, leveraging the flexible isocenter. The ARTIS pheno (VE2) system is equipped with a robotic multi-axis floor stand, C-arm, flat panel detector, x-ray tube, collimator, high voltage generator, patient table, and image post processing. The ARTIS pheno is partially coated with an optional anti-microbial coating. syngo Application Software is optional, available for the support of dedicated clinical workflows.

    The ARTIS pheno (VE2) covers the complete range of angiographic applications, cardiac angiography, neuro-angiography, general angiography, surgery and surgical angiography, multipurpose angiography, rotational angiography and radiographic/ fluoroscopic procedures.

    1. The following components are configured to create the ARTIS pheno VE2 system: Floor stand with C-arm, X-ray tube assembly and FD
    2. Patient table
    3. Display ceiling suspension with displays
    4. Footswitch for releasing radiation
    5. Control console for controlling the stand, patient table, collimator and imaging system.

    Images and operating elements are displayed on screens. Different display variants are used to visualize image and information content. Panoramic display configurations or large displays can be used, configurable to visualize multiple images and information content in various layouts.

    Post processing can be done in the exam room or in the control room that offers monitors as well, with a footswitch location in the exam room or the control room. The ARTIS pheno (VE2) System is capable of 2D and 3D imaging.

    Other systems and software syngo Application Software, syngo X Workplace, Sensis, and or third-party systems may also be integrated into the ARTIS pheno (VE2) screen configuration. Different screen configurations and layouts are possible in the examination room and in the control room.

    AI/ML Overview

    The provided text is a 510(k) summary for the ARTIS pheno (VE2) system, which is an angiography system. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than proving a medical device's performance against specific acceptance criteria for a new clinical claim or an AI algorithm.

    Therefore, the document does not contain the detailed information required to fill out a table of acceptance criteria and reported device performance as it would for a study proving a device meets a clinical performance claim (e.g., diagnostic accuracy of an AI algorithm). The information provided is about the safety and effectiveness of the updated angiography system in terms of hardware, software modifications, and compliance with general medical device standards.

    Specifically, the document lists modifications to the ARTIS pheno (VE2) system, such as improved roadmap features, new marketing terminology, updated user interface, and hardware changes. It states that non-clinical performance testing was conducted to assess functionality and compliance with various electrical, safety, and performance standards. However, it does not present clear, quantifiable acceptance criteria or reported performance results in the format requested for a study proving device performance against specific clinical metrics.

    Here's why the requested information cannot be extracted from this document:

    • No specific clinical performance metrics: The document discusses improvements like "increased image quality dose ratio" and "Improved Image Quality due to improved algorithm" but without quantifiable acceptance criteria (e.g., "sensitivity > X%", "specificity > Y%", "AUC > Z%") or reported numerical results against such criteria.
    • Focus on substantial equivalence: The core of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device, meaning it's as safe and effective. This is different from proving that a device (especially an AI-driven one making a new diagnostic claim) meets specific, pre-defined performance thresholds in a clinical study.
    • No mention of AI algorithm performance: While there's talk of "improved algorithms" and "syngo DynaCT Multiphase," the document doesn't define or evaluate an AI algorithm in the way you'd expect for a device making a new AI-based diagnostic claim. It's about overall system improvements, not a groundbreaking AI diagnostic tool.
    • Lack of test set details: There's no mention of a specific test set, its sample size, data provenance, or ground truth establishment for clinical performance evaluation of a new function (like an AI algorithm would require).
    • No human reader studies: There is no mention of MRMC studies or human-in-the-loop performance evaluations.
    • No training set details: Information about a training set for an AI algorithm (sample size, ground truth establishment) is absent.

    Therefore, I cannot fulfill your request for the table and study details based on the provided text. The document describes a traditional medical device modification clearance, not a study validating an AI algorithm against specific clinical performance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163286

    Validate with FDA (Live)

    Device Name
    ARTIS pheno
    Date Cleared
    2017-03-09

    (107 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

    Procedures that can be performed with the ARTIS family include cardiac angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.

    Additional procedures that can be performed include angiography in the operating room, image-guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions.

    ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.

    The ARTIS family include also the software option DynaCT with following indications for use:

    DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.

    DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

    Device Description

    The ARTIS pheno is a further development of the Artis Q and Artis Q.zen -Modular Angiography System. ARTIS pheno is equipped with C-arm, stand, flat panel detector, x-ray tube, high voltage generator, patient table, and image post processing. syngo Application Software is optional available for the support of dedicated clinical workflows.

    AI/ML Overview

    The provided text describes the ARTIS pheno device and its substantial equivalence to predicate devices, but it does not contain a detailed study with specific acceptance criteria and reported device performance metrics in the format requested.

    The document states that "Non-clinical tests were conducted for ARTIS pheno, during product development" and that "The modifications described in this Premarket Notification were supported with verification and validation testing." It also mentions "The performance data demonstrates continued conformance with special controls for medical devices containing software" and "The testing results support that all the software specifications have met the acceptance criteria."

    However, it does not provide the actual acceptance criteria or the specific numerical performance results achieved by the device in a table format. It only lists various standards the device complies with (e.g., ES60601-1:2005/(R)2012) and general statements about the tests being passed.

    Therefore, I cannot populate the requested table or provide detailed answers to questions 2 through 9 based on the provided text. The document refers to "non-clinical performance testing" and "verification and validation testing" but does not elaborate on the specific methodologies, sample sizes, ground truth establishment, or expert involvement for these tests.

    Summary of available information regarding performance:

    • Acceptance Criteria/Reported Performance: Not explicitly stated in a quantifiable manner or table. The document generally states that "all tests were passed" and that "all the software specifications have met the acceptance criteria."
    • Sample size for test set and data provenance: No information provided.
    • Number of experts and qualifications for ground truth: No information provided.
    • Adjudication method: No information provided.
    • MRMC comparative effectiveness study: No information provided.
    • Standalone (algorithm only) performance: No information provided the software is a component of the interventional fluoroscopic x-ray system, not a standalone AI device with specific algorithmic performance metrics.
    • Type of ground truth used: "Non-clinical performance testing" and "software specifications" are mentioned, suggesting internal validation against defined functional requirements and safety standards. No mention of expert consensus, pathology, or outcomes data for performance evaluation in the context of diagnostic accuracy.
    • Sample size for training set: No information provided.
    • How ground truth for training set was established: No information provided.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics, adherence to safety and performance standards for an interventional fluoroscopic x-ray system, and successful completion of verification and validation activities. It does not provide the kind of detailed study results one might expect for a novel AI/CAD device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1