Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K192441

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-17

    (41 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock Suture Anchors are intended to be used for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, hip, knee, hand/wrist, elbow, and hip in the following procedures:
    Shoulder: Rotator Cuff Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy.
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Quadriceps Tendon Repair and Meniscal root repair (4.75 SwiveLock C only). Secondary fixation for ACL/PCL reconstruction or repair (4.75 - 5.5. SwiveLock only).
    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Radial Collament Reconstruction.
    Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Unar or Radial collateral Ligament Reconstruction, Lateral Epicondylitis Repair.
    Hip: Capsular Repair, acetabular labral repair

    Device Description

    The Arthrex SwiveLock Anchor is a BioComposite two-component, knotless suture anchor comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is pre-mounted on a driver with the anchor body and eyelet physically separated on the driver shaft. Arthrex cleared suture may also be provided with the device.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Arthrex SwiveLock Anchors, specifically for the addition of a meniscal root repair indication. It's a regulatory submission, not a study report per se, and it pertains to a physical medical device (suture anchors), not a software or AI-driven device.

    Therefore, many of the requested fields are not applicable to this type of document or device. For example, there's no "device performance" in terms of accuracy metrics, no "test set" in the computational sense, no "ground truth" derived from expert consensus on images, no "training set," and no "AI assistance."

    However, I can extract information related to the performance data provided to support the new indication.

    Here's the breakdown of what can be extracted from the provided text, and where the requested information is not applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Meniscal root repair (new indication)Demonstrated that the ultimate load and displacement of the Arthrex SwiveLock are within the acceptable range for meniscal root repair.
    Pyrogen Limit SpecificationsDevice meets pyrogen limit specifications.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated numerically (e.g., how many anchors were tested, how many repetitions). The document refers to "testing" performed.
    • Data Provenance: The testing was conducted to demonstrate the mechanical properties and biocompatibility of the device for the new indication. It would be considered prospective testing for the purpose of this submission, performed by the manufacturer (Arthrex Inc.). Country of origin is not specified, but the manufacturer is based in Naples, Florida, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Not Applicable: This relates to a physical device, not an image-based diagnostic or AI system. The "ground truth" here is objective physical and chemical testing results, not expert interpretation.

    4. Adjudication Method for the Test Set

    • Not Applicable: As above, this is for physical/chemical testing, not expert adjudication of diagnostic outputs. The assessment of whether the ultimate load and displacement are "within the acceptable range" would be based on predefined engineering specifications, not human adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No: This is not relevant for a physical medical device like a suture anchor. MRMC studies are typically for diagnostic imaging systems or AI products.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not Applicable: This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    • Engineering/Mechanical Specifications and Biocompatibility Standards: The "ground truth" for the performance data involves established mechanical thresholds for ultimate load and displacement in meniscal root repair applications, as well as pyrogen limit specifications defined by standards like EP 2.6.14/USP <85>. It's based on objective physical and chemical measurements rather than expert consensus on interpretation.

    8. The Sample Size for the Training Set

    • Not Applicable: This is a physical device, and the concept of a "training set" for an algorithm does not apply.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable: As above, no training set for an algorithm exists here.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190728

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-04-17

    (27 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

    Shoulder: Rotator Cuff Repair, Bankart Repair, Siceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair and Bunionectomy.

    Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 SwiveLock only).

    Hand Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.

    Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction.

    Hip: Capsular repair, acetabular labral repair. Proximal Hamstring Repair (4/75-5.5 PEEK SwiveLock only).

    Device Description

    The Arthrex SwiveLock Anchor is a two-component, knotless suture anchor comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is premounted on a driver with the anchor body and eyelet physically separated on the driver shaft. Arthrex 510(k) cleared suture may also be provided with the device.

    AI/ML Overview

    This provides an analysis of the provided text regarding the Arthrex SwiveLock Anchors.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission for the Arthrex SwiveLock Anchors (K190728) describes a comparison to predicate devices, focusing on mechanical performance. The "acceptance criteria" can be inferred from the statement that the device performs "statistically equivalent" to the predicate.

    Acceptance Criteria (inferred)Reported Device Performance
    Statistically equivalent cyclic pull-out performance to predicate devices (K180768 Arthrex SwiveLock Suture Anchor and K150768 Zimmer Biomet JuggerKnot Soft Anchor).Cyclic pull-out testing demonstrated that the Arthrex SwiveLock Anchor performs statistically equivalent to the predicate devices.
    Meets pyrogen limit specifications per EP 2.6.14/USP <85>.Bacterial endotoxin testing per EP 2.6.14/USP <85> was conducted and demonstrated that the device meets pyrogen limit specifications.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not explicitly state the numerical sample size (n) used for the cyclic pull-out testing. It only mentions that "Cyclic pull-out testing was performed."
    • Data Provenance: Not explicitly stated, but given it's a 510(k) submission from a U.S. company (Arthrex Inc., Naples, FL), the testing would typically be conducted in a controlled lab environment, likely in the US or by a recognized testing facility. It would be considered prospective testing for the purpose of this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device and study. The "ground truth" for mechanical testing of a medical device like a suture anchor is based on objective, quantifiable physical measurements (e.g., force, displacement, material properties) rather than expert opinion or diagnostic interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving human interpretation (e.g., image reading) where there might be disagreement among experts. For mechanical testing, the results are typically derived from instrumentation, and judgment or adjudication in that sense is not usually employed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are typically performed for diagnostic imaging devices where human readers interpret results, and the AI's impact on their performance is evaluated. This submission is for a surgical implant (suture anchor), where performance is assessed through biomechanical testing, not reader interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone performance assessment was done. The "Performance Data" section describes "Cyclic pull-out testing" and "Bacterial endotoxin testing." These are laboratory-based, objective tests conducted on the device itself without human intervention in the performance measurement process.

    7. Type of Ground Truth Used

    The ground truth used for the performance data is based on objective, quantitative physical and biological measurements:

    • Biomechanical Metrics: For cyclic pull-out testing, the "ground truth" is the measured force-displacement characteristics and failure modes, compared against established performance benchmarks of the predicate devices.
    • Physicochemical Analysis: For bacterial endotoxin testing, the "ground truth" is the measured endotoxin level compared against predefined pyrogen limit specifications.

    8. Sample Size for the Training Set

    This information is not applicable as this device does not involve an AI algorithm that requires a "training set." The device is a physical medical implant whose performance is evaluated through mechanical and biological testing, not by learned patterns from data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no AI algorithm or training set involved with this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180768

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-12-14

    (266 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

    Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.

    Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 SwiveLock only).

    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.

    Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.

    Hip: Capsular repair, acetabular labral repair.

    Device Description

    The Arthrex SwiveLock Anchor is a two-component, knotless suture anchor comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is premounted on a driver with the anchor body and eyelet physically separated on the driver shaft. Arthrex 510(k) cleared suture may also be provided with the device.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: Arthrex SwiveLock Anchors. It is primarily focused on demonstrating substantial equivalence to predicate devices for the purpose of a regulatory clearance for an expanded indication (quadriceps tendon repair).

    Crucially, this document does not contain information about an AI/ML-driven device or study that would involve acceptance criteria for algorithms, human-in-the-loop performance, or the other specific details requested in your prompt (e.g., sample size for test/training sets, expert consensus, MRMC studies, ground truth establishment methods for AI).

    The "Performance Data" section of this 510(k) summary refers to:

    • Static pull-out testing of the Arthrex SwiveLock Anchor and a predicate device.
    • Cyclic pull-out testing of the subject device after degradation, compared to a non-resorbable predicate.
    • Bacterial endotoxin testing (pyrogen limit specifications).

    These are biomechanical and biocompatibility tests for a physical implantable device, not performance metrics for an AI algorithm.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves an AI device meets acceptance criteria based on this document. The document describes the regulatory submission for a physical medical device (suture anchors) and its associated mechanical and biological performance, not an AI product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171141

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-04

    (16 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:
    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis. Secondary fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 SwiveLock only).
    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
    Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.
    Hip: Capsular repair, acetabular labral repair.

    Device Description

    The Arthrex SwiveLock Anchor is a two-component, knotless suture anchor composed of an eyelet and a hollow anchor body. The SwiveLock Anchor is pre-mounted on a driver with the anchor body and eyelet physically separated on the driver shaft. Arthrex suture may also be provided with the device.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device, the Arthrex SwiveLock Anchors. It focuses on demonstrating substantial equivalence to a predicate device rather than undergoing a new, comprehensive clinical trial for a novel AI device. Therefore, the questions related to AI-specific study design (such as human reader improvement with AI assistance, ground truth for training sets, etc.) are not directly applicable to this document's content.

    However, I can extract the information relevant to the device's performance testing and how it meets acceptance criteria for substantial equivalence to its predicate.

    Here's the relevant information based on the provided text, framed as closely as possible to your request for acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    For a 510(k) submission for a modification, the "acceptance criteria" are typically defined by demonstrating that the modified device performs equivalently to the predicate device in relevant performance tests, or that the modifications do not introduce new safety or effectiveness concerns. The "reported device performance" is then the outcome of these tests.

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Performance: The modified device must perform equivalently to the predicate device in terms of:
    * Pull-out strengthTesting was performed to demonstrate that the SwiveLock Anchors with the proposed modified eyelet are substantially equivalent to the predicate.
    * Compressive load resistanceTesting was performed to demonstrate that the SwiveLock Anchors with the proposed modified eyelet are substantially equivalent to the predicate.
    * Mode of failureTesting was performed to demonstrate that the SwiveLock Anchors with the proposed modified eyelet are substantially equivalent to the predicate.
    Biocompatibility/Safety: The device meets pyrogen limits.Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate that the device meets pyrogen limit specifications.
    Overall Substantial Equivalence: Differences between the proposed device and predicate are minor and do not raise questions concerning safety or effectiveness.The Arthrex SwiveLock Anchors are substantially equivalent to the predicate device in which the basic design features and intended uses are the same. Any differences... are considered minor and do not raise questions concerning safety or effectiveness.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify the exact number of units or samples used for the pull-out, compressive load, and mode of failure testing, nor for the bacterial endotoxin testing. This level of detail is typically found in the full test reports, which are not part of this summary.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given it's a 510(k) for a manufacturing modification, the testing would generally be prospective laboratory testing conducted by the manufacturer (Arthrex Inc., located in Naples, Florida, USA).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not a study involving human interpretation of medical images or data requiring expert consensus for ground truth. The "ground truth" for mechanical testing is established by the physical testing equipment and methods themselves (e.g., measuring force, observing failure points).

    4. Adjudication method for the test set

    Not applicable. This is not a study involving human-based data interpretation or classification where adjudication would be necessary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device or a study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    For the mechanical tests (pull-out, compressive load, mode of failure), the ground truth is established through physical measurement and observation during standardized laboratory testing. For the bacterial endotoxin test, the ground truth is established by laboratory assay results against a pyrogen limit specification.

    8. The sample size for the training set

    Not applicable. This is not an AI device that requires a training set. The device is a modification of an existing mechanical anchor.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no AI training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151342

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-03-24

    (310 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:
    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis. Secondary fixation for ACL/PCL reconstruction or repair (4.75 – 5.5 SwiveLock only).Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.Hip: Capsular repair, acetabular labral repair.

    Device Description

    The Arthrex SwiveLock Anchor is a two-component, knotless suture anchor comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is pre-mounted on a driver with the anchor body and eyelet physically separated on the driver shaft. FiberWire suture may also be provided with the device.

    AI/ML Overview

    The provided document is a 510(k) summary for the Arthrex SwiveLock Anchors, specifically an amendment to include a new indication for secondary fixation in ACL/PCL reconstruction or repair. This document does not describe a study involving an AI device or a comparative effectiveness study with human readers.

    Instead, it focuses on demonstrating substantial equivalence of a medical device (surgical anchors) to previously cleared predicate devices through specific technical and biomechanical testing, rather than clinical performance based on diagnostic accuracy or AI-assisted interpretation.

    Therefore, many of the requested fields related to AI performance, human reader studies, and ground truth establishment for a diagnostic algorithm cannot be answered from this document.

    However, I can extract information related to the acceptance criteria and the study that demonstrates the device meets these criteria in the context of this 510(k) submission for the Arthrex SwiveLock Anchors.

    Here's a breakdown of the information that can be extracted or inferred from the document:

    1. A table of acceptance criteria and the reported device performance:

    The document states: "Biomechanical testing (ultimate load) was conducted to demonstrate that the use of the SwiveLock as a secondary fixation in ACL reconstruction is greater than a construct without it."

    While specific numerical acceptance criteria (e.g., "ultimate load must be > X Newtons") are not explicitly stated as a table in this summary, the acceptance criterion is implicitly "greater than a construct without it". The reported performance is that the SwiveLock did meet this criterion.

    Acceptance Criteria (Implicit)Reported Device Performance
    Ultimate load of construct with SwiveLock as secondary fixation in ACL reconstruction must be greater than a construct without SwiveLock.The biomechanical testing demonstrated that the ultimate load was "greater than a construct without it."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not explicitly stated in this 510(k) summary. Biomechanical testing usually involves a defined number of samples, but the exact count is not provided.
    • Data Provenance: The study is described as "biomechanical testing," implying laboratory-based testing rather than clinical data from a specific country. It's a study conducted specifically to support this regulatory submission, not a retrospective or prospective clinical data collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This question is not applicable. For a biomechanical test of a surgical anchor, there isn't "ground truth" derived from expert consensus on diagnostic images. The "truth" is established by the physical testing methods themselves, performed according to established engineering and mechanical testing standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as this refers to adjudication of expert opinions for ground truth establishment in studies involving human interpretation or subjective assessment, which is not the case here.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI device, and no MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this type of device and study is the direct measurement of physical properties (ultimate load) under controlled laboratory conditions, following validated biomechanical testing protocols. It's not based on expert consensus, pathology, or outcomes data in the traditional medical sense, but rather on direct empirical evidence from engineering tests.

    8. The sample size for the training set:

    Not applicable. This is not an AI device requiring a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101823

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-01-07

    (191 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

    Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy.
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.
    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction.
    Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair.
    Hip: Capsular Repair, acetabular labral repair

    Device Description

    The SwiveLock Anchor is a two-component, knotless suture anchor comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is pre-mounted on a driver with the anchor body and eyelet physically separated on the driver shaft. FiberWire suture may also be provided with the device.

    AI/ML Overview

    The provided 510(k) summary (K101823) describes the Arthrex SwiveLock Anchors, a medical device intended for fixing suture (soft tissue) to bone in various anatomical regions. The device is seeking substantial equivalence to previously marketed predicate devices.

    1. Acceptance Criteria and Reported Device Performance:

    The primary acceptance criteria for the Arthrex SwiveLock Anchors are based on demonstrating substantial equivalence to existing predicate devices. Since this is an equivalence submission, specific quantitative acceptance criteria for absolute performance metrics are not explicitly stated in the summary. Instead, the performance is compared against the known performance of the predicate devices.

    The study aimed to demonstrate that the SwiveLock Anchors perform comparably to the predicate devices in terms of:

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
    Degradation (for Bio/BioComposite materials)26-week degradation data demonstrated substantial equivalence.
    Mechanical Pull-Out (Tensile) StrengthMechanical pull-out (tensile) testing data demonstrated substantial equivalence to the pull-out forces of the predicate devices.
    Insertion CharacteristicsInsertion testing data demonstrated substantial equivalence.
    In-vivo Performance (Histology, Pull-Out, Imaging)Animal testing (including histology, in-vivo pull-out, and in-vivo imaging) data demonstrated no new issues of safety and effectiveness, implying comparable in-vivo performance.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes for each test mentioned (degradation, mechanical pull-out, insertion, animal testing). It only refers to "data" for each category.

    The data provenance is not specified. It is an industry standard to conduct these tests in a controlled laboratory environment or animal models. Neither the country of origin nor whether the data is retrospective or prospective is mentioned.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not applicable to this type of 510(k) submission. Establishing "ground truth" by experts is typically relevant for diagnostic or AI-driven devices where human interpretation is involved. For a mechanical device like a suture anchor, the "truth" is established through physical and biological testing against established standards or predicate device performance.

    4. Adjudication Method:

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human readers/experts in diagnostic studies, which is not relevant for the mechanical and biological testing of a suture anchor.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data, and the effect of AI assistance on their performance is evaluated.

    6. Standalone (Algorithm Only) Performance Study:

    No, a standalone (algorithm only) performance study was not done. This device is a physical medical implant, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. Type of Ground Truth Used:

    The "ground truth" in this context is established through physical and biological testing methodologies. This includes:

    • Degradation data: Measuring the breakdown of biomaterials over time.
    • Mechanical pull-out (tensile) testing data: Quantifying the force required to pull the anchor out of bone, adhering to engineering standards.
    • Insertion testing data: Assessing the ease and characteristics of device insertion.
    • Animal testing data: Observing the biological response (histology), in-vivo pull-out strength, and imaging of the device in a living system.

    The "truth" is rooted in the physical and biological properties and performance of the device as measured by these tests, compared against the known performance of predicate devices.

    8. Sample Size for the Training Set:

    This information is not applicable. "Training set" refers to data used to train machine learning algorithms. The Arthrex SwiveLock Anchor is a mechanical device, not an AI or algorithm-driven system.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1