    Intended Use

    The Amuchina Automatic Reprocessing Machine (ARM) is a device designed for both (a) reprocessing hemodialyzers for reuse, and (b) preprocessing hemodialyzers prior to first use. Reprocessed hemodialyzer will be reused on the same patient on who originally used the hemodialyzer. Both the reprocessing and preprocessing procedures use peracetic acid/hydrogen peroxide based disinfectant.

    Device Description

    The Amuchina ARM Automatic [Dialyzer] Reprocessing Machine is a stand alone device designed for the automated reprocessing of hemodialyzers for reuse and for the pre-processing of hemodialyzers prior to first use. The ARM Unit has 4 stations which can sequentially process up to 4 dialyzers at one time. The ARM Unit has no direct or indirect patient contact. The ARM Unit uses a peracetic acid/hydrogen peroxide based disinfectant as both a cleaning solution and a disinfectant. The disinfectant concentrate is diluted to the inuse strength with AAMI quality water. When reprocessing dialyzers, the ARM Unit uses the following cycles: Rinse, Cleaning, Flush, Volume & Leak Test, and Disinfection. When pre-processing dialvzers, the following cycles are used: Flush, Volume & Leak Test (only if instructed for by the user), and Disinfection. For regularly scheduled maintenance, the ARM Unit has the following system cycles: System Rinse, System Disinfect, and System Self Test. Other cycles which are included in the ARM Unit include: Prime Pump for priming the chemical pump with the disinfecting agent, Line Volume Calibration for use in determining the total cell volume of the dialyzer, and System Void to purge fluids from the circuits prior to moving the machine. The ARM Unit incorporates the feature of including a patient photograph on the dialyzer label, thus reducing the possibility of reused dialyzers being used on the wrong patient.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Amuchina ARM Automatic Reprocessing Machine:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary for the Amuchina ARM Automatic Reprocessing Machine (K013713) does not explicitly present a table of "acceptance criteria" in the typical sense of quantitative performance metrics with associated thresholds (e.g., sensitivity > 90%, specificity > 85%). Instead, the "acceptance criteria" are implied by the comparative features and the results of the non-clinical performance and in-vitro testing.

    The primary acceptance criterion appears to be functional equivalence to the predicate device (Seratronic DRS-4 Dialyzer Reprocessing System, K860674) for the intended uses of reprocessing and pre-processing dialyzers. The device's performance is demonstrated by functionality tests and in-vitro testing confirming proper dilution of the disinfectant.

    FeatureAcceptance Criteria (Implied by Predicate & Intended Use)Reported Device Performance (Amuchina ARM)
    Intended UseReprocessing hemodialyzers for reuseReprocessing hemodialyzers for reuse on the same patient.
    Pre-processing hemodialyzers prior to first usePre-processes hemodialyzers prior to first use.
    Number of Stations44
    Cleaning SolutionsCapable of using Peracetic Acid/Hydrogen PeroxideUses Peracetic Acid/Hydrogen Peroxide.
    DisinfectantsCapable of using Peracetic Acid/Hydrogen PeroxideUses Peracetic Acid/Hydrogen Peroxide.
    CyclesExpected cycles for reprocessing (Clean, Disinfect, Test)Rinse, Cleaning, Flush, Volume & Leak Test, Disinfection (reprocessing). Flush, Volume & Leak Test (optional), Disinfection (pre-processing). System Rinse, System Disinfect, System Self Test, Prime Pump, Line Volume Calibration, System Void (maintenance/utility).
    Test Cycle ContentsPressure Leak Test, Total Blood Cell VolumePressure Leak Test, Total Blood Cell Volume.
    Dialyzer LabelingComprehensive patient/dialyzer information for reuse safetyBar Code, Patient Name, Patient Photograph (digital), Dialyzer Lot Number, Dialyzer Serial Number, Number times reprocessed, Number times reused, Test Results, Dialyzer Status.
    Microprocessor ControlYesYes
    Interactive Touch ScreenYesYes
    Operator Defined ParametersYesYes
    Patient Photograph Feature(Not applicable to predicate, but an added feature for safety)Yes (reduces possibility of wrong patient use).
    Water QualityAAMI QualityAAMI Quality
    FunctionalityPerform all intended/programmed functions as designed."Each individual function of the ARM Unit was tested to see if they performed as intended/programmed. No errors or failures either-were detected."
    Dilution AccuracyProperly dilute cleaner/disinfectant concentrate."In-vitro testing was performed to assure the ARM Unit properly diluted the cleaner/disinfectant concentrate to the in-use concentrations of active ingredients. The results from these tests show that the ARM Unit performed as expected."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a sample size for a "test set" in the context of a clinical study or a large-scale data analysis. The testing described is "Non-Clinical Performance Data" and "In-vitro testing."

    • Test Set Sample Size: Not specified in terms of number of dialyzers or patient cases. The description implies a series of functional tests on the device itself and in-vitro tests on the dilution capabilities.
    • Data Provenance: The testing is described as non-clinical and in-vitro, meaning it was likely conducted in a controlled laboratory environment by the manufacturer (Amuchina International, Inc.). There is no mention of country of origin for data specifically, beyond the company's address in Gaithersburg, MD, USA. It is retrospective in the sense that the testing was completed before the 510(k) submission. No prospective clinical data is mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the summary. For non-clinical and in-vitro functional testing of a device like this, ground truth is typically established by engineering specifications, chemical analysis, and adherence to established standards (e.g., AAMI quality water). There's no mention of human experts interpreting results for a test set in the way a radiologist might for medical images.

    4. Adjudication Method for the Test Set

    This is not applicable as there's no mention of human interpretation of results requiring adjudication (e.g., by multiple experts). The tests described are objective functional and chemical tests.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done, nor is it applicable. This device is an automated reprocessing machine; it does not involve human "readers" or AI assistance for diagnostic interpretation. Its function is to mechanically and chemically prepare dialyzers for reuse.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a form of standalone performance was demonstrated through the non-clinical and in-vitro testing. The device's "individual functions" and its ability to "properly dilute the cleaner/disinfectant concentrate" were tested as an automated system without human intervention influencing the core performance metrics. The machine itself is designed to operate without a "human-in-the-loop" for its primary reprocessing functions, beyond operator setup and maintenance.

    7. The Type of Ground Truth Used

    The ground truth used for this device's evaluation is primarily based on:

    • Engineering Specifications and Design Requirements: Ensuring each function performs "as intended/programmed."
    • Chemical Analysis: Verifying "in-use concentrations of active ingredients" following dilution.
    • Comparison to Predicate Device's Performance Characteristics: Demonstrating substantial equivalence to the Seratronic DRS-4.
    • Adherence to Standards: Implicitly, AAMI quality water standard.

    There is no mention of pathology, outcomes data, or expert consensus in a clinical diagnostic sense.

    8. The Sample Size for the Training Set

    There is no mention of a training set in this 510(k) summary. This type of device does not utilize machine learning or AI models that require a "training set." Its functionality is based on programmed logic and mechanical/chemical processes.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set mentioned or applicable for this device, this question is not applicable.