Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K163359

    Validate with FDA (Live)

    Date Cleared
    2017-08-18

    (261 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of methotrexate in human serum or plasma on automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of methotrexate to help ensure appropriate therapy.

    Specimens from patients who have received glucarpidase (carboxypeptidase G2) as a high dose methotrexate rescue therapy should not be tested with the ARK Methotrexate Assay.

    Device Description

    The ARK Methotrexate Assay is a homogeneous immunoassay based on competition between drug in the specimen and Methotrexate labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay.

    The ARK Methotrexate Assay consists of reagents R1 anti-Methotrexate polyclonal antibody with substrate and R2 Methotrexate labeled with bacterial G6PDH enzyme. The ARK Methotrexate Calibrator consists of a six-level set to calibrate the assay, and the ARK Methotrexate Control consists of a six-level set used for quality control of the assay (tri-level calibration range set and tri-level high range set). The ARK Methotrexate Dilution Buffer is equivalent to zero calibrator (Calibrator A).

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study information:

    Acceptance Criteria and Device Performance for ARK™ Methotrexate Assay

    The provided document describes the acceptance criteria and the study that proves the ARK™ Methotrexate Assay, when used on the Beckman Coulter AU680 analyzer, meets these criteria. The study aims to demonstrate substantial equivalence to the assay's performance on the predicate device, the Roche/Hitachi 917 analyzer.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states: "The pre-determined Pass/Fail acceptance criteria were met for all of the above performance studies." However, the specific numerical acceptance criteria or performance values for each characteristic are not provided in the given text. It only lists the types of performance characteristics evaluated.

    Performance CharacteristicAcceptance Criteria (Not Detailed)Reported Device Performance
    Limit of BlankUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Limit of DetectionUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Limit of QuantitationUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    RecoveryUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    LinearityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Accuracy (Method Comparison)Undisclosed Pass/Fail CriteriaMet Acceptance Criteria
    PrecisionUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    SpecificityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Cross-reactivityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Carry-overUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    Dilution RecoveryUndisclosed Pass/Fail CriteriaMet Acceptance Criteria
    On-board StabilityUndisclosed Pass/Fail CriteriaMet Acceptance Criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the test set for any of the performance studies.

    The document refers to "human serum or plasma" as the sample type. The country of origin of the data is not explicitly stated. The study is a "validation and verification" activity, implying it was conducted specifically for this submission, likely making it a prospective study for the purpose of device validation.

    3. Number of Experts and Qualifications for Ground Truth

    The concept of "experts" and their qualifications for establishing ground truth, as typically seen in image analysis or diagnostic interpretation, is not applicable to this type of in vitro diagnostic device (immunoassay) study. The ground truth for this device's performance characteristics is established through analytical methods and reference standards, not expert interpretation.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (e.g., 2+1, 3+1) is not applicable to this type of in vitro diagnostic device (immunoassay) study. Performance is assessed through quantitative measurements against established analytical criteria, not through expert consensus on diagnostic outcomes.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this type of in vitro diagnostic device (immunoassay). This study design is typically used for imaging or diagnostic interpretation tasks where human readers' performance is being evaluated, often with and without AI assistance. This device is an automated assay, not an AI for human interpretive tasks.

    6. Standalone (Algorithm Only) Performance Study

    The study described is a standalone performance study for the algorithm (assay system) on the Beckman Coulter AU680 analyzer. The validation and verification activities listed (Limit of Blank, Limit of Detection, Accuracy, Precision, etc.) directly evaluate the performance of the device itself, without human interpretation in the loop. The purpose is to demonstrate that the assay system alone performs comparably to its predicate.

    7. Type of Ground Truth Used

    For an immunoassay like the ARK™ Methotrexate Assay, the "ground truth" for evaluating its performance characteristics (analytical accuracy, precision, linearity, etc.) would be established through:

    • Reference materials/standards: Calibrators and controls with precisely known concentrations of methotrexate.
    • Reference methods: Comparison with established, validated analytical methods (e.g., LC-MS/MS, or the predicate device's performance which was previously validated) for samples with unknown concentrations.
    • Spiking studies: Adding known amounts of analyte to a sample and measuring recovery to assess accuracy.

    The document implicitly refers to these as part of the "analytical performance studies" and "validation and verification activities."

    8. Sample Size for the Training Set

    The document does not mention a training set or its sample size. This is expected as the ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay kit, not a machine learning or AI model that requires a "training set" in the conventional sense. The "training" for such a system would involve validating its components and establishing its operational parameters on the target analyzer.

    9. How Ground Truth for the Training Set Was Established

    Since there is no "training set" in the context of machine learning, this question is not applicable. The assay is a chemical and biological system where parameters are set based on chemical principles and reagent characteristics, and then validated through performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111904

    Validate with FDA (Live)

    Date Cleared
    2011-10-18

    (105 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of Methotrexate in human serum or automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of methotrexate to help ensure appropriate therapy.

    The ARK™ Methotrexate Calibrator is intended for use in calibration of the ARK Methotrexate Assay.

    The ARKTM Methotrexate Control is intended for use in quality control of the ARK Methotrexate Assay.

    Specimens from patients who have received glucarpidase (carboxypeptidase G2) as a high dose methotrexate rescue therapy should not be tested with the ARK Methotrexate Assay.

    Device Description

    The ARK Methotrexate Assay is a homogeneous immunoassay based on competition between drug in the specimen and Methotrexate labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenyzme NAD functions only with the bacterial enzyme used in the assay.

    The ARK Methotrexate Assay consists of reagents R1 anti-Methotrexate polyclonal antibody with substrate and R2 Methotrexate labeled with bacterial G6PDH enzyme. The ARK Methotrexate Calibrator consists of a six-level set to calibrate the assay, and the ARK Methotrexate Control consists of a six-level set used for quality control of the assay (tri-level calibration range set and tri-level high range set). The ARK Methotrexate Dilution Buffer is equivalent to zero calibrator (Calibrator A).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies that prove the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Limit of Quantitation (LoQ)LoB < Result < LoQ: report "analyte detected; concentration < LoQ" Result ≥ LoQ: report the result as measuredLoB: 0.01 µmol/L LoD: 0.02 µmol/L LoQ: 0.04 µmol/L (LoQ-2 SD > LoD met)
    Accuracy (Analytical Recovery)Not explicitly stated as a numerical criterion in the "Acceptance Criteria" section for each data point, but typically implied to be close to 100% recovery.Mean percentage recovery: 102.1% (Individual recoveries ranged from 98.3% to 111.1%)
    LinearityPercent difference was ±10% between the predicted 1st and 2nd order regressed values for concentrations >0.10 µmol/L or ±0.01 µmol/L at concentrations ≤ 0.10 µmol/L.All observed differences for concentrations >0.10 µmol/L were within ±10% (-2.3% to 4.8%).All observed differences for concentrations ≤ 0.10 µmol/L were within ±0.01 µmol/L (-0.010 µmol/L to -0.004 µmol/L). The highest theoretical concentration tested was 1.30 µmol/L. Regression of assayed methotrexate concentrations was linear throughout the range of 2 to 1200 µmol/L when proportionally diluted.
    Assay RangeNot explicitly stated as an acceptance criterion, but the functional range.Measurement range: 0.04 - 1.20 µmol/L
    Method ComparisonNot explicitly stated as a numerical acceptance criterion (e.g., minimum correlation coefficient or confidence intervals for slope/intercept), but comparison to a predicate device is expected.Range 0.04 to 1.19 µM (102 samples): Slope: 1.00 (95% CI: 1.00 to 1.02) y-intercept: 0.01 (95% CI: 0.00 to 0.01) Correlation Coefficient (r²): 0.978 (95% CI: 0.968 to 0.985)Range 0.04 to 1440 µM (147 samples, including diluted): Slope: 0.99 (95% CI: 0.96 to 1.00) y-intercept: 0.01 (95% CI: 0.01 to 0.01) Correlation Coefficient (r²): 0.998 (95% CI: 0.997 to 0.998)
    Precision<10% total CV at >0.10 µmol/L SD ≤0.01 at ≤0.10 µmol/LAll control and patient pool samples met the criteria: - ARK Methotrexate Control (MID, HIGH, 5, 50, 500 µmol/L): Total CVs ranged from 3.8% to 7.0%. - ARK Methotrexate Control (LOW, 0.06 µmol/L): Total SD was 0.007, which is ≤0.01. - Patient Pool (MID, HIGH, 5, 50, 500 µmol/L): Total CVs ranged from 5.3% to 7.2%. - Patient Pool (LOW, 0.07 µmol/L): Total SD was 0.008, which is ≤0.01.
    Interfering SubstancesNot substantially affected by tested endogenous substances.Measurement of methotrexate was not substantially affected by clinically high concentrations of Albumin, Bilirubin (conjugated & unconjugated), Cholesterol, Gamma-Globulin, Hemoglobin, Intralipid®, Rheumatoid Factor, Triglycerides, and Uric Acid.
    Specificity - Crossreactivity< 0.07% crossreactivity with 7-Hydroxymethotrexate.7-Hydroxymethotrexate: ≤ 0.07% crossreactivity. DAMPA: 64.3% to 100% crossreactivity (significant). Triamterene: Crossreactivity ranged from 1.85% to 3.32%. Trimethoprim: Crossreactivity ranged from 0.12% to 0.54%. Folate analogs and other compounds (e.g., Adriamycin, Cyclophosphamide): ≤ 0.01% crossreactivity at ≥ 1000 µmol/L.
    AnticoagulantsNo significant difference in recovery between serum and plasma.No significant difference found in methotrexate recovery between serum and plasma samples.
    Sample StabilityAcceptable stability under various storage conditions.Stable frozen (at least 15 months), 48 hours at room temperature, refrigerated (at least 21 days), and after three (3) successive freeze/thaw cycles.
    On-Board StabilityAcceptable stability of calibration curve and reagents on the analyzer.Calibration curve effective up to at least 19 days. Reagents effective for up to at least 25 days in analyzer-specific containers. In-use stability of calibrator and controls also demonstrated.

    Study Details

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Limit of Quantitation (LoQ):
      • LoB, LoD: N = 60
      • LoQ: N = 40
      • Data provenance: Not specified, but generally analytical studies like this are conducted with prepared samples in a laboratory setting (prospective).
    • Accuracy: Concentrated methotrexate drug added into human serum negative for methotrexate. Six replicates of each sample were assayed.
      • Sample size: 5 target concentrations, each with 6 replicates (30 assays).
      • Data provenance: Not specified, but likely prospective, prepared samples in a lab.
    • Linearity: Dilutions of a 1.30 µmol/L serum sample and proportional dilutions of samples between 2 and 1200 µmol/L in pooled human serum.
      • Sample size: 13 different concentrations were tested for the initial linearity study. The second linearity study for higher concentrations (2-1200 µmol/L) stated "samples containing methotrexate... were prepared proportionally," implying multiple samples across this range.
      • Data provenance: Not specified, but likely prospective, prepared samples in a lab.
    • Method Comparison:
      • Sample size: 102 specimens within the measurement range (0.04 to 1.19 µM) and 147 specimens including those above the measurement range requiring dilution (total range 0.04 to 1440 µM).
      • Data provenance: Not specified, but these are likely clinical samples (human serum/plasma), which could be retrospective or prospective. The document does not specify country of origin.
    • Precision:
      • Sample size: Each of the 6 control levels and 6 patient pool levels were assayed in quadruplicate, twice a day for 20 days. This means 4 * 2 * 20 = 160 measurements per level.
      • Data provenance: Not specified, but likely prospective, prepared controls and pooled patient samples in a lab.
    • Interfering Substances: Multiple substances tested, each with samples at ~0.05 and ~0.50 µmol/L methotrexate.
      • Sample size: Each interfering substance was tested at two methotrexate levels, likely with duplicates or triplicates, plus control samples. The specific number of individual assays is not given but implies a significant number of tests.
      • Data provenance: Not specified, but likely prospective, prepared samples in a lab setting.
    • Specificity (Crossreactivity):
      • Sample size: Two concentrations of methotrexate (0.05 and 0.50 µmol/L) were tested with triamterene and trimethoprim. Various other compounds and folate analogs were tested at high concentrations (e.g., ≥ 1000 µmol/L), presumably just once for cross-reactivity.
      • Data provenance: Not specified, but likely prospective, prepared samples in a lab setting.
    • Anticoagulants:
      • Sample size: "Studies were conducted," implying multiple samples, but no specific number is given.
      • Data provenance: Not specified, but likely prospective, prepared samples in a lab setting.
    • Sample Stability:
      • Sample size: Not specified, but involved human specimens.
      • Data provenance: Not specified, but likely prospective.
    • On-Board Stability:
      • Sample size: Not specified, but utilized a calibration curve and reagents over time.
      • Data provenance: Not specified, but likely prospective, lab-based testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This device is an in vitro diagnostic assay for quantitative determination of methotrexate. The "ground truth" for such assays is typically established by:

    • Reference materials: Certified reference standards for methotrexate concentration (used in Accuracy, LoQ, Linearity, Crossreactivity).
    • Predicate device comparison: The "gold standard" or accepted method (Fluorescence Polarization Immunoassay - FPIA in this case, specifically the Abbott TDx®/TDxFLx® Methotrexate II assay) for method comparison.

    Therefore, the concept of "experts" establishing ground truth as it applies to image interpretation (like radiologists) or pathological diagnosis is not directly relevant here. The ground truth is analytical and based on established methodology and certified materials.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where subjective human interpretation (e.g., reading medical images) is involved and discrepancies between readers need to be resolved to establish a consensus ground truth. For in vitro diagnostic device performance studies, the results are quantitative and compared against analytical or predicate methods, not subjective expert consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an in vitro diagnostic assay, not an AI-powered image analysis tool or a device requiring human "readers" in the context of an MRMC study. Its performance is evaluated analytically, not by assessing human performance improvement with or without AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, all the performance studies described (analytical recovery, linearity, method comparison, precision, interference, specificity, stability) represent the standalone performance of the ARK Methotrexate Assay. There is no human-in-the-loop aspect described in these tests; the assay directly provides quantitative results.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The ground truth used for this in vitro diagnostic assay consists of:

    • Reference Standards: For accuracy, LoQ, linearity, and specificity, the ground truth was based on precisely known concentrations of methotrexate prepared using certified stock concentrates.
    • Predicate Device (Monoclonal FPIA): For method comparison, the ground truth was the results obtained from the legally marketed predicate device (Abbott TDx®/TDxFLx® Methotrexate II), which is an established method for methotrexate measurement.
    • Pre-defined Parameters/Conditions: For interference, specificity, and stability studies, the ground truth was the expectation of what the device should measure under specific well-defined conditions (e.g., no cross-reactivity with certain compounds, stability over a given period).

    8. The sample size for the training set:

    Not applicable. This document describes the validation of an immunoassay kit, not a machine learning or AI algorithm that requires a "training set." Immunoassays are based on biochemical reactions with antibodies, not pattern recognition learned from data.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no "training set" in the context of this immunoassay device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1