510(k) Data Aggregation

    K Number
    K133872

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-02

    (164 days)

    Product Code
    Regulation Number
    870.2770
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARIA WiFi Smart Scale is a body analyzer that measures body weight and uses bioelectrical impedance analysis (BIA) technology to estimate body fat percentage in generally healthy individuals 10 years of age or older. It is intended for home use only.

    Device Description

    ARIA is a body weight scale and a body fat analyzer that operates by using a low, safe, battery-generated electrical current through the body (using a bioelectrical impedance analysis technique) to provide body fat and body weight information. After the user registers their scale, the scale automatically recognizes the subject based on body weight and body fat readings. ARIA contains a WiFi module (802.11 module) that allows it to connect to the Internet in the user's home. The module provides a complementary interface to the Fitbit website. Body weight and body fat measurements are independent of internet communication after initial product registration.

    The ARIA scale automatically measures body weight and body fat composition. The scale recognizes the user based on previous weight readings, and can accept up to eight (8) different users. The 16 most recent readings are kept in memory on the scale and readings are also transmitted to the user's optional fitbit.com personal account for trending. If users have similar weight, the proper identity can be selected by tapping the scale.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Fitbit ARIA WiFi Smart Scale, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the ARIA WiFi Smart Scale are established not through explicit numerical thresholds but by demonstrating substantial equivalence to a predicate device (Withings Smart Body Scale K121971) and showing that its body fat measurements are not statistically different from the predicate, with variation within an acceptable range.

    Feature/MetricAcceptance Criteria (Implied from Predicate/Study)Reported Device Performance
    Substantial EquivalenceDemonstrates equivalence in technology, intended use, classification, product code, indication for use, device description, analysis method, operating parameters, number of electrodes, power source, IP connectivity, and measured parameters to the predicate device (Withings WBS01 Smart Body Scale K121971).The ARIA WiFi Smart Scale is listed as substantially equivalent to the Withings Smart Body Scale (K121971) across all listed features. Differences noted (e.g., age range, specific power source type, minor IP connectivity details) are presented as not impacting substantial equivalence.
    Body Fat MeasurementBody fat composition (%) measurements should not be statistically different (p>0.05) from the predicate device, and body fat measurements should vary by < 8% from one another when compared to the predicate device."Results of this study lead to the conclusion that the measurements from ARIA were not statistically different from the predicate device (p>0.05) and body fat measurements varied by <8% from one another."
    Safety and EMCCompliance with IEC 60601-1 (medical electrical equipment safety) and IEC 60601-1-2 (electromagnetic compatibility)."The ARIA WiFi Smart Scale has been tested according to IEC 60601-1, IEC 60601-1-2 and was found to meet all requirements."
    Reliability & Human FactorsMeet specified criteria as per internal testing."Performance data (reliability testing and human factors testing) also support that the ARIA device meet its specified criteria." (Specific criteria not detailed in the summary).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 50 subjects (25 male and 25 female).
    • Data Provenance: The study was a "small comparative clinical study" comparing the ARIA WiFi Smart Scale to the predicate device. The text does not specify the country of origin but implies it was conducted by the manufacturer or a contracted clinical research organization. The study design strongly suggests it was prospective as it involved collecting new data for direct comparison between the two devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The study does not establish an independent "ground truth" against which the device performance is measured in the classical sense (e.g., DEXA or underwater weighing). Instead, the performance of the ARIA device is compared directly against the predicate device (Withings Smart Body Scale K121971) as the reference. Therefore, there were no experts used to establish a separate ground truth for the test set. The predicate device's measurements serve as the comparator.


    4. Adjudication Method for the Test Set

    Not applicable. This was a direct comparison study between a new device and a predicate device. There was no complex labeling or interpretation by multiple human readers requiring adjudication. The study involved objective measurements of body fat percentage from both devices.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This device (Fitbit ARIA WiFi Smart Scale) is a standalone measurement device for body weight and body fat, not an AI-assisted diagnostic tool that aids human readers. The study performed was a direct comparison of its measurement accuracy against a predicate device.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done. The clinical performance testing compared the ARIA WiFi Smart Scale's measurements of body fat composition directly against the measurements from the predicate device. This is a standalone comparison as it assesses the device's output independently. The device's primary function is to measure and display these values, which are then transmitted to a user's account for tracking, but the core performance evaluation focuses on the accuracy of these direct measurements.


    7. The Type of Ground Truth Used

    The "ground truth" for the clinical performance study was the measurements obtained from the predicate device (Withings Smart Body Scale K121971). The study aimed to demonstrate that the ARIA device's measurements were not statistically different from those of the legally marketed predicate device. While BIA is an estimation method itself, for the purpose of demonstrating substantial equivalence, the predicate device's output serves as the comparative reference.


    8. The Sample Size for the Training Set

    The document does not mention a training set or any machine learning/AI model that would require a distinct training set. The ARIA WiFi Smart Scale uses Bioelectrical Impedance Analysis (BIA) technology, which is a well-established method, not typically relying on a separately described "training set" in the context of regulatory submissions for this type of device. The description focuses on its sensor technology and comparison to a predicate.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned or implied for a machine learning model, this question is not applicable. The device relies on physical principles of bioelectrical impedance.