Search Results
Found 8 results
510(k) Data Aggregation
(143 days)
Argos Infinity is intended for use in adult patients (18 years and older). It is designed to provide clinicians with continuous hemodynamic monitoring, including cardiac output and derived parameters, for patients who receive radial arterial blood pressure catheters.
The Argos Infinity software is used for the continuous measurement of cardiac output from an intravascular radial arterial blood pressure signal. This signal is derived from a blood pressure signal transmitted over a computer network from a vital signs monitor. The device is intended to be used by clinicians at the bedside or in remote settings to assess advanced hemodynamics of patients in an operating room or in an intensive care unit.
The proposed "Argos Infinity," (AI) is a "Software as a Medical Device" (SaMD) designed to calculate the hemodynamic status of a patient from their invasive blood pressure signal. The blood pressure signal is obtained from a digitized waveform source such as a data aggregator used for central waveform viewing in an intensive care unit or an operating room, or a monitoring device. The SaMD has an application programming interface that allows a digital waveform source to send 20 seconds of waveform data and obtain calculated hemodynamic data results for that data segment including cardiac output (CO), stroke volume (SV), systemic vascular resistance (SVR) and their respective indices (normalized to body surface area). In addition, pulse pressure variation is also calculated.
These results are pulled from the SaMD by a third party client application and appended to the patient's online medical record. The SaMD is to be deployed on a network server within the hospital's IT department that is accessible to a client application that can access the same hospital network.
Argos Infinity was clinically validated using bedside patient monitors configured to transmit invasive arterial blood pressure data via a data aggregation system, consistent with the Argos Infinity data acquisition configuration. The monitors included in clinical validation were Philips IntelliVue® patient monitors and GE CARESCAPE® patient monitors, each with invasive blood pressure monitoring capability. These monitors and acquisition configurations represent the complete set of systems included in the Argos Infinity clinical validation testing. Compatibility claims are limited to the validated configurations described above.
The Argos Infinity algorithm is identical to the algorithm implemented in the previously cleared Argos device, with no changes to the signal processing, calculation methods, or performance characteristics. The differences between Argos Infinity and the predicate device are limited to the intended operating environment and deployment model. Argos Infinity is designed for use within a hospital network and supports remote and centralized (command center) monitoring in addition to bedside use, without altering the underlying algorithm or its clinical performance.
N/A
Ask a specific question about this device
(27 days)
ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Clamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool.
It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.
The ARGOS is substantially equivalent to the predicate device identified previously:
- the ARGOS (Santec Corporation) that was cleared by the FDA on October 2nd, 2015(K150754) as the primary predicate as it has the most similar intended use and characteristics
- the IOLMaster 700 (Carl Zeiss Meditec) that was cleared by the FDA on June 29, ● 2015 (K143275, K170171) as an additional (secondary) predicate to support a new feature of reference image capture on the version of ARGOS in this submission.
The predicate devices are Class 2 devices to premarket notification, as defined per regulation number 21 CFR 886.1850. In addition, the predicate devices have product codes of MXK(ARGOS), and HJO(IOLMaster700).
The version of ARGOS in this submission is a modified version of the Argos cleared under K150754 which is substantially equivalent with regard to intended use, operating principle, function, materials, and energy source. The differences from the predicate ARGOS (K150754) that are subject of this 510(k) submission are:
- An additional feature of reference image capture function ●
- Labeling change including change in the intended use adding the feature of reference . image capture.
- . This image can be transferred to image guided devices in order to support the execution of preoperative plan.
The changes described in this submission do not affect how the hardware is used to acquire measurements as a biometer, nor do these changes affect the principle of operation of the device.
The provided document describes the ARGOS device, a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT), and its 510(k) submission for clearance. The document focuses on demonstrating substantial equivalence to predicate devices, particularly regarding the ARGOS ver1.5 which includes a new "Reference Image functionality."
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria and reported device performance" in terms of specific quantitative thresholds that the new ARGOS ver1.5 needs to meet for its added functionality. Instead, it argues for substantial equivalence primarily by:
- Comparing technological characteristics of the new ARGOS (ver1.5) with its primary predicate (ARGOS K150754) and a secondary predicate (IOLMaster 700 K143275, K170171).
- Stating that the new feature (reference image capture) is similar to a feature already present in a legally marketed predicate (IOLMaster 700).
- Referring to compliance with recognized consensus standards for performance and safety.
- Highlighting that the core biometry measurement capabilities did not change from the primary predicate. The in-vivo repeatability specifications for measured parameters (Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, Keratometry, Astigmatism, Pupil Size, White-to-White) are listed as identical to the primary predicate, implying these performance metrics were already accepted.
Implied Acceptance Criteria and Reported Performance (from comparison tables):
| Parameter / Characteristic | Acceptance Criteria (Implied - by being equivalent to Predicate ARGOS) | Reported Device Performance (ARGOS ver1.5) |
|---|---|---|
| Biometry Measurements | Identical to Predicate ARGOS (K150754) | Identical to Predicate ARGOS (K150754) |
| Axial Length Repeatability | SD 0.01mm (range 14-38mm) | SD 0.01mm (range 14-38mm) |
| Corneal Thickness Repeatability | SD 10um (range 200-120um) | SD 10um (range 200-120um) |
| Anterior Chamber Depth Repeatability | SD 0.01mm (range 0.7-8.0mm) | SD 0.01mm (range 0.7-8.0mm) |
| Lens Thickness Repeatability | SD 0.02mm (range 0.5-10.0mm) | SD 0.02mm (range 0.5-10.0mm) |
| Keratometry Repeatability | SD 0.02mm (range 5.5-10.0mm) | SD 0.02mm (range 5.5-10.0mm) |
| Astigmatism Repeatability | SD 5deg (Cylinder>1D) (range 0-180deg) | SD 5deg (Cylinder>1D) (range 0-180deg) |
| Pupil Size Repeatability | SD 0.09mm (range 2-13mm) | SD 0.09mm (range 2-13mm) |
| White-to-White Repeatability | SD 0.06mm (range 7-15mm) | SD 0.06mm (range 7-15mm) |
| Reference Image Functionality | Similar to IOLMaster 700 predicate | Implemented using white LED illumination |
| Electrical Safety | Compliance with ANSI/AAMI ES 60601-1, IEC 60601-1-2 | Compliant |
| Housing Protection | Compliance with IEC 60529 (IP20) | Compliant (IP20) |
| Light Hazard | Compliance with ANSI.Z80-36 (Group 2), IEC 62471 (Exempt group) | Compliant (Group 2 for SS-OCT, Exempt for LED illumination) |
| Laser Product Safety | Compliance with IEC 60825-1 (Laser Class 1) | Compliant (Laser Class 1) |
| Ophthalmometer Accuracy | Compliance with ISO 10343 | Compliant |
| Hardware Performance | Meet requirements for axial/lateral dist., SNR, depth attenuation | Met |
| Software Evaluation | Compliance with FDA guidance, IEC 62366, ISO 14971, cybersecurity guidance | Verified and Validated |
| Usability | Comparable to predicate Argos | Evaluated and found comparable |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size for any clinical test set in terms of patient data. It primarily relies on bench testing and software verification/validation. The comparison of in-vivo repeatability values references the existing performance of the predicate device, not new clinical data for the ARGOS ver1.5 to prove equivalent clinical performance.
- Sample Size for Test Set: Not explicitly stated for patient data. The "Performance Testing" section refers to "bench tests" and "software verification and validation," which typically involve internal testing and simulations rather than patient samples for the new features. The existing in-vivo repeatability data cited appears to be from the primary predicate, not new testing on the modified device regarding its core measurement functions.
- Data Provenance: The document does not provide details on the country of origin of data or whether it was retrospective or prospective. Given the focus on substantial equivalence through design and testing against standards rather than new clinical trials for the added feature, such details are not expected to be prominent.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
N/A. The document does not describe a clinical study involving human readers or expert-established ground truth for a test set, especially pertaining to the new "Reference Image functionality." The changes are assessed through engineering and software validation, and comparison to existing, already-cleared devices.
4. Adjudication Method for the Test Set
N/A. No clinical study with human readers and adjudication is described.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
N/A. No MRMC study is described. The device is not an AI/CADe device that assists human readers in diagnosis. It's a measurement device with an added image capture utility.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device itself is a standalone measurement instrument. Its "performance" is its ability to accurately measure ocular parameters and capture images. The "Performance Testing" section (12.1 and 12.2) effectively describes this standalone performance evaluation, focusing on:
- Bench tests: ISO standards, hardware specifications (axial/lateral distance, SNR, depth attenuation).
- Software verification and validation: According to FDA guidance and ISO standards.
- Cybersecurity review.
These tests demonstrate the device's capability to function as intended without human intervention for the measurement process itself, or for the image capture function. The usability evaluation ensures the human-device interaction is acceptable.
7. The Type of Ground Truth Used
- For biometric measurements: The "ground truth" for proving performance (cited from the predicate) would typically be established based on highly accurate reference devices or physical models, although the document doesn't detail the predicate's original ground truth methodology. For the new device, the claim is that its measurement capabilities have not changed and are thus equivalent to the predicate.
- For reference image functionality: The ground truth for this new feature would be its ability to capture and transfer an image. This is validated by functional testing and comparison to the secondary predicate (IOLMaster 700), which already has this function. There isn't a "ground truth" in the clinical diagnostic sense for an image capture tool beyond its intended function of capturing an image.
- For safety and performance standards: The "ground truth" is compliance with the specifications and limits defined by the international and national consensus standards (e.g., ISO, IEC, ANSI/AAMI).
8. The Sample Size for the Training Set
N/A. This is not an AI/ML device that requires a training set in the typical sense of machine learning. It's a measurement instrument. The "Vision Planner software package" and "Argos UI software" are traditional software applications that undergo verification and validation, not model training.
9. How the Ground Truth for the Training Set Was Established
N/A. No training set is involved.
Ask a specific question about this device
(204 days)
The Argos Cardiac Output monitoring device is intended for use on patients above the age of 18. It is intended to be used as a hemodynamic monitor for monitoring cardiac output and its derived parameters on patients in the intensive care unit or the operating room.
The Argos Monitor is a portable hemodynamic monitor that calculates Cardiac Output and other derived parameters, including cardiac index (CI), stroke volume (SV), stroke volume index (SVI), systemic vascular resistance (SVR), systemic vascular resistance index (SVRI), mean arterial pressure (MAP), heart rate (HR), and pulse pressure variation (PPV) based on a proprietary algorithm that analyzes the blood pressure waveform and user-entered patient demographic information (age, height weight and gender). The blood pressure waveform is input into the monitor via a connection with either a radial arterial catheter or the analog blood pressure signal output of a vital signs monitor. The scientific method that underlies the algorithm is based on a novel signal processing technique to determine the parameters of the well-established Windkessel model of the circulation in order to calculate cardiac output.
The Argos Monitor comes with a touchscreen monitor and computer system enclosed in a rigid plastic housing and a power cable. Cables to connect the monitor to a radial blood pressure transducer or to the analog blood pressure signal output of a vital signs monitor are also provided according to the setup and needs of the individual institution. The Monitor may be attached to a pole or a table stand via a standard screw interface pattern.
The requested information about the acceptance criteria and the study proving the device meets them is extracted from the provided text.
Based on the provided text, the device in question is the Retia Medical, LLC: Argos Monitor, which calculates Cardiac Output and other derived parameters.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of numerical acceptance criteria for specific performance metrics (e.g., accuracy thresholds for Cardiac Output). Instead, the acceptance is framed in terms of equivalence to a predicate device and meeting safety and performance standards.
| Acceptance Criteria Type | Acceptance Criteria (Implicit/Explicit) | Reported Device Performance |
|---|---|---|
| Clinical Performance | The device should demonstrate accuracy in calculating cardiac output comparable to or better than the predicate device when compared to a reference standard (pulmonary artery catheter). The document states the Argos algorithm should have "as low or lower errors than the predicate Edwards device" and "perform as well as or better than the predicate device." | "The accuracies of the Argos monitor and the predicate device were assessed using the pulmonary artery catheter as the reference for all measurements. The data demonstrated that the Argos monitor has the same or lower errors in measurement compared to the reference than the errors shown by the predicate device compared to the reference." |
| Safety and Effectiveness | The device should be as safe and as effective as the predicate device for its intended use. | "The Argos monitor passed all verification and validation testing and was shown to be safe, effective and substantially equivalent to the predicate device." And "The Retia Argos monitor has been shown to be safe and effective as the predicate." |
| Functional and Performance Testing | The device must pass a comprehensive set of engineering, software, and usability tests. This includes IEC 60601-1, IEC 60601-1-2, IEC 60601-2-34, IEC 60601-1-8, ISTA 2A, and ISO 10993-1 standards requirements. | "The Argos CO Monitor has successfully passed functional and performance testing, including electrical and mechanical testing, environmental testing, shipping tests, software verification and validation, clinical usability testing, and comparison testing with the predicate device on clinical data." And "The monitor was assessed and/or tested for and met the following standard requirements: 1. IEC 60601-1, 2. IEC 60601-1-2, 3. IEC 60601-2-34, 4. IEC 60601-1-8, 5. ISTA 2A, 6. ISO 10993-1." |
| Usability | Clinical users (physicians and nurses) should be able to successfully set up, connect, configure, and interpret information from the monitor. | "Clinical usability testing for the monitor was performed on 15 clinical users, comprising physicians and nurses who work in the ICU and OR. All users were successfully able to set up the monitor, connect the appropriate cables, enter the patient demographic information, configure the displayed parameters, and interpret the information provided, including alarms." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Clinical Performance Test Set: 40 patients (20 from the Operating Room (OR) and 20 from the Intensive Care Unit (ICU)).
- Data Provenance: The document does not explicitly state the country of origin. It describes the study as "clinical testing" and "clinical data," implying a prospective collection of data in a clinical setting for the purpose of this study. The timeframe is not specified beyond "No adverse effects or complications were noted during the study."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not specify the number of experts or their qualifications for establishing the ground truth. The ground truth (reference standard) used was an invasive method: "the pulmonary artery catheter as the reference for all measurements." This is a direct measurement, and thus expert interpretation for ground truth establishment might not be applicable in the same way it would be for an image-based diagnostic study requiring consensus reads.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method for the test set, as the ground truth was established by direct measurement via a pulmonary artery catheter, not by expert consensus. The comparison was between the device's output, the predicate device's output, and the pulmonary artery catheter's output.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC study was not conducted. This study focused on device performance against a reference standard and comparison to a predicate device, not on the improvement of human reader performance with AI assistance. The device is a "Single-Function, Preprogrammed Diagnostic Computer" that measures physiological parameters.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
Yes, a standalone performance assessment was conducted implicitly. The "accuracies of the Argos monitor and the predicate device were assessed using the pulmonary artery catheter as the reference." This directly measures the algorithm's output against the ground truth without human intervention in the loop of the measurement process. Human users were involved in setting up and interpreting the device, but the accuracy assessment was of the device's calculated parameters.
7. The Type of Ground Truth Used
The ground truth used for the clinical performance assessment was outcomes data in the form of direct measurements from a pulmonary artery catheter (PAC). The PAC is considered a gold standard for cardiac output measurement.
8. The Sample Size for the Training Set
The document does not specify the sample size for the training set. It only mentions that the algorithm is "proprietary" and based on a "novel signal processing technique."
9. How the Ground Truth for the Training Set was Established
The document does not describe how the ground truth for the training set was established. It focuses on the validation of the device's performance against the predicate and a reference standard.
Ask a specific question about this device
(193 days)
ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraccular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White (corneal diameter) and Pupil Size. It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.
Argos is a swept-source based biometer that provides the biometry and keratometry of the eye prior to cataract surgery and aids in the selection of the appropriate IOL.
Argos processes 3 OCT images (B-scans) and 3 CMOS camera images per measurement in seconds - reducing procedure time and minimizing patient discomfort. From the OCT images the biometry parameters are evaluated by segmenting the cornea, iris, lens, and retina. The operator can edit the biometry parameters by manual adjustment: Axial length, Corneal thickness, Anterior chamber depth (or aqueous depth), Lens thickness, White-towhite (corneal diameter), Pupil size. The keratometry values: K-values (flattest and steepest meridians), Astigmatism (angle of flattest meridian), are evaluated from the CMOS camera image, in combination with the OCT information.
All distance/thickness parameters (Axial length, Corneal thickness, Anterior chamber depth, Lens thickness, White, Pupil size) are simultaneously measured from 2dimensional OCT images.
All the boundaries of ocular segments are detected and distances and thicknesses are calculated by taking into account the refractive indices of each medium.
Axial length is the distance from the corneal apex to the fovea or, more specifically, to the ILM (Internal Limiting Membrane). The calculation is performed as the sum of the thicknesses (after refraction correction) of the cornea, aqueous humor, lens and vitreous.
Corneal thickness is the distance between the anterior and the posterior apexes of the cornea divided by the cornea refractive index (1.375).
Aqueous depth is evaluated as the distance between the posterior surface of the cornea and the anterior surface of the lens divided by the refractive index (1.336).
Anterior chamber depth is the sum of corneal and aqueous humor distances.
Lens thickness is the distance between the anterior and the posterior surfaces of the lens divided by its refractive index (1.410).
Pupil size is the lateral distance between the two inner boundaries of the iris region.
White-to-white (Corneal diameter) is the lateral distance between the inner boundaries of the cornea-sclera interface.
K-values (Radii of flattest and steepest meridians) and Astigmatism are evaluated by the size and distance of the reflected images of the infrared LED ring projected onto the cornea.
Safety protocol is well integrated into the device in both hardware and software to ensure the safety of both the patient and the operator.
The control program on the computer runs on Windows. The operation panel on the display attached to the PC provides all the functionalities.
IOL calculation uses widely recognized formulas; Hoffer Q, Haigis, Holladay1, SRK/T, etc.
The provided document describes the ARGOS device, an optical biometer, and its performance evaluation. Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal "acceptance criteria" for the clinical studies in the format of a predefined threshold. Instead, it presents the device's performance in relation to a predicate device (Lenstar LS900) through various statistical measures, implying that performance comparable to the predicate or within acceptable clinical ranges constitutes "acceptance." The "repeatability" values can also be considered a form of performance criteria.
| Parameter | Acceptance Criteria (Implied / Comparator Performance for Agreement) | Reported Device Performance (ARGOS) |
|---|---|---|
| Agreement Study (vs. Lenstar LS900) | ||
| Axial Length (AL) | Similar mean, small mean difference (0.00 mm), small SD (0.05 mm), LoA of [-0.10, 0.09] mm, Deming slope close to 1 (1.03), high Spearman correlation (1.00) compared to predicate (Lenstar: Mean 23.43 (1.01), SD N/A, Spearman 1.00 for comparison, Imp. Repeatability SD: 0.035mm). | Mean: 23.44 (0.98), Mean Difference (Argos-Lenstar): 0.00 (0.05) mm, LoA: [-0.10, 0.09] mm, Deming Regression: Intercept -0.68 [-0.93,-0.43], Slope 1.03 [1.02,1.04], Spearman Correlation: 1.00. Repeatability (Precision Study): SD 0.01 mm, CV 0.02%. Reproducibility: SD 0.02 mm, CV 0.09%. |
| Corneal Thickness (CCT) | Similar mean, small mean difference (0.00 mm), small SD (0.01 mm), LoA of [-0.02, 0.02] mm, Deming slope close to 1 (1.04), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 0.53 (0.04), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: 2.3um (0.0023mm)). | Mean: 0.53 (0.03), Mean Difference (Argos-Lenstar): 0.00 (0.01) mm, LoA: [-0.02, 0.02] mm, Deming Regression: Intercept -0.02 [-0.05,0.01], Slope 1.04 [0.99,1.10], Spearman Correlation: 0.97. Repeatability (Precision Study): SD 0.00 mm, CV 0.89%. Reproducibility: SD 0.01 mm, CV 1.19%. |
| Aqueous Depth (AD) | Similar mean, small mean difference (0.05 mm), small SD (0.16 mm), LoA of [-0.27, 0.37] mm, Deming slope close to 1 (1.06), high Spearman correlation (0.89) compared to predicate (Lenstar: Mean 2.66 (0.41), SD N/A, Spearman 0.89 for comparison, Imp. Repeatability SD: 0.04mm). | Mean: 2.63 (0.39), Mean Difference (Argos-Lenstar): 0.05 (0.16) mm, LoA: [-0.27, 0.37] mm, Deming Regression: Intercept -0.10 [-0.29,0.09], Slope 1.06 [0.98,1.13], Spearman Correlation: 0.89. Repeatability (Precision Study): SD 0.01 mm, CV 0.37%. Reproducibility: SD 0.03 mm, CV 0.90%. |
| Anterior Chamber Depth (ACD) | Similar mean, small mean difference (0.05 mm), small SD (0.17 mm), LoA of [-0.28, 0.37] mm, Deming slope close to 1 (1.07), high Spearman correlation (0.89) compared to predicate (Lenstar: Mean 3.19 (0.42), SD N/A, Spearman 0.89 for comparison, Imp. Repeatability SD: 0.04mm). | Mean: 3.16 (0.39), Mean Difference (Argos-Lenstar): 0.05 (0.17) mm, LoA: [-0.28, 0.37] mm, Deming Regression: Intercept -0.17 [-0.39,0.06], Slope 1.07 [1.00,1.14], Spearman Correlation: 0.89. Repeatability (Precision Study): SD 0.01 mm, CV 0.27%. Reproducibility: SD 0.03 mm, CV 0.77%. |
| Lens Thickness (LT) | Similar mean, small mean difference (-0.18 mm), small SD (0.22 mm), LoA of [-0.61, 0.24] mm, Deming slope close to 1 (1.09), high Spearman correlation (0.87) compared to predicate (Lenstar: Mean 4.46 (0.45), SD N/A, Spearman 0.87 for comparison, Imp. Repeatability SD: 0.08mm). | Mean: 4.62 (0.41), Mean Difference (Argos-Lenstar): -0.18 (0.22) mm, LoA: [-0.61, 0.24] mm, Deming Regression: Intercept -0.58 [-1.10,-0.05], Slope 1.09 [0.97,1.20], Spearman Correlation: 0.87. Repeatability (Precision Study): SD 0.02 mm, CV 0.38%. Reproducibility: SD 0.06 mm, CV 1.45%. |
| Pupil Size (PS) | Similar mean, small mean difference (-0.26 mm), small SD (0.46 mm), LoA of [-1.16, 0.64] mm, Deming slope close to 1 (0.81), high Spearman correlation (0.88) compared to predicate (Lenstar: Mean 4.10 (0.77), SD N/A, Spearman 0.88 for comparison, Imp. Repeatability SD: not specified). Note: Differences attributed to lack of ambient light control. | Mean: 4.38 (0.91), Mean Difference (Argos-Lenstar): -0.26 (0.46) mm, LoA: [-1.16, 0.64] mm, Deming Regression: Intercept 0.55 [-0.02,1.11], Slope 0.81 [0.68,0.95], Spearman Correlation: 0.88. Repeatability (Precision Study): SD 0.09 mm, CV 1.83%. Reproducibility: SD 1.01 mm, CV 20.31%. |
| Corneal Diameter (CD) | Similar mean, small mean difference (-0.28 mm), small SD (0.69 mm), LoA of [-1.62, 1.06] mm, Deming slope close to 1 (0.57 - lower than 1 but still within CI), high Spearman correlation (0.48 - lower correlation acknowledged due to different measurement principle) compared to predicate (Lenstar: Mean 11.86 (0.49), SD N/A, Spearman 0.48 for comparison, Imp. Repeatability SD: 0.04mm). Note: Accuracy not well confirmed due to different measurement principle. | Mean: 12.11 (0.68), Mean Difference (Argos-Lenstar): -0.28 (0.69) mm, LoA: [-1.62, 1.06] mm, Deming Regression: Intercept 4.91 [1.79,8.04], Slope 0.57 [0.31,0.83], Spearman Correlation: 0.48. Repeatability (Precision Study): SD 0.06 mm, CV 0.46%. Reproducibility: SD 0.10 mm, CV 0.77%. |
| Flattest Radius of Curvature (R1) | Similar mean, small mean difference (-0.01 mm), small SD (0.07 mm), LoA of [-0.15, 0.13] mm, Deming slope close to 1 (1.01), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 7.79 (0.30), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: 0.02mm). | Mean: 7.80 (0.29), Mean Difference (Argos-Lenstar): -0.01 (0.07) mm, LoA: [-0.15, 0.13] mm, Deming Regression: Intercept -0.09 [-0.45,0.26], Slope 1.01 [0.96,1.06], Spearman Correlation: 0.97. Repeatability (Precision Study): SD 0.01 mm, CV 0.14%. Reproducibility: SD 0.03 mm, CV 0.34%. |
| Steepest Radius of Curvature (R2) | Similar mean, small mean difference (0.00 mm), small SD (0.07 mm), LoA of [-0.14, 0.14] mm, Deming slope close to 1 (0.97), high Spearman correlation (0.98) compared to predicate (Lenstar: Mean 7.61 (0.27), SD N/A, Spearman 0.98 for comparison, Imp. Repeatability SD: 0.03mm). | Mean: 7.61 (0.28), Mean Difference (Argos-Lenstar): 0.00 (0.07) mm, LoA: [-0.14, 0.14] mm, Deming Regression: Intercept 0.22 [-0.05,0.48], Slope 0.97 [0.94,1.01], Spearman Correlation: 0.98. Repeatability (Precision Study): SD 0.02 mm, CV 0.21%. Reproducibility: SD 0.04 mm, CV 0.48%. |
| Average Radius of Curvature (Rav) | Similar mean, small mean difference (-0.01 mm), small SD (0.06 mm), LoA of [-0.12, 0.11] mm, Deming slope close to 1 (0.99), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 7.70 (0.28), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: N/A). | Mean: 7.70 (0.28), Mean Difference (Argos-Lenstar): -0.01 (0.06) mm, LoA: [-0.12, 0.11] mm, Deming Regression: Intercept 0.06 [-0.17,0.30], Slope 0.99 [0.96,1.02], Spearman Correlation: 0.97. Repeatability (Precision Study): SD 0.01 mm, CV 0.13%. Reproducibility: SD 0.03 mm, CV 0.35%. |
| Astigmatism (Cyl. < 1D) | Similar mean, small mean difference (8.29 degrees), small SD (48.04 degrees), LoA of [-85.87, 102.45] degrees, Deming slope close to 1 (1.11), Spearman correlation (0.56) compared to predicate (Lenstar: Mean 94.07 (45.94), SD N/A, Spearman 0.56 for comparison, Imp. Repeatability SD: 4deg). | Mean: 87.89 (46.85), Mean Difference (Argos-Lenstar): 8.29 (48.04) Degrees, LoA: [-85.87, 102.45] Degrees, Deming Regression: Intercept -1.27 [-24.26,21.71], Slope 1.11 [0.90,1.33], Spearman Correlation: 0.56. Repeatability (Precision Study, All AST): SD 4.39 degrees, CV 5.59%. Reproducibility (All AST): SD 11.13 degrees, CV 14.17%. |
| Astigmatism (Cyl. > 1D) | Similar mean, small mean difference (4.36 degrees), small SD (9.71 degrees), LoA of [-14.67, 23.39] degrees, Deming slope close to 1 (1.05), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 81.44 (46.96), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: 11deg). | Mean: 78.40 (46.34), Mean Difference (Argos-Lenstar): 4.36 (9.71) Degrees, LoA: [-14.67, 23.39] Degrees, Deming Regression: Intercept 0.07 [-6.02,6.15], Slope 1.05 [0.96,1.15], Spearman Correlation: 0.97. Repeatability (Precision Study, All AST): SD 4.39 degrees, CV 5.59%. Reproducibility (All AST): SD 11.13 degrees, CV 14.17%. |
| Enhanced Retinal Visualization (ERV) mode - Cataractous Eyes (vs. Lenstar for AL) | ||
| Axial Length (AL) (ERV) | Statistically insignificant differences (p-value < 0.05) in pairwise t-test, mean difference close to 0 (0.03 mm), small SD (0.04 mm), LoA of [-0.06, 0.12] mm, high Spearman correlation (1.00) compared to predicate. | Mean (Lenstar): 23.46 (1.08). Mean Difference (Lenstar-Argos ERV): 0.03 (0.04) mm, LoA: [-0.06, 0.12] mm, Spearman Correlation: 1.00. Repeatability (ERV mode): SD 0.01 mm, CV 0.06%. |
2. Sample Size Used for the Test Set and Data Provenance
-
Agreement Study (Table 17.2.1):
- Axial Length (AL): 100 eyes
- Corneal Thickness (CCT): 115 eyes
- Aqueous Depth (AD): 115 eyes
- Anterior Chamber Depth (ACD): 115 eyes
- Lens Thickness (LT): 111 eyes
- Pupil Size (PS): 108 eyes
- Corneal Diameter (CD): 115 eyes
- R1, R2, Rav: 115 eyes
- Astigmatism (Cyl. < 1D): 77 eyes
- Astigmatism (Cyl. > 1D): 36 eyes
- Data Provenance: A single U.S. clinical site; collected prospectively for the agreement study.
-
Precision Study (Repeatability and Reproducibility - Table 17.2.2):
- 43 healthy eyes (for both repeatability and reproducibility).
- Data Provenance: Not explicitly stated, but generally part of the same prospective clinical study.
-
Enhanced Retinal Visualization (ERV) mode Validation Study (Table 17.2.3 and 17.2.4):
- Cataractous eyes: 45 eyes (23 right eyes) for comparison with Lenstar.
- Healthy eyes: 43 consecutive healthy eyes for repeatability of ERV mode.
- Data Provenance: Not explicitly stated, but generally part of the same prospective clinical study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish ground truth for the test set beyond the clinical measurements themselves. The study primarily relies on the agreement between the ARGOS device and the predicate device (Lenstar LS900) as the reference for performance, and the precision of the ARGOS itself.
4. Adjudication Method for the Test Set
No adjudication method is described for the test set. The clinical studies compare measurements between devices and evaluate the precision of the ARGOS device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No MRMC comparative effectiveness study involving human readers' improvement with or without AI assistance was done or described. This device is an optical biometer for objective measurements, not an image interpretation AI system for diagnosis where human reader studies are typical.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the studies described are standalone performance evaluations of the ARGOS device. The ARGOS device is an automated measurement tool. The "Agreement Study" and "Precision Study" assess the device's inherent measurement capabilities. While an operator initiates measurements, the performance metrics (Axial Length, CCT, etc.) are generated solely by the device's algorithms and hardware.
7. The Type of Ground Truth Used
The "ground truth" for the clinical studies is implicitly established by:
- Predicate Device Measurements: The Lenstar LS900, an already legally marketed optical biometer, serves as the primary comparator. The assumption is that its measurements are an acceptable reference.
- Internal Consistency/Precision: The repeatability and reproducibility studies establish the device's consistency in its own measurements, which is a form of internal ground truth for precision.
There is no mention of "pathology" or external "outcomes data" being used as ground truth for parameter measurements in this context.
8. The Sample Size for the Training Set
The document does not specify a training set or its sample size. This is typical for medical devices that perform direct physical measurements rather than AI algorithms trained on large datasets for pattern recognition. The algorithms for biometry are based on physical principles (e.g., swept-source OCT) and established calculations, not machine learning model training in the conventional sense.
9. How the Ground Truth for the Training Set Was Established
As no training set is described for an AI model, there is no information on how its ground truth was established. The device relies on physical measurement principles and established formulas (e.g., Hoffer Q, Haigis, Holladay1, SRK/T) for IOL power calculation.
Ask a specific question about this device
(310 days)
Argos is intended as an aid to the surgeon for precisely locating anatomical structures on the human body in either open or percutaneous procedures. It links tracked probes to virtual computer image space on a patient's preoperative or intraoperative image data. The system is indicated for any medical condition in which the use of stereotactic surgery may appropriate and where the reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra can be identified relative to medical images such as CT, MR , fluoroscopic or ultrasound images. Example procedures include but are not limited to: Cranial Procedures: Cranial biopsies Tumor resection Craniotomies/Craniectomies Skull base procedures Thalamotomies/Pallidotomies Spinal Procedures: Spinal implant procedures such as pedicle screw placement ENT procedures: Transsphenoidal procedures Intranasal procedures Sinus procedures, such as Maximillary anstrostomies. Ethmoidectomies Sphenoidotomies/Sphenoid explorations, Turbinate resection and Frontal sinusotomies
Argos is a system for computer-assisted stereotactic surgery. It is designed to be used as an aid to the location of, and surgical approach to, lesions within the anatomical structures on the human body and to help define the exact anatomical relations of such lesions to surrounding structures. Prior to the brain surgery, the Argos system acquires contiguous MRI or CT scans of the area containing the lesion and uses them to construct an accurate three-dimensional representation of the patient's anatomy used initially to facilitate surgical planning. During the operation the Argos system continuously monitors the position of a pointing instrument, used by the surgeon, and visualizes it on a monitor screen in relation to the three-dimensional virtual anatomy of the patient. It is therefore possible to verify during the operation that the real position of various anatomical reference points, including the lesion, corresponding to those identified during surgical planning.
The provided text is a 510(k) Summary of Safety and Effectiveness for the ARGOS image-guided surgery system. It details the device's description, intended use, and a comparison to a predicate device (VECTORVISION). However, it does not contain any information about specific acceptance criteria, studies proving performance, sample sizes, ground truth establishment, expert qualifications, or MRMC studies.
The document focuses on establishing substantial equivalence to a legally marketed predicate device, which is a regulatory pathway for medical devices in the United States. This pathway typically relies on demonstrating that the new device is as safe and effective as a predicate device, rather than requiring extensive clinical trials with detailed performance metrics and statistical analyses.
Therefore, I cannot provide a table of acceptance criteria and reported device performance or answer most of your detailed questions based on the provided text.
Here's an overview of what can be extracted from the document:
1. Table of Acceptance Criteria and Reported Device Performance:
- Information not available in the provided text. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting specific performance metrics against defined acceptance criteria.
2. Sample size used for the test set and the data provenance:
- Information not available in the provided text. There is no mention of a test set or data provenance for performance evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Information not available in the provided text. There is no discussion of expert involvement in establishing ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Information not available in the provided text. No adjudication method is mentioned as a test set is not described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Information not available in the provided text. The ARGOS device is an image-guided surgery system, not an AI diagnostic tool primarily involving human readers interpreting images. An MRMC study is not relevant in this context and is not mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The ARGOS system is described as an "aid to the surgeon" that "continuously monitors the position of a pointing instrument, used by the surgeon, and visualizes it on a monitor screen." This clearly indicates a human-in-the-loop system. Therefore, a standalone algorithm-only performance study would not be applicable or described for this type of device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Information not available in the provided text. No ground truth establishment is described. The system utilizes "three-dimensional representation of the patient's anatomy" constructed from "MRI or CT scans," implying that these medical images serve as the basis for anatomical representation.
8. The sample size for the training set:
- Information not available in the provided text. There is no mention of a training set as the document does not describe a machine learning or AI model development process with a distinct training phase.
9. How the ground truth for the training set was established:
- Information not available in the provided text. As there is no mention of a training set, the establishment of its ground truth is also not discussed.
Summary of the Study (as described for regulatory purposes):
The "study" presented here is a Substantial Equivalence determination, which is a regulatory review process rather than a scientific performance study. The core of this determination is a comparison between the ARGOS device and a legally marketed predicate device, BRAINLAB VECTORVISION (K003589).
-
Methodology: The submission compares key characteristics of ARGOS to VECTORVISION, including:
- Indications for Use (found to be similar)
- Device main components (similar components, some differences in physical configuration like camera separability)
- Physical characteristics (differences in dimensions and weight acknowledged, but not considered to affect safety or effectiveness significantly for substantial equivalence)
- Localization technology (both use optical tracking based on infrared cameras and retro-reflective markers for 3D coordinate calculation by triangulation).
-
Conclusion: Based on this comparison, the FDA concluded that ARGOS is "substantially equivalent" to BRAINLAB VECTORVISION, implying it is as safe and effective as the predicate device for its intended use. This conclusion is documented in the FDA's letter (APR 5 - 2007).
Ask a specific question about this device
(69 days)
A. General Indications: The indication for use of the air conduction hearing aids in this submission is to amplify sound for individuals with impaired hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)): Severity: 2. Mild X, 3. Moderate X, 4. Severe X. Configuration: 1. High Frequency - Precipitously Sloping X, 2. Gradually Sloping X, 4. Flat X, 5. Other: steeply sloping X. Other: 1. Low Tolerance to Loudness X.
Programmable OnQue for HS, CCA, and CAMEO Models
The provided document is a 510(k) premarket notification letter from the FDA for a hearing aid device, not a study report. Therefore, it does not contain the detailed information about acceptance criteria or a study proving device performance as requested.
The document primarily focuses on establishing "substantial equivalence" to a predicate device, which is a regulatory pathway for certain medical devices. It confirms that the device, "Programmable OnQue for HS, CCA, and CAMEO Models," is substantially equivalent to devices marketed prior to May 28, 1976.
While it mentions "indications for use" and lists categories of hearing loss the device is intended for, it does not provide acceptance criteria, performance metrics, study details, or data provenance normally found in a clinical study report.
Therefore, I cannot populate the requested tables and information based on the provided text.
Here's a breakdown of why each point cannot be addressed:
- A table of acceptance criteria and the reported device performance: Not present. The document focuses on regulatory equivalence, not performance metrics like accuracy, sensitivity, or specificity.
- Sample size used for the test set and the data provenance: Not present. No test set or clinical study data is mentioned.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not present. No ground truth or expert review is detailed.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not present. No adjudication process is described.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not present. This document predates common AI applications in medical devices and there is no mention of such a study. The device is a hearing aid, not an AI-powered diagnostic tool.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not present. The device is a hearing aid, which works with a human (the wearer). No standalone algorithm performance is discussed.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not present.
- The sample size for the training set: Not present. No training data is mentioned.
- How the ground truth for the training set was established: Not present.
The document is a regulatory approval letter, not a scientific or clinical study report.
Ask a specific question about this device
(56 days)
A. General Indications:
The indication for use of the air conduction hearing aids in this submission is to amplify sound for individuals with impaired hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
Severity:
- Slight
X 2. Mild
X 3. Moderate
X 4. Severe - Profound
Configuration:
X 1. High Frequency - Precipitously Sloping
X 2. Gradually Sloping
3. Reverse Slope
X 4. Flat
X 5. Other Steeply sloping
Other
X 1. Low tolerance To Loudness
2.
3.
B. Specific Indications (Only if appropriate.): (Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.)
Argosy Dynameq Series Models: HS, CCA,Cameo, Passport, Passport Deluxe, Shado, Eclipse, CCAS, CCAS Deluxe, and CCA-Mini Canal
The provided text is a 510(k) clearance letter from the FDA for a series of hearing aids. It does not contain information about acceptance criteria or a study proving the device meets those criteria.
The letter primarily focuses on the device's substantial equivalence to previously marketed devices and general regulatory compliance. It does not include:
- A table of acceptance criteria and reported device performance.
- Details about sample sizes for test sets or data provenance.
- Information on experts used for ground truth or adjudication methods.
- Any mention of multi-reader multi-case (MRMC) comparative effectiveness studies.
- Results from standalone algorithm performance studies.
- Details regarding the type of ground truth used.
- Information on the training set size or how its ground truth was established.
The document only states the general indications for use of the hearing aids, primarily for amplifying sound for individuals with impaired hearing across various severity and configuration categories of hearing loss. It also mentions a general caution about potential electromagnetic interference but does not present any study data related to this or any other performance aspect of the device.
Ask a specific question about this device
(26 days)
Ask a specific question about this device
Page 1 of 1