Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K040489

    Validate with FDA (Live)

    Date Cleared
    2004-04-27

    (61 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The System 100 is indicated for temporary (up to eight hours) ultrafiltration treatment of patients with fluid overload.

    Device Description

    The Aquadex System 100 removes excess fluid from the patient in fluid overload by ultrafiltration of blood across a hollow-fiber hemofilter at the clinician selected rate. The system is comprised on a console mounted on a cart, proprietary software and accessories (venous access catheters, extensions and a blood pump circuit). Patient access is obtained via either peripheral or central venous veins.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Aquadex System 100, focusing on acceptance criteria and the supporting study:

    Aquadex System 100 Adjustable Blood Flow Rate - K04-0489

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly state quantitative acceptance criteria or detailed performance metrics from a study related to "adjustable blood flow rate" for the Aquadex System 100.

    However, based on the Safety & Performance and Conclusion sections, the implicit acceptance criteria are that the device remains safe and effective for its intended use, specifically with the software change to allow for adjustable blood flow rate and compatibility with various catheters.

    Acceptance Criterion (Implicit)Reported Device Performance
    Device safetyDemonstrated through bench testing, applying ISO 14971:2000 principles. Data demonstrated the Aquadex System 100 continues to be safe.
    Device effectivenessDemonstrated through bench testing. Data demonstrated the Aquadex System 100 continues to be effective.
    Software change validationBench testing was performed to validate the software change.
    Catheter compatibilityBench testing generated data in support of the labeling change which incorporates a catheter compatibility chart.
    Substantial EquivalenceThe software revision has been shown to be safe and effective for its intended use, making it substantially equivalent to the predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Bench testing was performed". This typically means testing conducted in a laboratory environment, not on human subjects.

    • Sample Size for Test Set: Not specified. Bench testing usually involves multiple iterations or units, but a specific number is not provided.
    • Data Provenance: The data is generated from bench testing. This implies a controlled laboratory setting. It is not patient or clinical data, and thus, provenance in terms of country of origin or retrospective/prospective is not applicable in the human-clinical sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. Bench testing typically relies on engineering specifications and validated test methods rather than expert consensus on a "ground truth" derived from human interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where human readers interpret data (e.g., medical images) and their agreements/disagreements need to be resolved to establish ground truth. Bench testing does not involve such human interpretation for its primary results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. The study described is bench testing, not a clinical study involving human readers or cases. Therefore, no effect size for human readers improving with AI vs. without AI assistance can be reported.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The device is a medical system (Aquadex System 100) with proprietary software, not an AI algorithm in the typical sense of standalone performance analysis. The "software change" was validated through bench testing, which evaluates the system's (including its software's) operational performance directly. Therefore, while the software's function was tested, it's not a "standalone performance" study of a diagnostic AI algorithm in the way that term is usually applied.

    7. Type of Ground Truth Used

    The ground truth for the bench testing would be based on engineering specifications, established physical principles, and measurement standards. For example, if testing blood flow rate, the ground truth would be the accurately measured flow rate using calibrated equipment, compared to the device's reported flow rate under various conditions. For catheter compatibility, the "truth" would be successful and safe operation with different catheters as defined by design specifications. It is not expert consensus, pathology, or outcomes data.

    8. Sample Size for the Training Set

    This information is not applicable/not provided. The "software change" was validated through bench testing. This typically involves software verification and validation activities against requirements, rather than an AI model "training set."

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable/not provided for the same reasons as point 8. The concept of a "training set" and establishing ground truth for it is relevant to machine learning/AI models, which is not the nature of the described software validation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1