510(k) Data Aggregation

    K Number
    K141220

    Validate with FDA (Live)

    Date Cleared
    2014-05-22

    (10 days)

    Product Code
    Regulation Number
    862.2570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument with the SDS Software is a real-time nucleic acid amplification and five color fluorescence detection system for use with FDA cleared or approved tests on human-derived specimens. The 7500 Fast Dx Real-Time PCR instrument and SDS Software are intended for use in combination with in vitro diagnostic tests labeled for use on this instrument. The 7500 Fast Dx instrument is intended for use by laboratory professionals trained in laboratory techniques, procedures, and on use of the system.

    Device Description

    The Applied Biosystems® 7500 Fast Dx Real-Time PCR instrument integrates a thermal cycler, a fluorimeter and application specific software. The instrument houses the thermal cycler and the fluorimeter, while the application software is run on a PC that is attached to the instrument. Samples are placed in a tube strip or 96-well low-head space plate that is moved to a Peltier-based thermal block and positioned relative to the optics using a tray loading mechanism. Excitation for all samples is provided by a halogen tungsten white source that passes through 5 switchable excitation filters prior to reaching the sample. Fluorescence emission is then detected through 5 color emissions filter wheel to a charge coupled device (CCD) camera. The instrument is designed to complete quantitative RT-PCR runs in about 40 minutes. The Sequence Detection Software (SDS) is used for instrument control, data collection and data analysis. The software provides a wizard for user-friendly set-up. The software measures cycleby-cycle real-time signals from the sample and provides a variety of tools to help the user analyze the data extracted from the samples. In addition, the software provides lamp-life monitoring and other instrument maintenance information. The software runs as an application on the Windows 7 platform.

    AI/ML Overview

    This document is a 510(k) summary for the Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software. It describes a Class II medical device used for real-time nucleic acid amplification and fluorescence detection.

    Here's an analysis of the provided information regarding acceptance criteria and the study:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state acceptance criteria or report specific device performance metrics (e.g., sensitivity, specificity, accuracy) in a quantitative manner. Instead, the focus of this 510(k) is on demonstrating substantial equivalence to a predicate device due to minor modifications.

    The summary states: "Based on the risk analysis evaluation results, verification testing was conducted to support the modifications in the instrument computer operating system. The verification testing report included in the submission supports substantial equivalence to the predicate device."

    This implies that the "acceptance criteria" are implicitly met by demonstrating that the updated device performs equivalently to the predicate, particularly in light of the operating system change. The performance of the predicate device (K082562) would have been established previously.

    A conceptual table based on the document's content would look something like this:

    Acceptance Criteria (Implied)Reported Device Performance
    Functional equivalence to predicate device (K082562)Verification testing supports substantial equivalence to the predicate device, especially regarding the operating system modification.
    No adverse impact on safety and effectiveness due to minor changes"The changes are not substantive changes in the use of the device and do not affect the safety and effectiveness of the device when used as labeled."
    Maintains real-time PCR functionality and five-color detectionDevice continues to operate as a "real-time nucleic acid amplification and five color fluorescence detection system."
    Compatibility with FDA cleared/approved testsIntended for "use with FDA cleared or approved tests on human-derived specimens" and "in combination with in vitro diagnostic tests labeled for use on this instrument."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "verification testing was conducted to support the modifications in the instrument computer operating system." However, it does not provide any specific sample sizes for this testing. It also does not specify the data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not describe the use of experts to establish ground truth for this specific verification testing. Given that the changes were primarily to the operating system and the focus is on substantial equivalence in functionality, this type of expert review is unlikely to have been a primary component of this particular submission.

    4. Adjudication Method for the Test Set:

    No information is provided regarding an adjudication method. This is not typically relevant for a submission focused on demonstrating functional equivalence of an instrument's operating system update.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not conducted or described in this document. This type of study is more common for diagnostic algorithms or imaging devices where human interpretation is a key component and the AI aims to assist or replace it. This document concerns a PCR instrument and its software, not an AI-assisted diagnostic interpretation tool.

    6. If a Standalone Performance Study was done:

    The document refers to "verification testing." While this testing would have evaluated the device's standalone performance after the operating system upgrade, it is presented in the context of demonstrating substantial equivalence to the predicate, rather than a de novo standalone performance study providing new, extensive performance metrics like sensitivity and specificity. The document states: "The verification testing report included in the submission supports substantial equivalence to the predicate device."

    7. The Type of Ground Truth Used:

    The document does not explicitly state the type of ground truth used for the verification testing. Given that it's a real-time PCR instrument, the ground truth for any performance validation would typically involve:

    • Reference standards/materials: Well-characterized samples with known concentrations of nucleic acid targets.
    • Comparison to a reference method: Performance compared against a gold standard or extensively validated method.
    • Known positive/negative controls: Established biological or synthetic samples with known outcomes.

    The testing would primarily confirm that the upgraded system continues to accurately detect and quantify nucleic acids as expected, without being negatively affected by the new operating system.

    8. The Sample Size for the Training Set:

    This device is an instrument and its software, not a machine learning model that requires a "training set" in the typical sense of AI development. Therefore, there is no concept of a "training set" as described for AI studies. The software handles instrument control, data collection, and data analysis using established algorithms, not by learning from a large dataset.

    9. How the Ground Truth for the Training Set was Established:

    As there is no "training set" in the context of AI for this device, this question is not applicable. The software's functionality is based on pre-programmed algorithms for signal processing and quantification inherent to real-time PCR, not on learning from labelled data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082562

    Validate with FDA (Live)

    Date Cleared
    2008-09-30

    (26 days)

    Product Code
    Regulation Number
    862.2570
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument with the SDS Software version 1.4 is a real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed RNA and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes. The 7500 Fast Dx is to be used only by technologists trained in laboratory techniques, procedures and on use of the analyzer.

    Device Description

    The AB 7500 Fast Dx RT-PCR instrument integrates a thermal cycler, a fluorimeter and application specific software. The instrument houses the thermal cycler and the fluorimeter, while the application software is run on a PC that is attached to the instrument. Samples are placed in a tube strip or 96-well low-head space plate that is moved to a Peltier-based thermal block and positioned relative to the optics using a tray loading mechanism. Excitation for all samples is provided by a halogen tungsten white source that passes through 5 switchable excitation filters prior to reaching the sample. Fluorescence emission is then detected through 5 color emissions filter wheel to a charge coupled device (CCD) camera. The instrument is designed to complete quantitative RT-PCR runs in about 40 minutes. The Sequence Detection Software (SDS) version 1.4 for the 7500 Fast Dx Instrument is used for instrument control, data collection and data analysis. The software can measure cycle-by-cycle real-time signals from the sample. The software provides a variety of tools to help the user analyze the data extracted from the samples. The software also provides lamp-life monitoring and other instrument maintenance information. The software runs as an application on Windows XP platform. Changes to the Dx software are subject to change control in accordance with 21 CFR Part 820.40.

    AI/ML Overview

    This 510(k) summary (K082562) for the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument does not contain the detailed information required to describe the acceptance criteria and the specific study that proves the device meets those criteria.

    The document primarily focuses on:

    • Device Identification: Name, classification, submitter information.
    • Predicate Device: Identifies the Affymetrix GeneChip Microarray instrument System as the predicate.
    • Device Description: Explains the physical components and basic functionality of the instrument and its software.
    • Intended Use/Indications for Use: States what the device is designed to do.
    • FDA Correspondence: The FDA letter confirming substantial equivalence.

    Crucially, it lacks the following information that would be necessary to answer the prompt directly:

    • Specific performance metrics: No mention of sensitivity, specificity, accuracy, precision, limit of detection, or any other quantitative measure.
    • Defined acceptance criteria: There are no thresholds or targets stated for any performance metric.
    • Study design details: No description of any clinical or analytical study, including sample sizes, data collection methods, ground truth establishment, or expert involvement.
    • Results of a study: No reported device performance values are provided against any criteria.

    Therefore, I cannot populate the requested table or answer most of the specific questions based on the provided text.

    Here's a breakdown of what cannot be extracted from the document:


    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance CriteriaReported Device Performance
    Not specifiedNot specifiedNot specified

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not specified. The document does not describe any test set or study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not specified. No information on ground truth establishment or expert involvement.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not specified. No test set described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is an RT-PCR instrument, not an AI-assisted diagnostic tool that would involve human readers interpreting images. The document does not mention any MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This device is an instrument with software for data analysis, not an "algorithm" in the sense of an AI model. The document does not provide details on standalone performance testing in this context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not specified. No ground truth information provided.

    8. The sample size for the training set:

    • Not applicable/Not specified. This is an instrument and software, not a machine learning model that would have a separate "training set" in the common sense. If it refers to data used to develop the algorithms for the SDS software, that information is not provided.

    9. How the ground truth for the training set was established:

    • Not applicable/Not specified. See point 8.

    Conclusion:

    The provided 510(k) summary is a high-level document intended to demonstrate substantial equivalence, not a detailed technical report of performance studies. To obtain the requested information, one would typically need to consult a more detailed submissions, such as an FDA review memo, or the full 510(k) submission, which would include the performance data and methodology.