    Intended Use

    The intended use of APC™ Flash-Free Adhesive is as a light cure orthodontic adhesive that is designed to be used in bonding orthodontic appliances for orthodontic treatment.

    APC™ Flash-Free Adhesive is indicated for use in bonding orthodontic appliances for orthodontic treatment.

    Device Description

    It consists of a resin-saturated mat that is attached to the apprimated for Craceets, ceramic brackets, and bondable buccal tubes. The relatively low viscosity of the resin allows it to form a fillet at the edges of the bracket which reduces the need to remove excess adhesive, also known as flash.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are described below:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Bond Strength TestProvide the minimum bond strength to hold a bracket to a tooth (comparable to predicate devices)APC™ Flash-Free Adhesive performs comparably to the predicate devices (Adhesive Precoated Brackets and APC™ Plus Adhesive) and provides the minimum bond strength to hold a bracket to a tooth.
    Primer Compatibility TestProvide the minimum bond strength to hold a bracket to a tooth when used with specified primers (comparable to predicate device)APC™ Flash-Free Adhesive performs comparably to APC™ Plus Adhesive and provides the minimum bond strength to hold a bracket to a tooth when used with Transbond MIP Primer and Transbond Plus SEP Primer.
    Accelerated Aging (Thermocycling)Provide the minimum bond strength to hold a bracket to a tooth following specified cycles (comparable to predicate device)APC™ Flash-Free Adhesive performs comparably to APC™ Plus Adhesive and provides the minimum bond strength to hold a bracket to a tooth following a specified number of cycles between hot and cold environments.
    Ambient Light Stability TestProvide the minimum bond strength to hold a bracket to a tooth following ambient light exposure (comparable to predicate device)APC™ Flash-Free Adhesive performs comparably to APC™ Plus Adhesive and provides the minimum bond strength to hold a bracket to a tooth following exposure to ambient light.

    2. Sample Size for Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each nonclinical performance test (bond strength, primer compatibility, accelerated aging, ambient light stability). However, these are bench tests, a common practice in dental material testing. The data provenance is from nonclinical, bench-top testing conducted by 3M Unitek Corporation. The country of origin for the data is not specified but implicitly assumed to be the United States, where the company is based and seeking FDA clearance. The data is prospective for the device being tested, in the sense that the tests were performed specifically to evaluate the performance of APC™ Flash-Free Adhesive.

    3. Number of Experts and Qualifications for Ground Truth

    No human experts were used to establish ground truth for the test set. All testing described is nonclinical bench testing to evaluate physical properties of the adhesive.

    4. Adjudication Method for the Test Set

    No adjudication method was used as the testing involved objective physical measurements (e.g., bond strength) rather than subjective human assessment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done, as this is a physical dental adhesive, not an imaging or diagnostic device requiring human interpretation of cases.

    6. Standalone (Algorithm Only) Performance Study

    No standalone (algorithm only) performance study was done, as this device is a physical product (an adhesive), not a software algorithm.

    7. Type of Ground Truth Used

    The ground truth used for these nonclinical tests was based on objective physical measurements (e.g., bond strength values) and comparison against the performance of legally marketed predicate devices, which are accepted as safe and effective.

    8. Sample Size for the Training Set

    This device is a physical product (adhesive), not an AI/ML algorithm requiring a training set. Therefore, there is no training set sample size.

    9. How Ground Truth for Training Set Was Established

    As there is no training set for a physical adhesive, this question is not applicable.