Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K142983

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-03-04

    (139 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the illiac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt PTA Scoring Balloon Catheter with HydroCross™ Coating is a standard twolumen catheter with a scoring balloon near the distal end of the catheter has a conventional nylon-blend balloon with a scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure. As shown below, the catheter has a segment which is coated with a hydrophilic coating (HydroCross™ Coating).

    AI/ML Overview

    This section describes the acceptance criteria and study proving the device meets said criteria for the "AngioSculpt® PTA Scoring Balloon Catheter with HydroCross™ Coating" based on the provided text.

    Based on the provided document, the device described is a medical device, and the information presented is a 510(k) summary for premarket notification to the FDA, demonstrating substantial equivalence to a predicate device. This document does not describe the acceptance criteria and a study in the context of an AI/algorithm-driven device's performance metrics (e.g., sensitivity, specificity) against a ground truth established by experts.

    Instead, the "acceptance criteria" discussed are related to the safety and effectiveness of a physical medical device, primarily through mechanical testing and biocompatibility testing, to demonstrate its substantial equivalence to a previously cleared predicate device.

    Therefore, many of the requested categories for AI/algorithm performance studies are not applicable to this document. I will fill in the applicable information based on the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with specific numerical targets and corresponding device performance values for each criterion in the manner typical for AI/algorithm validation. Instead, it describes categories of tests performed to verify design outputs met design requirements and to confirm proper function and durability. The general outcome is that the device "demonstrated adequate device performance," and is "non-sensitizing, non-irritating, not systemically toxic, non-hemolytic, and non-mutagenic."

    Acceptance Criteria Category (Evaluated Aspect)Reported Device Performance (Summary)
    Mechanical Performance
    Catheter Diameter and Balloon ProfileDesign outputs met design requirements; proper function and durability confirmed.
    Device Preparation, Deployment, and RetractionDesign outputs met design requirements; proper function and durability confirmed.
    Balloon Inflation and Deflation TimeDesign outputs met design requirements; proper function and durability confirmed.
    Catheter Bond (Tensile) StrengthDesign outputs met design requirements; proper function and durability confirmed.
    Tip Pull StrengthDesign outputs met design requirements; proper function and durability confirmed.
    Torque StrengthDesign outputs met design requirements; proper function and durability confirmed.
    Pushability, Trackability, Secure EdgesDesign outputs met design requirements; proper function and durability confirmed.
    Guidewire CompatibilityDesign outputs met design requirements; proper function and durability confirmed.
    Catheter Effective LengthDesign outputs met design requirements; proper function and durability confirmed.
    Catheter Surface AppearanceDesign outputs met design requirements; proper function and durability confirmed.
    Luer CompatibilityDesign outputs met design requirements; proper function and durability confirmed.
    Coating LengthDesign outputs met design requirements; proper function and durability confirmed.
    Coating ThicknessDesign outputs met design requirements; proper function and durability confirmed.
    Coating LubricityDesign outputs met design requirements; proper function and durability confirmed.
    Coating IntegrityDesign outputs met design requirements; proper function and durability confirmed.
    Particulate EvaluationDesign outputs met design requirements; proper function and durability confirmed.
    Rated Burst PressureDesign outputs met design requirements; proper function and durability confirmed.
    ComplianceDesign outputs met design requirements; proper function and durability confirmed.
    Device FatigueDesign outputs met design requirements; proper function and durability confirmed.
    Flexibility and KinkDesign outputs met design requirements; proper function and durability confirmed.
    Biocompatibility
    CytotoxicityNon-cytotoxic (MEM Elution Test using L-929 Mouse Fibroblast cells).
    SensitizationNon-sensitizing (Guinea Pig Maximization Sensitization Test).
    IrritationNon-irritating (Intracutaneous Reactivity/Irritation Test).
    Systemic Toxicity (Acute & Material-Mediated)Not systemically toxic (Acute Systemic Injection Test, Material Mediated Pyrogens).
    HemocompatibilityNon-hemolytic (Partial Thromboplastin Time (PTT), Hemolysis Direct Contact/Extract Method, Direct Contact Complement Activation).
    Thrombogenicity (in vivo)No thrombo-embolism observed in GLP animal study under simulated use conditions.
    GenotoxicityNon-mutagenic (Reverse Bacterial Mutation, In vitro Mouse Lymphoma Assay-Extended Treatment, Rodent bone Marrow Micronucleus Assays).

    Note: The document states that "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised."

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document refers to "Test articles consisted of finished sterilized catheters" for bench testing. It does not specify the exact number of devices tested for each mechanical or biocompatibility test. For biocompatibility, it indicates studies were conducted, including a GLP animal study for thrombogenicity.
    • Data Provenance: The studies were conducted by Spectranetics, Inc. (the submitter). The context implies these were internal tests conducted to support the 510(k) submission. No information on country of origin of data (beyond the applicant being US-based) or retrospective/prospective nature is explicitly stated, though a GLP animal study would be prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable to this type of device submission. The "ground truth" for a physical medical device's performance is established by standardized testing methods and regulatory requirements, not expert consensus on diagnostic interpretations.

    4. Adjudication method for the test set:

    • Not applicable for this type of device validation. Adjudication methods are typically for evaluating human or AI performance against a "truth" in diagnostic settings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical medical instrument (a balloon catheter), not an AI diagnostic tool that assists human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical medical instrument, not an AI algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the performance of this physical medical device is based on engineering specifications, established standards, and regulatory requirements for mechanical integrity, functional performance, and biocompatibility. For example, for biocompatibility, the ground truth is "non-toxic," "non-irritating," etc., as defined by ISO 10993-1:2009 and FDA GLP regulations. For mechanical tests, the ground truth is that the device must meet predefined design requirements and exhibit proper function and durability.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device requiring a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122685

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-01-11

    (133 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    AI/ML Overview

    This document is a 510(k) summary for the AngioSculpt® PTA Scoring Balloon Catheter, which is a medical device. As such, it describes various tests and comparisons to predicate devices to establish substantial equivalence for regulatory approval. However, it does not contain specific acceptance criteria, reported device performance metrics in the format of a table, nor does it describe a study explicitly designed to prove device acceptance criteria in the way a clinical trial for a new diagnostic or AI-powered device would.

    The document primarily focuses on bench testing, biocompatibility testing, and animal testing to demonstrate the device's safety and effectiveness in comparison to existing, already approved devices (predicate devices).

    Here's an analysis based on the information provided, recognizing that some of your requested points are not applicable to this type of device submission:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table with explicit acceptance criteria (e.g., minimum tensile strength, maximum inflation time) alongside reported numerical performance. Instead, it states that "Mechanical testing of the AngioSculpt catheter was conducted in accordance with AngioScore's Risk Analysis and all applicable FDA guidance documents and relevant standards. The following bench tests were conducted to verify that design outputs met design requirements and to confirm proper function and durability." The results are generally described as meeting requirements or demonstrating appropriate function.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Bench Testing: The specific sample sizes for each bench test (e.g., Catheter Diameter, Device Burst Strength) are not explicitly stated. The data provenance would be internal AngioScore, Inc. laboratory testing, which is prospective for the purpose of this submission.
    • Biocompatibility Testing: Biocompatibility tests were not conducted as part of this device modification because materials were shown to be equivalent to currently marketed AngioScore products. The original biocompatibility testing would have been prospective internal or contract lab testing.
    • Animal Testing: "An acute GLP and non GLP study was conducted to determine the safety and deliverability of the AngioSculpt catheter. The studies were conducted in the llio-femoral peripheral arteries of one experimental animal." (Note: this is a very small sample size for animal testing, likely acceptable for minor device modifications and predicate device comparison). This was a prospective study. The country of origin is not specified but presumed to be where AngioScore, Inc. conducts its R&D or contracts animal studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This point is not applicable as the device is a physical medical device (balloon catheter), not a diagnostic device or AI system that relies on expert interpretation for ground truth in a test set. The "ground truth" for this device's performance is determined by engineering specifications and direct physical measurements/observations in bench and animal testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This point is not applicable for the same reasons as #3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This point is not applicable as the device is a physical medical device, not an AI or diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This point is not applicable as the device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For bench testing, the "ground truth" is based on engineering specifications, established industry standards, and regulatory guidance documents. Performance is measured directly against these specified values (e.g., correct diameter, sufficient burst strength, acceptable inflation time).
    • For biocompatibility testing, the "ground truth" is based on the results of standardized biological assays (e.g., cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, thrombosis) against predefined acceptable limits according to ISO 10993 standards and FDA guidance.
    • For animal testing, the "ground truth" is based on direct observation of device performance (deliverability, trackability, passability, inflation/deflation parameters, radiopacity, ease of removal) and lack of adverse events (dissection, perforation, embolization, thrombosis, loss of components) within the animal model, as interpreted by the study investigators.

    8. The sample size for the training set

    This point is not applicable as the device is a physical medical device, not an AI/machine learning system that requires a training set.

    9. How the ground truth for the training set was established

    This point is not applicable for the same reasons as #8.

    In summary, this 510(k) submission demonstrates "substantial equivalence" of the AngioSculpt® PTA Scoring Balloon Catheter to its predicate devices through rigorous bench testing, biocompatibility evaluations, and acute animal studies. While it confirms that the device "met design requirements" and "demonstrated adequate device performance," it does so within the context of regulatory comparison to existing devices rather than proving a novel acceptance threshold through specific clinical performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112182

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-08-26

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    AI/ML Overview

    The provided text is a 510(k) summary for the AngioSculpt PTA Scoring Balloon Catheter, which is a medical device. This document describes the device, its intended use, and the testing performed to demonstrate its safety and effectiveness for substantial equivalence to predicate devices. It does not provide information relevant to the acceptance criteria and study design for a software algorithm/AI device. Therefore, I cannot extract the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110767

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-15

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    The modified AngioSculpt PTA catheter has very similar technological characteristics as the cleared AngioSculpt PTA catheter. The modified AngioSculpt PTA Catheter is exactly the same device as the cleared AngioSculpt PTA catheter, except for minor differences to the distal bond and intermediate bond of the 0.018" OTW platform, 20mm balloon length catheters.

    AI/ML Overview

    Here's the information regarding the acceptance criteria and the study for the AngioSculpt® PTA Scoring Balloon Catheter, based on the provided document:

    This document is a 510(k) summary for a modification to an already cleared device, the AngioSculpt PTA Scoring Balloon Catheter. Therefore, the "study" described focuses on demonstrating that the modified aspects of the device still meet expectations, not on a full-scale clinical trial to prove initial efficacy or safety for a new device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't provide specific numerical acceptance criteria (e.g., "Bond Tensile Strength >= X N") or precise numerical performance results (e.g., "Bond Tensile Strength = Y N"). Instead, it states that the modified design was evaluated against specifications and met them.

    Acceptance Criterion (Evaluated Product Attribute)Reported Device Performance
    Entry Profile (distal bond OD)Met specifications
    Profile (intermediate bond OD)Met specifications
    Fatigue StrengthMet specifications
    Bond Tensile StrengthMet specifications
    Visual inspection for distal bond imperfectionsNo observations of imperfections
    Visual inspection for intermediate bond imperfectionsNo observations of imperfections

    2. Sample Size Used for the Test Set and Data Provenance

    The document provides no explicit sample sizes for the non-clinical performance testing. It generally refers to "design verification testing" and "application of risk analysis and a change impact assessment."

    The data provenance is non-clinical bench testing conducted by the manufacturer, AngioScore, Inc. Details such as country of origin for the data are not specified beyond the manufacturer's location in Fremont, CA, USA. This testing is retrospective in the sense that it's evaluating a modification to an existing design.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This was non-clinical bench testing of device attributes, not a clinical study requiring expert assessment of patient data. The "ground truth" was established by engineering specifications and testing protocols.

    4. Adjudication Method for the Test Set

    Not applicable. This was non-clinical bench testing. Results would typically be evaluated against pre-defined engineering specifications, not through a consensus or adjudication process involving multiple human reviewers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This document describes non-clinical engineering testing, not a clinical study involving human readers or patient cases.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Not applicable. This device is a physical medical instrument (a balloon catheter), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this non-clinical testing was engineering specifications and established testing methodologies to measure physical properties like bond strength and profile.

    8. The Sample Size for the Training Set

    Not applicable. This involves non-clinical bench testing for a physical device, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As above, this is about a physical device, not an AI/ML system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101735

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-14

    (23 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to ald in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    AI/ML Overview

    The provided document is a 510(k) summary for the AngioSculpt® PTA Scoring Balloon Catheter. It focuses on demonstrating substantial equivalence to a previously cleared predicate device (K100303) rather than presenting a de novo study with specific clinical acceptance criteria for effectiveness.

    Therefore, the document does not contain the acceptance criteria or a study proving the device meets those criteria for clinical effectiveness in the way typically expected for a new medical device seeking marketing authorization based on clinical performance.

    Instead, the document details bench testing and animal testing to demonstrate safety and performance characteristics for substantial equivalence to a predicate device. These tests are designed to show that the new device performs as expected and does not raise new safety or effectiveness concerns compared to the predicate.

    Here's a breakdown of what is in the document related to testing, formatted to address your questions where applicable, but highlighting that no clinical acceptance criteria or human clinical study results are presented:


    Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

    Disclaimer: This 510(k) summary focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing (bench and animal studies), rather than establishing clinical effectiveness against specific clinical acceptance criteria in a human study. Therefore, the requested information regarding human-in-the-loop performance, multi-reader multi-case studies, effect sizes, specific clinical ground truth, and training set details are not applicable and not found in this regulatory submission.

    1. A table of acceptance criteria and the reported device performance

    Since this is a 510(k) for substantial equivalence based on bench and animal testing, there are no "acceptance criteria" in the sense of clinical performance metrics for a human study. Instead, the document lists various bench tests and their purpose (e.g., to verify design outputs met design requirements, confirm proper function and durability) and describes the findings from animal testing.

    Acceptance Criterion (Test Objectives)Reported Device Performance (Summary of Results)
    Bench Testing:
    Balloon Rated Burst PressureConducted to verify that design outputs met design requirements and to confirm proper function and durability. (Specific quantitative results not provided in this summary, but implied to meet requirements.)
    Balloon Inflation and Deflation
    Balloon and Scoring Element Fatigue
    Bond Strength
    Crossing Profile
    Trackability and Pushability
    Over-the-Arch Torque Strength
    Catheter Body Integrity
    Balloon Compliance
    Surface Appearance
    Secure Edges
    Freedom from Interference
    Scoring Element Creates Focal Force
    Biocompatibility Testing:
    CytotoxicityNon-cytotoxic
    SensitizationNon-sensitizing
    IrritationNon-irritating
    Systemic ToxicityNot systemically toxic
    Hemocompatibility (PTT, Hemolysis)Non-hemolytic
    Hemocompatibility (Thrombosis in-vivo)No thrombo-embolism observed under simulated use conditions.
    Animal Testing (Acute GLP Study):
    Safety & DeliverabilitySuccessfully introduced using standard guide wires and expanded in targeted tissue. No dissection, perforation, or embolization. All devices maintained integrity with no loss of components. All procedures performed with ease, no adverse events. Clinically acceptable.
    Trackability & PassabilityDemonstrated as satisfactory. (Specific quantitative metrics not provided)
    Device Inflation & DeflationDemonstrated as satisfactory. (Specific quantitative metrics not provided)
    RadiopacityDemonstrated as satisfactory. (Specific quantitative metrics not provided)
    Device RemovalDemonstrated as satisfactory. (Specific quantitative metrics not provided)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Bench Testing: "Test articles consisted of finished sterilized catheters." (Specific sample sizes for each test are not provided in this summary). Data provenance is from AngioScore, Inc. (Fremont, CA, USA). These are prospective tests performed as part of device development and verification.
    • Biocompatibility Testing: Specific sample sizes for each in vitro or in vivo test are not explicitly provided in this summary. The thrombogenicity evaluation was part of a GLP animal study.
    • Animal Testing: The study involved one experimental animal (a single animal). The study was acute and conducted in the femoral peripheral arteries. It was a prospective, GLP (Good Laboratory Practice) study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not Applicable. No human clinical study or test set requiring expert ground truth for diagnostic accuracy (common in AI/imaging devices) was conducted. Evaluations were based on physical measurements, biological assays, and observational findings in an animal model by study personnel.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. No human clinical study or test set requiring adjudication was conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a physical medical device (balloon catheter), not an AI/imaging device that would involve human readers or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a physical medical device (balloon catheter), not an AI/algorithm device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For bench testing, ground truth is based on established engineering principles, material specifications, and regulatory standards.
    • For biocompatibility testing, ground truth is based on standard ISO 10993-1 methods and 21 CFR Part 58 GLP regulations, relying on established biological endpoints.
    • For animal testing, the "ground truth" (i.e., evaluation of device performance, integrity, and lack of adverse events) was established through direct observation, examination (e.g., for dissection, perforation, embolization), and assessment by the trained GLP study personnel.

    8. The sample size for the training set

    • Not Applicable. This device is a physical medical device. There is no concept of a "training set" as would be used for an AI/machine learning model.

    9. How the ground truth for the training set was established

    • Not Applicable. As above, no training set for an AI model. For physical device development, "ground truth" is typically defined by design specifications and regulatory standards against which performance is measured.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100303

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-03-22

    (47 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the AngioSculpt® PTA Scoring Balloon Catheter, seeking substantial equivalence to previously cleared devices. It describes the device, its intended use, and compares it to predicate devices. However, the document does not contain specific acceptance criteria, study details, or performance data in the format requested.

    Therefore, I cannot provide the requested table and detailed information based on the provided text. The document focuses on demonstrating substantial equivalence rather than presenting a performance study with acceptance criteria.

    Information not available in the provided document:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes used for test sets.
    • Data provenance (country of origin, retrospective/prospective).
    • Number and qualifications of experts for ground truth.
    • Adjudication method for the test set.
    • Results of a multi-reader multi-case (MRMC) comparative effectiveness study or any effect size of human reader improvement with AI assistance (as this is a medical device, not an AI product).
    • A standalone algorithm performance study (as this is a medical device, not an AI algorithm).
    • Type of ground truth used (pathology, outcomes data, etc.).
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    The document states: "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised." However, it does not provide the specifics of these tests, their acceptance criteria, or the results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091966

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-07-30

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The sconng element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    AI/ML Overview

    This document is a 510(k) summary for the AngioSculpt PTA Scoring Balloon Catheter (K091966). It does not contain the detailed acceptance criteria or the study results that would be needed to fill out a table of acceptance criteria and reported device performance. It focuses on demonstrating substantial equivalence to previously cleared devices rather than presenting novel clinical study data against defined acceptance criteria.

    Therefore, I cannot provide the requested information from the given text. The provided document states:

    • No specific acceptance criteria are listed. The document focuses on demonstrating substantial equivalence to predicate devices.
    • No specific study is described with performance metrics. The document mentions "Design verification and validation testing demonstrated adequate device performance," but does not detail the nature, sample size, or specific results of these tests, particularly in a way that aligns with "device performance" as might be measured in a clinical study (e.g., success rates, complication rates).

    The remaining sections of your request (sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, type of ground truth, training set sample size, how training set ground truth was established) require information that would typically be found in a detailed clinical study report or a more comprehensive premarket submission, which is not present in this 510(k) summary.

    A 510(k) summary is a high-level overview intended to demonstrate substantial equivalence, not a full clinical study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082059

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-08-11

    (21 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon. slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stennsis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    AI/ML Overview

    Unfortunately, the provided document K082059 is a 510(k) premarket notification correcting a previous letter for the AngioSculpt® PTA Scoring Balloon Catheter. This type of document, particularly for a device that is substantially equivalent to a predicate, typically focuses on demonstrating equivalence rather than detailing a specific clinical study with acceptance criteria and a comprehensive study report as you've requested.

    Therefore, the document does not contain the following information:

    • A table of acceptance criteria and the reported device performance.
    • Sample size used for the test set and the data provenance.
    • Number of experts used to establish the ground truth for the test set and their qualifications.
    • Adjudication method for the test set.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or its effect size.
    • If a standalone performance (algorithm only) study was done.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc).
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    The document states that "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised." However, it does not provide the specifics of these tests, including the acceptance criteria or detailed results in a format that would allow me to populate the requested table or answer the specific questions about study design and ground truth.

    In summary, this document is a regulatory approval letter based on substantial equivalence to existing devices, not a detailed clinical study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080151

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-04-25

    (94 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, illio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    AI/ML Overview

    This document describes the AngioSculpt® PTA Scoring Balloon Catheter, a device intended for dilating lesions in various arteries and for treating obstructive lesions in arteriovenous fistulae. While it details the device's intended use, design, and comparison to predicate devices, it does not contain information about specific acceptance criteria, device performance metrics, or a study design with sample sizes, expert involvement, or adjudication methods.

    This document is a 510(k) Summary, a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. This type of submission typically focuses on demonstrating equivalence through design and functional performance similarities and verification/validation testing, rather than reporting detailed clinical study results against pre-defined acceptance criteria.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, ground truth establishment, or multi-reader studies.

    Here's what can be extracted from the document:

    • Device Name: AngioSculpt® PTA Scoring Balloon Catheter
    • Intended Use/Indications for Use:
      • Dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries.
      • Treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
      • Not for use in the coronary or neuro-vasculature.
    • Regulatory Class: Class II (two)
    • Product Codes: PNO, DQY, and LIT (the first page mentions PNO, the summary mentions DQY and LIT)
    • Predicate Devices:
      • Cordis Aviator® PTA Dilatation Catheter (K013581)
      • Cordis Savvy® PTA Dilatation Catheter (K971010)
      • AngioSculpt® Scoring Balloon Catheter (K050629)
      • AngioSculpt® Scoring Balloon Catheter (K072225)
    • Device Description: A standard two-lumen catheter with a scoring balloon near the distal tip. The distal end has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force to minimize balloon slippage and assist with luminal expansion. The balloon has radiopaque markers.
    • Statement on Performance: "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised."

    Missing Information (as per your request):

    1. A table of acceptance criteria and the reported device performance: Not present in this document. The document states "adequate device performance" but provides no specific metrics or criteria.
    2. Sample sized used for the test set and the data provenance: Not present. The document mentions "design verification and validation testing" but does not detail the testing methodology, sample sizes, or data provenance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable or not present. This type of detail is usually found in clinical study reports, which are not part of this 510(k) summary. Given the device type, "ground truth" would likely refer to physical/mechanical testing results, not expert consensus on medical images.
    4. Adjudication method for the test set: Not applicable or not present.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No, this is an angioplasty catheter, not an imaging AI device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is an angioplasty catheter.
    7. The type of ground truth used: For this device, ground truth would refer to engineering specifications and performance metrics (e.g., balloon burst pressure, trackability, pushability, material compatibility, dimensional accuracy). This document does not detail the specific "ground truth" used for its verification and validation.
    8. The sample size for the training set: Not applicable. This document is for a medical device (catheter), not an AI algorithm that requires a training set.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072225

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-10-12

    (63 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt Scoring Balloon Catheter is intended for balloon dilatation of lesions in infrapopliteal arteries. Not for use in the coronary or neuro-vasculature.

    Device Description

    The modified AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diagneter and length at a specific pressure.

    AI/ML Overview

    This document is a 510(k) premarket notification for the AngioSculpt® PTA Scoring Balloon Catheter, specifically addressing a modification. The document primarily focuses on establishing substantial equivalence to previously cleared devices rather than providing detailed clinical study results or acceptance criteria in the format requested.

    Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" cannot be fully extracted from the provided text. The document states that "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised," but it does not detail those tests or their specific acceptance criteria.

    Here's a breakdown of what can be inferred or explicitly stated based on the provided text, and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided. The document states that "Design verification and validation testing demonstrated adequate device performance," but it does not specify what those performance criteria were or the quantitative results of those tests. It focuses on demonstrating substantial equivalence in fundamental design, scientific technology, functional performance, and materials to a predicate device (AngioSculpt® Scoring Balloon Catheter K050629) and similar dimensions/size configurations to another predicate (Cordis Savvy® PTA Dilatation Catheter K971010).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be determined from the provided text. The document mentions "Design verification and validation testing" but does not provide details on sample sizes, study design (retrospective/prospective), or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Cannot be determined. This document describes a medical device submission, not a study evaluating an AI algorithm with expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Cannot be determined. This document describes a medical device submission, not a study involving expert adjudication of data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI device. The document describes a physical medical device (a scoring balloon catheter).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is not an AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable. The document refers to "Design verification and validation testing," which for a physical device would involve engineering tests (e.g., burst pressure, fatigue, dimensional accuracy, force profiles) rather than establishing ground truth like pathology or expert consensus for clinical disease states.

    8. The sample size for the training set

    • Not applicable. This is not an AI device.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI device.

    Summary based on the provided document:

    The provided document is a 510(k) submission for a modified AngioSculpt® PTA Scoring Balloon Catheter. The primary method for demonstrating the device's acceptability is through substantial equivalence to predicate devices. The document explicitly states:

    • "The modified AngioSculpt catheter shares the same fundamental design, scientific technology (operating principle), functional performance, and similar materials as the original AngioSculpt catheter."
    • "The modified AngioSculpt catheter has similar dimensions and size configurations as the Savvy catheter."
    • "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised."

    However, the specific "acceptance criteria" (such as quantitative thresholds for performance metrics) and the detailed results of the "design verification and validation testing" are not included in this summary document. The document functions as a high-level notification to the FDA about a device modification and its substantial equivalence, not a detailed technical report of testing data.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1