Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K111870

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-07-29

    (28 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications.

    Device Description

    The Amsco Warming Cabinet is designed to store and warm sterile IV solutions, surgical irrigation solutions, linens and/or blankets to an acceptable level for hospital and surgical outpatient center applications. The 24" Wide Pass-thru model will be mounted in a wall opening. Doors (upper and lower compartment) added to the rear of the cabinet will allow users to load supplies in compartments from outside the operating room, while doors on the front side will allow access to compartments when inside the operating room. The upper compartment of this model holds up to 20( 1-liter) liquid bottles, or 6 ( 1liter) liquid (IV or Irrigation solution) bags; the lower compartment holds up to 40 (1-liter) liquid bottles or 12 (1 liter) liquid bags.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text, formatted to match your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test DescriptionAcceptance CriteriaReported Device Performance
    Electrical Testing
    Dielectric WithstandAchieve passing results through automated dielectric strength tester.PASS
    Earth and Enclosure Leakage CurrentMust not exceed requirements listed in UL 61010-1 clause 6.3.1b (must not exceed .5mA normal condition and 3.5mA single fault condition).PASS
    Ground bondMust not exceed 0.1 ohm and maintain ground continuity.PASS
    Power Input & Line VoltageThe power input of equipment at rated voltage and steady state current shall not exceed the marked rating by 10%.PASS
    System Functions
    Main Power ON indicatorMain Power ON indicator at rear upper door is illuminated when Main Power Switch is set to the ON position.PASS
    Exhaust fans and blowers operationExhaust fans and blowers must operate continuously when the main power is set to the ON position. Opening or closing the doors shall not cause fans or blowers to turn off.PASS
    Input currentInput current must be within 11.4 - 14.25 Amps.PASS
    Over-temp alarmOver-temp alarm must activate and displayed temperature flashes alternately with the error "Hi" when compartment temperature is more than 10°F above set point temperature.PASS
    Door lockDoor must remain locked when 40 pounds of force is applied.PASS
    Thermal Test
    Temperature Limits (recessed mounted, normal condition)Temperatures of easily touched surfaces and identified components must comply with UL 61010-1 clauses 10.1 and 10.2, UL 61010-2-010 clause 10.1, temperature limits under normal conditions. Maintain normal operation without faults occurring.PASS
    Single Fault Conditions
    Cooling (single fault condition)Pass dielectric strength test per UL 61010-1, section 6, clause 6.8. Cabinet outer surface must not exceed 105°C at 40°C ambient. Transformer, blower and exhaust fan windings must not exceed 150°C under single fault condition at 40°C ambient. Voltage measurement of cabinet outer surfaces must not exceed test voltages per UL 61010-1, section 6, clause 6.3.2a. No signs of molten metal, burning insulation, flaming particles, etc. No signs of charring, glowing, or flaming of tissue paper or cheesecloth.PASS
    Heating Devices (single fault condition)Pass dielectric strength test per UL 61010-1, section 6, clause 6.8. Cabinet outer surface must not exceed 105°C at 40°C ambient. Transformer, blower and exhaust fan windings must not exceed 150°C under single fault condition at 40°C ambient. Voltage measurement of cabinet outer surfaces must not exceed test voltages per UL 61010-1, section 6, clause 6.3.2a. No signs of molten metal, burning insulation, flaming particles, etc. No signs of charring, glowing, or flaming of tissue paper or cheesecloth.PASS
    Protection against Hazards from FluidsPass dielectric strength test per UL 61010-1, section 6, clause 6.8. Voltage measurement of cabinet outer surfaces must not exceed test voltages per UL 61010-1, section 6, clause 6.3.2a.PASS
    Heating Performance (empty compartment)Temperature reading of each thermocouple must not vary from set-point temperature by more than $\pm$ 3°F in upper compartment and $\pm$ 5°F in lower compartment. At a 160°F set-point temperature, the upper compartment display must reach set-point within 35 minutes and lower compartment display must be within 65 minutes. Normal system operation is maintained without over-temperature condition or faults occurring.PASS
    Heating Performance (Full IV solution load in upper compartment and full blanket load in lower compartment)Temperature reading of each thermocouple must not vary from set-point temperature by more than $\pm$ 3°F in upper compartment and $\pm$ 5°F in lower compartment. Upper and lower compartment controls temperature display must reach set-point within 12 hours. Normal system operation is maintained without over-temperature condition or faults occurring.PASS
    Heating Performance (Full irrigation solution load in upper and lower compartments)With a fully loaded compartment of Irrigation solution, bottles of solution shall not overbalance and fall off shelf when doors are opened. Temperature reading of each thermocouple must not vary from set-point temperature by more than $\pm$ 3°F in upper compartment and $\pm$ 5°F in lower compartment. Upper and lower compartment controls temperature display must reach set-point within 12 hours. Normal system operation is maintained without over-temperature condition or faults occurring.PASS
    ETL/cETL Code ComplianceMeet UL 61010-1 Standard for Safety Electrical Equipment for Measurement, Control, and Laboratory Use. Meet CAN/CSA C22.2 61010-1 Standard for Safety Electrical Equipment for Measurement, Control and Laboratory Use.PASS

    2. Sample size used for the test set and the data provenance

    The document describes verification and validation activities performed on the physical device, not an algorithm using a dataset. Therefore, terms like "sample size for the test set" and "data provenance" (country of origin, retrospective/prospective) are not directly applicable in the context of this device's testing. The "test set" in this case refers to the physical Amsco Warming Cabinet, 24" Wide Pass-thru model, itself undergoing various engineering and functional tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical device (warming cabinet) and the ground truth for its performance is established through objective engineering tests against established safety and performance standards (UL, CAN/CSA). There are no "experts" establishing ground truth in the sense of clinical interpretations or diagnoses.

    4. Adjudication method for the test set

    Not applicable. The tests are objective measurements and evaluations against predefined quantitative and qualitative acceptance criteria (e.g., current limits, temperature ranges, successful operation of alarms). There is no ambiguity requiring adjudication by multiple readers or decision-makers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (warming cabinet), not an AI algorithm for diagnostic imaging or similar applications where MRMC studies or AI assistance for human readers would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical warming cabinet, not an algorithm. The "standalone" performance is the device operating as intended based on the tests described.

    7. The type of ground truth used

    The ground truth for this device is based on established engineering and safety standards (UL 61010-1, CAN/CSA C22.2 61010-1), which define acceptable performance limits for electrical safety, temperature control, and mechanical integrity.

    8. The sample size for the training set

    Not applicable. This is a medical device undergoing physical and functional testing, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as no training set for an algorithm was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110769

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-06-03

    (77 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications.

    Device Description

    The Amsco Warming Cabinet is designed to store and warm sterile IV solutions, surgical irrigation solutions, linens and/or blankets to an acceptable level for hospital and surgical outpatient center applications. The mid-size model consists of a single chamber unit and is available in (4) saleable configurations all with premium features (24" deep, glass door, electronic door lock, data recording and 120 or 230Vac with or without option mobile base). Optional pull-out wire baskets can be interchangeable with the existing wire shelf. Each of the two wire baskets have a 30L maximum capacity. The external measurements are 36-3/8"H x 27-1/8"D x 30"W (without mobile base) and the internal heat compartment is 22-3/4"H x 22-7/8"D x 24"W (7.2 cu ft).

    AI/ML Overview

    This document describes the acceptance criteria and the verification and validation study for the Amsco Warming Cabinet, Mid-size Model.

    1. Table of acceptance criteria and the reported device performance

    Test DescriptionAcceptance CriteriaResults
    Electrical Testing• Achieve passing results through automated dielectric strength tester• Earth and Enclosure Leakage Current must not exceed requirements listed in UL 61010-1 clause 6.3.1b (must not exceed .5mA normal condition and 3.5mA single fault condition)• Ground bond must not exceed 0.1 ohm and maintain ground continuity.• The power input of equipment at rated voltage and steady state current shall not exceed the marked rating by 10%.PASS
    System Functions: Controls, Heating, Door Lock System Test• Exhaust fans and blowers must operate continuously when the main power is set to the ON position• Opening or closing the doors shall not cause fans or blowers to turn off• Input current must be within 5.9 - 7.12 Amps for 120Vac unit.• Over-temp alarm must activate and displayed temperature flashes alternately with the error "Hi" when compartment temperature is more than 10°F above set point temperature• Warming cabinet must communicate with laptop or personal-computer to transfer stored temperature data to the computer• Numeric code entered on keypad must unlock the door within 4 seconds.• Door must remain locked when 40 pounds of force is appliedPASS
    Spillage• Pass dielectric strength test per UL 61010-1, section 6, clause 6.8• Voltage measurement of cabinet outer surfaces must not exceed test voltages per UL 61010-1, section 6, clause 6.3.2a.PASS
    Heating Performance (empty compartment)• Temperature reading of each thermocouple must not vary from set-point temperature by more than ±3°F for temperature settings between 90-110°F and ±5°F for temperature settings between 110-160°F.• Normal system operation is maintained without over-temperature condition or faults occurring.PASS
    Heating Performance (Full IV solution load - upper and lower wire baskets)• Temperature reading of each thermocouple must not vary from set-point temperature by more than ±3°F for temperature settings between 90-110°F.• Controls temperature display must reach set-point within 12 hours.• Normal system operation is maintained without over-temperature condition or faults occurring.PASS
    Heating Performance (Full irrigation solution load - upper and lower wire baskets)• Temperature reading of each thermocouple must not vary from set-point temperature by more than ±3°F for temperature settings between 90-110°F and ±5°F for temperature settings between 110-160°F• Controls temperature display must reach set-point within 12 hours.• Normal system operation is maintained without over-temperature condition or faults occurring.PASS
    Heating Performance (Full irrigation solution load - upper wire basket) (Full blanket load - bottom shelf)• Temperature reading of each thermocouple must not vary from set-point temperature by more than ±3°F for temperature settings between 90-110°F and ±5°F for temperature settings between 110-160°F• Controls temperature display must reach set-point within 12 hours.• Normal system operation is maintained without over-temperature condition or faults occurring.PASS
    Drop Test• Pass dielectric strength test per UL 61010-1, section 6, clause 6.8• No damage that allows access to electrical live parts• No damage to the enclosure that could cause a safety hazard• No damage that could cause spread of fire• No damage to insulation of internal wiringPASS
    Stability Test• No overbalance during testPASS
    Wire Basket Load and Force Test• No catastrophic failure at 125 pounds.• No catastrophic failure at 105 pounds with the sliding of wire basket in a back and forward horizontal motion for at least 100 cycles.PASS
    ETL/cETL Code Compliance• Meet UL 61010-1 Standard for Safety Electrical Equipment for Measurement, Control, and Laboratory Use.• Meet CAN/CSA C22.2 61010-1 Standard for Safety Electrical Equipment for Measurement, Control and Laboratory Use.PASS

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific number of units used for each test. However, it details "verification and validation activities" conducted on the "Mid-size Single Compartment Warming Cabinet". The nature of these tests (electrical, system functions, heating performance, mechanical tests) suggests that the testing was performed on the device itself (product prototypes or final units) rather than on a dataset of external data. This would be considered prospective testing of the device hardware and software. The country of origin of the data is not specified, but the manufacturer (STERIS Corporation) is based in Mentor, OH, USA, implying the testing was likely conducted in the US or by a contracted lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a medical device (warming cabinet) validation, not an AI or diagnostic imaging study. Ground truth in this context is established by engineering specifications, safety standards (UL, CAN/CSA), and performance requirements for heating and mechanical integrity. There are no "experts" in the clinical sense establishing ground truth; rather, the device's performance is measured against predefined objective physical and electrical standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As described above, the acceptance criteria are objective physical and electrical measurements against established standards. There is no subjective interpretation or "adjudication" required in the typical sense of a clinical study or AI model evaluation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI or diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI or algorithm-based device in the context of diagnostic or interpretive functions. The "software" mentioned is for control and data recording within the device's operational functions, not for independent diagnostic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device's performance is based on:

    • Engineering Specifications: Specific temperature ranges, electrical parameters, power draw.
    • Safety Standards: Compliance with UL 61010-1 and CAN/CSA C22.2 61010-1 for electrical safety.
    • Performance Requirements: Functional operation of fans, alarms, door locks, and heating profiles under various load conditions.
    • Mechanical Integrity Standards: Resistance to physical stress (drop test, load test).

    These are objective, measurable criteria, not subjective expert consensus or clinical outcomes data.

    8. The sample size for the training set

    Not applicable. This is not an AI or machine learning device that requires a training set. The device operates based on predefined engineering and control logic.

    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI/ML device, there is no training set or associated ground truth for such a set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092823

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-12-18

    (95 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amsco Warming Cabinet is designed to raise the temperature of blankets, linens and sterile surgical irrigation solutions and IV solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications.

    Device Description

    The Amsco Warming Cabinet is designed to store and warm sterile IV solutions, surgical irrigation solutions, linens and/or blankets to an acceptable level for hospital and surgical outpatient center applications. The upper compartment of the 18" (457mm) deep model holds up to 24(1-liter) liquid bags or bottles and 14 (1-liter) IV solution bags; the lower compartment holds up to 72 (1-liter) liquid bags or bottles. The upper compartment of the 24" (610mm) deep model holds up to 30 (1-liter) surgical flasks; the lower compartment holds up to 90 (1-liter) surgical flasks.

    AI/ML Overview

    I am sorry, but based on the provided text, there is no information about acceptance criteria or a study that proves a device meets those criteria.

    The document is a 510(k) premarket notification for a medical device called the "Amsco Warming Cabinet." It primarily discusses the substantial equivalence of this device to predicate devices and its intended use. While it includes a table comparing technological characteristics of the proposed device against predicate devices, this table focuses on features and specifications rather than acceptance criteria for performance studies.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them, nor can I provide details related to sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set information.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1