510(k) Data Aggregation

    K Number
    K082898

    Validate with FDA (Live)

    Date Cleared
    2008-12-19

    (80 days)

    Product Code
    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amedica Home Drug Test Cup is an in vitro diagnostic test for the rapid detection of the following drugs in human urine.
    Marijuana, Cocaine, Amphetamine, Methamphetamine, Opiates, Opiates300, Pencyclidine, Barbiturates, Benzodiazepines, Methadone, Oxycodone, MDMA, Tricyclic Antidepressants.
    This test is intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing the second step in the process, is provided in the package labeling.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) summary for the "Amedica Home Drug Test Cup," a device intended for over-the-counter (OTC) consumer use for detecting various drugs in human urine.

    Acceptance Criteria and Device Performance:

    The document does not explicitly state "acceptance criteria" in a table format with performance metrics. However, the core of the submission is to demonstrate substantial equivalence to legally marketed predicate devices, which implies that the performance of the Amedica Home Drug Test Cup is comparable to these predicates for the stated indications.

    The key performance aspect mentioned is the detection of specific drugs at defined cutoff levels. The device is intended to provide a preliminary positive or negative result based on these cutoffs.

    DrugAnalyteCutoffReported Device Performance (Implied)
    MarijuanaTHC50 ng/mLDetects THC at or above 50 ng/mL
    CocaineBenzoecgonine300 ng/mLDetects Benzoecgonine at or above 300 ng/mL
    AmphetamineAmphetamine1000 ng/mLDetects Amphetamine at or above 1000 ng/mL
    MethamphetamineMethamphetamine1000 ng/mLDetects Methamphetamine at or above 1000 ng/mL
    OpiatesMorphine2000 ng/mLDetects Morphine at or above 2000 ng/mL
    Opiates300Morphine300 ng/mLDetects Morphine at or above 300 ng/mL
    PencyclidinePhencylidine25 ng/mLDetects Phencylidine at or above 25 ng/mL
    BarbituratesSecobarbital300 ng/mLDetects Secobarbital at or above 300 ng/mL
    BenzodiazepinesOxazepam300 ng/mLDetects Oxazepam at or above 300 ng/mL
    MethadoneMethadone300 ng/mLDetects Methadone at or above 300 ng/mL
    OxycodoneOxycodone100 ng/mLDetects Oxycodone at or above 100 ng/mL
    MDMAMDMA500 ng/mLDetects MDMA at or above 500 ng/mL
    Tricyclic AntidepressantsNortriptyline1000 ng/mLDetects Nortriptyline at or above 1000 ng/mL

    Note: The document only lists the cutoffs and states the device's purpose is to detect these drugs at these cutoffs. It does not provide specific sensitivity, specificity, or accuracy data for the Amedica Home Drug Test Cup. The substantial equivalence claim implies that its performance is equivalent to legally marketed predicate devices, meaning it meets the generally accepted performance standards for point-of-care drug tests.

    Study Information:

    The provided text does not contain any details about a specific study (like a clinical trial or performance evaluation study) that proves the device meets acceptance criteria. A 510(k) submission primarily relies on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting full clinical trial data as would be required for a PMA (Premarket Approval) application.

    Therefore, the following information points cannot be directly extracted from the provided text:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not mentioned.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not mentioned. The "ground truth" for these types of tests is typically established through a more specific alternate chemical method (like GC/MS or HPLC), not expert consensus in the diagnostic imaging sense.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not mentioned.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not relevant or mentioned, as this is a device for drug detection and not an AI-assisted diagnostic imaging tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device itself is standalone in terms of generating a preliminary result. There's no "algorithm only" performance separate from the physical test cup.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The document specifies that for preliminary positive results, a "more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography / Mass Spectrometry (GC/MS) or High Performance Liquid Chromatography (HPLC) is the preferred confirmatory method." This indicates that GC/MS or HPLC are the established ground truth methods.
    7. The sample size for the training set: Not mentioned. This type of device relies on chemical reactions, not machine learning or AI with training sets.
    8. How the ground truth for the training set was established: Not applicable, as there's no training set in the AI/ML sense.