Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K173653

    Validate with FDA (Live)

    Date Cleared
    2018-05-02

    (155 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Alere i Strep A 2 is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.

    Device Description

    Alere™ i Strep A 2 is a rapid, instrument-based isothermal test for the qualitative detection of Streptococus pyogenes Group A Strep from throat swab specimens. The Alere™ i Strep A 2 System utilizes isothermal nucleic acid amplification technology and is comprised of:

    • . Sample Receiver - single use, disposable containing the elution buffer
    • Test Base – single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet
    • Transfer Cartridge single use, disposable for transfer of the eluted sample to the Test Base, and ●
    • . Alere™ i Instrument – repeat use reader

    The reaction tubes in the Test Base contain the reagents required for Streptococcus pyogenes Group A Strep bacterial lysis and the subsequent amplification of the target nucleic acid and an internal control. Alere™ i Strep A 2 utilizes a pair of templates (similar to primers) for the specific amplification of DNA from Streptococcus pyggenes, Group A Strep and fluorescently labeled molecular beacons designed to specifically identify the amplified nucleic acid targets. Alere™ i Strep A 2 is performed within the confinement of the Test Base, and no other part of the Alere™ i Instrument has contact with the sample during the amplification process. This reduces the risk of instrument contamination and sample carry-over between measurements.

    To perform the assay, the Sample Receiver and Test Base are inserted into the Alere™ i Instrument. The sample is added to the Sample Receiver and transfer Cartridge to the Test Base, resuspending the lyophilized pellets contained within the Test Base and initiating bacterial lysis and target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.

    Results are displayed by the Alere™ i Instrument and are also stored in an on-board archive and are assigned to a sample ID that has been entered into the Alere™ i Instrument by the operator, and the date/time the test was performed. Data can be retrieved and downloaded by the operator at any time after testing. An external Alere™ Universal Printer can be attached via USB to the Alere™ i Instrument to print test results.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Alere™ i Strep A 2 device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance document doesn't explicitly state all "acceptance criteria" as clear numerical thresholds for performance metrics. However, it presents the clinical performance of the device against a comparator method (bacterial culture), and these reported values implicitly demonstrate that the device met the necessary performance expectations for clearance.

    Performance MetricAcceptance Criteria (Implied by Clearance)Reported Device Performance (Alere™ i Strep A 2 vs. Culture)
    Clinical SensitivityHigh (specific threshold not stated)98.5% (95% CI = 95.6%, 99.5%)
    Clinical SpecificityHigh (specific threshold not stated)93.4% (95% CI = 91.4%, 94.9%)
    Positive Predictive ValueHigh (specific threshold not stated)78.9% (95% CI = 74.3%, 83.6%)
    Negative Predictive ValueHigh (specific threshold not stated)99.6% (95% CI = 98.3%, 99.9%)
    Initial Invalid RateLow (specific threshold not stated)0.9%
    Invalid Rate (after retest)Very Low (specific threshold not stated)0.4%
    Analytical Sensitivity (LOD)Specific concentrations for each strainATCC 12344: 147 cells/mL (100% Detected) ATCC 19615: 25 cells/mL (95% Detected)
    ReactivityPositive results for tested strainsAll listed strains produced positive reactions
    Analytical Specificity (Cross-Reactivity)Negative results for tested microorganismsAll listed microorganisms and yeast produced negative results
    Interfering SubstancesNo effect on test performanceMost substances showed no effect; few instances of false-negative/positive at higher concentrations
    Reproducibility (low/moderate positive)100% agreement with expected results100% (90/90) agreement
    Reproducibility (negative)100% negative results100% (90) negative results

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 981 evaluable throat swab specimens.
    • Data Provenance: Multi-center, prospective clinical study conducted at nine (9) US trial sites in 2017.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The document states that the Alere™ i Strep A 2 performance was evaluated "in comparison to bacterial culture." Bacterial culture is typically performed in a laboratory by trained microbiologists using established protocols. The document does not specify the number of experts, nor their specific qualifications (e.g., years of experience), but implies standard laboratory practices for culture results.
    • For discrepancies, a "laboratory developed real-time PCR assay" was used for confirmation. This also implies expert analysis within a laboratory setting, but specific expert details are not provided.

    4. Adjudication Method for the Test Set

    • The primary ground truth for the clinical study was bacterial culture. In cases of discordance between Alere™ i Strep A 2 and bacterial culture, a "laboratory developed real-time PCR assay" was used for further investigation:
      • 38 of 52 samples positive by Alere™ i Strep A 2 and negative by bacterial culture were positive by PCR.
      • 1 of 3 samples negative by Alere™ i Strep A 2 and positive by bacterial culture was negative by PCR.
    • This suggests an implicit adjudication based on a tertiary, highly sensitive method (PCR) for resolving some discrepancies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not conducted. This device is an in-vitro diagnostic (IVD) molecular test, not an imaging or diagnostic assistant used by human readers in the traditional sense. The output is an automated positive/negative result.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, the clinical performance study directly reflects the standalone performance of the device (Alere™ i Strep A 2) without human interpretation affecting the result. The device is "instrument-based" and provides "automated" results. The operator's role is to perform the assay steps, but the diagnostic determination is made by the instrument/algorithm.

    7. The Type of Ground Truth Used

    • The primary ground truth used for the clinical performance study was bacterial culture of throat swab specimens.
    • For discordant results, a laboratory-developed real-time PCR assay was used as a confirmatory method.

    8. The Sample Size for the Training Set

    • The document does not provide information regarding a specific training set size. For IVD devices, especially molecular diagnostic kits, the "training" (development and optimization) process typically involves internal analytical studies rather than a distinct, prospectively collected "training set" of clinical samples with established ground truth in the same way an AI/ML algorithm might. Clinical studies are primarily for validation.

    9. How the Ground Truth for the Training Set Was Established

    • As a molecular diagnostic test, the "ground truth" for its development (analogous to a training set for AI) would primarily rely on well-characterized clinical samples and reference strains with known Streptococcus pyogenes status confirmed by methods like culture and sequencing during the assay development and optimization phases. However, the document does not detail this. The provided clinical study serves as the primary validation of the device against bacterial culture as the ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141757

    Validate with FDA (Live)

    Device Name
    ALERE I STREP A
    Date Cleared
    2015-03-31

    (274 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Alere™ i Strep A is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.

    All negative test results should be confirmed by bacterial culture because negative results do not prection with Group A Streptococcus and should not be used as the sole basis for treatment.

    Device Description

    Alere™ i Strep A is a rapid, instrument-based isothermal test for the qualitative detection of Group A Strep from throat swab specimens. The Alere™ i Strep A System utilizes isothermal nucleic acid amplification technology and is comprised of:

    • . Sample Receiver – single use, disposable containing the elution buffer
    • . Test Base – single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet
    • Transfer Cartridge – single use, disposable for transfer of the eluted sample to the Test Base, and
    • Alere™ i Instrument – repeat use reader

    The reaction tubes in the Test Base contain the reagents required for Group A Strep bacterial lysis and the subsequent amplification of the target nucleic acid and an internal control. Alere™ i Strep A utilizes a pair of templates (similar to primers) for the specific amplification of DNA from Group A Strep and a fluorescently labeled molecular beacon designed to specifically identify the amplified nucleic acid target. Alere™ i Strep A is performed within the confinement of the Test Base, and no other part of the Alere™ i Instrument has contact with the sample during the amplification process. This reduces the risk of instrument contamination and sample carry-over between measurements.

    To perform the assay, the Sample Receiver and Test Base are inserted into the Alere™ i Instrument and the elution buffer is automatically heated by the instrument. The sample is added to the Sample Receiver and transferred via the Transfer Cartridge to the Test Base, resuspending the lyophilized pellets contained within the Test Base and initiating bacterial lysis and target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.

    Results are displayed by the Alere™ i Instrument and are also stored in an on-board archive and are assigned to a sample ID that has been entered into the Alere™ i Instrument by the operator, and the date/time the test was performed. Data can be retrieved and downloaded by the operator at any time after testing. An external Alere™ Universal Printer can be attached via USB to the Alere™ i Instrument to print test results.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device (Alere™ i Strep A) meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Strict "acceptance criteria" are not explicitly stated in a quantitative manner (e.g., "Sensitivity must be >90%"). However, the clinical study's performance metrics act as the de facto demonstration of acceptable performance for regulatory purposes. The reported performance is what was considered sufficient for substantial equivalence.

    MetricReported Device Performance (95% CI)
    CLINICAL PERFORMANCE
    Sensitivity95.9% (91.4%, 98.1%)
    Specificity94.6% (91.6%, 96.6%)
    Positive Predictive Value88.7% (82.8%, 92.7%)
    Negative Predictive Value98.1% (96.0%, 99.1%)
    Initial Invalid Rate4.8% (3.3%, 7.1%)
    Invalid Rate (after re-test)2.8% (1.7%, 4.8%)
    ANALYTICAL PERFORMANCE
    Limit of Detection (ATCC 12344)4.2 CFU/mL (95% detected)
    Limit of Detection (ATCC 19615)41.8 CFU/mL (95% detected)
    Reactivity Test (GAP Strains)All tested strains produced positive results at or near LOD
    Cross-Reactivity (33 organisms)All produced negative results at specified concentrations
    Reproducibility (Moderate Pos)100% agreement (90/90)
    Reproducibility (Low Pos)91.1% agreement (82/90)
    Reproducibility (Negative)100% negative results (90/90)
    Reproducibility (High Negative)94.4% negative results (85/90)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical): 481 evaluable throat swab specimens.
    • Data Provenance: Multi-center, prospective clinical study conducted at 8 US trial sites in 2014.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not specify the number or qualifications of experts used to establish the ground truth. It states that the device performance was "compared to bacterial culture," implying independent laboratory analysis, but details about the personnel involved in interpreting these cultures are not provided.

    4. Adjudication Method for the Test Set

    The document doesn't explicitly describe an "adjudication method" in the typical sense (e.g., 2+1 reader adjudication for imaging studies). For samples with discordant results between the Alere™ i Strep A and the initial bacterial culture, a "laboratory developed real-time PCR assay" was used for further investigation:

    • Of 18 samples positive by Alere™ i Strep A and negative by bacterial culture, 13 were positive by RT-PCR.
    • Of 6 samples negative by Alere™ i Strep A and positive by bacterial culture, 4 were negative by RT-PCR.

    This suggests that the PCR assay acted as a secondary confirmation method for discordant results, but it's not a formal "adjudication method" agreed upon by multiple human readers in the context of interpretation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) test, where performance is typically evaluated against a gold standard (like bacterial culture or PCR), not by comparing human reader performance with and without AI assistance, as would be common for imaging AI.

    6. Standalone (Algorithm Only) Performance Study

    Yes, the clinical study (and analytical studies) assess the standalone performance of the Alere™ i Strep A device. It's an automated, instrument-based test, meaning its performance is inherently "algorithm only" in the sense that the instrument provides the result without human interpretation of the primary signal. The comparator (bacterial culture) serves as the ground truth.

    7. Type of Ground Truth Used

    • Clinical Study: Bacterial culture was the primary ground truth. For discordant results, a laboratory developed real-time PCR (RT-PCR) assay was used as a secondary method.
    • Analytical Sensitivity: Defined by the concentration of Group A Strep bacteria producing positive results 95% of the time, determined by testing known bacterial concentrations.

    8. Sample Size for the Training Set

    The document does not provide information about a separate "training set" for the Alere™ i Strep A. This is typical for IVD devices where the "training" (development and optimization) of the assay is conducted internally by the manufacturer, rather than through publicly documented machine learning training datasets. The presented clinical and analytical studies are validation studies, not training studies.

    9. How the Ground Truth for the Training Set Was Established

    As no separate training set details are provided, the method for establishing its ground truth is also not elaborated. The development of such assays involves extensive internal research, optimization, and verification using characterized bacterial strains and clinical samples, but these are not typically documented as a "training set" in the same way an AI model's training data would be.

    Ask a Question