Found 2 results

510(k) Data Aggregation

    K Number
    K122445

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-12

    (63 days)

    Product Code
    Regulation Number
    868.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.

    Device Description

    The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The systems are to be used only by trained and qualified medical professionals.

    The Aespire 7900 and Aespire View supply set flows of medical gases to the breathing system. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. They are available in trolley and pendant models, with two or three gases, two vaporizer positions and up to three cylinder connections. All models connect to oxygen and can additionally connect with up to two optional gases (air, N2O). The Aespire systems accept Tec 4. Tec 5. Tec 6. Tec 6+ and Tec 7 vaporizers on a Selectatec manifold. Safety features are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aespire View product provides optional electronic Total Fresh Gas Flow (TFS) monitoring. The Aespire View also features a color display, while the Aespire 7900 uses a monochromatic display.

    The Datex-Ohmeda 7900 Anesthesia Ventilator is used in this family of Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control (VCV) Mode, Pressure Control (PCV) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Support Ventilation (SIMV/PSV) Mode, Pressure Support with Apnea Backup.(PSVPro) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Control (SIMV-PC) Mode (Optional), and Pressure Control Ventilation – Volume Guaranteed (PCV-VG) mode (Optional on Aespire View variant only).

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device. As such, it primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving the device meets specific performance acceptance criteria in a clinical setting.

    Therefore, many of the requested items related to clinical studies, such as sample size, ground truth establishment, expert qualifications, and comparative effectiveness studies (MRMC), are not applicable or not provided in this type of submission.

    Here's an analysis of the provided text based on your request, highlighting what information is available and what is not:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like a typical clinical or standalone performance study. Instead, it focuses on verifying specifications and safety against existing standards and the predicate device.

    Acceptance Criteria (Implied)Reported Device Performance
    Risk analysis conductedCompleted
    Design reviews conductedCompleted
    Performance testing (Verification)Completed (e.g., environmental, electromagnetic compatibility)
    Safety testing (Verification)Completed
    No change to intended useConfirmed
    No change to fundamental scientific technologyConfirmed
    Functional equivalence to predicate device (K092864)Stated as "substantially equivalent," "as safe," and "as effective"
    Airway overpressure detectionMechanical switch replaced with electronic monitoring circuit; existing mechanical relief valve remains

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided. The submission focuses on verification testing (engineering/bench testing) of the updated circuit board and overall system against specifications, not a clinical "test set" in the context of diagnostic or AI performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. There is no mention of a "test set" requiring expert ground truth in the context of clinical performance evaluation for this 510(k) submission.

    4. Adjudication Method for the Test Set

    Not applicable, as there is no mention of a "test set" requiring expert-based ground truth adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "The modification to the family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator did not require clinical testing." This indicates no clinical study comparing human readers (or human readers with AI assistance) was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable in the context of this device. The changes were to an internal electronic sensor interface circuit board within an anesthesia system, not a diagnostic algorithm. The "performance testing (Verification)" and "safety testing (Verification)" would be akin to standalone engineering tests to ensure the updated component functions correctly and safely within the system, but not "algorithm only without human-in-the-loop performance" in the sense of an AI/diagnostic device.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission is based on engineering specifications, safety standards, and validated performance characteristics of the device and its components, verified through various engineering and quality assurance measures (e.g., risk analysis, design reviews, environmental testing, electromagnetic compatibility testing). There is no "clinical ground truth" (like pathology or outcomes data) mentioned because no clinical study was performed.

    8. The Sample Size for the Training Set

    Not applicable. This device is an anesthesia system with an updated circuit board, not an AI or machine learning algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092864

    Validate with FDA (Live)

    Date Cleared
    2010-01-12

    (117 days)

    Product Code
    Regulation Number
    868.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation. The devices are not suitable for use in a MRI environment.

    Device Description

    The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The systems are to be used only by trained and qualified medical professionals.

    The Aespire 7900 and Aespire View supply set flows of medical gases to the breathing system. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. They are available in trolley and pendant models, with two or three gases, two vaporizer positions and up to three cylinder connections. All models connect to oxygen and can additionally connect with up to two optional gases (air, N2O). The Aespire systems accept Tec 4, Tec 5, Tec 6, Tec 6+ and Tec 7 vaporizers on a Selectatec manifold. Safety features are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aespire View family member provides optional electronic Total Fresh Gas Flow (TFS) monitoring. The Aespire View also features a color display, while the Aespire 7900 uses a monochromatic display.

    The Datex-Ohmeda 7900 Anesthesia Ventilator is used in this family of Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control (VCV) Mode, Pressure Control (PCV) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Support Ventilation (SIMV/PSV) Mode, Pressure Support with Apnea Backup (PSVPro) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Control (SIMV-PC) Mode (Optional), and Pressure Control Ventilation - Volume Guaranteed (PCV-VG) mode (Optional on Aespire View variant only),

    AI/ML Overview

    This document is a 510(k) summary for the GE Datex-Ohmeda Aespire Anesthesia System, specifically introducing the Aespire View variant. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than detailing extensive clinical trials or acceptance criteria for a new device type. As such, much of the requested information regarding acceptance criteria, study details, and expert involvement is not present in the provided text.

    Here is an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not provide a table of acceptance criteria or reported device performance in the way typically associated with a new algorithm or diagnostic device. Instead, it states that the device was thoroughly tested through verification of specifications and validation. The performance is generally implied by the claim of substantial equivalence to predicate devices (K050626 and K090233), meaning it is expected to perform comparably.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a separate "test set" in the context of clinical performance evaluation. The "testing" mentioned refers to nonclinical verification and validation. Therefore, sample size and data provenance for a clinical test set are not applicable here.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not present as there was no clinical study described that involved establishing a ground truth by experts from a test set.

    4. Adjudication Method:

    This information is not present as there was no clinical study described that would require an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document explicitly states: "The modifications to the family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator did not require clinical testing." Therefore, an MRMC comparative effectiveness study was not done.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This concept is not applicable to an anesthesia system. The Aespire Anesthesia System is a medical device operated by trained professionals, not a standalone algorithm. The document describes non-clinical testing for the device's functionality.

    7. Type of Ground Truth Used:

    The document does not describe any clinical ground truth (e.g., pathology, outcomes data) for performance evaluation, as no clinical testing was deemed necessary for the modifications. The "ground truth" for the device's functionality would be its adherence to engineering specifications and safety standards through nonclinical testing.

    8. Sample Size for the Training Set:

    The concept of a "training set" is not applicable here as this is a hardware/software medical device, not a machine learning algorithm being trained on data in the traditional sense. The software validation is mentioned, but "training set" doesn't fit the context.

    9. How the Ground Truth for the Training Set Was Established:

    This question is not applicable for the reasons stated above.

    Summary of the Study that Proves the Device Meets Acceptance Criteria (as described in the document):

    The document highlights that the GE Datex-Ohmeda Aespire View Anesthesia System underwent "thoroughly tested through verification of specifications and validation, including software validation." Additionally, biocompatibility testing was conducted for the total flow sensor module, and electrical safety and electromagnetic compatibility testing were completed.

    The conclusion is that based on these nonclinical tests and the comparison to legally marketed predicate devices, there are "no new questions of safety and effectiveness for the introduction of the Aespire View". This indicates that the device met its internal design and safety specifications, and its performance was considered to be equivalent to existing, cleared devices. This represents a substantial equivalence claim under 510(k) regulations, rather than a de novo clinical study proving novel acceptance criteria.