510(k) Data Aggregation

    K Number
    K042015

    Validate with FDA (Live)

    Date Cleared
    2004-09-29

    (64 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bayer ADVIA® 1650 Glucose Hexokinase II assay is an in vitro diagnosic device for use in the quantitative determination of glucose in human cerebrospinal fluid (CSF), serum, plasma (lithium heparin), and urine on the ADVIA® 1650 Chemistry system. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia and insulin overdose.

    The Bayer ADVIA® 1650 Glucose Oxidase assay is an in vitro diagnostic device for use in the quantitative determination of glucose in human cerebrospinal fluid (CSF), serum, plasma and urine on the ADVIA® 1650 Chemistry system. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia and insulin overdose.

    Device Description

    Not Found

    AI/ML Overview

    Acceptance Criteria and Study for Bayer ADVIA® 1650 Glucose Assays

    This document describes the acceptance criteria and the studies performed to demonstrate the performance of the Bayer ADVIA® 1650 Glucose Hexokinase II and Glucose Oxidase assays.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document focuses on demonstrating substantial equivalence to a predicate device rather than defining explicit acceptance criteria with specific numerical thresholds for accuracy, sensitivity, or specificity. Instead, the performance is evaluated through imprecision and method correlation studies against a legally marketed predicate device (Roche Glucose Hexokinase). The "acceptance criteria" can be inferred from the successful outcome of these studies, demonstrating comparable performance to the predicate.

    Performance MetricAcceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (ADVIA® 1650 Glucose Oxidase)Reported Device Performance (ADVIA® 1650 Glucose Hexokinase II)
    Imprecision (Within-run CV%)Comparable to Roche Glucose Hexokinase (e.g., ≤ 1.8% at similar levels)
    - Level 1≤ 1.6% (Roche at 31 mg/dL)1% (at 58.0 mg/dL)1.2% (at 60.2 mg/dL)
    - Level 2≤ 1.8% (Roche at 59 mg/dL)1.1% (at 34.2 mg/dL)1.7% (at 36.9 mg/dL)
    Method CorrelationStrong correlation with predicate device (high r-value, regression close to Y=X)
    - Regression Equation(Implicitly close to Y=X)$Y = 0.97x + 4.47$$Y = 1.03x - 1.25$
    - Syx (Standard Error of Y on X)(Implicitly low)7.66.4
    - Correlation Coefficient (r)(Implicitly high, e.g., > 0.95)0.9940.987
    Analytical RangeSufficient for intended use0-750 mg/dL0-700 mg/dL

    2. Sample Size and Data Provenance for Test Set

    • Glucose Oxidase:
      • Sample Size: N = 56
      • Data Provenance: Not explicitly stated, but the comparison is made against a "Roche (On Integra)" system, implying clinical samples were tested on both systems. The document does not specify the country of origin, nor whether the samples were retrospective or prospective.
    • Glucose Hexokinase:
      • Sample Size: N = 55
      • Data Provenance: Not explicitly stated, but the comparison is made against a "Roche (On Integra)" system, implying clinical samples were tested on both systems. The document does not specify the country of origin, nor whether the samples were retrospective or prospective.
    • Imprecision Study:
      • The sample sizes for the imprecision study are not explicitly stated as 'N' for each level, but typical laboratory imprecision studies involve multiple replicates (e.g., 20 or more) over several days for each level. The reported "Within-run CV(%)" suggests a specific set of measurements, but the exact number of samples or runs is not detailed.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. This device is a diagnostic assay for quantitative determination of glucose. The "ground truth" is established by the reference measurement on the predicate device itself, which is a chemical analyzer. No human experts are involved in establishing the "ground truth" for the test results.

    4. Adjudication Method for the Test Set

    Not applicable. Since the device performs a quantitative chemical analysis, the direct output of the instrument (glucose concentration) is used for comparison. There is no qualitative assessment or interpretation by observers requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an imaging device or a subjective interpretation device where human readers are involved. It is a quantitative chemical analyzer. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed.

    6. Standalone Performance Study

    Yes, standalone performance was done. The studies (imprecision and method correlation) directly assess the performance of the Bayer ADVIA® 1650 Glucose assays in isolation, by comparing its results to a predicate device. This is inherently a standalone performance evaluation of the algorithm/assay without human intervention in the result determination.

    7. Type of Ground Truth Used

    The ground truth for the method correlation studies was established by measurement on a legally marketed predicate device (Roche Glucose Hexokinase on an Integra system). For the imprecision study, the "ground truth" for the mean glucose level was the determined mean value from the measurements themselves, evaluated against the variability of those measurements.

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" or "training set sample size." This is typical for a traditional in vitro diagnostic device like a glucose assay evaluated through method comparison and imprecision studies. These types of devices do not typically undergo a development process involving distinct "training" and "test" sets in the way AI/Machine Learning models do. The performance characteristics are derived from direct analytical testing.

    9. How Ground Truth for the Training Set was Established

    Not applicable, as a distinct training set (in the context of AI/machine learning) is not described or implied for this type of medical device submission. The analytical performance is characterized through direct experimental measurement.