Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K090822

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-04-27

    (32 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The Advance® 14Rx Rapid Exchange PTA Balloon Dilatation Catheter is a rapid exchange catheter indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The device with 4.0 French nylon tubing compatible with a 0.014-inch guidewire. It will be supplied sterile, intended for one-time use.

    AI/ML Overview

    The provided document describes the Advance® 14Rx Rapid Exchange PTA Balloon Dilatation Catheter, a medical device used for percutaneous transluminal angioplasty (PTA). This submission is a 510(k) Premarket Notification, which aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device.

    The document does not contain information about acceptance criteria or a study proving that the device meets those criteria in the context of AI/machine learning performance. Instead, it details the safety and performance testing of a physical medical device. Therefore, many of the requested bullet points, which are relevant to AI/ML device studies, cannot be answered from the provided text.

    Here's a breakdown of what can and cannot be answered:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Physical Performance TestsThe device was subjected to the following tests:
    Tensile strengthResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    Balloon inflation/deflation reliabilityResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    Balloon burst pressureResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    Balloon complianceResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    Balloon fatigue resistanceResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    Bond strengthResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    Biological/Sterility Tests
    BiocompatibilityResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    SterilityResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    Durability/Stability Tests
    Shelf lifeResults provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.
    In Vivo Testing
    Animal test (performance in biological system)Results provide reasonable assurance of conformance for use as a PTA dilatation balloon catheter.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The tests performed are engineering, material, and biological tests for a physical device, not an AI model.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the document describes the testing of a physical medical device, not an AI/ML diagnostic or prognostic system requiring expert-established ground truth from medical images or clinical data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for the same reasons as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the document describes the testing of a physical medical device, not an AI assistance system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the document describes the testing of a physical medical device, not a standalone algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the physical device, the "ground truth" would be the engineering specifications and performance standards established for such devices (e.g., specific tensile strength values, burst pressures, biocompatibility standards). The document states that the tests provide "reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a PTA dilatation balloon catheter," implying compliance with these established standards. However, the specific quantitative "ground truth" values are not explicitly listed.

    8. The sample size for the training set

    This information is not applicable as there is no mention of an AI/ML training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1