510(k) Data Aggregation

    K Number
    K071104

    Validate with FDA (Live)

    Date Cleared
    2008-01-14

    (270 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Flier's Quality Water Systems, Inc. systems are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001. They remove organic and inorganic substances and microbial contaminants from water used to dilute dialysis concentrate to form dialysate, in the reprocessing of hemodialyzers, and equipment rinse and disinfection.

    The Acute System #1 consists of two each carbon and mixed bed DI tanks with pre & post filtration. It is intended to be placed on the back of dialysis machine or on a handcart for single patient dialysis. Acute System #2 consists of just carbon tanks and is intended to be used as pretreatment to a single patient RO system.

    The Central System #1 and #2 consists of two or more carbon and mixed bed D1 tanks with pre & post filtration. They are used as either a temporary system to feed multiple dialysis machines in normal use, or for a system to feed a biomedical technician service station. For permanent, multiple station dialysis clinics Flier's recommends using central water treatment systems including reverse osmosis.

    The Back-up System #1 and #2 consists of two or more mixed bed DI tanks as a component in a "back up" condition when a RO system or major component on a water purification system for hemodialysis is in need of repair. It may also be used in DI polishing mode to remove ions from an RO stream to meet AAMI RD62:2001 requirements.

    Upon exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or with new resin altogether.

    These systems are components of a larger water treatment system employing adequate pre-treatment and post-treatment sections. Flier's systems are not to be used alone.

    Device Description

    The systems are primary or temporary devices used to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001.

    The Flier's Quality Water Systems, Inc. systems purify potable feed water through deionization. Deionization is used to remove 99.9% of ions from water. Deionization alone does not remove particulates, organics, bacteria, viruses or endotoxins. The systems require pretreatment and post-treatment. Deionizers are designed in a hemodialysis application to treat RO water or carbon filtered potable water. Improper use can result in the formation of nitrosamines in the effluent of the deionizer.

    The purpose of the pretreatment section of the system is to condition that feed water supplying the deionizers. Conditioning the feed water will include: cartridge filters to reduce particulates and sediment and carbon filtration tarks to remove chlorine/chloramine residual.

    The purpose of the post-treatment section of the system is to remove bacteria and endotoxins or lower them to acceptable levels as required by AAMI standards. The post-treatment section of the system will include: submicron/ultrafilters after the deionizers, sample ports to check for microbial contamination and to test for comprehensive AAMI analysis, a 1 megohm quality control light at mid tank and a temperature compensated audible and visual resistivity alarm set at 1M ohm or greater for final water quality. A remote alarm will be installed if the deionization system is not located in the patient treatment area. A distribution section of the system is necessary to deliver product water that meets ANSI/AAMI standards to the points of use. During use, continual monitoring of the effluent through the use of the temperature compensated audible and visual resistivity alarm is required. Routine monitoring of the mid tank quality indicator is also required on intervals during the day, at least before and after each patient treatment.

    Should final water quality fall at or below 1 megohm, the alarm will sound and all patient treatments must be discontinued. The deionization tanks must be replaced per tank exchange procedure prior to resuming and patient treatment.

    Divert to drain option will be provided when necessary per AAMI quidelines. Divert to drain is not supplied for acute systems located in patient room. Monitor and divert to drain not supplied for back-up systems with existing monitor/alarm/divert.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Flier's Quality Water Systems, Inc. line of deionization systems for hemodialysis. It is a submission to the FDA seeking clearance for these medical devices. Unlike a research paper that presents a study to prove a device meets acceptance criteria, this document declares substantial equivalence to a predicate device already on the market. Therefore, the information requested in your prompt regarding acceptance criteria and a study proving performance is largely not present in this type of regulatory submission in the way you're asking.

    Here's an analysis based on the provided text, explaining why certain information is not available and what can be inferred:


    Acceptance Criteria and Study for Flier's Quality Water Systems, Inc. (K071104)

    This 510(k) submission primarily focuses on demonstrating substantial equivalence to a predicate device (USFilter Corporation's/Siemens, K980182) rather than presenting a standalone study with defined acceptance criteria and performance metrics. Therefore, explicit tables of acceptance criteria for the new device and a study proving it meets those criteria are not provided in this document. Instead, the "acceptance criteria" are implicitly met by demonstrating that the new device functions in a manner substantially equivalent to the predicate device, thereby fulfilling the requirements for water purification in hemodialysis as specified by ANSI/AAMI RD62:2001.

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria (Inferred): The primary "acceptance criteria" for the Flier's systems are to achieve water purity levels suitable for hemodialysis applications, specifically by removing 99.9% of ions and effectively reducing particulates, chlorine/chloramine, bacteria, viruses, and endotoxins to acceptable ANSI/AAMI RD62:2001 standards. This is achieved through deionization combined with pre- and post-treatment.
      • Specific criteria mentioned:
        • Remove 99.9% of ions.
        • Remove chlorine/chloramine residual (minimum empty bed contact time of 10 minutes, activated carbon with Iodine number of 900 or greater).
        • Submicron/ultrafilters after deionizers to remove bacteria and endotoxins or lower them to acceptable AAMI levels.
        • Temperature compensated audible and visual resistivity alarm set at 1M ohm or greater for final water quality.
        • Mid-tank quality indicator for routine monitoring.
        • Divert to drain option when necessary.
    • Reported Device Performance (Inferred): The document states that "The Flier's Quality Water Systems, Inc. systems purify potable feed water through deionization. Deionization is used to remove 99.9% of ions from water." It also highlights that "The systems require pretreatment and post-treatment" to address other contaminants. The performance is essentially claimed to be equivalent to the predicate device, which also uses carbon and deionization with pre- and post-filtration for water purification in hemodialysis. The conclusion explicitly states: "The core water purification components and technology are exactly the same. The predicate device components utilize the exact same water purification principles."
    Acceptance Criteria (Inferred from Predicate Equivalence & AAMI Standards)Reported Device Performance (Inferred/Stated)
    Removal of 99.9% of ions from water"Deionization is used to remove 99.9% of ions from water."
    Removal of chlorine and chloramines to AAMI standardsCarbon filtration utilized, two carbon filters in series, minimum 10 min EBCT, Iodine number ≥ 900
    Reduction of bacteria and endotoxins to AAMI standardsPost-treatment includes submicron/ultrafilters.
    Final water quality monitoring and alarm at ≥ 1 M OhmTemperature compensated audible and visual resistivity alarm set at 1M ohm or greater.
    Mid-tank quality indicator for routine monitoring1 megohm quality control light at mid-tank.
    Compliance with ANSI/AAMI RD62:2001Systems "are primary or temporary devices used to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This document is a 510(k) premarket notification for substantial equivalence, not a clinical or performance study report. There is no "test set" in the context of clinical trial data or a formal performance evaluation with a specified sample size. The claim of equivalence is based on the technical characteristics and operating principles being the same as a legally marketed predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. As there is no "test set" used in the manner of a diagnostic device requiring ground truth establishment by experts, this information is not provided. The "ground truth" here is the established standards for water quality in hemodialysis (ANSI/AAMI RD62:2001) and the demonstrated performance of the predicate device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. This information is relevant to studies involving human interpretation or subjective assessments, which is not the nature of this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a water purification system, not a diagnostic imaging device or an AI-powered system designed to assist human readers. Therefore, an MRMC study is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device does not involve an algorithm or AI. It is a physical water purification system. Its performance is inherent in its engineering and components, designed to meet physical and chemical purity standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is the physicochemical requirements for water purity in hemodialysis as defined by established standards, specifically ANSI/AAMI RD62:2001. The performance of the predicate device (K980182) in meeting these standards serves as the benchmark for substantial equivalence.

    8. The sample size for the training set

    • Not Applicable. This device does not utilize machine learning or AI, and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no "training set," this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K042018

    Validate with FDA (Live)

    Date Cleared
    2005-04-15

    (262 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Systems are intended for use with a hemodialysis system to remove organic and inorganic substances resulting in contaminants from water used to dilute dialysis concentrate to form dialysate, reprocessing of hemodialyzers, equipment rinse and disinfection.

    The Acute PEDI System consists of a sediment filter, carbon filtration, mixed bed deionizers, and system monitoring with remote alarm functions. It is intended for use in a location with a portable dialysis machine for the dialysis of a single patient.

    The Central PEDI System consists of a booster pump / pressure tank assembly, sediment filter, carbon filtration, mixed bed deionizers, resin trap filter, endotoxin filter and system monitoring with flow isolation/divert and remote alarm functions. It is intended for use in a central location for the dialysis of multiple patients.

    The Back-up PEDI System consists of mixed bed deionizers, intermediate water quality (QC) indicator (light) and connection hoses. It is intended for use to provide a "back-up" for reverse osmosis filtration and system monitoring with flow isolation / divert and remote alarm functions are provided when required to meet ANSI/AAMI RD62:2001 requirements.

    Device Description

    The PEDI systems are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001.

    The systems consist of a booster pump / pressure tank assembly, sediment filter, carbon filtration, mixed bed deionizers, resin trap filter, endotoxin filter and system monitoring with flow isolation/divert and remote alarm functions. Refer to Operating Manuals for flow diagrams.

    The booster pump / pressure tank assembly (central system only) consists of a multistage pump, and pressure tank. The multistage pump construction has minimum 304 SST wetted parts with no copper or iron wetted parts. The pressure (bladder) tank construction is an airtight shell inside of coil rolled steel with a 1 27 approved butyl rubber diaphragm to separate water from tank inside. The assembly requires 115V/1P/60Hz power to operate a multistage pump to raise pressure up to 80 psig.

    The sediment filter consists of a polypropylene housing with pressure relief valve and a replaceable filter cartridge. The filter cartridge consists of spun polypropylene fiber wound on a polypropylene core and with a filtration rating of 10 microns. (Back-up system uses existing sediment filtration).

    The carbon filtration consists of ABS lined fiberglass pressure tank containing granular activated carbon and PVC distributors. The size and number of tanks is dependent on flowrate required with carbon and PVC distributors listed in Appendix C. Back-up system uses facilities existing carbon filters if available, Acute system single tank.

    The mixed bed deionizers consist of ABS lined fiberglass pressure tank with PVC distributors containing mixed bed ion exchange resin. The size and number of tanks is dependent on flowrate required and inlet water conductivity maximum values listed in Operating Manual.

    The resin trap filter consists of a polypropylene housing with pressure relief valve and a replaceable filter cartridge. The filter cartridge consists of spun polypropylene fiber wound on a polypropylene core and has a filtration rating of 5 microns. (Not used on acute systems).

    The endotoxin filter consists of a polypropylene housing with a replaceable filter cartridge. The filter cartridge consists of an element with 222 o-ring/closed type seals and has a reduction factor of 10' for bacteria and 10 for endotoxins.

    System monitoring consists of pressure gauges, quality control indicator and resistivity monitor. Pressure gauges prior to deionization minimum brass body with phosphorous bellows, gauge after deionization 304 SST stem with SST bellows. The resistivity monitor has a digital readout, indicators for above and below water quality, push buttons for alarm setpoint indication and alarm function test. Connections to the monitor provide under low water quality: remote alarm (horn), isolation of flow to use (valve) and diversion of flow to drain (valve). Inline sensor construction stainless element with polypropylene body. Divert to drain not supplied for acute systems located in patient room. Monitor and divert to drain not supplied for back-up systems with existing monitor/alarm/divert. Refer to Operating Manuals for Control Diagram.

    AI/ML Overview

    This 510(k) summary (K042018) does not describe the acceptance criteria and the study that proves the device meets the acceptance criteria. It focuses on establishing substantial equivalence with a predicate device and describing the device's components and indications for use.

    Therefore, I cannot provide the requested information from the provided text. The document does not contain:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes used for test sets or their data provenance.
    3. Number of experts used to establish ground truth or their qualifications.
    4. Adjudication method for test sets.
    5. Information on a multi-reader multi-case (MRMC) comparative effectiveness study or effect sizes.
    6. Details on a standalone algorithm performance study.
    7. The type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    The document is a regulatory submission for a water purification system for hemodialysis, which typically involves performance testing against industry standards (e.g., ANSI/AAMI RD62:2001 for water quality parameters) rather than clinical studies with expert readers or ground truth establishment in the way AI/imaging devices might. The "performance" in this context would likely refer to water quality parameters (e.g., conductivity, removal of specific contaminants) which are not detailed as specific acceptance criteria or study results here beyond general compliance with RD62:2001.