510(k) Data Aggregation

    K Number
    K130881

    Validate with FDA (Live)

    Date Cleared
    2013-10-11

    (196 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S2000 and S3000 ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

    The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement: Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

    The Acuson Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

    Device Description

    The S2000 and S3000 Ultrasound Systems are multi-purpose mobile, software controlled diagnostic ultrasound systems with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or Harmonic Imaging and 3D/4D Imaging on a Flat Panel Display.

    AI/ML Overview

    The provided document is a 510(k) summary for the Acuson S2000 and S3000 Diagnostic Ultrasound Systems, focusing on their Virtual Touch IQ (VTIQ) shearwave elasticity imaging software. The document primarily establishes substantial equivalence to predicate devices and outlines the device's technical characteristics and intended uses.

    Here's an analysis of the requested information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" in terms of performance metrics (e.g., sensitivity, specificity, accuracy) for the VTIQ feature. Instead, it argues for substantial equivalence based on technological characteristics and a comparison with the predicate device (Supersonic Imagine Aixplorer).

    Feature / CharacteristicPredicate Device (Aixplorer)Submission Device (Siemens VTIQ)Comments / Performance
    Shear Wave Elasticity Imaging SoftwareAixplorerVTIQComparable exam types and functionality. Siemens is requesting clearance for point measurements in addition to displaying shear wave velocity on the color bar. Siemens VTIQ is stated to achieve higher signal to noise ratio and more consistent shearwave velocity estimation based on in-vivo studies.
    Comparable Exam TypesBreast, ThyroidBreast, Thyroid-
    Transducer for same Exam Types10L2, 15L4 Linear Array9L4 Multi-D Linear Array-
    Operating ModeLow Frame Rate Real TimeSingle FrameVTIQ's single-frame acquisition allows for higher SNR without image compounding, contrasted with the predicate's lower frame rate and need for compounding.
    Cool Down Period before Live Imaging ResumesNoYes (variable, ~3-6 sec)A difference noted, but not tied to specific acceptance criteria.
    Push Pulse Beam DensityLow (variable with FOV size)High (variable with FOV size)Explained as a factor contributing to higher SNR in VTIQ.
    Multiple Push Pulse Focal Zones"Mach Cone"YesBoth use for increased radiation force uniformity.
    Two Dimensional Display with Elasticity Region of InterestYesYes-
    Localized Quantitative Shear Wave Velocity MeasurementYes - Predicate not cleared in U.S.A.YesSiemens VTIQ requests clearance for this.
    Localized Quantitative Young's Modulus MeasurementYes - Predicate not cleared in U.S.A.NoSiemens chose not to include Young's Modulus due to assumptions needed (tissue density, viscosity) that may not be correct, preferring direct shear wave velocity.
    Shear Wave Velocity Display ModeYesYes-
    Shear Wave Quality Display ModeNoYesAssists user in interpreting possible shear wave velocity estimate artifacts.
    Shear Wave Travel Time Display ModeNoYesImproves dynamic range of shear wave image in focal regions of high shear wave velocity.
    Shear Wave Displacement Display ModeNoYesProvides information on relative shear wave amplitudes, useful for identifying lesion boundaries and correlating with attenuation.
    Shear Wave Propagation "Movie" (clip)YesNoSiemens uses its own display modes (Quality, Time, Displacement) to visualize shear wave characteristics.
    Shear Wave Velocity Measurement Range0.1-10 m/sec0.5-10 m/secSiemens opted for a higher lower cutoff (0.5 m/s) to avoid potential artifacts, stating that velocities below this are not encountered in vivo.
    "See through to B mode" when no shear wave detectedYesYes-
    Color Coded Shear Wave Velocity DisplayYesYes-
    Adjustable Maximum and Minimum Velocity ScaleYesYes-
    In Color Code. Red is 'stiff' and blue is 'soft'YesYes-
    Color map is transparent over B mode displayYesYes-

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly mention a formalized "test set" with a specified sample size for validating the VTIQ feature against acceptance criteria. It mentions:

    • "Using VTIQ and Supersonic Aixplorer in the same lesions on the same subjects, we have encountered many cases where maximum shearwave velocity with VTIQ was higher and more completely filled in with shearwave information throughout the lesion than with the predicate device in biopsy proven breast cancers."
    • "25 example cases submitted (refer to tab 1.7.5.10 D)."

    This suggests some form of comparative evaluation on a limited set of cases (at least 25), but details regarding the design of this study (e.g., sample size, prospective/retrospective, country of origin) are missing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document mentions "biopsy proven breast cancers" as a reference for comparison in the aforementioned "many cases." This implies pathology as ground truth. However, the exact number of experts (e.g., pathologists) involved in establishing this ground truth for the "25 example cases" or more generally for the "many cases" is not specified, nor are their specific qualifications.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set or the "many cases" mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance

    The document does not mention a Multi Reader Multi Case (MRMC) comparative effectiveness study. It describes a comparison between the VTIQ device and a predicate device, focusing on technical performance differences rather than human reader performance with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The document implies a standalone comparison of the VTIQ algorithm's output (shear wave velocity, image quality) against the predicate device's output. The statement, "Using VTIQ and Supersonic Aixplorer in the same lesions on the same subjects, we have encountered many cases where maximum shearwave velocity with VTIQ was higher and more completely filled in with shearwave information throughout the lesion than with the predicate device in biopsy proven breast cancers," highlights an algorithm-only comparison of the imaging capabilities and derived measurements.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The primary ground truth mentioned that is relevant to disease state is histopathological diagnosis ("biopsy proven breast cancers").

    8. The Sample Size for the Training Set

    The document does not provide information on a specific training set or its sample size. This filing is a 510(k) for an ultrasound system with new software features, implying that the VTIQ algorithm was likely developed internally and possibly trained on proprietary data which is not described in this summary.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set information is provided, how its ground truth was established is also not available in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130739

    Validate with FDA (Live)

    Date Cleared
    2013-06-06

    (80 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S2000 and S3000 ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

    The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transrectal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculoskeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

    The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the .* ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

    The Acuson Acunav Ultrasound Catheter is intended for intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

    Virtual Touch Imaging and eSie Touch Elastography provide qualization of relative stiffness between tissues with different elastic properties.

    Device Description

    The S2000 and S3000 Ultrasound Systems are multi-purpose mobile, software controlled diagnostic ultrasound systems with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or Harmonic Imaging and 3D/4D Imaging on a Flat Panel Display. It is substantially equivalent to the S2000 (K111674) and S3000 system (K121138) which are legally marketed devices.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the Siemens Medical Solutions Acuson S2000 and S3000 Diagnostic Ultrasound Systems:

    The document is a 510(k) Summary, which primarily focuses on demonstrating substantial equivalence to predicate devices, rather than establishing de novo performance criteria or conducting extensive clinical trials for new applications. As such, the information provided is geared towards proving that the new device performs as well as or similarly to existing cleared devices.

    Based on the provided text, the key "device" being discussed is the Acuson S2000 and S3000 Diagnostic Ultrasound Systems, including various transducers and software features like Virtual Touch Imaging (VTI) and eSie Touch Elastography.


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: This document does not explicitly state quantitative acceptance criteria in the way one might expect for a new performance claim (e.g., "sensitivity must be >X%"). Instead, the primary "acceptance criterion" demonstrated throughout this 510(k) submission is substantial equivalence to predicate devices. The "reported device performance" is essentially that the new device operates similarly and safely.

    The document highlights the features and indications for use as being equivalent to previously cleared devices.

    Acceptance Criterion (Implicit: Substantial Equivalence via Feature/Performance Comparison)Reported Device Performance (as demonstrated by comparison)
    Indications for Use (General Ultrasound Applications)Identical to predicate devices (Acuson S2000 K111674, S3000 K121138, Acuson Antares K063138, Acuson S3000 K122825). Covers Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
    Technological Characteristics (Modes, Frequencies, Features)Confirmed to be consistent with or improved from predicate devices. Specifically, modes (B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined) and a range of center frequencies (2.0 MHz to 10.0 MHz) are available, matching or expanding on predicates. Features like Native™ tissue harmonic imaging, SieScape™ panoramic imaging, 3-Scape™ real-time 3D, TEQ™ ultrasound technology, etc., are also present.
    Safety (Acoustic Output, Biocompatibility, Electrical, Thermal, Mechanical, EMC)Device evaluated and found to conform with applicable medical device safety standards (e.g., UL 60601-1, IEC 60601-2-37, CSA C22.2 No. 601-1, AIUM/NEMA UD-3, UD-2, 93/42/EEC, EN/IEC 60601-1-1/2, IEC 1157, ISO 10993-1). Cleared patient contact materials, electrical, and mechanical safety are unchanged.
    Performance for Elastography (Virtual Touch Imaging - VTI)"Additional testing was performed to compare EI to VTI by validating intra-operator and inter-operator reproducibility of VTI using EI as a benchmark. Visual comparison of phantom images generated with EI and VTI demonstrate more consistent contrast to noise ratio and clearer inclusion visualization with VTI than with EI." (This implies a qualitative improvement, not a quantitative acceptance metric.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a number of cases or patients. The document refers to "phantom images" for the elastography comparison.
    • Data Provenance: Not specified for the phantom study (e.g., no country of origin, no mention of retrospective or prospective data collection from human subjects). It's a non-clinical test.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: Not applicable, as expert involvement for establishing ground truth on a test set (clinical data) is not mentioned. The primary comparison for elastography was qualitative visual comparison of phantom images.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. No mention of multiple readers or an adjudication process for clinical ground truth. The comparison for elastography (EI vs VTI) was described as "visual comparison of phantom images."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done?: No. The document states: "The S2000 and S3000 use the same technology and principles as existing devices, clinical data is not required." This explicitly indicates that a comparative effectiveness study involving human readers and clinical data was not performed or submitted for this 510(k).
    • Effect Size of Human Readers with vs. without AI Assistance: Not applicable, as no MRMC study was conducted.

    6. Standalone (Algorithm Only) Performance

    • Standalone Performance Done?: Not in the typical sense of a quantitative diagnostic system evaluated against clinical ground truth. The "additional testing" for Virtual Touch Imaging (VTI) involved a "visual comparison of phantom images generated with EI and VTI." This is a qualitative assessment of the image quality and reproducibility on a phantom, not a standalone diagnostic performance metric on a clinical dataset. The system itself is a diagnostic ultrasound, implying direct human operation rather than a standalone algorithm providing a diagnostic output independently.

    7. Type of Ground Truth Used

    • Type of Ground Truth:
      • For the overall device, the implicit ground truth is substantial equivalence to legally marketed predicate devices. This means the new device is compared to existing devices known to be safe and effective.
      • For the specific feature of Virtual Touch Imaging (VTI), the "ground truth" for comparison was the Established Imaging (EI) technique on phantoms, and the assessment was "visual comparison" for "more consistent contrast to noise ratio and clearer inclusion visualization." This isn't a pathological or outcome-based ground truth, but a comparative image quality assessment on a controlled phantom.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This document describes a medical device undergoing premarket notification (510(k)), not an AI/ML algorithm development with distinct training and test sets as typically understood. The device is an ultrasound system with advanced imaging features.

    9. How Ground Truth for the Training Set Was Established

    • How Ground Truth Was Established: Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device