510(k) Data Aggregation

    K Number
    K042044

    Validate with FDA (Live)

    Date Cleared
    2004-08-13

    (14 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diagnostic imaging or fluid flow analysis of the human body as follows: Fetal, Abdominal, Intraoperative, Intraoperative Neurological, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Cardiac, Transesophageal, Peripheral vessel, Laparoscopic, Musculo-skeletal Conventional, Musculo-skeletal Superficial applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

    Device Description

    The CV70 system is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to our current product that is already proprietary boxware, under the following 510(k) Premarket Notification number: K032111 (July 18, 2003) cleared as ACUSON CV70™ Cardiovascular System. The CV70 functions in the same manner as other diagnostic ultrasound systems, in that they transmit ultrasonic energy into the body via a transducer. In the body, acoustic impedance of different tissues reflect different amounts of ultrasound energy back to the transducer, where post processing of received echoes is performed to generate two-dimensional images of structures and fluid flow within the body. Doppler principles are used to process reflected ultrasound energy to display moving blood as a spectrum, or as color-coded two-dimensional images. The predicate device provides specialized measurements of structures and flow, and provide various calculations functions.

    AI/ML Overview

    This document is a 510(k) submission for the ACUSON CV70™ Cardiovascular System, an ultrasound device. It primarily focuses on demonstrating substantial equivalence to a previously cleared device (K032111) and detailing its intended uses and compliance with safety standards.

    Based on the provided text, there is NO information about acceptance criteria, study design, sample sizes for test or training sets, ground truth establishment, expert qualifications, adjudication methods, or multi-reader multi-case (MRMC) comparative effectiveness studies regarding the device's performance.

    The document is not a study report demonstrating the device meets performance criteria in the way described in your request (i.e., evaluation of AI/algorithm performance against a defined ground truth). Instead, it's a regulatory submission affirming that the updated model of an existing device is substantially equivalent to its predicate.

    However, I can extract the intended uses of the device and its various transducers as a proxy for what the device is designed to do, which effectively serves as the "acceptance criteria" for its functionality as an ultrasound system in a regulatory context when demonstrating substantial equivalence.


    Summary of Acceptance Criteria and Reported Device Performance (as demonstrated by intended uses and substantial equivalence):

    The "acceptance criteria" in this context are the previously cleared intended uses and performance of the predicate device (ACUSON CV70™ Cardiovascular System, K032111). The reported device performance is that the new ACUSON CV70™ Cardiovascular System is substantially equivalent and performs in the same manner as the predicate, transmitting and processing ultrasonic energy to display real-time images and perform measurements.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a 510(k) submission for a diagnostic ultrasound system, the "acceptance criteria" are defined by the cleared Intended Uses and modes of operation for each transducer, which are considered substantially equivalent to the predicate device. The "reported device performance" is implicitly that the device successfully performs these functions as intended and is as safe and effective as the predicate.

    Below is a combined table summarizing the overarching intended uses and modes of operation for the ACUSON CV70™ Cardiovascular System across its various transducers. "P" indicates previously cleared indications, and "E" indicates new indications added under Appendix E for a specific transducer. "N" indicates new indications for a specific transducer. "BMDC" stands for B-mode, M-mode, Doppler (PWD-Pulsed Wave, CWD-Continuous Wave), and Color Doppler.

    Clinical ApplicationMode of Operation (P/N/E)Other Features (P/N/E)
    FetalB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, B&W SieScape, Power SieScape
    AbdominalB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, B&W SieScape, Power SieScape
    Intraoperative (Abdominal, Neurological)B, M, PWD, Color, Amplitude Doppler3D Imaging
    PediatricB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, B&W SieScape, Power SieScape
    Small OrganB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, B&W SieScape, Power SieScape
    Neonatal CephalicB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, B&W SieScape, Power SieScape
    Adult CephalicB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging
    CardiacB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, Contrast Agent Imaging, B&W SieScape, Power SieScape
    TransesophagealB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging
    Peripheral VesselB, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, B&W SieScape, Power SieScape
    LaparoscopicB, M, PWD, Color, Amplitude Doppler3D Imaging, B&W SieScape, Power SieScape
    Musculo-skeletal (Conventional & Superficial)B, M, PWD, CWD, Color, Amplitude DopplerEnsemble Tissue Harmonic Imaging, 3D Imaging, B&W SieScape, Power SieScape

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not applicable. This document does not describe a clinical study with a "test set" in the context of an algorithm or AI performance evaluation. The submission demonstrates substantial equivalence to a predicate device, relying on technical comparison, safety standards compliance, and previous clearances.
    • Data Provenance: Not applicable. There is no mention of specific clinical data or patient data used for performance testing in this 510(k) summary. The comparison is primarily against the established performance of the legally marketed predicate device (K032111) and compliance with recognized standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not applicable. No specific "ground truth" establishment by experts for a test set is described in this regulatory submission.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method for the test set:

    • Adjudication Method: Not applicable. No test set requiring expert adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This document does not present any MRMC comparative effectiveness studies. The device is a diagnostic ultrasound system, not an AI-assisted interpretation tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Standalone Performance Study: No. This document does not describe a standalone algorithm performance study.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: Not applicable. The submission for substantial equivalence relies on the established performance and safety profile of the predicate device, not on a new ground truth established for a specific clinical outcome or diagnostic accuracy study.

    8. The sample size for the training set:

    • Sample Size for Training Set: Not applicable. This document does not describe the development or training of an AI algorithm; therefore, there is no training set.

    9. How the ground truth for the training set was established:

    • Ground Truth Establishment for Training Set: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K032111

    Validate with FDA (Live)

    Date Cleared
    2003-07-18

    (9 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diagnostic imaging or fluid flow analysis of the human body as follows: Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

    Device Description

    The CV70 is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in: B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or Harmonic Imaging, or 3D imaging, on a CRT display.

    AI/ML Overview

    The provided text is a 510(k) summary for the ACUSON CV70 Cardiovascular System, a diagnostic ultrasound system. It primarily focuses on demonstrating substantial equivalence to predicate devices and detailing the intended uses and safety standards. It does not contain information about acceptance criteria or a study proving the device meets specific performance criteria through a clinical evaluation.

    Therefore, I cannot fulfill your request for the following information based on the provided text:

    1. A table of acceptance criteria and the reported device performance: This document does not specify quantitative acceptance criteria or detailed device performance metrics from a study.
    2. Sample size used for the test set and the data provenance: There is no mention of a test set or clinical study data. The document focuses on regulatory compliance and device description.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: As no test set or ground truth establishment is described, this information is not present.
    4. Adjudication method for the test set: No test set means no adjudication method is discussed.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: There is no indication of any MRMC study or AI assistance capability for this device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device is described as a diagnostic ultrasound system with various imaging modes, implying human operation, not a standalone algorithm.
    7. The type of ground truth used: No ground truth is mentioned.
    8. The sample size for the training set: There is no mention of a training set as no machine learning/AI component is described.
    9. How the ground truth for the training set was established: Not applicable, as no training set is mentioned.

    Summary of what the document does provide regarding the device validation:

    The document primarily establishes the device's substantial equivalence to previously cleared predicate devices (SONOLINE G50 & G60 S, and ACUSON Sequoia) by comparing its technological characteristics and intended uses. It also lists the product safety standards the device has been designed to meet (UL 2601-1, CSA C22.2 No. 601-1, AIUM/NEMA UD-3, AIUM/NEMA UD-2, 93/42/EEC Medical Devices Directive, EN/IEC 60601-1, EN/IEC 60601-1-1, EN/IEC 60601-1-2, IEC 1157, ISO 10993).

    The FDA's clearance letter requests a "postclearance special report" containing "complete information, including acoustic output measurements based on production line devices." This indicates that performance data (specifically acoustic output) would be collected post-clearance to confirm safety, but it's not a pre-market study establishing efficacy or meeting performance criteria in the way your questions imply.