Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210045

    Validate with FDA (Live)

    Date Cleared
    2021-09-22

    (257 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACTIVA PRESTO PACK is intended to be used by dental professionals as a dental filling material for pits, root surface cavities, and Class I, II, III, IV and V restorations include, but are not limited to, direct anterior and posterior restorations (including occlusal surfaces), core build-ups, splinting, and indirect restorations including crowns, inlays, onlays and veneers.

    Device Description

    ACTIVA PRESTO™ PACK is an aesthetic, light cure material that contains calcium, phosphate, and fluoride in a durable, wear-resistant resin matrix indicated for all restorative procedures. The material is packable, highly polishable and contains no Bisphenol A, no BisGMA and no BPA derivatives. It is available in various shades, has low solubility, low water sorption, high physical properties, and is bioactive as demonstrated within this 510(k) submission.

    AI/ML Overview

    The document provided is a 510(k) Premarket Notification from Pulpdent Corporation to the FDA for their device, ACTIVA PRESTO PACK. This submission focuses on demonstrating substantial equivalence to a predicate device, ACTIVA Presto (K153249), rather than presenting a study where a device's performance is measured against acceptance criteria in the format typically used for AI/ML devices.

    Therefore, many of the requested categories (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth for training) are not applicable to the information contained in this document, as it describes a physical dental material, not an AI/ML diagnostic or assistive device.

    However, I can extract the comparative performance data between the proposed device and its predicate, which serves a similar function to "acceptance criteria" in that the new device must perform comparably to the predicate.

    Here's the information extracted in a modified format to fit the request where applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, "acceptance criteria" are implied by the performance of the legally marketed predicate device. The new device "ACTIVA™ PRESTO™ PACK" is compared against "Pulpdent ACTIVA™ PRESTO™ (K153249)". Substantial equivalence is demonstrated by comparable physical properties and intended use.

    CharacteristicPredicate Device (ACTIVA™ Presto™ K153249)Proposed Device (ACTIVA™ PRESTO™ PACK)Result/Comparison
    ClassificationClass II Dental Device, CFR 872.3690, EBFClass II Dental Device, CFR 872.3690, EBFEquivalent
    Common NameTooth shade resin material; restorativeTooth shade resin material; restorativeEquivalent
    DescriptionAesthetic, light cure material, urethane-based resin, flowable, stackable, contains calcium, phosphate, fluoride, no Bisphenol A, no BisGMA, no BPA derivativesAesthetic, light cure material, urethane-based resin, packable, sculptable, contains calcium, phosphate, fluoride, no Bisphenol A, no BisGMA, no BPA derivativesSimilar, with difference in handling characteristic (flowable vs. packable)
    Intended UseLight-cure, flowable composite used by dental professionals as a restorative.Light-cure, packable composite used by dental professionals as a restorative.Similar, with difference in handling characteristic (flowable vs. packable)
    Indications for UseFilling material for pits, root surface cavities, Class I, II, III, IV and V restorations.Filling material for pits, root surface cavities, Class I, II, III, IV and V restorations. Includes direct anterior/posterior (occlusal), core build-ups, splinting, indirect restorations (crowns, inlays, onlays, veneers).Expanded clarity in indications but functionally equivalent.
    ContraindicationsNot for direct placement on exposed pulp.Not for direct placement on exposed pulp.Equivalent
    CompositionResins: Diurethane dimethacrylate, Bis(2-(Methacryloyloxy) Ethyl) Phosphate; Fillers: Barium glass, MCP; Photoinitiator: CamphorquinoneResins: Diurethane dimethacrylate, Bis(2-(Methacryloyloxy) Ethyl) Phosphate; Fillers: Barium glass, MCP; Photoinitiator: CamphorquinoneEquivalent
    TechnologyLight-cure polymerization, releases Ca, P, F ions.Light-cure polymerization, releases Ca, P, F ions.Equivalent
    Light Cure20 seconds in 2 mm increments20 ± 2 seconds in 2 mm incrementsEquivalent
    Dispensing SystemMulti-dose, push syringes (1ml, 3 ml) + applicator tips; Unit-dose capsulesMulti-dose, screw-type composite syringes; Unit-dose capsulesSimilar
    StandardsISO 14971, ISO 10993-1, ISO 7405, ISO 4049ISO 14971, ISO 10993-1, ISO 7405, ISO 4049Equivalent
    AppearancePaste, various tooth shadesPaste, various tooth shadesEquivalent
    OdorMild, characteristicMild, characteristicEquivalent
    Filler70.0%80.0%Different
    Resins30.0%20.0%Different
    Specific Gravity1.820 g/ml2.165 g/mlDifferent
    Depth of Cure2.74 mm2.42 mmSimilar
    Flexural Strength100.1 MPa105.6 MPaHigher (better)
    Water Sorption36.3 µg/mm³34.5 µg/mm³Lower (better)
    Solubility4.7 µg/mm³7.2 µg/mm³Higher (worse)
    Radio-opacity2.3 mm2.92 mm AlHigher (better)
    Flexural Modulus5.8 GPa7.4 ± 0.5 GPaHigher (better)
    Deflection at Break0.70 mm0.54 mmLower (worse)
    Compressive Strength327.4 MPa253 ± 25 MPaLower (worse)
    Diametral Tensile Strength61 MPa47.6 ± 5.7 MPaLower (worse)
    Polymerization Shrinkage2.1%2.1%Equivalent

    Study Conclusion: The submission concludes that ACTIVA™ PRESTO™ PACK is substantially equivalent to ACTIVA™ Presto™ (K153249) based on comparable design, composition, performance, and intended use. While some physical properties show variation, the overall performance profile is considered acceptable for substantial equivalence for a dental restorative material. The key difference noted is the change from a "flowable" to a "packable/sculptable" composite.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as the document describes a physical medical device (dental filling material), not software or an AI/ML diagnostic. The performance data presented are physical and chemical properties measured in a lab setting, not clinical data from a test set of patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. Ground truth, in the context of AI/ML, refers to definitively established diagnoses or classifications. For a dental material, "ground truth" would be established by standardized laboratory testing methods, not expert consensus on diagnostic images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are relevant for resolving discrepancies in expert interpretations of images or data, which is not the case for material property testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the device is a dental material, not an AI/ML assistive device for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable as the device is a dental material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance data in this submission is derived from standardized laboratory testing protocols as per international standards like ISO 4049:2019 for Polymer-based restorative materials. These tests define how to rigorously measure properties like flexural strength, depth of cure, water sorption, etc., ensuring reproducible and objective results.

    8. The sample size for the training set

    This information is not applicable as the device is a physical material, not an AI/ML device that undergoes "training."

    9. How the ground truth for the training set was established

    This information is not applicable as the device is a physical material, not an AI/ML device.