510(k) Data Aggregation

    K Number
    K981698

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-07-10

    (57 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chiron Diagnostics ACS:180 BR is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the Chiron Diagnostics ACS:180® Automated Chemiluminescence Systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.

    Device Description

    The Chiron Diagnostics ACS:180 BR assay is a fully automated, competitive, chemiluminescent assay. One reagent, designated Lite Reagent, is composed of a mouse monoclonal antibody specific for CA 27.29, labeled with acridinium ester. The antibody used in the assay, MAb B27.29, binds to a peptide epitope in the tandem repeat region of the MUC-1 gene product. The Solid Phase is composed of purified breast cancer antigen (CA 27.29) which is covalently coupled to paramagnetic particles (PMP). After onboard pretreatment, the patient serum sample is incubated with both reagents simultaneously for 7.5 minutes.

    The ACS: 180 system automatically performs the following steps:

    • dispenses sample and Pretreatment Reagent into a cuvette .
    • dispenses Lite Reagent and Solid Phase and incubates for 7.5 minutes u
    • separates, aspirates and washes the cuvettes
    • dispenses reagents which initiate the chemiluminescent reaction
    • = reports results

    An inverse relationship exists between the concentration of CA27.29 in a sample and the relative light units (RLU) detected by the system.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modified in vitro diagnostic device, the ACS:180 BR, which is an automated chemiluminescence system for detecting cancer antigen CA 27.29 in human serum.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" but rather focuses on demonstrating "substantial equivalence" to a predicate device. The primary performance metric reported is the correlation between the modified assay and the predicate device.

    Performance MetricAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device PerformanceComments
    Correlation ($r^2$)A high coefficient of determination, ideally close to 1, demonstrating a strong linear relationship with the predicate device.$>0.99$This indicates a very strong positive linear correlation between the modified and predicate assays.
    SlopeA slope close to 1, indicating that the modified assay measures values very similarly to the predicate device across the range.$1.0408$Close to 1, suggesting good agreement in the rate of change between the two methods.
    Y-interceptA Y-intercept close to 0, indicating minimal constant bias between the modified assay and the predicate device.$0.9907$ U/mLRelatively close to 0, suggesting a small positive bias but overall good agreement with the predicate.
    Assay RangeMust match the predicate device to ensure comparable applicability.$3.5$ U/mL to $450$ U/mLMatches the predicate device.
    Upper Limit of NormalMust match the predicate device, as it defines the clinical cutoff.$38.6$ U/mLMatches the predicate device, with a note that individual labs should determine their own reference ranges.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Over 875 specimens.
    • Data Provenance: The document does not specify the country of origin. It does not explicitly state whether the data was retrospective or prospective, but it implies a retrospective comparison of previously collected specimens, given the statement "These specimens had CA 27.29 levels that spanned the range from 3.68 U/mL to 420 U/mL."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The study described is a method comparison study between a modified device and a predicate device, not typically requiring expert-established "ground truth" in the same way an imaging or diagnostic AI device would. The "ground truth" here is the measurement provided by the existing, legally marketed predicate device. Therefore, no information is provided regarding expert qualifications or the number of experts used to establish ground truth in this context.

    4. Adjudication Method for the Test Set

    Not applicable. This was a direct comparison of measurements between two diagnostic assays, not a study requiring adjudication of interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This is a comparison between two automated diagnostic assays, not a study involving human readers or AI assistance for interpretation.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this was a standalone performance comparison study of the device (an automated chemiluminescence system) against its predicate device. The results reported are for the device (modified assay) performing on its own, without human intervention in the measurement process beyond standard laboratory procedures.

    7. The Type of Ground Truth Used

    The "ground truth" for this comparative study was the results obtained from the predicate device (the original ACS:180 BR assay). The objective was to show that the modified assay yielded results substantially equivalent to the established predicate.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI. This is a traditional in vitro diagnostic device modification, where performance is typically validated through comparison to a well-established method, not through machine learning training and testing paradigms. Therefore, information on a training set size is not applicable/provided.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set (in the machine learning sense) is described or implied for this type of device validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970695

    Validate with FDA (Live)

    Date Cleared
    1997-08-08

    (164 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Chiron Diagnostics ACS:180 BR is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum. The test is intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.

    Device Description

    The Chiron Diagnostics ACS:180 BR assay is a competitive, chemiluminescent assay intended for use on the Chiron Diagnostics Automated Chemiluminescent System (ACS). One reagent, designated Lite Reagent, is composed of a mouse monoclonal antibody specific for CA 27.29, labeled with acridinium ester. The Solid Phase is composed of purified breast cancer antigen covalently coupled to paramagnetic particles (PMP). The patient serum sample is incubated with both reagents simultaneously for 7.5 minutes. An inverse relationship exists between the concentration of CA 27.29 in a sample and the relative light units (RLU) detected by the ACS:180 system. Automatic dilution is available on the ACS:180® for the ACS:180 BR assay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Chiron Diagnostics ACS:180 BR device, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Analytical Sensitivity< 3.5 U/mL
    Sensitivity to Change in Disease Status73%
    Specificity to Change in Disease Status71%
    Positive Predictive Value to Change in Disease Status77%
    Negative Predictive Value to Change in Disease Status67%
    Upper Limit of Normal (ULN)38.6 U/mL
    Correlation with predicate device (r)0.96 (for all 203 specimens)
    Correlation with predicate device (r)0.96 (for 103 breast cancer patients)
    Recovery of Interfering SubstancesRange: 93.3% (Ondansetron) to 105.8% (Methotrexate)
    Recovery of Cross-reacting AntigensRange: 103.6% (CA 19.9) to 105.9% (CA125)

    Study Information

    2. Sample size used for the test set and the data provenance:

    • Test Set for Clinical Trial Results: The exact sample size for the "Clinical Trial Results" (Sensitivity, Specificity, PPV, NPV) is not explicitly stated, but it is described as a "Prospective Clinical Trial." The country of origin is not mentioned.
    • Test Set for Normal Range/ULN: 314 healthy volunteer donors were tested at three laboratories. Data provenance is not specified.
    • Test Set for Specificity (Non-breast Malignancies):
      • Colon: 43
      • Liver: 20
      • Lung: 47
      • Ovary: 50
      • Pancreas: 45
      • Prostate: 34
      • Stomach: 29
      • Uterus: 30
        Data provenance is not specified.
    • Test Set for Specificity (Other Diseases/Conditions):
      • Benign Breast Disease (Breast Adenoma): 61
      • Benign Breast Disease (Fibrocystic Breasts): 68
      • Cirrhosis: 25
      • Endometriosis: 24
      • Lactating Woman: 37
      • Mild Chronic Hepatitis: 28
      • Ovarian Cyst: 50
      • Pregnancy: 49
      • Renal Impairment: 20
      • Severe Chronic Hepatitis: 20
        Data provenance is not specified.
    • Test Set for Potentially Interfering Substances: Five serum pools containing CA 27.29 at different levels.
    • Test Set for Correlation with Predicate Device: 203 specimens overall, including a subset of 103 women with histologically confirmed breast cancer. Data provenance is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the number or qualifications of experts used to establish ground truth for any of the clinical or specificity studies. For the breast cancer patients used in the correlation study, it mentions "histologically confirmed breast cancer," implying pathology involvement, but details are absent.

    4. Adjudication method for the test set:

    • The document does not mention any adjudication method for establishing ground truth for the test sets.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This device is an in vitro diagnostic test (an automated chemiluminescence system) for measuring a cancer antigen in human serum. This type of device does not involve "human readers" or "AI assistance" in the typical sense of imaging or diagnostic interpretation. Therefore, an MRMC comparative effectiveness study in the context of human readers improving with AI would not be applicable and was not performed.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    • Yes, the entire study focuses on the standalone performance of the ACS:180 BR automated system. It is designed to perform the quantitative determination of CA 27.29 without human intervention in the assay process itself, beyond sample loading and results interpretation by a clinician.

    7. The type of ground truth used:

    • Clinical Trial Results (Sensitivity/Specificity to Change in Disease Status): Implied to be clinical disease progression or regression based on established clinical practice, though the exact method (e.g., imaging, clinical assessments, pathology) is not detailed.
    • Specificity Studies (Non-breast Malignancies, Other Diseases/Conditions): Based on the diagnosis of the malignancy or condition (e.g., "Colon," "Liver," "Benign Breast Disease").
    • Normal Range/ULN: Based on results from healthy volunteer donors.
    • Correlation with Predicate Device: The ground truth for the concentration of CA 27.29 was presumably the measurement from the predicate device (Biomira TRUQUANT® BR™ RIA), against which the new device was compared. For the breast cancer patient subset, the "histologically confirmed breast cancer" provides an additional clinical ground truth for the patient population itself.
    • Interfering Substances/Antigens: Based on known concentrations of spiked substances and antigens in serum pools with established CA 27.29 levels.

    8. The sample size for the training set:

    • The document does not explicitly mention a separate "training set" or its sample size. This is a common characteristic of in vitro diagnostics developed and evaluated prior to the widespread use of deep learning AI, where training might involve assay optimization rather than large-scale data-driven model training.

    9. How the ground truth for the training set was established:

    • As a training set is not explicitly mentioned, the method for establishing its ground truth is also not described. The performance characteristics discussed relate to the validation or test of the final assay.