    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACON™ hCG One Step Pregnancy Test Device (Urine) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid for the early detection of pregnancy. This test is for professional use.

    Device Description

    ACON™ hCG One Step Pregnancy Test Device (Urine) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid for the early detection of pregnancy. The test utilizes a combination of monoclonal and polyclonal antibodies to selectively detect elevated levels of hCG. The assay is conducted by adding specimen to the test device and observing for the formation of colored lines. The specimen migrates via capillary action along the membrane to react with the colored conjugate. Positive specimens react with the specific antibody-hCG-colored conjugate and form a colored line in the Test Area of the membrane. Absence of this colored line suggests a negative result. To serve as a procedural control, a colored line in the Control Area will always appear regardless of the presence or absence of hCG.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the ACON™ hCG One Step Pregnancy Test Device (Urine), based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategoryAcceptance Criteria (Implied/Direct)Reported Device Performance
    Qualitative Detection of hCGDetection of hCG in urine as an aid for early pregnancy detection.Detects hCG. Positive results correspond to pregnancy.
    Concordance with Predicate DeviceHigh concordance (ideally 100%) with another commercially available urine membrane test.100% concordance with the commercially available test.
    Detection Limit (Sensitivity)To detect hCG concentrations of 25 mIU/ml and greater.Successfully detects hCG concentrations of 25 mIU/ml and greater.
    Cross-Reactivity (Specificity)No cross-reactivity with hLH (300mIU/ml), hFSH (1000 mIU/ml), and hTSH (1000uIU/ml).No cross-reactivity observed with specified hormones.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Test Set Sample Size: 159 urine specimens.
    • Data Provenance: The document states "A multi-center clinical evaluation was conducted," implying the data was collected from multiple locations. It does not explicitly state the country of origin, but the submission is to the US FDA, and the company is based in San Diego, CA, USA, suggesting it's likely US-based or international data submitted for US approval. The study appears to be prospective in comparing the new device against a predicate.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    • The document describes a comparative study against "another commercially available urine membrane test." The results of this predicate device were used as the comparator, essentially serving as a surrogate for ground truth.
    • The number of experts involved in determining the "truth" for the predicate device's results is not specified. The qualifications of anyone interpreting the predicate device's results are also not specified.

    4. Adjudication Method for the Test Set:

    • The text describes a direct comparison: "comparing the results obtained using ACON™ hCG One Step Pregnancy Test Device (Urine) and another commercially available urine membrane test."
    • There is no mention of an adjudication method (e.g., 2+1, 3+1) because the comparison is directly between the new device and an established predicate device's results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    • No, an MRMC comparative effectiveness study was not done. The study's design was a direct comparison of the new device's performance against a predicate device, not an evaluation of how human readers improve with or without AI assistance. The device itself is a diagnostic test, not an AI-assisted diagnostic tool for human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    • Yes, a standalone study was done. The ACON™ hCG One Step Pregnancy Test Device (Urine) is a rapid chromatographic immunoassay. Its performance, as reported in the study, refers to the device's ability to detect hCG independently, without human interpretation beyond reading the lines on the test strip according to instructions. The comparison is the device's output against the predicate device's output.

    7. The Type of Ground Truth Used:

    • The "ground truth" for the comparative study was based on the results obtained from another commercially available urine membrane test (predicate device). This serves as a clinical reference or comparator standard.
    • In terms of the device's intrinsic capabilities (sensitivity and specificity testing), the ground truth for sensitivity was based on known hCG concentrations (25 mIU/ml) standardized to the World Health Organization Third International Standard, and for specificity, it was based on known concentrations of interfering hormones.

    8. The Sample Size for the Training Set:

    • The document does not specify a training set sample size. This type of immunoassay device is typically developed and validated through laboratory testing and then clinical evaluation, rather than through a machine learning paradigm that relies on distinct training and test sets in the same way. The 159 urine specimens were the test set for clinical performance.

    9. How the Ground Truth for the Training Set Was Established:

    • As a traditional immunoassay, the concept of a "training set" and associated "ground truth" derived from expert consensus or pathology, etc., is not applicable in the same way it would be for AI/machine learning algorithms.
    • The development and calibration of such a device primarily involve optimizing antibody-antigen reactions, membrane characteristics, and visual signal development to achieve desired sensitivity and specificity targets in laboratory settings, often using controlled samples with known hCG concentrations. The "ground truth" in this context would be the precisely measured hCG concentrations in spiked samples or reference materials used during assay development and calibration.