510(k) Data Aggregation

    K Number
    K981377

    Validate with FDA (Live)

    Date Cleared
    1998-06-10

    (55 days)

    Product Code
    Regulation Number
    862.1700
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACE® T4 Reagent is intended for use in the quantitative determination of thyroxine in human serum.

    ACE® T4 Reagent is used in the diagnosis and treatment of thyroid disorders. It is intended for the quantitative determination of thyroxine in serum using the ACE clinical chemistry analyzer.

    Device Description

    The ACE® T4 Reagent contains two reagents, an Antibody/Substrate reagent and an Enzyme Coniugate reagent. The assay uses 8-anilino-1-naphthalene sulfonic acid (ANS) to dissociate thyroxine from the plasma-binding proteins. The dissociated thyroxine in the sample competes with an enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled thyroxine for a fixed amount of anti-thyroxine specific antibody binding sites. In the absence of thyroxine from the sample, the thyroxine-labeled G6PDH is bound by the specific antibody and the enzyme activity is inhibited. This phenomenon creates a relationship between thyroxine concentration in the sample and the enzyme activity. The enzyme G6PDH activity is determined bichromatically on the ACE® at 340/505 nm by measuring its ability to convert NAD* to NADH.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the ACE® T4 Reagent are implicitly defined by its substantial equivalence to the predicate device, the DRI Thyroxine (T4) Enzyme Immunoassay. The reported device performance is compared directly to the predicate device's performance characteristics.

    ParameterAcceptance Criteria (Predicate Device)Reported Device Performance (ACE® T4 Reagent)
    Assay Range0.7 µg/dL to 20 µg/dL0.47 µg/dL to 20 µg/dL
    Precision
    - Within Run5.4 %CV3.7 %CV
    - Between Run4.3 %CV4.6 %CV
    Correlation vsCommercial thyroxine assayHitachi 717
    - Slope1.021.070
    - Intercept-0.63-0.405
    - r0.9930.9824
    - N10850

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size (Correlation Study):

      • Predicate Device: 108 samples (N=108)
      • Proposed Device: 50 samples (N=50)
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given the context of a 510(k) submission for an in-vitro diagnostic, it is generally assumed to be prospective clinical studies or laboratory studies conducted in a controlled environment as part of performance validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document describes an in-vitro diagnostic (IVD) device for quantitative determination of thyroxine in human serum. For such devices, "ground truth" is typically established by reference methods or validated comparative assays, rather than expert consensus on images or clinical assessments. The comparison is made against existing commercial thyroxine assays (for the predicate) and the Hitachi 717 analyzer (for the proposed device). Therefore, the concept of "experts" to establish ground truth as would be relevant for image-based diagnostic aids does not directly apply here. The expertise lies in the established accuracy and reliability of the comparative analytical methods.

    4. Adjudication Method for the Test Set

    Not applicable. As described in point 3, the ground truth is established by comparative analytical methods, not by human adjudication of observations.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is an in-vitro diagnostic reagent, not an imaging device or an AI-assisted diagnostic tool that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this is effectively a standalone performance study. The device (ACE® T4 Reagent) is an analytical tool that quantitatively determines thyroxine levels. Its performance (precision and correlation) is evaluated directly, without human interpretation of results as part of the primary measurement.

    7. The Type of Ground Truth Used

    The ground truth for the test set (correlation study) was established by:

    • For the predicate device: A "Commercial thyroxine assay."
    • For the proposed device: The "Hitachi 717" analyzer, which itself is a widely accepted clinical chemistry analyzer.

    This is a form of comparative assay reference method or reference instrument data used to establish equivalency and accuracy for IVDs.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning, as this device is a reagent for an enzyme immunoassay, not an AI software. The performance assessment section focuses on validation data (precision and correlation).
    For traditional IVDs, "training" might refer to the development and optimization of the assay itself using various samples, but specific sample sizes for such a "training set" are not typically reported in 510(k) summaries in this manner, nor are they relevant in the same way as for AI/ML models.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable in the context of machine learning. The "ground truth" for developing a chemical assay typically involves using calibrators and control materials with known concentrations, or comparing against established reference methods during the assay development phase. These details are not provided in this 510(k) summary.