Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201112

    Validate with FDA (Live)

    Date Cleared
    2020-05-27

    (30 days)

    Product Code
    Regulation Number
    870.3535
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AutoCAT®2 Intra-Aortic Balloon Pump (IABP) and AC3™ Series Intra-Aortic Balloon Pump (IABP) arc clinically indicated for use for the following conditions:

    1. Acute Coronary Syndrome

    2. Cardiac and Non-Cardiac Surgery

    3. Complications of Heart Failure

    Device Description

    The AutoCAT Series IABP System, including the AutoCAT®2 and AC3™ Series IABP's, provide counter-pulsation therapy to adult patients with impaired left ventricular (LV) function. They provide hemodynamic support of blood pressure and reduced cardiac work through volume displacement principles. The IABP is attached to an intra-aortic balloon (IAB) catheter which is inserted into the femoral artery and positioned in the descending thoracic aorta.

    The IABP delivers helium (He) into the IAB catheter during diastole to displace blood above and below the IAB. increasing blood pressure and perfusion to organs close to the IAB catheter. The IABP deflates or removes He from the IAB catheter just prior to or in the early phase of systole, reducing the pressure in the aorta and therefore the LV must generate to open the aortic valve and eject its contents into the circulatory system. This results in a decrease in work and oxygen demand.

    The AutoCAT Series IABP Systems consists of two main components:

    • The pump control/display module which incorporates a touch screen and keypad for . system operation, and
    • . The pneumatic drive module which is incorporated into the body of the device

    The AutoCAT Series IABP are designed to be used with 30, 40 and 50cc IAB catheters with the appropriate connectors. The primary difference between the AutoCAT®2 IABP and AC3TM series IABP is the user interface, where the display is a touchscreen.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the AutoCAT®2 and AC3™ Series Intra-Aortic Balloon Pump (IABP) systems. The clearance is based on the substantial equivalence to predicate devices, and the document primarily focuses on design modifications and verification testing rather than an AI-driven diagnostic or prognositc device.

    Therefore, many of the requested categories related to medical AI applications, such as expert consensus, ground truth establishment, sample sizes for training sets, multi-reader multi-case studies, and specific performance metrics like sensitivity, specificity, and AUC, are not applicable to this document. This document describes a medical device (an intra-aortic balloon pump) and its mechanical/electrical performance, not an AI/ML algorithm.

    However, I can extract information related to the acceptance criteria and study that proves the device meets the criteria based on the provided text, focusing on the device's functional integrity as a medical pump.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document states:

    • "The proposed modification to the IABP systems were verified through design verification testing as outlined on the following page."
    • "The results of the verification tests met the specified acceptance criteria and performed similar to the predicate device."

    While specific numerical acceptance criteria (e.g., pressure ranges, flow rates) and the exact reported performance values are not detailed within this document, the statement confirms that such criteria existed and were met. The "performance" is implicitly stated as "similar to the predicate device."

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified in the provided document.
    • Data Provenance: Not specified. Given it's a device modification verification for a US-based manufacturer, it's highly likely the testing was conducted internally or with US-based facilities. The document describes "design verification testing" which implies lab or bench testing, not clinical studies with patient data.
    • Retrospective or Prospective: Not applicable as this refers to a device's performance verification, not analysis of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This is a medical device (pump) clearance, not an AI/ML algorithm requiring expert interpretation of medical images or data to establish ground truth. The "ground truth" for a pump is its physical and electrical reliability and performance metrics, which would be established by engineering specifications and testing.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are relevant for human interpretation of medical data (e.g., radiologists reviewing images). For device verification, engineers perform tests against predefined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an IABP, a mechanical pump, not an AI product designed to assist human readers in tasks like image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance validation is derived from engineering specifications and established performance characteristics of the predicate devices. The testing verifies that the modified device performs according to these engineering and functional standards, similar to the predicate device. It's about meeting designed-for-purpose operational parameters, not diagnostic accuracy against a clinical reference.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML driven device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML driven device.

    Summary relevant to the document:

    The device clearance relies on the "design verification testing" performed on the modified IABP systems. The document states that "The results of the verification tests met the specified acceptance criteria and performed similar to the predicate device." The primary "study" proving the device met acceptance criteria was this internal engineering and performance testing against established technical and functional specifications. The acceptance criteria themselves are implicitly the functional and safety requirements for an intra-aortic balloon pump, ensuring the modified parts do not introduce new safety concerns or degrade performance compared to the previously cleared predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1