510(k) Data Aggregation

    K Number
    K214116

    Validate with FDA (Live)

    Date Cleared
    2022-10-11

    (285 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AAMI4 Isolation Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. The proposed AAMI4 Isolation Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB 70:2012, Liquid Barrier Performance and Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities. The AAMI4 Isolation Gown is a single use, disposable medical device provided non-sterile.

    Device Description

    The proposed device, AAMI4 Isolation Gown, is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. The proposed device is a single use, disposable medical device provided non-sterile. There is one model of AAMI4 Isolation Gown: 4015T. And the proposed AAMI4 Isolation Gown is available in one product size: U. The barrier protection level for AAMI4 Isolation Gown meets AAMI Level 4. The proposed device consists of five components: 1) Basic Gown, 2) Belt, 3) Cuff, 4) Hook-and-loop Velcro and 5) Seam Sealing Tape. The gown is made of polypropylene and polyethylene laminating. The gown is blue in color.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "AAMI4 Isolation Gown." This document focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It does not describe a study involving human readers or AI assistance. Therefore, information regarding human readers, AI assistance, effect size, standalone algorithm performance, and sample size for training sets is not available in this document.

    Here's the information that can be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Test MethodologyPurposeAcceptance CriteriaResult
    FlammabilityTo evaluate the flammability of the test sample in accordance with 16 CFR Part 1610.Meets Class 1 requirementsClass 1
    Hydrostatic pressureTo determine the hydrostatic pressure of the test sample in accordance with AATCC 127: 2017.>50 cmGown sleeve: 188 cmH₂OSleeve seam: 80 cmH₂OBody-sleeve seam: 65 cmH₂OGown body: 199 cmH₂OBelt attachment: 74 cmH₂O
    Water impactTo determine the hydrostatic pressure of the test sample in accordance with AATCC 127: 2017. (Note: The purpose description seems to be a typo and should likely refer to water impact.)≤1.0 gGown sleeve: 0gSleeve seam: 0gBody-sleeve seam: 0gGown body: 0gBelt attachment: 0g
    Breaking strengthTo evaluate the breaking strength of the test sample in accordance with ASTM D 5034:2009(2017).>20 NMD: Mean 87.3 NCD: Mean 66.7 N
    Tearing strengthTo evaluate the tearing strength of the test sample in accordance with ASTM D5587:2015(2019).>20 NMD: Mean 31.6 NCD: Mean 22.9 N
    LintingTo evaluate the linting of the test sample in accordance with ISO 9073-10:2003.Log10(particle count) < 4Log10(lint count): Mean 2.5
    Seam strengthTo evaluate the seam strength of the test sample in accordance with ASTM D1683/D1683M: 2017(2018).>50 NMean 70.9 N
    Viral barrier (resistance to bacteriophage Phi-X174)To evaluate the resistance to bacteriophage penetration in accordance with ASTM F1671/F1671M-13.No bacteriophage penetrationNo bacteriophage penetration
    CytotoxicityTo evaluate the cytotoxicity of the test sample in accordance with ISO 10993-5 Third edition 2009-06-01.Viability ≥ 70% of blank; 50% extract viability ≥ 100% extract viability.The viability was ≥70% of the blank. The 50% extract had higher viability than 100% extract. Proposed device was non-cytotoxic.
    SensitizationTo evaluate the sensitization of the test sample in accordance with ISO 10993-10 Third Edition 2010-08-01.Non-sensitizingThe proposed device was non-sensitizing.
    IrritationTo evaluate the irritation of the test sample in accordance with ISO 10993-10 Third Edition 2010-08-01.Non-irritatingThe proposed device was non-irritating.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes for each test. It refers to "test sample" for each methodology. The tests were performed in accordance with international and US standards. The manufacturer is Hubei Xinxin Non-woven Co., Ltd. from Xiantao, Hubei, China. The data provenance is non-clinical laboratory testing; it does not involve human data, thus the terms "retrospective" or "prospective" are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The ground truth for these non-clinical tests is established by the specified industry and regulatory standards, not by human experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a non-clinical performance and biocompatibility study, not a clinical study requiring human adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This document does not describe studies involving human readers or AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This document describes the testing of a physical medical device (an isolation gown), not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth for the performance tests (Flammability, Hydrostatic pressure, Water impact, Breaking strength, Tearing strength, Linting, Seam strength, Viral barrier) is defined by the technical specifications and requirements within the referenced industry standards (e.g., 16 CFR Part 1610, AATCC 127, ASTM D5034, ISO 9073-10, ASTM D1683/D1683M, ASTM D5587, ASTM F1671/F1671M-13).

    The ground truth for the biocompatibility tests (Cytotoxicity, Sensitization, Irritation) is defined by the requirements and methodologies within the ISO 10993 series of standards (ISO 10993-5, ISO 10993-10).

    8. The sample size for the training set

    Not applicable. This document describes the testing of a physical medical device, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for a physical medical device.