Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K070266

    Validate with FDA (Live)

    Date Cleared
    2007-02-22

    (24 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Smith & Nephew HD Camera System is indicated for use in endoscopic surgical procedures to allow visualization of the articular cavities, body cavities, hollow organs and canals when used with an appropriately indicated endoscope.

    Additionally, when used in conjunction with a Smith & Nephew light source and light cable the Smith & Nephew HD Camera System is indicated for use in endoscopic surgical procedures in the thoracic cavity when used with an appropriately indicated thoracoscope.

    Device Description

    The Smith & Nephew HD Camera System is used in endoscopic surgical procedures to capture and transmit video images. The Smith & Nephew HD Camera System consists of a camera control unit and a camera head. The camera system has been designed to be capable of outputting a true high definition picture. The high definition picture is maintained by continuous transmission of high resolution ourput from the camera head to the camera control unit.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Smith & Nephew HD Camera System, focusing on acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document (K070266) is a 510(k) Summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific quantifiable acceptance criteria and their direct performance results.

    However, based on the text, the implicit acceptance criteria are that the device performs as well as the legally marketed predicate device and that any differences do not raise new issues of safety and efficacy.

    Acceptance Criteria (Implied)Reported Device Performance
    Safety: Conformance to relevant electrical safety standards.The Smith & Nephew HD Camera System conforms to: - IEC 60601-1 (1998) + Amendment 1 (1991) + Amendment 2 (1995) (UL 2601-1) - IEC 60601-1-1 (1992) + Amendment 1 (1995), (2000) - UL 60601-1 (2003) - CAN/CSA 22.2 No. 601.1 (1990)
    Electromagnetic Compatibility (EMC): Conformance to relevant EMC standards.The Smith & Nephew HD Camera System conforms to: - IEC 60601-1-2 (2001-09) 2nd Edition
    Efficacy/Performance: Capability to capture and transmit video images for endoscopic surgical procedures.The performance testing, EMDEMC testing, UL Safety testing, and software verification and validation demonstrate substantial equivalence to the DYONICS Vision 325Z DV 3-CCD Hermes-Ready Camera System (K031379).
    No New Safety/Efficacy Issues: Any differences from the predicate device do not introduce new risks."The testing also demonstrates that any differences in the new device and the predicate device do not raise any new issues of safety and efficacy..."
    Substantial Equivalence: Similar in intended use, operating principle, basic design, and technology."The Smith & Nephew HD Camera System similarities to the predicate device are: - has the same Indications for Use - utilizes the same operating principle - incorporates the same basic design - incorporates the same technological characteristics as the predicate"

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a sample size for a test set. This type of submission (510(k)) primarily demonstrates equivalence to a predicate device through engineering tests, rather than clinical efficacy studies with patient data. The "performance testing" mentioned refers to engineering and validation tests rather than clinical trials with patient samples. Therefore, there's no information on data provenance (country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. As noted above, the validation relies on engineering and regulatory compliance testing, not a clinical study requiring expert-established ground truth from medical images/data.

    4. Adjudication Method for the Test Set:

    This information is not applicable and therefore not provided. No clinical test set requiring adjudication is mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or described in this 510(k) summary. This type of study would compare human reader performance with and without AI assistance, which is not relevant for a camera system simply providing visualization.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The "performance testing" mentioned in the summary is, by definition, a standalone evaluation of the device's technical capabilities (e.g., image quality, electrical safety, EMC). This is not an algorithm performance study as would be typical for an AI/ML device, but rather a functional and safety evaluation of the hardware system itself. The camera system provides an image; it doesn't process data with an algorithm to make a diagnostic assessment.

    7. The Type of Ground Truth Used:

    For the engineering and regulatory compliance testing, the "ground truth" would be established by technical specifications, industry standards, and validated test methodologies. For example:

    • Electrical safety: Measured electrical parameters must fall within specified ranges defined by standards like IEC 60601-1.
    • EMC: Emissions and immunity must meet the limits defined by IEC 60601-1-2.
    • Image output: The system's output would be compared against a high-definition standard to confirm it meets "true high definition."

    There is no mention of ground truth established by expert consensus, pathology, or outcomes data, as this is a hardware device for visualization, not a diagnostic interpretive tool.

    8. The Sample Size for the Training Set:

    This information is not provided and is not applicable. The Smith & Nephew HD Camera System is a hardware device; it does not involve machine learning algorithms that require a "training set" of data. Its design and manufacturing follow established engineering principles and a Quality System.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided and not applicable, for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1