    Intended Use

    The C32SP+/C32S+ 3MP Color LCD Display is intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners.

    Device Description

    The C32SP+/C32S+ 3MP Color LCD Display is a display system for medical viewing, with high resolution 2048 x 1536, built-in brightness stabilization circuit and ambient light sensor, stable brightness and persistent calibration can be guaranteed. The display can support both landscape and portrait mode. The anti-reflection coated protective screen can prevent display from damage under hard using conditions, make the clean and disinfect easier.

    AI/ML Overview

    The provided document is a 510(k) Summary for a 3MP Color LCD Display (C32SP+/C32S+) and primarily focuses on establishing substantial equivalence to a predicate device. It does not contain information about a study proving the device meets acceptance criteria in the manner of an AI/human performance study. The document states that "All of parameters comply with 21CFR1020.33 and related IEC standards" and that the differences "do not affect the safety and effectiveness (SE)." This suggests that acceptance criteria would be defined by adherence to these standards and the comparison tables to the predicate device.

    Given the nature of the document as a 510(k) submission for an LCD display, it focuses on technical specifications and substantial equivalence rather than clinical performance or human-in-the-loop studies typically associated with AI-powered devices. No information is present regarding:

    • Sample sizes for test sets or training sets
    • Data provenance
    • Number or qualifications of experts
    • Adjudication methods
    • Multi-reader multi-case (MRMC) comparative effectiveness studies
    • Standalone algorithm performance
    • Type of ground truth used

    Therefore, the requested information elements cannot be fully populated as they are not within the scope of this document.

    However, based on the comparative tables provided, we can infer the acceptance criteria are alignment with the predicate device's specifications and compliance with relevant standards. The reported device performance is its technical specifications as listed.

    Here's a breakdown based on the available information:

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria are implicitly defined by the parameters of the predicate device and relevant standards (21CFR1020.33, IEC 60601-1, IEC 60601-1-2). The reported device performance is its specifications, which are presented as comparable or superior to the predicate.

    Acceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (C32SP+/C32S+)
    Panel Size and Type: 21.3", TFT LCD display21.3", TFT LCD display
    Pixel Pitch: 0.2115 mm0.2115 mm
    Native Resolutions: 2048 x15362048 x1536
    Brightness: 1700 cd/m² (or comparable)1700 cd/m²
    Contrast Ratio: 1300:1 (or comparable/better)1400:1 (considered better)
    Network Interface: USB(1 Up, 2 Downstream)USB(1 Up, 2 Downstream)
    Active Display Size: 433.2mm x324.9mm (or comparable)433.152mm x324.864mm
    Operating Temperature: 0°C ~ 40°C0°C ~ 40°C
    Transport/Storage Temperature: -20°C ~ 60°C-20°C ~ 60°C
    Operating Relative Humidity: 8% ~80% (non-condensing)15%~85%
    Transport/Storage Relative Humidity: 5% ~95%10%~90%
    Power Capacity: <40W (or comparable)<80W
    Input Voltage: 100~240v (or comparable)DC12V/7.0A
    Usability: Button operation, LED indicatorButton operation, LED indicator
    Mode of operation: Continuous operationContinuous operation
    Biocompatibility: Not neededNot needed
    Sterilization: Not neededNot needed
    Electrical & Mechanical safety: Class I, Ordinary equipment, Complies with IEC 60601-1Class I, Ordinary equipment, Complies with IEC 60601-1
    EMC Evaluation: Complying with IEC 60601-1-2Complying with IEC 60601-1-2

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not applicable and not mentioned. The evaluation is based on technical specifications and comparison to a predicate device, not a diagnostic performance study with patient data.
    • Data provenance: Not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. There is no mention of human experts establishing ground truth for a test set in this context. The "ground truth" for this device would be its conformity to engineering specifications and safety standards.

    4. Adjudication method for the test set:

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is an LCD display, not an AI-powered diagnostic tool. MRMC studies are not relevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device does not have an algorithm in the sense of AI or image processing. Its performance is purely hardware-based.

    7. The type of ground truth used:

    • The "ground truth" is adherence to established technical specifications, safety standards (e.g., 21CFR1020.33, IEC 60601-1, IEC 60601-1-2), and the technical characteristics of the legally marketed predicate device.

    8. The sample size for the training set:

    • Not applicable. This document is not about an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.