510(k) Data Aggregation
(102 days)
The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients.
A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy. Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection. Laparoscopically assisted hysterectomy. Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT/neurosurgeons and urologists.
The 1788 4K Camera System with Advanced Imaging Modality is an endoscopic camera system that produces live video in the surgical field during surgical endoscopic procedures. The system is sensitive in the visible, infrared spectrums. The optical image is transferred from the surgical site to the camera head by a variety of rigid and flexible to the camera head. The 1788 4K Camera System consists of three main components: (1) a camera control unit (CCU); (2) a camera head with an integral cable that connects to the CCU; and (3) a coupler for attaching an endoscope to the camera head.
I am sorry, but the provided text does not contain any information about acceptance criteria or a specific study proving the device meets acceptance criteria. The document is an FDA 510(k) clearance letter for the Stryker Endoscopy 1788 4K Camera System with Advanced Imaging Modality.
It primarily focuses on:
- Regulatory Clearance: Affirming that the device is substantially equivalent to legally marketed predicate devices.
- Device Description: A brief overview of the camera system and its components.
- Indications for Use: The specific medical procedures and patient populations for which the device is intended.
- Technological Comparison: Stating that the subject device has similar technological characteristics to its predicate, with the addition of an optical shutter.
- Non-Clinical Testing Summary: Listing various non-clinical bench, software, EMC, and electrical safety tests performed to standards like IEC 62304, IEC 60601, and ANSI AAMI ES 60601. It concludes that these tests successfully completed to conform with recognized safety standards and design specifications.
- Absence of Clinical Studies: Explicitly stating, "The subject devices do not require clinical studies to support the determination of substantial equivalence."
Therefore, I cannot extract the requested information such as:
- A table of acceptance criteria and reported device performance.
- Sample size and data provenance for a test set.
- Number and qualifications of experts for ground truth.
- Adjudication method.
- MRMC comparative effectiveness study results.
- Standalone (algorithm only) performance.
- Type of ground truth used.
- Training set sample size.
- How ground truth for the training set was established.
This document indicates that substantial equivalence was determined based on non-clinical testing against recognized safety standards and a comparison of technologies, rather than through a clinical study with detailed performance metrics and ground truth establishment against an AI algorithm.
Ask a specific question about this device
The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ endoscope/ arthroscope/ sinuscope is indicated for use. The 1788 4K Camera System with Advanced Imaging Modality is indicated for use in adults and pediatric patients.
A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repar, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the 1788 4K Camera System with Advanced Imaging Modality are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, orthopedic surgeons, ENT/heurosurgeons and urologists.
Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used with SPY AGENT GREEN to provide real-time endoscopic visible and near infrared fluorescence imaging. The L12 Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients aged one month and older, and visualization of at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct) in adults and pediatric patients 12 to 17 years of age, using near-infrared imaging.
Fluorescence imaging of biliary ducts with the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
Additionally, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable enable surgeons to perform minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
Upon interstitial administration of SPY AGENT GREEN, the L12 LED Light Source with Advanced Imaging Modality and SafeLight Cable are used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Upon administration and use of pafolacianine consistent with its approved label, the L12 LED Light Source with Advanced Imaging Modality and SafeLight™ Cable are used to perform intraoperative fluorescence imaging of tissues that have taken up the drug.
The L12 LED Light Source with Advanced Imaging Modality is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.
Stryker's Advanced Imaging Modality (AIM) System is an endoscopic real-time 4K visible white light, near-infrared illumination and transillumination, and cyan spectral imaging system. Near-infrared illumination is used for fluorescence imaging using SPY AGENT™ GREEN (indocyanine green for injection, USP) or CYTALUX™ (pafalocianine) injection. Near-infrared illumination is also intended for use during transillumination of the ureters using the IRIS Ureteral Kit during minimally invasive and open surgical procedures. Cyan Spectral Imaging is intended as an alternative mode of visualization that uses narrow bands for illumination during endoscopic surgery to provide greater visualization of surface structures and fine capillary patterns of the mucosal membrane.
Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Stryker 1788 4K Camera System with Advanced Imaging Modality and L12 LED Light Source with AIM:
A table of acceptance criteria and the reported device performance cannot be fully constructed from the provided document. The document primarily lists regulatory standards and general performance categories, indicating that the device "passed" these tests. It does not provide specific quantitative acceptance criteria or detailed reported performance values for each criterion.
However, based on the Performance Data section (page 9), we can infer the types of tests conducted and their general outcome:
Inferred Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Reported Device Performance | Specific Standard/Test (if mentioned) |
|---|---|---|
| Safety Testing | ||
| Electromagnetic Compatibility | Pass | IEC 60601-1-2:2014, IEC 60601-4-2:2016 |
| Electrical Safety | Pass | ANSI AAMI ES 60601-1:2005 + A1:2012 + A2:2021, IEC 60601-1-6:2010 + A1:2013 + A2:2020, IEC 60601-2-18:2009 |
| Laser Safety | Pass | IEC 60825-1:2014, IEC 62471 / Comparative testing to legally marketed predicate device |
| System Functionality | ||
| Sterilization | Pass | ISO 17664-1:2021, ISO 17664-2:2021, ISO 14937:2009, AAMI TIR12:2020, AAMI TIR30:2011, AAMI ST58:2013/(R)2018 |
| Software Verification & Validation | Pass | IEC 62304:2015 |
| Usability | Pass | IEC 62366:2015 + A1:2020 |
| Imaging Performance | ||
| Bench Performance Testing | Pass (against device input specifications and predicate devices) | Spatial Uniformity, Minimum Detectable Fluorescence, Fluorescence Detection Depth, Clinically Meaningful Limits of Detection, Signal to Noise, Photobiological Safety (IEC 62417), Contrast |
| General Performance Testing | Pass (against user needs and intended uses) | (Not specified beyond general user needs and intended uses) |
Here's the requested information based on the provided document:
-
A table of acceptance criteria and the reported device performance
- (See table above)
-
Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- The document states: "NOTE: The Advanced Imaging Modality (AIM) System does not require clinical studies to support the determination of substantial equivalence." This implies that no clinical test set (patient data) was used in the assessment for substantial equivalence. The testing described appears to be primarily bench and laboratory-based. Therefore, sample size and data provenance in the context of clinical data are not applicable based on this document.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Since no clinical studies were performed, there's no mention of experts establishing ground truth for a clinical test set. The "Pass" results are based on compliance with engineering standards and comparisons to predicate devices in bench testing.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. No clinical test set demanding adjudication was used.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done. The document explicitly states: "The Advanced Imaging Modality (AIM) System does not require clinical studies to support the determination of substantial equivalence." There is also no mention of AI assistance in the context of human readers for this device. The "Advanced Imaging Modality" refers to specialized illumination and visualization techniques, not AI-powered image analysis for diagnosis.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is an imaging system (camera and light source) that provides visual information to a surgeon. It is not an AI algorithm that operates standalone without human interpretation or intervention for diagnostic or therapeutic decisions.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the performance testing, the implied "ground truth" or reference was primarily device input specifications and the performance of currently legally marketed predicate devices. For regulatory compliance, the ground truth was adherence to FDA-recognized voluntary consensus standards (e.g., IEC standards for electrical safety, usability, software, etc.).
-
The sample size for the training set
- Not applicable. This document describes a medical imaging hardware system, not a machine learning model that requires a training set.
-
How the ground truth for the training set was established
- Not applicable. As there is no machine learning component requiring a training set, the establishment of ground truth for such a set is not mentioned.
Ask a specific question about this device
(28 days)
The 1788 Video Camera is indicated for use in general laparoscopy, ear endoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ arthroscope/ sinuscope is indicated for use. The 1788 Video Camera is indicated for use in adults and pediatric patients.
The 1788 4K Camera System with Advanced Imaging Modality is an endoscopic camera system that produces live video in the surgical field during surgical endoscopic procedures. The system is sensitive in the visible and infrared spectrums. The optical image is transferred from the surgical site to the camera head by a variety of rigid and flexible endoscopes, which are attached to the camera head. The 1788 4K Camera System consists of three main components: (1) a camera control unit (CCU); (2) a camera head with an integral cable that connects to the CCU; and (3) a coupler for attaching an endoscope to the camera head.
The Stryker 1788 4K Camera System with Advanced Imaging Modality (K230605) is an endoscopic camera system.
Here's an analysis of its acceptance criteria and supporting studies based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance:
| Test | Acceptance Criteria (Method) | Reported Device Performance |
|---|---|---|
| Electromagnetic Compatibility | IEC 60601-1-2:2014 | Pass |
| Electrical Safety | ANSI/AAMI ES60601-1:2005 + A1:2012, IEC 60601-1-6:2010 + A1:2013, IEC 60601-2-18:2009 | Pass |
| Software | IEC 62304:2015 | Pass |
| Performance Testing - Bench: | ||
| • Camera Head Soaker Cap Tether Pull Force | (Method not explicitly detailed but implied by "Pass") | Pass |
| • Camera Head to Camera Control Unit Connector Insertion/Removal Force | (Method not explicitly detailed but implied by "Pass") | Pass |
| • Camera Head Button to Camera Control Unit Activation Timing | (Method not explicitly detailed but implied by "Pass") | Pass |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not provide information about sample sizes for a 'test set' in the context of clinical or image-based evaluation. The performance data presented consists of engineering and software compliance testing. The data provenance is not specified beyond being generated by Stryker for its 510(k) submission.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
This information is not applicable as the provided documentation details engineering and software performance testing for a camera system, not the evaluation of a diagnostic or interpretive algorithm that would require expert-established ground truth.
4. Adjudication Method for the Test Set:
This information is not applicable for the same reasons mentioned in point 3.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size:
No, an MRMC comparative effectiveness study was not explicitly conducted or referenced in the provided document. The device is a camera system providing visual input, not an AI-assisted diagnostic tool that would typically be evaluated with MRMC studies to measure human reader improvement. The document explicitly states: "The 1788 4K Camera System does not require clinical studies to support the determination of substantial equivalence."
6. If a Standalone (algorithm only without human-in-the-loop performance) was done:
This information is not applicable as the device described is an endoscopic camera system, which provides visual information to a human user in a clinical setting; it is not a standalone algorithm performing a diagnostic task.
7. The Type of Ground Truth Used:
For the performance tests conducted, the "ground truth" refers to the established standards or specifications being tested against (e.g., IEC 60601-1-2:2014 for EMC, product design specifications for bench tests). It's not an expert consensus, pathology, or outcomes data in the sense of clinical decision-making.
8. The Sample Size for the Training Set:
This information is not applicable. The document describes a medical device (camera system), not an AI/ML algorithm that would typically require a training set.
9. How the Ground Truth for the Training Set was Established:
This information is not applicable for the same reasons mentioned in point 8.