Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172468

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-06

    (113 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 091 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasulature.

    Device Description

    The InNeuroCo, Inc. 091 Long Sheath is a variable stiffness catheter that has a catheter shaft reinforced with a stainless steel double coil. It has a radiopaque Platinum/lridium marker band on the distal end. The distal 10 cm of the 091 Long Sheath Catheter has a hydrophilic coating. The catheter has a nominal outer diameter of 0.109 inches and a nominal inner diameter of 0.091 inches. It is available in three working lengths: 70 cm, and 90 cm. The 091 Long Sheath has a PTFE-lined lumen. The 091 Long Sheath is inserted at a vascular access point to provide access to the target site and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst Hemostasis Valve and a Vessel Dilator. The 091 Long Sheath is supplied sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    The provided text is a 510(k) summary for the InNeuroCo 091 Long Sheath peripheral, coronary, and neuro vasculature catheter. It outlines the device's characteristics and its comparison to a predicate device, along with the results of performance testing to demonstrate substantial equivalence.

    Based on the provided document, here's a description of the acceptance criteria and the study proving the device meets these criteria:

    This document describes the validation of a medical device, the InNeuroCo 091 Long Sheath, for regulatory approval. It is not an AI/ML device, and therefore several of the requested sections (e.g., number of experts, adjudication methods, MRMC study, standalone performance, training set details) are not applicable as they relate to the validation of AI/ML algorithms, not traditional medical devices like catheters.


    Acceptance Criteria and Device Performance Study for the InNeuroCo 091 Long Sheath

    The InNeuroCo 091 Long Sheath is a percutaneous catheter intended for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The study conducted was a series of non-clinical (bench and animal) tests to demonstrate substantial equivalence to a legally marketed predicate device (Stryker AXS Infinity LS, K152876).

    1. Table of Acceptance Criteria and Reported Device Performance

    TestTest Method SummaryAcceptance CriteriaReported Device Performance (Conclusions)
    Biocompatibility-Material Mediated Pyrogen TestingCompleted per ISO 10993-11The test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject.Test results for the InNeuroCo Super Distal Access (K161262) and the Zenith Catheter (K171672) were leveraged as materials and manufacturing processes are equivalent. Met acceptance criteria.
    Biocompatibility-Cytotoxicity MEM ElutionCompleted per ISO 10993-5The cultures treated with the test article must not have a reactivity grade greater than 2.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Hemolysis ASTM Method, extract human bloodCompleted per ISO 10993-4The hemolytic index above the negative control article must be less than 5%.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Hemolysis, ASTM method, direct contact (human blood)Completed per ISO 10993-4The hemolytic index above the negative control article must be less than 5%.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Unactivated Partial Thromboplastin TimeCompleted per ISO 10993-4There must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Complement ActivationCompleted per ISO 10993-4There must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Dog ThrombogenicityCompleted per ISO 10993-4The test articles must receive a thrombus formation score less than or equal to that of the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Maximization SensitizationCompleted per ISO 10993-10The test article must elicit a positive response in less than 10% of the test animals.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Intracutaneous Toxicity/ReactivityCompleted per ISO 10993-10The test article extracts must not induce a significantly greater biological reaction than the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Acute Systemic Toxicity TestCompleted per ISO 10993-11The test article extracts must not induce a significantly greater biological reaction than the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Animal Testing-Angiographic EvaluationProduct evaluated within a porcine model for clinically significant injury.No units tested can cause vessel injury.Test article met the acceptance criteria for Animal Testing.
    Bench Testing - Chemical CompatibilityCatheter exposed to chemicals readily available in a clinical setting.Chemicals have no negative effect on the catheter or accessories.Leveraged results from InNeuroCo Super Distal Access (K161262). Met acceptance criteria.
    Bench Testing - Visual InspectionCompleted per ISO 10555-1Catheter shall appear free from damage, including a rounded tip and smooth transition points.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - DimensionsCompleted per ISO 10555-1Test samples should be within existing dimensional specifications.091 Long Sheath test samples met the acceptance criteria. Leveraged results for Hemostasis Valve from InNeuroCo Super Distal Access (K161262), which also met the criteria.
    Bench Testing - Hub compatibilityCompleted per ISO 594-1 and ISO 594-2Hub shall meet existing Luer specifications.Leveraged results for the 091 Long Sheath and Dilator from InNeuroCo Intermediate Catheter (K152202) and for the Hemostasis Valve from Super Distal Access (K161262). All tested samples met the acceptance criteria.
    Bench Testing - Leak - AirCompleted per ISO 10555-1Test samples should be within existing Air – Leak specifications.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - Leak - LiquidCompleted per ISO 10555-1Test samples should be within existing Leak - Liquid specifications.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - ParticulatesCompleted per USP 788Test samples should be within existing Particulate specifications.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - Simulated UseLS underwent simulated use testing by a physician in a benchtop model.Test samples must meet predetermined user needs.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - Catheter BurstCompleted per ISO 10555-1Test sample burst pressures must meet or exceed existing minimum burst pressure specification.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - TensileCompleted per ISO 10555-1Test sample ultimate tensile strength must meet or exceed existing tensile strength specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - CorrosionCompleted per ISO 10555-1Test samples shall exhibit no evidence of corrosion.Leveraged results from InNeuroCo Intermediate Catheter (K152202). All samples met the acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Packaging - Dye LeakCompleted per ASTM F1929-12Test sample shall not exhibit any visual leaks or channels.091 Long Sheath test samples met the acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Packaging - PeelCompleted per ASTM F88-09Test sample tensile strength must meet or exceed existing tensile strength specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Kink ResistanceSamples are subjected to different diameters until kink is observed.Test sample kink resistance must meet or exceed existing Kink Resistance specifications.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - TorqueConditioned samples are torqued to failure.Test sample torque results must meet or exceed existing torque specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - PTFE Liner inspectionLS was challenged to demonstrate liner adherence.Test sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Hydrophilic Coating IntegrityConditioned samples were repeatedly exposed to friction to demonstrate that the hydrophilic coating is not affected.Test sample results must meet or exceed existing Hydrophilic Coating Integrity specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Labeling LegibilityLabel is legible after printing.Test samples shall demonstrate text legibility.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - BarcodeBarcode is readable with a standard barcode reader.Test samples shall demonstrate readily readable barcodes.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    SterilizationISO 11135 and AAMI TIR 28Sterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads.091 Long Sheath sterilization load met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bacterial EndotoxinAAMIST72Test samples must meet the bacterial endotoxin acceptance criteria.The 091 Long Sheath samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Shelf LifeASTM F1980Aged test samples must meet or exceed existing specifications.091 Long Sheath test samples met the acceptance criteria. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The exact sample sizes for each specific test are not explicitly enumerated in the provided summary table. The statements typically indicate "Test samples" or "All samples," implying that a sufficient number were tested according to the referenced ISO or ASTM standards.
    • Data Provenance: The data is a mix of:
      • Direct testing: For many bench tests (e.g., Visual Inspection, Leaks, Particulates, Simulated Use, Catheter Burst, Tensile, Kink Resistance, Torque, PTFE Liner inspection, Hydrophilic Coating Integrity, Labeling Legibility, Barcode, Sterilization, Bacterial Endotoxin, Shelf Life). These tests were conducted on the 091 Long Sheath.
      • Leveraged data: For all biocompatibility tests, Chemical Compatibility, Hub compatibility, and Corrosion, data was leveraged from previously approved InNeuroCo devices (Super Distal Access K161262, Zenith Catheter K171672, and Intermediate Catheter K152202). This leveraging is justified by stating that the materials and manufacturing processes are equivalent.
      • Animal Testing: Product evaluated within a porcine model for angiographic evaluation.
    • Retrospective/Prospective: All described testing appears to be prospective (new tests conducted for this device and leveraged results from other devices), not retrospective analysis of existing clinical data.
    • Country of Origin: Not specified in the document, but given the FDA 510(k) submission, it would be expected to follow US or internationally recognized standards. The company address is in Florida, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Not Applicable. This is a traditional medical device (catheter) validation, not an AI/ML algorithm. Ground truth for device performance is established through standardized physical, chemical, biological, and animal testing, not human expert interpretation of data.
    • For "Simulated Use" testing, it mentions "by a physician," indicating one or more physicians were involved in that specific assessment, but their number and specific qualifications are not detailed beyond "physician."

    4. Adjudication Method for the Test Set

    • Not Applicable. As this is not an AI/ML study, there is no need for adjudication of conflicting expert opinions on a dataset. Test results are objective measurements against defined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. This type of study is specifically designed for evaluating the impact of AI on human reader performance, which is not relevant for the validation of a physical medical device like a catheter.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. There is no AI algorithm being validated in this submission. The "device" is a physical catheter.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance is established by:
      • Standardized Test Methods: Adherence to established international (ISO, ASTM, USP) and industry standards for material properties, mechanical performance, biocompatibility, and sterility.
      • Physical Measurements: Objective measurements of dimensions, burst pressure, tensile strength, etc.
      • Chemical/Biological Assays: Results from biocompatibility tests (e.g., cytotoxicity, hemolysis, pyrogenicity, sensitization) and bacterial endotoxin tests.
      • Functional Demonstrations: Successful performance in simulated use environments and in an animal model (porcine model for angiographic evaluation, showing no vessel injury).
      • Predetermined Specifications: Comparison against established product specifications and the performance of the predicate device.

    8. The Sample Size for the Training Set

    • Not Applicable. This is not an AI/ML device, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As there is no AI/ML training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1