510(k) Data Aggregation

    K Number
    K222573

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-03-03

    (190 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The imani i2Plus Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breast. The imani i2Plus Breast Pump is a single user device.

    Device Description

    The imani i2Plus Breast Pump is an electrically powered wearable single breast pump consisting of the following key components: a funnel, insert, milk collector, valve, protector, charging cable, adaptor, cap, and iBox accessory. It is designed to work in the user's bra and has a rechargeable battery so it can be used hands-free. The main body includes a press-button user interface, pump body, and LCD display. Pumping can be performed on one breast (single pumping) or two breasts (double pumping) when configured with the iBox accessory. The user interface allows the user to switch from massage to pumping mode and control the vacuum levels within those modes. Massage and pumping modes consist of 5 vacuum levels and fixed cycle speeds when used standalone. Massage mode consists of 5 vacuum levels and 1 cycle speed and pumping mode consists of 9 vacuum levels and 3 cycle speeds for each level. The i2Plus in standalone configuration is capable of providing vacuum levels from 50-150 mmHg at a fixed cycling rate of 65 cycles per minute in massage mode and vacuum levels from 80-270 mmHg with cycling rates from 24-55 cycles per minute in pumping mode. The i2Plus with the iBox accessory is capable of providing vacuum levels from 150-200 mmHg at a fixed cycling rate of 60 cycles per minute in massage mode and vacuum levels from 150-270 mmHg with cycling rates from 24-55 cycles per minute in pumping mode. The imani i2Plus is charged with a 5 V DC USB type C adaptor and powered by a 3.7 V, 1400mAh internal rechargeable lithium-ion battery. The motor unit operates on embedded software updates by endusers are not supported. The subject device is for repeated use by a single user in a home environment. The device is provided non-sterile. The breast pump expresses breast milk by creating a seal around the nipple using a flange and applying and releasing suction to the milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system. All other components (i.e., motor unit) of the subject device are not in contact with the breast. The i2Plus operates on a rechargeable battery when standalone and does not function when charging. When configured with iBox, the pump is only operable when connected with the AC adapter. The rechargeable battery can be charged using the external USB adapter if the motor unit is not in operation. All milk contacting components are compliant with 21 CFR 174-179.

    AI/ML Overview

    This document is a 510(k) premarket notification for a breast pump, not a clinical study or effectiveness study for a medical device that uses AI. Therefore, it does not contain information about acceptance criteria or studies that prove an AI device meets acceptance criteria.

    The document discusses the substantial equivalence of the imani i2Plus Breast Pump to a predicate device (imani i2) based on non-clinical performance testing.

    Here's a breakdown of the information that is available, and why the requested AI-specific information is not present:

    Information available in the document (and why it's not AI-specific):

    1. A table of acceptance criteria and the reported device performance:

      • The document provides a "Comparator Table for Subject and Predicate Devices" (Table 1) which compares technological characteristics like suction levels, cycles per minute, power supply, etc.
      • This table highlights differences and states that these differences "do not raise different questions of safety and effectiveness." This implicitly serves as a form of acceptance criteria (i.e., the differences are acceptable because they don't compromise safety or effectiveness compared to the predicate).
      • Performance data reported: Ranges for suction levels (e.g., 50-150 mmHg for massage, 80-270 mmHg for pumping) and cycles per minute (e.g., 65 CPM for stimulation, 24-55 CPM for pumping) are detailed for both the subject device (i2Plus, with and without iBox) and the predicate (i2).
      • Why it's not AI-specific: This is a comparison of physical and functional specifications of a "powered breast pump," not an AI-powered diagnostic or therapeutic device.
    2. Sample size used for the test set and the data provenance:

      • The document describes "Performance Testing" which included "Vacuum level verification testing," "Backflow protection testing," "Use life testing," "Battery performance testing," and "Battery status indicator testing."
      • It also mentions "Biocompatibility studies" and "Electrical Safety" and "Electromagnetic Compatibility" testing.
      • No specific sample sizes for these tests are provided. The document generally states that "testing was conducted."
      • Why it's not AI-specific: These are engineering and safety tests for a physical device, not a performance evaluation of an AI algorithm on a medical dataset.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The "ground truth" for a breast pump's performance involves mechanical measurements (vacuum, cycles per minute), material safety, electrical safety, and battery life, not expert interpretation of medical images or data.
      • Why it's not AI-specific: There is no medical data interpretation involved in evaluating the breast pump's function in this context.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This refers to methods for resolving discrepancies in expert interpretations, which is not relevant for the type of non-clinical tests described for a breast pump.
      • Why it's not AI-specific: Not relevant for this type of device submission.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This document describes a 510(k) submission for a breast pump. MRMC studies are typically for diagnostic imaging devices where human readers (e.g., radiologists) interpret images, sometimes with AI assistance.
      • Why it's not AI-specific: Not relevant to a breast pump.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. There is no AI algorithm in this device that would have a "standalone performance."
      • Why it's not AI-specific: Not relevant to a breast pump.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for this device's performance is based on engineering specifications and recognized safety standards. For example, vacuum levels are measured against specified mmHg ranges, biocompatibility is evaluated against ISO standards, and electrical safety against IEC standards.
      • Why it's not AI-specific: This is a fundamental difference in the nature of "ground truth" for a mechanical/electrical device versus an AI algorithm.
    8. The sample size for the training set:

      • Not applicable. This device does not use machine learning or AI, and therefore has no "training set."
      • Why it's not AI-specific: No AI, no training set.
    9. How the ground truth for the training set was established:

      • Not applicable. As there is no training set for an AI algorithm, there is no ground truth established for one.
      • Why it's not AI-specific: No AI, no training set, no ground truth for a training set.

    In summary, this document pertains to the regulatory clearance of a standard powered breast pump and not an AI-powered medical device. Therefore, the information requested regarding AI-specific acceptance criteria and study details is not present.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202037

    Validate with FDA (Live)

    Device Name
    imani i2
    Manufacturer
    Date Cleared
    2021-03-01

    (221 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The imani i2 breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The imani i2 breast pump is a single user device.

    Device Description

    The imani i2 is a powered breast pump that is intended to express and collect mills from the breasts of lactating women. This breast pump is intended for use by a single user and can be used to pump breast milk from a single breast (single pumping). The imani i2 breast pump is a wearable breast pump comprised of a main body (pump), charqing cable, collection bottle, valve, diaphragm for backflow protection, and a breast funnel. It is designed to work in the user's nursing bra and is battery operated so it can be used hands-free without external power cords. The only user-contacting material is polypropylene (breast funnel). The device is provided non-sterile. The imani i2 breast pump allows the user to adjust the vacuum levels. Two suction patterns, massage and breast pumping mode, are pre-programmed with variable levels and cycle speeds. The subject device is powered by a rechargeable Li-polymer battery (3.7 VDC, 1400 mAh). The imani i2 provides the following user features: Two displays for level of vacuum and battery. Five buttons for power on/off, mode selection (Massage Mode, Pumping Mode), decrease vacuum, and increase vacuum Pumping Mode: This mode is used to express and collect milk from the breast. This mode includes five vacuum and cycle speed levels. Massage Mode: This mode is used to massage the breast before pumping. This mode includes five vacuum levels and a consistent single cycle speed across all vacuum levels.

    AI/ML Overview

    The provided text is a 510(k) summary for the imani i2 breast pump, detailing its substantial equivalence to a predicate device. However, it does not contain any information regarding acceptance criteria, device performance metrics (like sensitivity, specificity, accuracy, etc.), or details of a study that would typically prove such criteria for a diagnostic or AI-driven device.

    This document describes a medical device, specifically a powered breast pump, which is a mechanical device, not an AI or diagnostic device that would require the type of performance evaluation criteria you listed (e.g., sensitivity, specificity, accuracy, multi-reader multi-case studies, ground truth establishment, etc.).

    The "Non-Clinical Test summary" section (page 6) describes bench performance testing related to:

    • Electrical Safety/Electromagnetic Compatibility in accordance with various IEC and AAMI standards.
    • Software Validation (for a "moderate level of concern" software, likely related to device control, not diagnostic algorithms).
    • Biocompatibility of user-contacting materials (polypropylene).
    • Bench Performance Testing of vacuum level settings, cycle speed settings, and backflow/cross-contamination protection.
    • Battery Testing (use-life, charging time, indicator/operating time).
    • Simulated Use Testing to confirm users can properly use the device and their satisfaction.

    These are all standard engineering and safety performance tests for a mechanical/electrical medical device like a breast pump. They are not the types of studies or criteria used to validate AI or diagnostic performance in medical imaging or pathology.

    Therefore, I cannot provide the requested table or answer the questions related to AI/diagnostic device validation based on this document. The document simply doesn't contain that information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202045

    Validate with FDA (Live)

    Device Name
    imani i1
    Manufacturer
    Date Cleared
    2021-02-26

    (218 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The imani il is a powered breast pump to be used by lactating women to express and collect milk from their breasts. This device is intended for multiple users.

    Device Description

    The imani i1 breast pump is a powered breast pump that is intended to express and collect milk from the breasts of lactating women. The pump component of this device can be used by multiple users using separate collection accessories. The user has the option to pump breast milk from a single breast (single pumping) or from both breasts (double pumping). The imani i1 breast pump is comprised of a main pump body, diaphragm and cap for backflow protection, valve, milk bottle connector, a breastmilk storage packset (30pcs), and air hose hub. The only user-contacting material is polypropylene (breast shield). The device is provided non-sterile. The imani i1 breast pump allows the user to adjust the vacuum levels. Two suction patterns, massage and breast pumping mode, are pre-programmed with variable levels and cycle speed (pump speed). The subject device is AC-powered (AC adaptor: 100 - 240 V, 50/60 Hz, 550 mA [Output: 12VDC, 2A]). The imani i1 breast pump provides the following user features: Three displays for level of cycle, time, and vacuum, Eight buttons for power on/off, mode selection, decrease vacuum and cycle, increase vacuum and cycle, and light control, Breast Pumping Mode: This mode is used to express and collect milk from breast. This mode includes 12 vacuum levels and 6 cycle speed levels. Massage Mode: This mode is used to massage the breast before pumping. This mode includes 6 vacuum levels and 2 cycle speed levels.

    AI/ML Overview

    This document describes the regulatory submission for the imani i1 powered breast pump (K202045) and compares it to a predicate device (OPERA, OPERA Eco; K190810). The information provided pertains to the device's technical specifications and non-clinical testing to demonstrate substantial equivalence, rather than a study proving performance against specific acceptance criteria in a clinical setting with human readers or AI.

    The document does not contain information about a clinical study comparing the imani i1 to established acceptance criteria in the manner requested. Specifically, it does not include:

    • A table of acceptance criteria and reported device performance in a clinical context.
    • Sample sizes used for a test set, data provenance, ground truth establishment by experts, or adjudication methods for a clinical study.
    • Information regarding Multi-Reader Multi-Case (MRMC) comparative effectiveness studies, effect sizes of human readers with/without AI assistance, or standalone algorithm performance.
    • The type of ground truth used in a clinical study (e.g., pathology, outcomes data).
    • Sample size for a training set or how ground truth for a training set was established.

    The provided text focuses on the following non-clinical testing and comparisons:

    1. Acceptance Criteria and Device Performance (Non-Clinical Bench Testing):

    The document broadly states that "Device specifications were met for all tests conducted" under the "Bench Performance Testing" section. It does not provide a detailed table of explicit acceptance criteria or specific numerical results for these non-clinical tests beyond confirming compliance.

    Acceptance Criteria CategoryReported Device Performance
    Electrical SafetyMet AAMI / ANSI ES 60601-1 /A1:2012, IEC 60601-1-2:2014, and IEC 60601-1-11:2015 standards.
    Electromagnetic CompatibilityMet IEC 60601-1-2:2014 and IEC 60601-1-11:2015 standards.
    Software ValidationConducted per 2005 FDA guidance for "Moderate" level of concern.
    BiocompatibilityPassed Cytotoxicity (ISO 10993-5:2009), Sensitization (ISO 10993-10:2010), and Irritation (ISO 10993-10:2010), showing user-contacting materials are biocompatible.
    Bench Performance TestingDemonstrated device performance for vacuum level settings, cycle speed settings, and backflow/cross-contamination protection. "Device specifications were met for all tests conducted."

    2. Sample Size and Data Provenance (for the non-clinical tests mentioned):

    Not applicable in the context of clinical "test sets" as described in the prompt. The testing performed is laboratory-based and does not involve human subjects or clinical data in the sense of patient cohorts and outcomes.

    3. Number of Experts and Qualifications for Ground Truth:

    Not applicable. The ground truth for electrical safety, EMC, software validation, biocompatibility, and bench performance testing is based on established industry standards, test protocols, and engineering specifications, not expert clinical consensus on patient data.

    4. Adjudication Method:

    Not applicable to the described non-clinical testing.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    No MRMC study was conducted or reported in this document. The device is a physical breast pump, not an AI or imaging diagnostic tool that would typically undergo such a study.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. The device is a physical powered breast pump, not an algorithm or AI system.

    7. Type of Ground Truth Used (for non-clinical tests):

    The ground truth for the non-clinical tests (electrical safety, EMC, software validation, biocompatibility, and bench performance) is based on:

    • Established national and international standards: e.g., AAMI / ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993 series.
    • FDA guidance documents: e.g., "Guidance for the Content of Premarket Submissions for Software in Medical Devices," "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing."
    • Device specifications: as defined by the manufacturer for parameters like vacuum levels, cycle speed, and backflow protection.

    8. Sample Size for Training Set:

    Not applicable. This document describes non-clinical testing for a physical medical device, not the training of an AI algorithm or model.

    9. How Ground Truth for Training Set was Established:

    Not applicable, as there is no training set mentioned for an AI algorithm.