Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K193087

    Validate with FDA (Live)

    Device Name
    Rapid ICH
    Date Cleared
    2020-03-31

    (146 days)

    Product Code
    Regulation Number
    892.2080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RAPID ICH is a radiological computer aided triage and notification software indicated for use in the analysis of nonenhanced head CT images.

    The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of pathologies in head CT images, namely Intracranial Hemorrhage (ICH). RAPID ICH uses an artificial intelligence algorithm to analyze images and highlight cases with detected ICH on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH findings. Notifications include compressed preview images, that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of RAPID ICH are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    RAPID ICH is a clinical module which operates within the integrated RAPID Platform to provide triage and notification of suspected intra-cranial hemorrhage (ICH). The RAPID ICH module consists of the core RAPID Platform software which provides the administration and services for the RAPID; and the RAPID ICH module which functions as one of many image processing modules hosted by the platform.

    The RAPID platform is a software package that provides for the visualization and study of changes in tissue using digital images captured by diagnostic imaging systems including CT (Computed Tomography), CTA, XA3D and MRI (Magnetic Image Resonance), as an aid to physician diagnosis. RAPID can be installed on a customer's Server or it can be accessed online as virtual system. It provides viewing, quantification, analysis and reporting capabilities. The RAPID platform has multiple modules a clinician may elect to run and provide analysis for decision making.

    RAPID ICH provides an automatic analysis of received NCCT scan data for the triage and notification of ischemic hemorrhage (ICH). The application is selected via DICOM encoding which is processed by the DICOM handler within the RAPID Platform. Once the DICOM server identifies the selected ICH processing modality, the Job Manager initiates image processing using the ICH Module.

    Upon processing a case with suspicion of ICH, the clinical team is notified via messaging, the case has a suspicion of ICH. The notifications are sent to the PACS/Workstation, via email and to a mobile application. The notification provides the attending clinical team physician that suspicion of hemorrhage has been identified and immediate attention to the case should be given. The messaging provides notification and compressed image data of the case. The compressed preview is informational only and labeling identifies it as not to be used for diagnostic use and to review the data within the PACS/Workstation prior to making any diagnostic decisions. No additional information markings are added to the case.

    The notifications are pop-up messages or email with the appropriate case information and suspicion or non-suspicion of ICH labelled. In all cases, the normal standard of care workflow is adhered to, this ensures the case is still reviewed when non-suspicion is determined.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for RAPID ICH:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and the Reported Device Performance

    MetricAcceptance Criteria (Performance Goal)Reported Device Performance95% Confidence Interval
    Sensitivity> 80%0.899(0.847 - 0.935)
    Specificity> 80%0.943(0.895 - 0.970)
    Time to NotificationNot explicitly stated as a hard acceptance criterion, but a secondary endpoint aim.2.28 minutes(2.24 - 2.33)

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 336 cases
    • Data Provenance: Retrospective, blinded, multicenter, multinational study.
      • Countries of Origin: US (4 Sites and 1 Multisite Study) and OUS (1 site, 1 multisite).
      • Nature of Study: Retrospective.
    • Case Distribution: Approximately an equal number of positive cases (images with ICH) and negative cases (images without ICH) were included in the analysis.

    3. Number of Experts and Qualifications for Ground Truth

    The document states "reader truthing of the data" was used, but it does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") for establishing the ground truth of the test set.

    4. Adjudication Method for the Test Set

    The document mentions "reader truthing," but it does not specify the adjudication method used (e.g., 2+1, 3+1, none) for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done or reported. The study focused on the standalone performance of the AI algorithm. Therefore, there is no information on how human readers improve with AI vs without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone (algorithm only) performance study was conducted. The entire "Performance Data" section describes the RAPID ICH software's ability to identify ICH independently, reporting its sensitivity and specificity.

    7. Type of Ground Truth Used

    The ground truth for the test set was established through "reader truthing of the data." This implies expert consensus or expert interpretation as the basis for ground truth, rather than pathology or outcomes data.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size used for the training set. It only mentions that the device uses a "deep learning algorithm trained on medical images."

    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly detail how the ground truth for the training set was established. It generally states the algorithm was "trained on medical images," implying that these images were annotated or labeled with ground truth information (e.g., presence or absence of ICH) by experts, but the process or expert involvement is not described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1