Found 4 results

510(k) Data Aggregation

    K Number
    K251085

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-12

    (247 days)

    Product Code
    Regulation Number
    866.3987
    Age Range
    0.5 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iHealth Flu A&B/COVID-19/RSV Rapid Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, influenza B, SARS-CoV-2, and respiratory syncytial virus (RSV) protein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to influenza, SARS-CoV-2, and RSV can be similar.

    This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged six (6) months or older.

    All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, RSV, or other pathogens.

    Individuals who test negative and/or experience continued or worsening symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.

    Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

    The iHealth Flu A&B/COVID-19/RSV Rapid Test Pro is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, influenza B, SARS-CoV-2, and respiratory syncytial virus (RSV) protein antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to influenza, SARS-CoV-2, and RSV can be similar.

    This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged six (6) months or older.

    All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, RSV, or other pathogens.

    Individuals who test negative and/or experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.

    Positive results do not rule out co-infection with other respiratory pathogens.

    Test results should not be used as the sole basis for treatment or other patient management decisions.

    Device Description

    The iHealth Flu A&B/COVID-19/RSV Rapid Test and iHealth Flu A&B/COVID-19/RSV Rapid Test Pro (both versions hereafter also referred to as the candidate device) are lateral flow immunoassay device intended for the qualitative detection and differentiation of influenza A, influenza B, SARS-CoV-2, and respiratory syncytial virus (RSV) protein antigens.

    Two versions are available for this over-the-counter (OTC) test: one is labeled for lay-user use and one is labeled for professional use, both with identical designs. The candidate device detects antigens from influenza A, influenza B, SARS-CoV-2, and RSV in anterior nasal swabs collected from individuals exhibiting signs and symptoms of respiratory infection within the first six (6) days of symptom onset, when tested at least twice over a three-day period with a minimum of 48 hours between tests.

    The device is validated for testing direct anterior nasal samples (ANS) without the use of transport media. The device does not utilize biotin-streptavidin or avidin chemistry.

    The test card contains two test strips, each with a nitrocellulose membrane, enclosed in a plastic case. One test strip has four lines: three test lines (Flu A, Flu B, and CoV) and a control line (Ctrl), while the other test strip has two lines: one test line (RSV) and a control line (Ctrl).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251092

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-12

    (246 days)

    Product Code
    Regulation Number
    866.3987
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iHealth Flu A&B/COVID-19 Rapid Test:

    The iHealth Flu A&B/COVID-19 Rapid Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid protein directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar.

    This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.

    All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.

    Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.

    Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

    iHealth Flu A&B/COVID-19 Rapid Test Pro:

    The iHealth Flu A&B/COVID-19 Rapid Test Pro is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid protein directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar.

    This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.

    All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.

    Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.

    Positive results do not rule out co-infection with other respiratory pathogens.

    Test results should not be used as the sole basis for treatment or other patient management decisions.

    Device Description

    The iHealth Flu A&B/COVID-19 Rapid Test and iHealth Flu A&B/COVID-19 Rapid Test Pro is a lateral flow immunoassay device intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens.

    This over the counter (OTC) test has two versions, one labeled for self-testing use, the (iHealth Flu A&B/COVID-19 Rapid Test), and one labeled for professional use (iHealth Flu A&B/COVID-19 Rapid Test Pro) (generically referred to as iHealth Flu A&B/COVID-19 Rapid Test for the remainder of this document). Both versions of the iHealth Flu A&B/COVID-19 Rapid Test have an identical general design and are intended to separately detect antigen from influenza A, influenza B, and SARS-CoV-2 in anterior nares swabs from individuals with signs and symptoms of respiratory infection within the first five (5) days of symptom onset. The iHealth Flu A&B/COVID-19 Rapid Test is validated for testing direct anterior nares samples (ANS) without transport media. The iHealth Flu A&B/COVID-19 Rapid Test does not use biotin-streptavidin/avidin chemistry.

    The test card in the test kit is assembled with a test strip in a plastic housing that contains a nitrocellulose membrane with four lines: three test lines (Flu A line, Flu B line and SARS CoV-2 line) and a control line (Ctrl line).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243146

    Validate with FDA (Live)

    Device Name
    iCare APP
    Manufacturer
    Date Cleared
    2025-02-03

    (126 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iCare App is intended for use in the home and clinical settings as and their healthcare professionals to view test results which are measured by iHealth devices to better manage user's health and get feedback from their professional care team.

    The iCare App can also connect to medical devices and or non-medical devices and get data from devices during measurement or from the data stored in memory of the device for enhanced data managements. Data can be transmitted, displayed, and stored in the App.

    Device Description

    The iCare APP is a mobile application on both Android and iOS platforms.iCare allows users to better manage their own health by enabling them to measure their vital signs, access their results and relevant health information with just their smart device and internet connection, and receive feedback from their professional care team.

    iCare includes a patient darshboard featuring the Home, Health, Plus, Education, and Profile tabs. Accessory devices can be connected to the system to allow for collection of blood sugar, blood pressure, blood oxygen, and/or weight measurements. The patient darshboard functionality includes the ability to start measuring, allows users to view and track measurements, and export testing schedules for blood sugar, blood pressure, blood oxygen, and weight measurements; send messages to their professional care team; view previous appointment history information; view medication instructions; add entries to the food diary and review feedback from their registered dietician; set timers; and access articles and videos about health knowledge.

    AI/ML Overview

    The provided text is a 510(k) Summary for the iCare App, focusing on its substantial equivalence to a predicate device. It primarily details regulatory information, device description, and non-clinical test summaries. It does not contain information about a study that proves the device meets specific performance acceptance criteria for a medical diagnostic or screening function.

    The iCare App is classified as a "Medical Device Data System" (MDDS) that transmits, displays, and stores data from connected medical devices. Its function is to aid users and healthcare professionals in viewing test results for health management. It explicitly states: "Both devices make no interpretation, evaluation, medical judgments, or recommendations for treatment." This means the app itself doesn't perform diagnostic functions that would require specific performance metrics like sensitivity, specificity, or AUC against a ground truth.

    Therefore, many of the requested criteria, such as acceptance criteria for diagnostic performance, a test set, expert ground truth establishment, MRMC studies, or standalone algorithm performance, are not applicable or not provided in this document because the device is a data management system, not a diagnostic algorithm.

    Here's a breakdown of the applicable information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a table of quantitative performance acceptance criteria for diagnostic accuracy, sensitivity, or specificity, because the iCare App is an MDDS for data management, not a diagnostic tool. Instead, acceptance criteria are implied through the successful completion of non-clinical tests that demonstrate the basic functionality, safety, and effectiveness for its intended use as a data display and storage system.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Software Verification & ValidationCompliance with FDA guidance for "moderate" level of concern software; no minor injury to patient/operator due to failure or latent flaw."Software verification and validation has been performed according to FDA guidance... The iCare App software was considered a 'moderate' level of concern...". All tests passed.
    Wireless Coexistence TestAbility to be used in intended environments without harmful interference."Wireless coexistence test has been performed to verify that the subject device can be used in intended environments." All tests passed.
    CybersecurityAdherence to FDA guidance for cybersecurity; appropriate risk-based assessment and testing."Cybersecurity activities were conducted in accordance with FDA Guidance... The iCare App underwent appropriate risk-based cybersecurity assessment and testing..." All tests passed.
    Usability TestingSafe and effective use by lay users with provided labeling."Usability testing was conducted in accordance with FDA guidance... The test result demonstrates that the iCare App can be used by lay users with only provided labeling, the device is safe and effective for the intended use." All tests passed.

    2. Sample sized used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable for diagnostic performance as the device is not a diagnostic algorithm. The document mentions non-clinical testing (software, wireless, cybersecurity, usability) but does not specify "test set" sizes in the context of clinical data for diagnostic performance.
    • Data Provenance: Not applicable in the context of clinical diagnostic data. The document focuses on the technical aspects of the software.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable, as the device does not perform diagnostic interpretations requiring expert-established ground truth for clinical cases.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable, as there is no clinical test set requiring ground truth adjudication for diagnostic performance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC study was not done. The iCare App is an MDDS and does not involve AI assistance for human readers in a diagnostic capacity. It makes "no interpretation, evaluation, medical judgments, or recommendations for treatment."
    • Effect Size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: No, a standalone performance study in the context of diagnostic accuracy was not done. The device's function is data transmission, display, and storage, not diagnostic algorithm performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Not applicable for clinical diagnostic performance. For the software verification and validation, the "ground truth" would be the successful execution against specified requirements and accepted software engineering practices and FDA guidance.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable. This document does not describe a machine learning model that was trained on a dataset. The iCare App is a software application for data management, not an AI/ML algorithm requiring a training set of clinical data for diagnostic purposes.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment for Training Set: Not applicable, as there is no mention of a training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233842

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-31

    (179 days)

    Product Code
    Regulation Number
    866.3984
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iHealth COVID-19 Antigen Rapid Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19 within the first 6 days of symptom onset. This test is for nonprescription home use by individuals aged 15 years or older testing themselves, or adults testing individuals aged 2 years or older.

    The iHealth COVID-19 Antigen Rapid Test does not differentiate between SARS-CoV-2.

    All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as for treatment. Positive results do not rule out co-infection with other respiratory pathogens.

    This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider.

    The performance characteristics for SARS-CoV-2 were established from October 2022 to June 2023 when the COVID-19 variant Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.

    Device Description

    The iHealth COVID-19 Antigen Rapid Test employs lateral flow immunoassay technology. Using this test allows for the rapid detection of nucleocapsid protein from SARS-CoV-2.

    To begin the test, a self-collected anterior nares swab sample in individuals aged 15 and older or individuals between the age of 2 to 14 with a swab collected by a parent or quardian is inserted into a tube that has been pre-filled with reagent. The reagent in the tube interacts with the specimen and facilitates exposure of the appropriate viral antigens to the antibodies used in the test. The liguid in the tube, now containing the specimen, is added to the Sample Port of the COVID-19 Test Card.

    If the extracted specimen contains SARS-CoV-2 antigens, a pink-to-purple T Line, along with a pink-to-purple C Line will appear on the COVID-19 Test Card indicating a positive result. If SARS-CoV-2 antigens are not present at very low levels, only a pinkto-purple C Line will appear.

    AI/ML Overview

    Acceptance Criteria and Study to Prove Device Meets Criteria: iHealth COVID-19 Antigen Rapid Test

    The iHealth COVID-19 Antigen Rapid Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19 within the first 6 days of symptom onset. This test is for nonprescription home use by individuals aged 15 years or older testing themselves, or adults testing individuals aged 2 years or older.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit from provided data on clinical performance)Reported Device Performance (Clinical Study)
    Positive Percent Agreement (PPA): High agreement with highly sensitive molecular SARS-CoV-2 assays for positive samples.88.9% (95% Confidence Interval: 81.9% to 93.4%)
    Negative Percent Agreement (NPA): High agreement with highly sensitive molecular SARS-CoV-2 assays for negative samples.99.9%; (95% Confidence Interval: 99.3% to 100.0%)
    PPA by Days Since Symptom Onset: Expected performance across different days since symptom onset, especially higher PPA earlier in symptom onset.Day 1: 50.0% (Note: Low sample size); Day 2: 100.0%; Day 3: 91.9%; Day 4: 88.0%; Day 5: 75.0%; Day 6: 71.4%
    Precision/Reproducibility: Consistent and reliable results across different lots, operators, and sites, particularly for negative, low positive (1x LoD), and moderate positive (3x LoD) samples.100% agreement for all samples (negative, 1x LoD, 3x LoD) across 3 sites, 3 operators, and 3 lots over 5 non-consecutive days. (95% CI: 99.1-100%)
    Limit of Detection (LoD): Ability to detect low concentrations of SARS-CoV-2 virus.1.33x10^4 TCID50/mL (665 TCID50/swab) for Omicron Variant. 4.00x10^2 IU/mL (20 IU/swab) for WHO Standard.
    Cross-Reactivity (Analytical Specificity): No false positives or interference from common respiratory pathogens and commensal organisms.No cross-reactivity or interference observed with 53 commensal and pathogenic microorganisms (22 bacteria and 31 viruses) tested.
    Endogenous Interfering Substances: Performance not affected by common substances found in respiratory specimens or artificially introduced to the nasal cavity.No cross-reactivity or interference observed with 31 potentially interfering substances tested at relevant concentrations.
    Hook Effect: No decrease in signal at very high concentrations of the analyte.No Hook effect observed at 3.98x10^6 TCID50/mL (300x LoD), the highest concentration tested.
    Inclusivity (Analytical Reactivity): Ability to detect various SARS-CoV-2 variants.Successfully detected 14 SARS-CoV-2 strains including Alpha, Beta, Gamma, Delta, Omicron, and several Omicron sub-lineages (BA.2, BA.2.3, BA.4, BA.4.6, BA.5, BQ.1, BQ.1.1, XBB), with specified lowest variant concentrations for 5/5 positive replicates.
    Flex Studies: Robustness of the test under variations in test procedure and environmental conditions.Studies support that the test is robust with an insignificant risk of erroneous results under various conditions (swab extraction time, sample drops, development time, sample analysis delay, agitation, temperature, humidity, cassette disturbance, sample/reagent temperature, lighting).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 915 individuals.
    • Data Provenance:
      • Country of Origin: United States (investigational sites throughout the U.S.).
      • Retrospective or Prospective: Prospective. The clinical performance characteristics were evaluated in a study where subjects with signs and symptoms of COVID-19 were enrolled, collected samples, and the test was performed. Data was collected between October 2022 to June 2023.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of "experts" used for ground truth or their specific qualifications for the clinical study.

    • Ground Truth Method: The iHealth COVID-19 Antigen Rapid Test results were compared to a highly sensitive molecular FDA EUA Authorized SARS-CoV-2 assays (RT-PCR) to determine test performance. This molecular test is considered the gold standard for SARS-CoV-2 detection, establishing the definitive ground truth for the presence or absence of the virus.

    4. Adjudication Method for the Test Set

    The document indicates that for discordant results (false positives or false negatives against the initial comparator method), a "second FDA EUA high sensitivity molecular SARS-CoV-2 assay" was used. This suggests a form of 2+1 adjudication where the initial comparator result is one "read," the iHealth test result is the other, and a second molecular assay acts as the tie-breaker or confirmatory truth.

    • Specifically:
      • Of 13 false negative samples (iHealth negative, comparator positive), 1 was confirmed negative by a second molecular assay, and 12 were confirmed positive.
      • Of 1 false positive sample (iHealth positive, comparator negative), it was confirmed positive by a second molecular assay.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not explicitly done in the context of comparing human readers with and without AI assistance, as this is an antigen rapid test that is visually read by lay users. The study evaluated the standalone performance of the test as interpreted by lay users (self-testers or adult testers for children).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    For the clinical performance evaluation, the test was performed by lay users (individuals aged 15+ testing themselves, or adults testing individuals aged 2-14). Therefore, the clinical performance data presented (PPA, NPA) represents the device's performance with human-in-the-loop interpretation by the intended user. It is not a purely "algorithm-only" or "standalone" performance without any human involvement in interpreting the visual result. However, the device itself is a diagnostic test, not an AI/algorithm for image interpretation. The "algorithm" here (lateral flow immunoassay) is inherent to the device's physical mechanism.

    Other analytical studies (precision, LoD, cross-reactivity, etc.) represent technical performance without human interpretation error as these are typically performed in a controlled lab setting by trained personnel.

    7. The Type of Ground Truth Used

    The primary ground truth for the clinical study (testing individuals) was established using a highly sensitive molecular FDA EUA Authorized SARS-CoV-2 assay (RT-PCR). For discordant results, a second molecular assay was used for confirmation.

    8. The Sample Size for the Training Set

    This document describes a premarket notification for a medical device (an antigen rapid test). It does not explicitly mention a "training set" in the context of machine learning. The term "training set" is typically used for AI/ML models. If the underlying assay development involved any statistical modeling or calibration, the data used for that would precede this type of submission. However, for a lateral flow immunoassay, the "training" involves optimizing the biochemical reagents and visual readout.

    9. How the Ground Truth for the Training Set Was Established

    Given that this is a lateral flow immunoassay device and not an AI/ML diagnostic, the concept of a "training set" with established ground truth as commonly understood in AI/ML is not directly applicable. The "ground truth" during the development and optimization phases would involve rigorous analytical validation using known positive and negative samples, purified viral cultures, and clinical samples characterized by gold standard molecular tests (like RT-PCR), similar to the analytical studies detailed (e.g., LoD, inclusivity, cross-reactivity) but performed during product development before clinical trials.