510(k) Data Aggregation

    K Number
    K232163

    Validate with FDA (Live)

    Date Cleared
    2024-02-22

    (216 days)

    Product Code
    Regulation Number
    884.6100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lotus Single Lumen Ovum Aspiration Needle is used for ultrasound guided transvaginal aspiration of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Lotus Double Lumen Ovum Aspiration Needle is used for ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Device Description

    The subject devices are single use sterile devices used for ultrasound-guided transvaginal aspiration of oocytes from ovarian follicles.

    The Lotus Single Lumen Ovum Aspiration Needle consists of a stainless steel needle with a handle, which is connected to an aspiration tube with a silicone stopper at the other end. The stopper can fit onto a test tube where the oocytes are collected. The stopper is connected to a vacuum tube with a vacuum tubing connector at the other end. The vacuum tubing connector is a standard luer lock connector that enables connection to a vacuum pump to provide aspiration suction. The Lotus Single Lumen Ovum Aspiration Needle is 35 cm in length and has device variations from 21 to 16 gauge with aspiration tubing of 600 or 900 mm in length.

    The Lotus Double Lumen Ovum Aspiration Needle consists of a stainless steel needle with a Y shaped handle. The Y-shaped handle has two ports: the central port through which oocytes are aspirated via the aspiration lumen, and a secondary side port to allow flushing of follicles via the flushing lumen. The central port is connected to an aspiration tube with a silicone stopper at the other end. The stopper can fit onto a test tube where the oocytes are collected. The stopper has a vacuum connector that is a standard luer lock connector that enables connection to a vacuum pump to provide aspiration suction. The side port is connected to a flushing tube with a flushing connector at the other end, which is a standard luer lock connector that can be connected to a compatible syringe for flushing of follicles. The Lotus Double Lumen Ovum Aspiration Needle is 17 gauge with an aspiration tubing length of 900 mm and with device variations of 30 or 35 mm needle length and flushing tubing of 700 or 1000 mm in length.

    AI/ML Overview

    This document is an FDA 510(k) Premarket Notification for medical devices, specifically ovum aspiration needles. It does not describe a study involving an AI/Machine Learning device. Therefore, it is not possible to extract information about acceptance criteria or study details relevant to AI/ML device performance from this text.

    The document discusses traditional medical device testing, such as:

    • Biocompatibility testing: Cytotoxicity, Sensitization, Intracutaneous Irritation.
    • Sterilization testing: Ethylene Oxide (EO) sterilization, residual testing.
    • Simulated transportation: Adherence to ASTM D4169-22.
    • Stability and Shelf Life testing: accelerated aging, package integrity, mechanical performance (appearance, dimensions, needle stiffness, breakage resistance, corrosion resistance, tip penetration force, needle bond strength, flow rate, ultrasound detectability, tubing bond strength, leak, aspiration test, component compatibility).
    • Mouse Embryo Assay (MEA): This is a biological test to assess the non-toxicity of materials to embryos, not an AI performance metric. The acceptance criterion for 1-Cell MEA was ≥80% embryos developed to expanded blastocyst at 96 hours after 30 minute exposure.
    • Endotoxin evaluation: Acceptance criterion was ≤ 20 EU/device.

    Since the request specifically asks for details related to AI/Machine Learning device performance, and this document pertains to a traditional, non-AI medical device, I cannot fulfill the request as stated.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223265

    Validate with FDA (Live)

    Date Cleared
    2023-05-11

    (199 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CryoX™ Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.

    CryoX™ Vitrification Thaw Kit is intended for use in the thawing of virified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.

    Device Description

    CryoX™ Vitrification Freeze Kit is designed to facilitate dehydration of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos before vitrification via rapid cooling in liquid nitrogen. CryoX™ Vitrification Freeze Kit contains two solutions to be used sequentially during vitrification. Both solutions consist of Medium 199 (M199), human serum albumin (HSA) and gentamicin sulfate. They also contain varying levels of cryoprotectants, including dimethyl sulfoxide (DMSO), ethylene glycol (EG), and sucrose.

    CryoX™ Vitrification Thaw Kt contains three solutions to be used sequentially during oocyte and embryo thawing procedures. All three solutions contain M199, HSA, and gentamicin sulfate. They also contain decreasing concentrations of cryoprotectant.

    The five solutions in the CryoX™ Vitrification Freeze Kit and Thaw Kt are aseptically filtered and provided in 4.5 mLPETG vials. The solutions in these kits are single-use only. They have a shelf-life of 6 months when stored at 2-8℃

    AI/ML Overview

    The document you provided is a 510(k) Premarket Notification from the FDA for a device called "CryoX™ Vitrification Freeze Kit / Thaw Kit." It details the device's indications for use, comparison to a predicate device, and a summary of non-clinical performance testing.

    Based on the provided text, here's the information regarding acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are the product specifications that were tested during shelf-life testing. The reported device performance is indicated by the statement that these specifications "were met at time 0 and after accelerated aging."

    ParameterAcceptance CriteriaReported Device Performance
    AppearanceClean, transparent, pink; no impuritiesMet (implicitly, as specifications were met)
    pH (per USP <791>)7.2-7.6Met (implicitly, as specifications were met)
    Osmolality (mOsmol/kg)ES: 8551042 (1:2 dilution) VS: 19162477 (1:2 dilution) TS: 16532430 DS: 8711025 WS: 307~318Met (implicitly, as specifications were met)
    Sterility (per USP <71>)No microbial growthMet (implicitly, as specifications were met)
    Bacterial Enodtoxin (per USP <85>)< 0.5 EU/mLMet (implicitly, as specifications were met)
    Mouse Embryo Assay (1-Cell MEA)≥ 80% embryos developed to expanded blastocyst at 96 hoursMet (implicitly, as specifications were met)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set for any of the performance tests (e.g., pH, osmolality, sterility, bacterial endotoxin, MEA).

    The data provenance is from non-clinical performance testing conducted by the manufacturer, Zhejiang Horizon Medical Technology Co., Ltd. The document does not specify the country of origin of the data beyond the manufacturer's location in China, nor does it explicitly state whether the studies were retrospective or prospective, though performance testing for product validation is typically prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish ground truth for the performance tests. These tests are primarily laboratory-based measurements of chemical and biological properties according to established standards (e.g., USP, ASTM, FDA guidance). The Mouse Embryo Assay (MEA) results are quantitative and follow a specified threshold rather than expert consensus.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set. Since the tests are largely objective measurements against defined numerical or qualitative criteria (e.g., pH range, endotoxin limit, percentage of blastocyst development), human adjudication in the typical sense (e.g., 2+1 reading) is not applicable.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically conducted for diagnostic imaging devices or other AI-driven tools where human interpretation plays a significant role. The CryoX™ Vitrification Freeze Kit / Thaw Kit is a reproductive media product, and its evaluation focuses on its chemical and biological properties and performance in preserving and thawing oocytes/embryos, not on human interpretive tasks.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    This question is not applicable to the device described. The CryoX™ Vitrification Freeze Kit / Thaw Kit is a biological reagent kit, not an algorithm or AI-driven device. Its performance is intrinsic to the chemical formulation and its biological effect on cells, not an algorithmic output.

    7. The Type of Ground Truth Used

    The ground truth for the performance tests is based on defined quantitative and qualitative specifications derived from established standards and biological efficacy requirements. For example:

    • pH, Osmolality, Bacterial Endotoxin: Numerical ranges or limits according to USP monographs.
    • Sterility: Absence of microbial growth as per USP <71>.
    • Mouse Embryo Assay (MEA): A specific developmental endpoint (≥ 80% expanded blastocyst development) at 96 hours, as per FDA guidance.
      These are objective, rather than subjective, ground truths typically established by consensus or pathology.

    8. The Sample Size for the Training Set

    The concept of a "training set" is not applicable to this device. Training sets are relevant for machine learning or AI models. This product is a medical device in the category of reproductive media, and its development and validation involve formulation, manufacturing, and performance testing, not algorithmic training.

    9. How the Ground Truth for the Training Set Was Established

    As the concept of a "training set" does not apply to this device, the question of how its ground truth was established is not applicable.