510(k) Data Aggregation

    K Number
    K240563

    Validate with FDA (Live)

    Date Cleared
    2024-10-08

    (222 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    This Power Wheelchair, model: JJW-6001, is a motor driven, indoor and outdoor transportation vehicle, which a device for assisting action handicapped people and disabled people to move. It is suitable for disabled people with mobility difficulties and elderly people. The device consists of front wheel, drive wheel, frame, controller, motor, armrest, backrest, seat cushion, seatbelt, pedal, battery box and charger. The device is powered by Li-ion Battery pack with 16 Km range, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use. The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the U.S. FDA for a Power Wheelchair (JJW-6001). It outlines the device's characteristics, comparison to a predicate device, and the non-clinical tests performed.

    This document pertains to the physical performance and safety of a medical device (a power wheelchair), not the performance of an AI/ML algorithm or software. Therefore, the questions related to AI/ML specific criteria (such as MRMC studies, ground truth for AI, training/test set sample sizes for AI, number of experts for AI ground truth, or adjudication methods for AI performance) are not applicable to the information contained in this document.

    The document describes non-clinical tests conducted to confirm the device's safety and effectiveness compared to a predicate device, adhering to various ISO standards relevant to wheelchairs and electrical medical equipment.

    Here's the relevant information based on the provided text, addressing the applicable questions:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally implied by demonstrating compliance with various ISO standards. The reported device performance is that it met the design specifications by passing these tests. Specific numerical acceptance criteria and reported values for each performance metric are not explicitly presented in a single table as distinct "acceptance criteria" vs. "reported performance" for each individual test, but rather as statements of compliance.

    However, a comparison table (Table 1) provides some performance specifications for both the subject device and the predicate device.

    Table of Device Performance (from Table 1, "General Comparison")

    Element of ComparisonPredicate Device (K220747)Subject Device (JJW-6001)Remark / Compliance
    Max Speed forwardUp to 6 km/h (1.6 m/s), adjustableUp to 6 km/h (1.6 m/s), adjustableS.E. (Substantially Equivalent)
    Max Speed backwardLess than 3 km/h (0.5 m/s)Less than 3 km/h (0.5 m/s)S.E.
    Braking distance$\leq$ 1.5 m$\leq$ 1.5 mS.E.
    Maximum safe operational incline degree10°Analysis (presumably found acceptable after testing)
    Maximum distance of travel on the fully charged battery15 km16kmAnalysis (presumably found acceptable after testing)
    Maximum obstacle climbing40 mm25mmAnalysis (presumably found acceptable after testing)

    Implied Acceptance Criteria for Safety and Performance:
    The document states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:" followed by a list of ISO and IEC standards (e.g., ISO 7176 series for wheelchairs, ISO 10993 series for biocompatibility, IEC 60601-1-2 for electromagnetic compatibility). The acceptance criteria for these tests are the requirements specified within these standards. The device's reported performance is that it met these standards.

    2. Sample size used for the test set and the data provenance

    The document does not specify a "sample size" in terms of number of patients or data records, as it refers to engineering and non-clinical performance testing of a physical medical device. The "test set" would be the physical JJW-6001 device itself. The data provenance is from non-clinical laboratory testing performed to demonstrate compliance with international standards. The origin of the testing data is implied to be in China, as the manufacturer and submission correspondent are based there. The tests are "non-clinical," meaning they are not performed on human subjects and thus do not have retrospective or prospective patient data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a physical device submission, not an AI/ML or diagnostic imaging submission. Ground truth for device performance is established by standardized test procedures (e.g., ISO 7176, ISO 10993).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is a physical device and not an AI/ML algorithm requiring human expert adjudication of output. Performance is determined by compliance with standard measurements and specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical device, not an AI-assisted diagnostic tool. No human reader study information is provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical non-AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is compliance with established international standards (e.g., ISO 7176 series for wheelchair performance, ISO 10993 series for biocompatibility). These standards define the test procedures and acceptance criteria for various physical and safety aspects of the device.

    8. The sample size for the training set

    Not applicable. This is a non-AI physical device. No training set is described.

    9. How the ground truth for the training set was established

    Not applicable. No training set or associated ground truth establishment is described for this physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240255

    Validate with FDA (Live)

    Date Cleared
    2024-08-07

    (189 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    This Electric Wheelchair (JLE-W01A-2023), is a motor driven, indoor and outdoor transportation vehicle, which a device for assisting action handicapped people and disabled people to move. It is suitable for disabled people with mobility difficulties and elderly people.

    The device consists of front wheel, drive wheel, frame, controller, motor, armrest, backrest, seat cushion, seatbelt, pedal, battery box, charger and anti-tip wheels.

    The device is powered by Li-ion Battery pack with 27.5 Km range, which can be recharged by an offboard battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use.

    The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick, the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

    Anti-tip wheel is a safety device to prevent the wheelchair from turning over when driving on the slope. When the operator needs to use the subject electric wheelchair, he must pull down the anti-tip wheels.

    AI/ML Overview

    This document is a 510(k) summary for an Electric Wheelchair (JLE-W01A-2023). It describes the device, its intended use, and substantial equivalence to a predicate device (K220747).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Electric Wheelchair are primarily based on compliance with various ISO standards for wheelchairs and related components, as well as biocompatibility and EMC standards. The reported device performance is that the device "meets its design specification" or "meets the requirements" of these standards.

    Acceptance Criteria (Standard)Description of CriterionReported Device Performance (JLE-W01A-2023)
    Biocompatibility
    ISO 10993-5: 2009Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro CytotoxicityAll user directly contacting materials are in compliance with ISO10993-5 requirements.
    ISO 10993-10: 2021Biological Evaluation of Medical Devices Part 10: Tests For Irritation And Skin SensitizationAll user directly contacting materials are in compliance with ISO10993-10 requirements.
    ISO 10993-23: 2021Biological evaluation of medical devices - Part 23: Tests for irritationAll user directly contacting materials are in compliance with ISO 10993-23 requirements.
    Wheelchair Performance
    ISO 7176-1: 2014Wheelchairs - Part 1: Determination of static stabilityThe static stability has been determined after the testing according to ISO 7176-1, and test results meet its design specification.
    ISO 7176-2: 2017Wheelchairs Part 2: Determination of dynamic stability of Powered WheelchairsThe dynamic stability has been determined after the testing according to ISO 7176-2, and test results meet its design specification.
    ISO 7176-3: 2012Wheelchairs - Part 3: Determination of effectiveness of brakesThe effectiveness of brakes has been determined after the testing according to ISO 7176-3, and test results meet its design specification. (Note: Braking distance of 1.2m stated in comparison table).
    ISO 7176-4: 2008Wheelchairs Part 4: Energy consumption of electric wheelchairs and wheelchairs for determination of theoretical distance rangeThe theoretical distance range has been determined after the testing according to ISO 7176-4, and test results meet its design specification. (Note: Max distance of travel on fully charged battery: 27.5 km stated in comparison table).
    ISO 7176-5: 2008Wheelchairs Part 5: Determination of overall dimensions, mass and manoeuvring spaceThe dimensions, mass has been determined after the testing according to ISO 7176-5. (Note: Overall and Folded dimensions, and Turning Radius stated in comparison table).
    ISO 7176-6: 2018Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of Powered WheelchairsThe dimensions, mass has been determined after the testing according to ISO 7176-6. (Note: Max speed forward and backward stated in comparison table).
    ISO 7176-7Wheelchairs - Part 7: Measurement of seating and wheel dimensionsThe seating and wheel dimensions has been determined after the testing according to ISO 7176-7. (Note: Wheel sizes stated in comparison table).
    ISO 7176-8: 2014Wheelchairs Part 8: Requirements and test methods for static, impact and fatigue strengthsAll test results meet the requirements in Clause 4 of ISO 7176-8.
    ISO 7176-9: 2009Wheelchairs Part 9: Climatic tests for Powered WheelchairsThe test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9.
    ISO 7176-10: 2008Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairsThe obstacle-climbing ability of device has been determined after the testing according to ISO 7176-10. (Note: Maximum obstacle climbing of 35mm stated in comparison table).
    ISO 7176-11: 2012Wheelchairs Part 11: Test dummiesThe test dummies used in the testing of ISO 7176 series meet the requirements of ISO 7176-11.
    ISO 7176-13: 1989Wheelchairs - Part 13: Determination of coefficient of friction of test surfacesThe coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved.
    ISO 7176-14: 2008Wheelchairs Part 14: Power and control systems for electrically powered wheelchairs and wheelchairs - Requirements and test methodsAll test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14.
    ISO 7176-15: 1996Wheelchairs - Part 15: Requirements for information disclosure, documentation and labelingThe test results shown that information disclosure, documentation and labeling of device meet the requirements of ISO 7176-15. (Also stated: Conforms to FDA Regulatory).
    ISO 16840-10: 2021 (or ISO 7176-16)Wheelchair seating Part 10: Resistance to ignition of postural support devices Requirements and test method. (Note: Predicate refers to ISO 7176-16).The performance of resistance to ignition meet the requirements of ISO 16840-10.
    Electromagnetic Compatibility
    ISO 7176-21: 2009 / IEC 60601-1-2Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and wheelchairs, and battery chargersThe EMC performance results meet the requirements of ISO 7176-21 and IEC 60601-1-2.
    Battery and Charger
    ISO 7176-25: 2022Wheelchairs Part 25: Lead-acid batteries and chargers for powered wheelchairs - Requirements and test methods (Note: The proposed device uses Li-ion battery, but the standard title mentions Lead-acid. It's likely the standard covers general battery/charger safety).The Lead-acid batteries and chargers results meet the requirements of ISO 7176-25.
    IEC 62133-2Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium SystemsThe device complies with IEC 62133-2. (Specifically for the Li-ion battery).

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample size used for the test set in the non-clinical tests. It indicates that "Non-clinical tests were conducted to verify that the proposed device met all design specifications" and mentions "test records" and "test results" without quantifying the number of devices or components tested.

    The data provenance is from China, as the applicant (Zhejiang Jianrui Technology Company Limited) is located in Jinhua City, Zhejiang Province, China, and the tests were conducted to support their 510(k) submission to the FDA. The tests are prospective in nature, as they were conducted on the specific device to demonstrate its performance against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The "ground truth" for these types of device performance tests is established by adherence to globally recognized engineering and medical device safety standards (like the ISO 7176 series and ISO 10993 series). The standards themselves define the test methodologies and acceptance criteria. The execution and interpretation of these tests are typically performed by qualified test engineers and technicians at accredited testing laboratories, rather than by medical "experts" in the clinical sense to establish a "ground truth."

    4. Adjudication Method for the Test Set

    This is not applicable for the type of non-clinical, standards-based testing described. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or image interpretation studies where expert consensus is needed to establish a definitive diagnosis or assessment for a complicated case. Here, the "adjudication" is implicitly covered by the adherence to the published test methodologies and criteria of the ISO/IEC standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    An MRMC study was not done. This submission is for an Electric Wheelchair, a physical medical device. It does not involve AI or human readers for diagnostic interpretation, so an MRMC comparative effectiveness study is not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    This is not applicable. The device is an Electric Wheelchair and does not feature an AI algorithm that would have standalone performance.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical performance and safety tests is based on compliance with established international standards (ISO 7176 series, ISO 10993 series, IEC 62133-2). These standards define objective, measurable parameters and acceptable limits for various aspects of the device's function, stability, safety, and biocompatibility.

    8. The Sample Size for the Training Set

    This is not applicable. The Electric Wheelchair is a physical device and not a machine learning or AI model, so there is no "training set" in the context of data.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132257

    Validate with FDA (Live)

    Date Cleared
    2014-02-13

    (209 days)

    Product Code
    Regulation Number
    890.3850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. It is designed for use indoors and outdoors, over smooth surfaces (all standard indoor flooring surfaces, concrete, asphalt and packed dirt) that are free of large obstacles and inclines greater than 8 degrees.

    Device Description

    A mechanical wheelchair is a chair with wheels, designed to be a replacement for walking, where it is propelled by the seated occupant turning the rear wheels by hand. There are also handles behind the seat for someone else to do the pushing. Wheelchairs are used by people for whom walking is difficult or impossible due to illness, injury, or disability.

    The Manual Wheelchairs Model SY 100-MA02 (02A & 02B) and Model SY 100-MA06 (or other clients private labeling) are user propelled, manually operated folding wheelchairs that are indicated for use indoors and outdoors, over smooth surfaces (all standard indoor flooring surfaces, concrete, asphalt and packed dirt) that are free of large obstacles and inclines greater than 8 degrees. Each consists of four wheels, a mechanical steel frame and nylon upholstery that is ignition resistant. The device is designed to be lightweight and foldable, and can be dissembled. In both models, both rear and caster wheels use solid tires.

    The model SY100-MA02 has two sub-models (02A & 02B) which both has a physical dimension of 1025 (depth) x 661 (width) x 920 (height) mm, with the seat itself has a dimension of 408 (depth) x 440 (width) x 455 (height) mm. The device has a weight capacity of 130 kilograms.

    Sub-model MA02A uses hard plastic back wheels and weighs approximately 16.5 kilograms. Sub-model MA02B uses wired steel wheels and weighs 13.8 kilograms. Both sub-models are dark green in color.

    Model SY100-MA06 has a physical dimension of 1026 (depth) x 651 (width) x 860 (height) mm. with the seat itself has a dimension of 394 (depth) x 374 (width) x 482 (height) mm, and a weight capacity of 100 kilograms, and weighs approximately 15.2 kilograms.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a manual wheelchair and does not contain information about an AI/ML powered device. Therefore, many of the requested categories in the prompt are not applicable.

    However, I can extract information related to the acceptance criteria and the study performed for this manual wheelchair.

    1. Table of Acceptance Criteria and Reported Device Performance

    DescriptionAcceptance Criteria (Test Standard)Reported Device Performance
    Static StabilityMeets ISO 7176-1:1999Meets ISO 7176-1:1999
    Effectiveness of BrakesMeets ISO 7176-3: 2003Meets ISO 7176-3: 2003
    Static StrengthMeets ISO 7176-8: 1998Meets ISO 7176-8: 1998
    Impact StrengthMeets ISO 7176-8: 1998Meets ISO 7176-8: 1998
    Fatigue StrengthMeets ISO 7176-8:1998Meets ISO 7176-8:1998
    Resistance to IgnitionMeets ISO 7176-16:1997Meets ISO 7176-16:1997
    Weight Capacity (Model SY100-MA02)130 Kg130 Kg
    Weight Capacity (Model SY100-MA06)100 Kg100 Kg

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document refers to "Our Devices" (Manual Wheelchair Model SY100-MA02 (02A & 02B) and Model SY100-MA06) as having undergone testing. It's implied that at least one of each model/sub-model was tested.
    • Data Provenance: The tests were conducted following recommended procedures outlined in consensus standards (ISO 7176 series). The document does not specify the country of origin for the testing or whether the data was retrospective or prospective, but it's non-clinical testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is a non-clinical performance assessment of a mechanical device against international standards, not an AI/ML device requiring expert ground truth for image interpretation or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is a technical performance test, not a subjective assessment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI-powered device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for this medical device is adherence to established international performance standards (ISO 7176 series) for mechanical wheelchairs. These standards define the acceptable physical and functional characteristics.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device that requires a training set.