Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K051723

    Validate with FDA (Live)

    Date Cleared
    2005-07-15

    (18 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves (white, yellow) that meets all of the requirements of ASTM standard D 5250-0084.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on what is stated and noting areas where the information is not present:

    Acceptance Criteria and Device Performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM standard D 5250-00e4Meets
    Physical PropertiesASTM standard D 5250-00e4Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-00e4<2mg/glove
    Biocompatability - Primary Skin Irritation in rabbitsISO10993-10Passes (Not a Primary Skin Irritation)
    Biocompatability - Dermal sensitization in the guinea pigISO10993-10Passes (Not a Dermal sensitization)

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Sample size: Not explicitly stated. The document refers to meeting standards (ASTM, FDA) which would imply a defined sample size within those standards, but the specific quantity tested for each characteristic is not provided in this summary.
      • Data provenance: Not explicitly stated. The studies are non-clinical, likely conducted by the manufacturer or a third-party lab to demonstrate compliance with the referenced ASTM and ISO standards. The country of origin for the data (beyond the manufacturer's location in China) is not mentioned. It can be inferred that these are prospective tests performed on the device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable/Not stated. For this type of device (patient examination gloves) and the tests conducted (physical and biocompatibility), "experts" in the sense of clinical reviewers establishing ground truth is not typically relevant. The ground truth for these tests is defined by the objective measurement criteria within the ASTM and ISO standards (e.g., maximum allowable pinholes, specific tensile strength values, observed skin reactions in animal models).
    3. Adjudication method for the test set:

      • Not applicable/Not stated. Adjudication methods (like 2+1, 3+1) are typically used in studies where human interpretation of medical images or data is involved to establish a consensus ground truth. For the physical and biocompatibility testing of gloves, the results are objectively measured against predefined criteria in the standards.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a physical patient examination glove, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies or AI assistance metrics are not relevant.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a physical product, not an algorithm, so standalone performance in the context of AI is not relevant.
    6. The type of ground truth used:

      • Objective Measurement/Standard Compliance: The ground truth for the performance criteria (dimensions, physical properties, freedom from pinholes, powder residual) is established by the specific, objective measurement criteria and threshold values defined within the ASTM standard D 5250-00e4 and 21 CFR 800.20.
      • Biological Response in Models: For biocompatibility, the ground truth is based on the observed biological responses in the animal models (rabbits for skin irritation, guinea pigs for dermal sensitization) according to the protocols outlined in ISO10993-10. "Passing" the test means the observed reaction falls below predefined thresholds for irritation or sensitization.
    7. The sample size for the training set:

      • Not applicable. This device is a physical product, not a machine learning model, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K051148

    Validate with FDA (Live)

    Date Cleared
    2005-06-21

    (48 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-0064.

    AI/ML Overview

    The provided document describes the "Powder Free Vinyl Patient Examination Gloves, white (non-colored)" and its premarket notification (510(k)) summary. It focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive study of device performance against specific acceptance criteria using a test set. This type of submission relies on established standards and non-clinical testing.

    Here's an analysis of the available information based on your requested categories:

    1. A table of acceptance criteria and the reported device performance

    CharacteristicsAcceptance Criteria (Standard)Device Performance
    DimensionASTM standard D 5250-00e4Meets
    Physical PropertiesASTM standard D 5250-00e4Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-00e4 and D6124-01Meets (<2mg/glove)
    BiocompatibilityPrimary Skin Irritation in rabbitsPasses (Not a Primary Skin Irritation)
    BiocompatibilityDermal sensitization in guinea pigPasses (Not a Dermal sensitization)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for the testing mentioned (e.g., for dimension, physical properties, pinholes, powder residual, or biocompatibility). It only indicates that the device "Meets" the respective standards. The data provenance is not specified, but the device manufacturer is ZIBO SINOCARE PLASTIC PRODUCTS CO., LTD in China. The testing would have been conducted by or for this company. This testing would be considered prospective, as it's conducted to demonstrate compliance for a new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the device's performance is assessed against established ASTM and CFR standards, not through expert consensus on a subjective test set. The standards themselves define objective metrics. For biocompatibility, animal models (rabbits and guinea pigs) were used, not human experts to establish "ground truth" in the way one would for image interpretation, for example.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Device performance is measured directly against technical standards, not through expert adjudication of subjective results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a disposable medical glove, not an AI-powered diagnostic or assistive technology that would involve human readers or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable to a physical medical device like a glove. The "standing alone" performance refers to the device's inherent characteristics as measured against specified standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for evaluating the performance of these gloves is defined by:

    • Established ASTM Standards: These provide quantitative and qualitative specifications for dimensions, physical properties (e.g., tensile strength, elongation), and powder residual.
    • Regulatory Standards (21 CFR 800.20): Specifically for freedom from pinholes, which dictates an Acceptable Quality Level (AQL) for water leak testing.
    • Biocompatibility Standards (ISO10993-10): These guide the animal testing for primary skin irritation and dermal sensitization. The "ground truth" for these tests is the observed biological response in the animal models according to the standard's interpretation guidelines (e.g., absence of irritation).

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of this device. The physical characteristics and biological safety of gloves are determined through standardized testing, not machine learning or algorithm training.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051152

    Validate with FDA (Live)

    Date Cleared
    2005-06-03

    (30 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00-4.

    AI/ML Overview

    This document is a 510(k) Premarket Notification summary for medical gloves, not a study of an AI device. As such, it does not contain the information requested in your prompt regarding AI device acceptance criteria and study details.

    However, I can extract the relevant information about the device performance and acceptance criteria for the medical gloves as presented in this document.

    1. A table of acceptance criteria and the reported device performance:

    CharacteristicsStandard / Acceptance CriteriaDevice Performance
    DimensionASTM standard D 5250-00e4Meets
    Physical PropertiesASTM standard D 5250-00e4Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder AmountASTM standard D 5250-00e4<10mg/dm2
    Biocompatibility (Primary Skin Irritation)Primary Skin Irritation in rabbitsPasses (Not a Primary Skin Irritation)
    Biocompatibility (Dermal Sensitization)Dermal sensitization in the guinea pigPasses (Not a Dermal Sensitization)

    Regarding the AI-specific questions (2-9), this document does not contain that information because it pertains to a medical glove, not an AI device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1