510(k) Data Aggregation

    K Number
    K233859

    Validate with FDA (Live)

    Device Name
    BASIS FLOW II
    Date Cleared
    2024-03-05

    (90 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BASIS FLOW II is intended for fabrication or repair of the denture base.

    Device Description

    BASIS FLOW II is supplied as components in powder and liguid form. The powder is primarily a polymer of polymethyl methacrylate (PMMA) with small quantities of initiator and color pigments. The liquid is primarily the monomer methyl methacrylate (MMA) with small quantities of a cross-linking agent and activator.

    To fabricate the denture base, the powder and liguid materials are mixed and stirred to create a dough state that is packed or poured into a mold, saddle, or core. The resin then cures in a self-curing process (pressure vessel or via quick-setting).

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding a dental resin, not an AI/ML device. Therefore, the requested information about acceptance criteria, study details, ground truth, and expert involvement for an AI/ML device is not applicable and cannot be extracted from this document.

    The document discusses the substantial equivalence of BALANCE FLOW II (subject device) to BALANCE FLOW (predicate device) based on physical and chemical properties and adherence to ISO standards for denture base polymers.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172683

    Validate with FDA (Live)

    Date Cleared
    2018-01-16

    (132 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Yamahachi Pink CAD/CAM Disk is used for the fabrication of dentures.

    PMMA Disks are polymethyl methacrylate blanks used to mill dentures in various CAD/CAM systems.

    Device Description

    Yamahachi PMMA Disks are acrylic blanks used in dental CAD/CAM milling systems by professional dental technicians to fabricate dentures. The restoration process begins in the dental office with an intraoral scan of the affected tooth and the surrounding teeth. Proprietary software takes this digital image and creates a replacement part for the missing areas of the tooth, creating a virtual restoration. The software sends this virtual data to a milling machine where the replacement part is carved out of a Yamahachi PMMA Disk. The optical impression system, software, and milling machine are not included in this submission.

    The device is composed of hot-cured polymethyl methacrylate (PMMA) and pigments.

    Yamahachi Pink Disks are available in different shapes and shades, with disks of all shades available in different dimensions (diameter, thickness and profile margin). It is intended for use in both open, closed and block type CAD/CAM systems. The circular disk model fits open milling systems, while the configurations for closed systems include Amann Girrbach, Zirkonzahn and block type can be used for any milling machine with the appropriate and compatible CAM system.

    AI/ML Overview

    This document is a 510(k) premarket notification for a dental device, the Yamahachi Pink CAD/CAM Disk. It focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the device meets acceptance criteria derived from a study that quantifies specific performance metrics in a clinical context (e.g., accuracy, sensitivity, specificity for a diagnostic device).

    Therefore, I cannot provide a direct answer to all parts of your request as it pertains to a medical device software approval process involving performance metrics, expert reads, and clinical study data. This document describes a material (PMMA disk) used for fabricating dentures, and its approval relies heavily on demonstrating equivalent physical and chemical properties and intended use to existing, approved materials.

    However, I can extract the relevant information and explain why other sections of your request are not applicable for this type of device submission based on the provided text.

    Here's a breakdown of what can be extracted and what cannot:

    Device: Yamahachi Pink CAD/CAM Disk
    Regulation Number: 21 CFR 872.3760
    Regulation Name: Denture relining, repairing, or rebasing resin
    Product Code: EBI
    Regulatory Class: Class II


    1. A table of acceptance criteria and the reported device performance

    For a material like this, the "acceptance criteria" are typically the standards conformance and physical/chemical property equivalence to a predicate, not performance metrics like sensitivity or specificity.

    Acceptance Criteria (from ISO Standards / Predicate Equivalence)Reported Device Performance (Yamahachi Pink CAD/CAM Disk)
    Physical Properties (ISO 10477:2004)Met acceptance criteria of this standard.
    Appearance (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Size/Diameter (min to max) (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Thickness (min to max) (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Color of Shades (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Flexural Strength (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Flexural Modulus (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Water Sorption (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Solubility (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Color Stability (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Translucency (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    Residual MMA (ISO 20795-1:2013)Tested and evaluated (implied acceptable as part of meeting standard).
    BiocompatibilityPreviously tested and established with FDA with cleared device K131036 (reference predicate).
    Shelf Life10 Years

    Note on "Acceptance Criteria" for this device: The primary "acceptance criteria" for a 510(k) submission like this are demonstrating:

    • Same intended use as a predicate.
    • Same technological characteristics (or differences that do not raise new questions of safety or effectiveness).
    • Conformance to relevant performance standards (ISO standards in this case).

    Regarding the "Study that proves the device meets the acceptance criteria":

    The "study" here is non-clinical performance testing and comparison to predicates. It's not a clinical trial in the sense of a diagnostic AI device.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in the document for the material testing. ISO standards often specify minimum sample sizes for various tests, but the report only states "The subject and predicate have both completed ISO standardized testing and have passed."
    • Data Provenance: The material is manufactured by Yamahachi Dental Mfg., Co. in Japan. The testing was non-clinical (laboratory-based) and conducted to ISO standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is not a diagnostic device where "ground truth" is established by expert readers. The "ground truth" for a material like this is defined by adherence to validated ISO standards for physical and chemical properties.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No human adjudication of results in the context of clinical reads is relevant for this material.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not a diagnostic AI device subject to MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not a medical device algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" is adherence to international consensus standards (ISO 10477:2004, ISO 20795-1:2013) for dental polymeric materials, combined with biocompatibility data from a previously cleared device (K131036) with identical chemical composition.

    8. The sample size for the training set

    • Not Applicable. This device is a material, not an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.