510(k) Data Aggregation

    K Number
    K170709

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-06-06

    (90 days)

    Product Code
    Regulation Number
    878.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CURVE Laser System is indicated for non-invasive aesthetic treatment for the temporary reduction in circumference of the waist.

    Device Description

    The CURVE Laser System consists of a main console and 4 treatment paddles. The console houses the main electronics, controls and embedded software. The liquid crystal display (LCD) identifies all key treatment parameters. The main console is also equipped with a micro controller that provides automatic calculation of energy output for a specific set of treatment parameters. The treatment paddles are constructed so that each paddle contains 8 laser emission diode sources at a power output of 40mW per laser diode. These treatment paddles are non-thermal and non-invasive, at a wavelength of 630-680.

    AI/ML Overview

    The provided document is a 510(k) summary for the CURVE Laser System and details the device's indications for use, substantial equivalence, and performance data, including a clinical study.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Primary Efficacy Endpoint)Reported Device Performance
    Achievement of at least 1 inch loss from baseline in averaged waistline measurement at the final visit.100% of subjects receiving treatment with the CURVE device lost at least 1 inch in their waistline.

    (Note: The document doesn't explicitly state the acceptance criteria as a separate "acceptance criteria" section but directly presents the primary efficacy endpoint of the clinical study as the measure for success.)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size:
      • CURVE Laser System treatment group: 19 subjects
      • Control (sham) device group: 18 subjects
      • Total: 37 subjects
    • Data Provenance: The document does not explicitly state the country of origin of the data. The study was a "randomized, double blind clinical study," indicating it was prospectively collected.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The document does not mention the use of experts to establish ground truth for the test set in the context of image analysis or diagnostic interpretation. The primary outcome was a physical measurement (waistline circumference).

    4. Adjudication Method for the Test Set:

    Not applicable. The study involved direct physical measurements of waistline circumference, not interpretive assessments requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study involved a comparison between a device and a sham, not a comparison of human readers with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    A standalone study was not done. The CURVE Laser System is a physical device that delivers laser treatment, not an algorithm, and its performance is evaluated in direct application to subjects.

    7. Type of Ground Truth Used:

    The ground truth used was direct physical measurement of waistline circumference.

    8. Sample Size for the Training Set:

    The document does not mention a training set. This is a clinical study for a physical device, not an AI/algorithm where a separate training set is typically used to develop the model.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there was no mention of a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143741

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-24

    (114 days)

    Product Code
    Regulation Number
    878.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lipofina Laser System is indicated for non-invasive aesthetic treatment for the temporary reduction in circumference of the waist.

    Device Description

    The Lipofina Laser System (Figure 1) consists of a main console and 8 treatment paddles. The console houses the main electronics, controls and embedded software. The liquid crystal display (LCD) identifies all key treatment parameters. The main console is also equipped with a micro controller that provides automatic calculation of energy output for a specific set of treatment parameters. The treatment paddles are constructed so that each paddle contains 12 laser emission diode sources at a power output of 35mW per laser diode. These treatment paddles are non-thermal and noninvasive, at a wavelength of 658 (central).

    AI/ML Overview

    The provided document is a 510(k) summary for the Lipofina Laser System, which is indicated for non-invasive aesthetic treatment for the temporary reduction in circumference of the waist. The document details a clinical study conducted to demonstrate the device's safety and effectiveness.

    Here's an analysis of the acceptance criteria and the study, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Lipofina Laser System)
    Primary Efficacy Endpoint: Achievement of at least 1 inch loss from baseline in averaged waistline measurement at the final visit.95% of subjects treated with the Lipofina Laser System lost at least 1 inch in their waistline.
    Safety Endpoint: No device-related Adverse Events (AEs) or skin color changes.No device-related AEs or skin color changes were reported for any subject.

    2. Sample Size and Data Provenance

    • Sample Size (Test Set):
      • Lipofina Laser System treated group: 20 subjects
      • Control (sham) device group: 21 subjects
    • Data Provenance: The study was "at the request of CDRH," implying it was conducted as part of the FDA regulatory process. The geographical origin of the data (country) is not explicitly stated, but the sponsor, YOLO Medical Inc., is located in British Columbia, Canada, and the submission correspondent Rockin' Regulatory, Inc. is in California, USA, suggesting a North American context. The study was prospective as it was a "randomized, double blind clinical study."

    3. Number of Experts and Qualifications for Ground Truth

    The document does not mention the use of experts to establish ground truth for the test set in the conventional sense of medical image interpretation (e.g., radiologists reviewing images). The ground truth here is directly measured waistline circumference loss and observed adverse events.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this study. The "ground truth" was objective measurement (waistline circumference) and direct observation for adverse events.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study typically involves multiple human readers evaluating medical images or data, with and without AI assistance, to assess the impact of AI on diagnostic accuracy or efficiency. The Lipofina Laser System is a physical device used for aesthetic treatment, not an AI-powered diagnostic tool.

    6. Standalone Performance Study

    Yes, a standalone performance study was done in the sense that the device's effect was directly measured on subjects without human intervention in the treatment application that would alter the device's output. The clinical study was a randomized, double-blind trial demonstrating the device's efficacy on its own when applied as intended.

    7. Type of Ground Truth Used

    The ground truth used was objective measurement (waistline circumference loss) and clinical observation (for adverse events and skin color changes).

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning. The Lipofina Laser System is a low-level laser system, not an AI/ML-based device that would typically have a training set. The clinical study described served as the primary validation set to demonstrate the device's safety and efficacy.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention or indication of an AI/ML training set for this device.