510(k) Data Aggregation

    K Number
    K183357

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-01-21

    (48 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XableCath Catheter is intended for use with a guidewire to access discrete regions of the peripheral vasculature. The XableCath™ Crossing Catheter is intended for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic peripheral lesions (including chronic total occlusions) prior to placement of other interventional devices.

    Device Description

    The XableCath Crossing Catheter device is an over-the-wire (OTW) single-lumen catheter, with blunt and abrasion tip design. The catheter is offered in a product family configuration, with size models as outlined below:
    Catalog Number: AbrasionTip (A), Blunt Tip (B)
    Dimensions: Working Length (cm), Recommended Guidewire Inches (mm)
    XC-014-065A, XC-014-065B, 65, 0.014" (0.36 mm)
    XC-014-090A, XC-014-090B, 90, 0.014" (0.36 mm)
    XC-014-145A, XC-014-145B, 145, 0.014" (0.36 mm)
    XC-018-065A, XC-018-065B, 65, 0.018" (0.46 mm)
    XC-018-090A, XC-018-090B, 90, 0.018" (0.46 mm)
    XC-018-145A, XC-018-145B, 145, 0.018" (0.46 mm)
    XC-035-065A, XC-035-065B, 65, 0.035" (0.89 mm)
    XC-035-090A, XC-035-090B, 90, 0.035" (0.89 mm)
    XC-035-145A, XC-035-145B, 145, 0.035" (0.89 mm)

    The catheter shaft consists of a cortically reinforced co-polymeric material blend, with radiopague cobalt chromiumtunqsten-nickel distal tip. The proximal portion is comprised of an olefin strain relief and polycarbonate hub, which includes a female luer connection with the catheter lumen. The catheter pass-through lumen is used to advance the catheter over an appropriately sized guidewire for intravascular advancement to the target site. Component connections are ensured through inter-mechanical locks and utilization of medical grade adhesives. Catheter quidewire compatibility size/length are declared on the XableCath Crossing Catheter strain relief. All provided family member models are compatible with ≥4Fr introducer sheaths. All product configurations are provided sterile, single-use only.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the XableCath Crossing Catheter, based on the provided document:

    Acceptance Criteria and Device Performance

    The document does not present a formal table of distinct, quantitative acceptance criteria for device performance. Instead, it details various non-clinical and clinical tests performed and states that the device was deemed "substantially equivalent" to a predicate device, and that specific endpoints were met in the clinical assessment.

    However, based on the narrative, we can infer some key performance achievements:

    Acceptance Criteria (Inferred from study outcomes)Reported Device Performance
    Non-Clinical Performance:
    Substantially equivalent to predicate device in various physical and functional areas.Direct product bench in-vitro comparison testing showed the subject & predicate products to be substantially equivalent in areas such as visual & technical inspections, dimensional assessment, luer syringe compatibility, guidewire compatibility, sheath compatibility, leak testing, guidewire retraction/reinsertion, catheter kink-resistance, relative radiopacity comparison, tensile testing, torque transmission & capability, corrosion testing, simulated use (iliac model), and general packaging/shelf-life.
    BiocompatibilityFull panel biocompatibility successfully performed under GLP rigors, demonstrating all materials and construction methods passed biocompatibility (including cytotoxicity, sensitization, irritation/intracutaneous reactivity, systemic toxicity, hemolysis, thromboresistance, complement activation, partial thromboplastin time, platelet and leukocyte count, and materials mediated pyrogenicity).
    Packaging integrity and transport challengeApplied and successfully completed in accordance with established acceptance criteria demonstrating configurational adequacy.
    Clinical Performance (Real-world assessment):
    Primary Endpoint: Successful passage within the target lesion, enabling subsequent therapy (angioplasty and/or stenting).98% primary endpoint achievement with successful PTA/Stenting & technical success with lesion/claudication crossing.
    Secondary Endpoint: Absence of major adverse event attributable to XableCath Catheter product use (perforation, thrombosis, or distal embolization).100% secondary endpoint achievement. (Routine minor dissection in 10% of treatment subjects experiencing an access site complication, both successfully treated with standard remedial maneuvers, without any residual clinical sequelae at 30-45 day post-treatment follow-up.) No recorded minor adverse events were caused by or attributable to the XableCath Crossing Catheter deployment.
    Human factors engineering and usabilityAssessments performed to ensure subject product features could be safely and effectively used.

    Study Details

    This submission describes a pre-market notification (510(k)) for the XableCath Crossing Catheter, leveraging substantial equivalence to a predicate device (eV3/Medtronic Viance Crossing® Catheter (K120533)). The study components include non-clinical (bench) testing and a real-world clinical assessment.

    1. A table of acceptance criteria and the reported device performance:

    • See table above. The acceptance criteria are largely inferred from the successful completion of various tests and the achievement of primary and secondary endpoints in the clinical assessment.

    2. Sample sized used for the test set and the data provenance:

    • Non-Clinical Test Set: The document does not specify exact sample sizes for each non-clinical bench test but indicates that "Direct product bench in-vitro comparison testing" was performed on the subject and predicate devices. No specific number of devices tested is provided.
    • Clinical Test Set:
      • Sample Size: 49 patients underwent lesion modifying therapeutic intravascular treatment with XableCath Catheters.
      • Data Provenance: This was a "real-world assessment," implying prospective data collection in clinical settings where the device was used. No specific country of origin is mentioned, but the sponsoring company location (Salt Lake City, Utah) suggests US-based or international clinical sites.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Non-Clinical: Not specified. Testing would typically involve qualified engineers and lab personnel.
    • Clinical: The "real-world assessment" involved "physicians" administering the treatment and assessing outcomes. While their specific qualifications (e.g., interventional cardiologists, radiologists) are not detailed, it can be assumed they are qualified medical professionals experienced in peripheral arterial disease re-canalization and intervention. The document mentions "clinical outcomes," which would be determined by these treating physicians.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Non-Clinical: Not applicable in the context of expert adjudication. Bench testing results are typically based on objective measurements and established test protocols.
    • Clinical: No formal external adjudication method (like 2+1 or 3+1 expert review) is mentioned for the clinical outcomes. The outcomes (primary and secondary endpoints, adverse events) were reported from the real-world treatment context, presumably as recorded and assessed by the treating clinicians.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a physical medical catheter, not an AI or imaging software, so MRMC studies involving human readers and AI assistance are not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical medical catheter, not an algorithm. Its standalone performance is demonstrated through the non-clinical and clinical studies where it is used as a tool by a human operator.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Non-Clinical: The "ground truth" for non-clinical tests was based on objective measurements, adherence to ISO standards (ISO 10555-1, ISO 10993), and engineering performance criteria.
    • Clinical: The "ground truth" for the clinical study was based on patient outcomes data, specifically:
      • Successful passage within the target lesion and enablement of subsequent therapy (angioplasty and/or stenting).
      • Absence of major adverse events attributed to the device (perforation, thrombosis, distal embolization).
      • Resolution of minor adverse events without residual clinical sequelae.

    8. The sample size for the training set:

    • Not applicable. This submission describes a medical device, not an AI model, so there is no "training set" in the context of machine learning. The non-clinical and clinical data serve as validation/verification of the device's performance.

    9. How the ground truth for the training set was established:

    • Not applicable. As a physical medical device, there is no "training set" or corresponding ground truth establishment in the machine learning sense. The device's design and manufacturing are based on engineering principles and preclinical testing, and its clinical efficacy is evaluated through the real-world patient outcomes mentioned above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180986

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-05-15

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices.

    Device Description

    Catheter, Percutaneous

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device Name: XableCath Support Catheter

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than explicitly listing quantitative acceptance criteria and then demonstrating performance against them in a tabular format. Instead, it details various tests and their successful completion, implying that "passing" each test constitutes meeting an implicit acceptance criterion.

    However, based on the Performance Testing section and the Summary of non-clinical tests conducted, we can infer the acceptance criteria and reported performance:

    Acceptance Criteria (Inferred from tests conducted)Reported Device Performance (Implied "Pass" or "Acceptable")
    Visual & Technical InspectionsEquivalent to predicate
    Dimensional Assessment & ComparisonsEquivalent to predicate
    Luer Syringe CompatibilityEquivalent to predicate
    Guidewire CompatibilityEquivalent to predicate
    Sheath CompatibilityEquivalent to predicate
    Leak TestingEquivalent to predicate
    Guidewire Retraction/ReinsertionEquivalent to predicate
    Catheter Kink-ResistanceEquivalent to predicate
    Relative Radiopacity ComparisonEquivalent to predicate
    Tensile Testing (proximal & distal)Equivalent to predicate
    Torque Transmission & CapabilityEquivalent to predicate
    Corrosion TestingEquivalent to predicate
    Simulated Use (iliac model)Equivalent to predicate
    General Packaging, Shelf-Life/ExpirySuccessfully completed; configurational adequacy demonstrated
    Biocompatibility (Cytotoxicity, Sensitization, Irritation/Intracutaneous, Reactivity, Hemolysis, Systemic Toxicity (Acute), Complement Activation, Thromboresistance, Platelet and Leukocyte Count, Partial Thromboplastin Time, Materials Mediated Pyrogenicity)All utilized materials and methods of construction/processing passed biocompatibility rigors (successfully performed in accord with product classification, under GLP rigors).
    Packaging Integrity & Transport Challenge TestingSuccessfully completed in accordance with established acceptance criteria
    Clinical Safety (Intra-procedural complications)No procedural complications attributable to the XableCath Support Catheter in 27 uses across 26 patients. No thrombosis, arterial rupture, or distal embolization.
    Clinical Effectiveness (Performance as intended)All arterial access sites and segments typically used for lower extremity angiography were used/traversed safely. The device performed as intended as an over-the-wire PAD catheter for guidewire and other interventional product exchange. (80 arterial segments traversed, no new questions of safety and effectiveness, no intra-procedural complications observed).
    Post-procedure follow-up (6-8 weeks)No complications noted at a mean follow-up of 53 days post-procedure office visit attributable to the XableCath Support Catheter. No XableCath Support Catheter related vessel perforation, major dissection, or distal debris embolization, nor abrasion tip post-procedural clinical sequelae were observed.
    Infusion Pressure/Burst Strength≥300 psi (Stated as a characteristic, implies meeting a minimum)

    2. Sample size used for the test set and data provenance

    • Sample Size for Clinical (In-vivo) Testing: 26 patients (20 male, 6 female), using 27 XableCath Support Catheters.
    • Data Provenance: The study was a "real-world actual-use assessment" and "real-world validation study cohort" conducted as part of product design validation. The document does not explicitly state the country of origin, but given it's an FDA submission, it's highly likely to be U.S.-based or from a country adhering to similar regulatory standards. It is a prospective collection of data from actual patient procedures for validation purposes.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts

    • Number of Experts: One operating physician was responsible for medically observing patients in follow-up. It can be inferred that multiple experts (interventional cardiologists/radiologists) were involved in the procedures themselves, as the collective actions and observations during the "real-world actual-use assessment" formed the basis of the safety and performance evaluation. However, only one "operating physician" is explicitly mentioned for follow-up. Ground truth for device performance and complications during the procedure would be established by the proceduralists.
    • Qualifications of Experts: The document refers to "operating physician(s)" and the context of procedures like angiography and PAD interventions implies that these are interventional cardiologists or radiologists with expertise in peripheral vascular interventions. Specific years of experience are not mentioned.

    4. Adjudication method for the test set

    The document does not explicitly state an adjudication method (like 2+1 or 3+1). The "operating physician" observed patients in follow-up, and findings during the procedure would have been documented by the medical team. The conclusions drawn are from "real-world actual-use assessment" and "real-world validation," suggesting that the medical community's standard practices for evaluating device performance and adverse events during and after such procedures would apply. This typically involves the proceduralist's assessment and follow-up clinical evaluation, but not a formal, independent adjudication panel as might be seen for subjective AI interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. This device is a physical medical instrument (a support catheter), not an AI algorithm for diagnostic imaging interpretation. Therefore, the concept of human readers improving with AI assistance is not applicable.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. As stated above, this is a physical medical device, not a standalone AI algorithm. The performance described is its standalone mechanical and functional operation.

    7. The type of ground truth used

    For the clinical study:

    • Expert Clinical Assessment and Outcomes Data:
      • Intra-procedural observations: Adequacy of passage, procedural complications (e.g., dissection, thrombosis, rupture, embolization) as observed by the operating physician(s) and documented during the angiography procedures.
      • Post-procedural follow-up: Clinical sequelae (complications attributable to the device) observed at a mean of 53 days post-procedure office visit by the operating physician. This constitutes direct observation of patient outcomes.

    8. The sample size for the training set

    • Not applicable. This document describes a physical medical device validation, not an AI/machine learning model that requires a training set. The "design validation" and "verification tests" relate to engineering principles and clinical observational assessments rather than model training.

    9. How the ground truth for the training set was established

    • Not applicable. (See point 8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K170041

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-09-07

    (245 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices.

    Device Description

    The XableCath Support Catheter device is an over-the-wire (OTW) single-lumen catheter, with blunt tip design. The catheter is offered in a product family configuration, with size models as outlined in the table below:

    Catalog Number Blunt Tip (B)French SizeWorking Length (cm)Recommended Guidewire Inches (mm)
    XC-014-065B5Fr650.014" (0.36 mm)
    XC-014-090B5Fr900.014" (0.36 mm)
    XC-014-145B5Fr1450.014" (0.36 mm)
    XC-018-065B6Fr650.018" (0.46 mm)
    XC-018-090B6Fr900.018" (0.46 mm)
    XC-018-145B6Fr1450.018" (0.46 mm)
    XC-035-065B7Fr650.035" (0.89 mm)
    XC-035-090B7Fr900.035" (0.89 mm)
    XC-035-145B7Fr1450.035" (0.89 mm)

    The catheter shaft consists of a cortically reinforced co-polymeric material blend, with radiopaque cobalt chromiumtungsten-nickel distal tip. The proximal portion is comprised of an olefin strain relief and polycarbonate hub, which includes a female luer connection with the catheter lumen. The catheter pass-through lumen is used to advance the catheter over an appropriately sized guidewire for intravascular advancement to the target site. Component connections are ensured through inter-mechanical locks and utilization of medical grade adhesives. Catheter quidewire compatibility size/length are declared on the XableCath Support Catheter strain relief. All provided family member models are compatible with >=4Fr introducer sheaths. All configurations are provided sterile, single-use only.

    AI/ML Overview

    The document describes the XableCath Support Catheter and its substantial equivalence determination to a predicate device, the Merit Medical SureCross® Catheter (K123311). This is a 510(k) premarket notification, which relies on demonstrating substantial equivalence rather than independent proof of safety and effectiveness through clinical trials like a PMA.

    Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of this 510(k) submission refer to the non-clinical (bench) testing performed to demonstrate that the XableCath Support Catheter is as safe and effective as the predicate device. There is no mention of an AI-powered device or a study involving human readers or expert ground truth for image interpretation.

    Here's a breakdown based on the provided text, focusing on the device's performance relative to the predicate rather than AI-specific criteria:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal "acceptance criteria" table with specific quantitative thresholds. Instead, it describes various non-clinical (bench) tests and states that the subject device was found to be "substantially equivalent" to the predicate in these areas. The performance is assessed comparatively.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Direct Product Bench In-vitro Comparison Testing
    Visual & Technical InspectionsComparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Luer Syringe CompatibilityCompatible with standard luer syringes, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Sheath CompatibilityCompatible with introducer sheaths, comparable to predicate. All models compatible with ≥4Fr introducer sheaths.Subject & predicate products shown to be substantially equivalent. All provided family member models are compatible with ≥4Fr introducer sheaths.
    Guidewire Retraction/ReinsertionSmooth and effective, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Relative Radiopacity ComparisonAdequate visibility under fluoroscopy, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Torque Transmission & CapabilityEffective torque transmission, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Simulated Use (Iliac model)Effective performance in simulated use, comparable to predicate. Ease of insertion and tracking yields equivalent product delivery performance.Subject & predicate products shown to be substantially equivalent.
    Dimensional Assessment & ComparisonsWithin acceptable variations, comparable to predicate."Similarly offered working lengths, and identical dimensional guidewire compatibility (i.e., 0.014", and 0.035")." Subject & predicate products shown to be substantially equivalent. Outer Diameters vary by model but are compared.
    Leak TestingNo leaks, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Catheter Kink-ResistanceResistant to kinking, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Tensile Testing (proximal & distal)Adequate tensile strength, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    Corrosion TestingNo significant corrosion, comparable to predicate.Subject & predicate products shown to be substantially equivalent.
    General Packaging, Shelf-life/ExpiryStable and effective packaging over shelf-life, comparable to predicate.Subject & predicate products shown to be substantially equivalent. Packaging integrity, transport challenge testing, and shelf-life testing were applied and successfully completed in accordance with established acceptance criteria demonstrating configurational adequacy.
    Infusion Pressure/Burst Strength≥300 psi burst strength.Reported: ≥300 psi, matching predicate.
    BiocompatibilityMeet GLP rigors for biocompatibility (e.g., non-cytotoxic, non-irritating).Full panel biocompatibility successfully performed under GLP rigors, demonstrating all utilized materials and methods of construction/processing passed biocompatibility rigors. (Specific tests: Cytotoxicity, Irritation/Intracutaneous Reactivity, Hemolysis, Complement Activation, Platelet and Leukocyte Count, Sensitization, Systemic Toxicity (Acute), Thromboresistance, Partial Thromboplastin Time, Materials Mediated Pyrogenicity).

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes (e.g., number of catheters) used for each bench test. The provenance is internal laboratory testing ("bench in-vitro comparison testing") likely conducted by the manufacturer, XableCath, Inc., to support their 510(k) submission. The data is non-clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical device clearance based on substantial equivalence to a predicate, not an AI/imaging device requiring expert interpretation for ground truth. The "ground truth" for these tests comes from objective physical measurements and established biological safety testing protocols (e.g., ISO standards, GLP practices).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is not an imaging study requiring human adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical medical catheter, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests would be established by:

    • Physical standards and measurements: e.g., using calipers for dimensional checks, force gauges for tensile strength, calibrated pressure sensors for burst strength.
    • Standardized test methods/protocols: e.g., ISO 10555-1 for intravascular catheters, and GLP (Good Laboratory Practice) guidelines for biocompatibility tests.
    • Analytical chemistry/biology results: For biocompatibility, results from established lab assays (e.g., cell culture for cytotoxicity, animal models for systemic toxicity).
    • Comparison to predicate: A key "ground truth" in a 510(k) is often the performance of the predicate device, with the new device needing to demonstrate comparable performance.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/machine learning device.