Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K213049

    Validate with FDA (Live)

    Date Cleared
    2022-10-16

    (389 days)

    Product Code
    Regulation Number
    868.2775
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is a nerve stimulation device designed to be used by an anesthetist during:

    1. General anesthesia, for the purpose of establishing the efficacy of a Neuromuscular Blocking Agent using non-invasive surface electrodes.

    2. Regional anesthesia, for the purpose of:

    i. nerve mapping using the non-invasive Nerve Mapping Probe (supplied)

    ii. nerve locating using invasive needles (not supplied)

    Device Description

    The STIMPOD NMS450 is a quantitative Neuromuscular Transmission (NMT) Monitor which provides real-time quantitative feedback utilizing tri-axial accelerometery.

    The STIMPOD NMS450 is also a precision nerve locating tool used for localizing specific neural pathways. Localization of nerves by electrical stimulation involves connecting the nerve stimulator to a conducting needle through which local anaesthetics can be injected. The distance of the needle (cathode) from the nerve can be estimated by establishing the minimum threshold current required, to facilitate a neuromuscular response.

    The STIMPOD NMS450 currently makes use of 3 dimensional acceleromyography (AMG) as the measurement technology of choice for objective Neuromuscular Transmission (NMT) monitoring in surgery. Although AMG is still the de-facto standard in industry, the use of electromyography (EMG) has gained renewed interest in NMT monitoring applications where the introduction of robotic surgeries, new surgical techniques, and changes in workflow, for instance, made the use of AMG impracticable.

    In order to address this steadily increasing market demand for EMG sensor technology in NMT monitoring applications, Xavant identified the need to design and develop a dedicated NMT Monitoring Cable (EMG) that could be used interchangeably with the existing NMT Monitoring Cable (AMG) already in use with the STIMPOD NMS450.

    The NMT Monitoring Cable (EMG) was specifically designed for use with the STIMPOD NMS450 to assist anaesthesiologists in theatre with monitoring the efficacy of Neuromuscular Blocking Agents (NMBAs).

    The NMT Monitoring Cable (EMG) was designed with an intelligent EMG signal processing module (henceforth referred to as the µEMG) on one end of the cable for connection to an EMG electrode, and with a standard FireWire connector on the other end for connection to the STIMPOD NMS450.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the STIMPOD NMS450 Nerve Stimulator, focusing on its substantial equivalence to a predicate device (STIMPOD NMS450, K102084) and a reference device (Senzime, Tetragraph, K190795) with the addition of a new NMT Monitoring Cable (EMG) accessory.

    However, the document does not provide specific acceptance criteria or an explicit study proving the device meets those criteria with detailed performance metrics. Instead, it states that "The subject device met or exceeded the stated acceptance criteria for each performance test." It then lists the types of performance tests conducted.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I detail specific sample sizes, ground truth establishment, or multi-reader multi-case study results, as this information is not present in the provided text.

    Based on the available information, here's what can be extracted:

    Acceptance Criteria and Reported Device Performance:

    The document broadly states that the device met or exceeded acceptance criteria, but it does not provide the specific criteria or quantitative performance results.

    Acceptance Criteria CategoryReported Device Performance
    Functional PerformanceMet or exceeded
    Electrical Safety (IEC 60601-1)Met or exceeded
    Electromagnetic Compatibility (IEC 60601-1-2, IEC 60601-2-40)Met or exceeded
    Biocompatibility (ISO 10993-1)Met or exceeded
    UsabilityMet or exceeded
    Battery LifeMet or exceeded
    Tensile Strength of ElectrodeMet or exceeded
    Stimulation Performance of Electrode (Charge Density, Current Density, Power Density)Met or exceeded

    Study Information (Based on available text):

    1. Sample size used for the test set and the data provenance:
      This information is not provided in the document. The text only lists the types of performance tests conducted, but not the specific methodologies, sample sizes, or data provenance (e.g., country of origin, retrospective/prospective nature).

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This information is not provided in the document. The performance tests mentioned are primarily technical and engineering-focused (e.g., electrical safety, biocompatibility), which typically rely on standardized measurements rather than expert-established ground truth in the context of clinical interpretation.

    3. Adjudication method for the test set:
      This information is not provided in the document.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      This information is not applicable/provided. The device is an electrical peripheral nerve stimulator for monitoring neuromuscular blockade, not an AI-assisted diagnostic imaging tool with human readers. No MRMC study or AI-assistance evaluation is mentioned.

    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
      This information is not applicable/provided in the context of an algorithm's performance. The device itself (STIMPOD NMS450 with the new EMG cable) performs measurements and stimulation; its "performance" is evaluated through the technical tests listed.

    6. The type of ground truth used:
      For the technical performance tests mentioned (functional, electrical safety, EMC, biocompatibility, battery life, tensile strength, stimulation parameters), the ground truth would be established by engineering standards, specifications, and physical measurements rather than clinical ground truth like pathology or outcomes data. The document states that the device "met or exceeded the stated acceptance criteria," implying these criteria are based on defined technical standards and specifications.

    7. The sample size for the training set:
      This information is not applicable/provided. The STIMPOD NMS450 is a hardware device with an accessory. There is no mention of a "training set" in the context of machine learning or AI models, as this is not an AI/ML powered device.

    8. How the ground truth for the training set was established:
      This information is not applicable/provided as there is no mention of a training set or AI/ML components.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161091

    Validate with FDA (Live)

    Date Cleared
    2017-01-18

    (275 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STIMPOD NMS460 Nerve Stimulator is a Transcutaneous Electrical Nerve Stimulation (TENS) device used for symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment of post-surgical pain, post traumatic acute pain problems, as well as an adjunct for pain control due to rehabilitation.

    Device Description

    The STIMPOD NMS460 Nerve Stimulator is a hand held low frequency neuromodulation Transcutaneous Electrical Nerve Stimulation (TENS) device, used for symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical pain, post traumatic acute pain, as well as an adjunct for pain control due to rehabilitation exercises. The STIMPOD NMS460 Nerve Stimulator offers two types of waveforms for the management of pain. The first is a Monophasic Square Wave, which is typical of normal TENS machines. The second waveform is a Hybrid RF waveform which consists of a Monophasic Square Wave with a superimposed Radio Frequency waveform. This waveform is proprietary and is unique to STIMPOD NMS460 Nerve Stimulator.

    AI/ML Overview

    The provided document is a 510(k) summary for the STIMPOD NMS460 Nerve Stimulator. It focuses on demonstrating substantial equivalence to a predicate device (Acticare, K081835) rather than presenting a study to prove acceptance criteria in the traditional sense of clinical performance measures.

    Here's an analysis of the requested information based on the document provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in terms of clinical performance metrics that the device had to achieve (e.g., sensitivity, specificity, accuracy for a diagnostic device). Instead, it demonstrates an equivalence to a predicate device through technical characteristics and adherence to recognized electrical safety and EMC standards.

    Acceptance Criteria Category (Derived from document's focus)Demonstrated Performance / Comparison
    Intended UseThe STIMPOD NMS460 has a similar intended use (symptomatic relief and management of chronic intractable pain, adjunctive treatment of post-surgical/traumatic pain, and pain control due to rehabilitation exercises) to the predicate device.
    Technical Characteristics EquivalenceThe device's range of amplitudes, pulse widths, frequencies, and polarities are within the ranges delivered by the predicate device.
    Electrical SafetyComplies with IEC 60601-1: 2005
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2: 2007
    Safety of Nerve and Muscle StimulatorsComplies with IEC 60601-2-10: 2015
    BiocompatibilityMaterials (stainless steel, ABS plastic) have been previously used in medical devices (K093591) and are safe.
    Worst-case Scenarios (Output)For critical output parameters (e.g., peak current, average phase current, net charge, current density), the predicate device (Acticare) showed significantly higher values than the NMS460, suggesting the NMS460 operates within safer or comparable ranges.
    RF Parameters (Hybrid RF Waveform)The RF parameters of the NMS460 are within the range of RF parameters offered by the Acticare TSE.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document describes non-clinical testing for electrical safety, EMC, and safety of nerve/muscle stimulators. It does not mention a "test set" in the context of patient data or clinical performance evaluation. The data provenance would be laboratory testing results against specified standards. Countries of origin for the device submission are South Africa and the USA (FDA).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the submission relies on non-clinical engineering and safety standards testing, not on clinical performance evaluated against expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the submission relies on non-clinical engineering and safety standards testing, not on clinical performance evaluation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a nerve stimulator for pain relief, not an AI-assisted diagnostic or interpretation device that would involve human "readers" or benefit from an MRMC study in this context.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is hardware-based for electrical nerve stimulation, not an algorithm, and its use inherently involves human interaction for application and control.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests, the "ground truth" implicitly refers to the requirements and specifications within the cited international standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10) regarding electrical safety, EMC, and specific nerve/muscle stimulator safety. There is no clinical "ground truth" such as pathology or outcomes data presented in this 510(k) summary.

    8. The sample size for the training set

    This information is not applicable. The device is not an AI/Machine Learning model that requires a training set. The submission focuses on device characteristics and compliance with standards.

    9. How the ground truth for the training set was established

    This information is not applicable. As stated above, there is no training set for an AI/Machine Learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140853

    Validate with FDA (Live)

    Date Cleared
    2014-11-06

    (217 days)

    Product Code
    Regulation Number
    868.2775
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STIMPOD ST2-3010 is a nerve stimulation device designed to be used for the purpose of nerve locating using invasive electrodes or needles (not supplied)

    Device Description

    The STIMPOD ST2-3010 device is a battery powered peripheral nerve stimulator that can be used for the purpose of nerve locating using invasive electrodes or needles (not supplied). The STIMPOD ST2-3010 is a precision nerve locating tool used for localizing specific neural pathways. Localization of nerves by electrical stimulation involves connecting the nerve stimulator to a conducting needle. The stimulus is generated by a constant current source. The waveform is a monophasic square wave with 3 options for pulse width. These are: 0.1, 0.2 and 0.3 milliseconds. The unit dimensions are 145mm x 90mm x 30mm with a weight of 130 g

    AI/ML Overview

    I'm sorry, but the provided text does not contain the information needed to answer your request about acceptance criteria and study details for a medical device. The document is a 510(k) premarket notification for a nerve stimulator, focusing on demonstrating substantial equivalence to a predicate device. It details the device's technical characteristics, intended use, and comparison to a predicate, but it does not describe specific acceptance criteria and the results of a study (like a clinical trial or performance study) to prove the device meets those criteria.

    In particular, the document lacks information on:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for a test set or data provenance.
    • Number of experts or their qualifications for establishing ground truth.
    • Adjudication method for a test set.
    • Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done, or its effect size.
    • Whether a standalone (algorithm only) performance study was done.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The document discusses "safety testing" and references IEC standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10), indicating that the device passed these electrical and electromagnetic compatibility tests. However, these are general device safety standards, not specific performance criteria for a diagnostic or AI-based device's clinical accuracy that would require the kind of study details you've requested.

    Therefore, I cannot fill in the table or provide the requested study details based on the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1