Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K160549

    Validate with FDA (Live)

    Date Cleared
    2017-04-07

    (403 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "VELASI -15D" is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

    Device Description

    Medical diode laser systems are a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source. The medical diode laser systems for this unit is GaAlAs diode bar, and the wavelength is 1470nm of VELAS II-15D. It features impact structure, high efficiency and long lifetime. The beam divergence angle is too large lead to poor beam quality. In order to better meet the needs of Laser Diode practical applications, Laser Diode carried beam shaping is very necessary. With the GIGAA'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber. To start the laser unit, turn the main switch ON and turn the key switch clockwise to "I" position. The power indicator will turn green immediately with the system fans working. At the same time, the LCD screen lights up. After the parameters setting is finished, press the "Ready" button and the system will remind you to wear the protective eyewear (protective wavelength is from 800nm to 1700mm). At this time when you press down the footswitch, the laser will emit.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device, specifically a Medical Diode Laser System (VELAS II -15D). It focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than defining specific performance acceptance criteria for the novel device itself.

    Therefore, the document does not explicitly state specific acceptance criteria in terms of numerical performance thresholds (e.g., accuracy, sensitivity, specificity) for the device's clinical efficacy in treating reflux of saphenous veins. Instead, the "acceptance criteria" are implied by the demonstration of substantial equivalence to the predicate device, meaning the new device should perform as safely and effectively as the predicate.

    The "study that proves the device meets the acceptance criteria" refers to the non-clinical testing performed to demonstrate adherence to relevant safety and performance standards, and a comparison presented in the "Summary of Substantial Equivalence" table.

    Here's an analysis based on the information provided, though it will not contain specific performance metrics for clinical efficacy as none are present.


    1. Table of Acceptance Criteria and the Reported Device Performance

    As specific clinical performance acceptance criteria (e.g., success rate of vein closure, reduction in vein diameter) are not provided in this regulatory submission for the novel device, the "acceptance criteria" are inferred from the comparison to the predicate device, primarily focusing on technical specifications and safety standards.

    Acceptance Criterion (Inferred from Substantial Equivalence)Reported Device Performance (VELAS II -15D)
    Indications for Use"The 'VELAS II -15D' is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities."
    Laser TypeDiode laser
    Wavelength$1470nm \pm 10nm$
    Output power1-15W
    Operation modeCW, single pulse, repeat pulse
    Pulse width10ms-2.5s
    Pulse repetition rate0.2Hz-50Hz
    Application / Light delivery systemRecommended disposable sterile fiber (K124003, MED-Fibers, Inc.), bare fiber, 3m, 600μm fiber core diameter, NA ≥0.22, SMA905 connector, single use.
    Aiming BeamDiode laser of 635/532nm, power max. < 5mW, adjustable brightness.
    Laser Class4
    Operation interfaceColor LCD touch screen
    Power Supply100-240VAC, 50/60Hz, 350VA
    Safety classificationClass I Type B
    Waterproof levelIPX1 (device) / IPX8 (footswitch)
    Standards ComplianceIEC60601-1, IEC60601-1-2, IEC60601-2-22, IEC60825-1, FDA Guidance for Software.
    Non-sterileFiber is sterilized by EO.
    Microprocessor ControlYes
    Substantial Equivalence Conclusion"The proposed device of VELAS II -15D has the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device."

    Comparison to Predicate (K082225):

    • Indications for Use: Same.
    • Laser Type: Same.
    • Components: Similar (Proposed device does not equip with fiber and handpiece initially, but recommends a specific disposable fiber set; predicate includes these components directly). The submission argues this does not affect safety and effectiveness.
    • Wavelength: Similar ($1470nm \pm 10nm$ vs $1470nm \pm 30nm$). Proposed device has better accuracy.
    • Output power: Same (1-15W).
    • Operation mode: Same (CW, single pulse, repeat pulse).
    • Pulse width: Similar (10ms-2.5s vs 0.01s-60s). Proposed device's range is contained within the predicate, deemed sufficient for intended use.
    • Pulse repetition rate: Similar (0.2Hz-50Hz vs 0.01Hz-50Hz). Deemed not to affect safety and effectiveness.
    • Application / Light delivery system: Similar. Proposed device is more detailed, fiber core diameter range is contained within predicate's general description, deemed sufficient for intended use.
    • Aiming Beam: Similar. Proposed device's power range contains predicate's, deemed sufficient for intended use.
    • Laser Class: Same (4).
    • Operation interface: Same (Color LCD touch screen).
    • Power Supply: Similar (350VA vs 450VA). Rated power difference not affecting safety and effectiveness.
    • Safety classification: Same (Class I Type B).
    • Dimensions & Weight: Similar, proposed device is bigger and heavier. Not noted to affect safety or effectiveness.
    • Waterproof level: Same (IPX1 device, IPX8 footswitch).
    • Standard Compliance: Same listed standards.
    • Non-sterile: Same (Fiber sterilized by EO).
    • Microprocessor Control: Same (Yes).

    2. Sample size used for the test set and the data provenance

    The document specifies non-clinical tests (compliance with standards) and software verification. It does not describe a clinical study or a test set in the sense of patient data. Therefore, information regarding "sample size used for the test set" and "data provenance" (country of origin, retrospective/prospective) is not applicable or not provided in this regulatory submission, which focuses on substantial equivalence based on technical specifications and international standards compliance.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As no clinical test set/study is described, the concept of establishing ground truth by experts is not applicable or not provided in this document. The "ground truth" for the device's technical specifications and safety is established by adherence to international standards and internal quality assurance procedures, as overseen by the manufacturer's technical staff and verified through testing.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set or study requiring adjudication is described.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a surgical laser system, not an AI-assisted diagnostic tool or an imaging device involving "human readers." No MRMC study was mentioned or required for this type of submission.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a laser surgical instrument, meaning it inherently has a "human-in-the-loop" (the surgeon) performing the procedure. No standalone algorithm performance is relevant or described.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the technical performance aspects, the "ground truth" relies on:

    • International standards compliance: Demonstrated by testing against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
    • Software verification: According to FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
    • Predicate device's established safety and effectiveness: The core of the 510(k) submission is to argue that the new device is "substantially equivalent" to an already approved predicate device, implying its "ground truth" for safety and effectiveness is comparable to that of the predicate, which has presumably demonstrated its clinical efficacy through other means (e.g., initial clinical trials for its approval, or long-standing safe use).

    No specific clinical "outcomes data" or "expert consensus" related to a new discrete clinical study for this specific device is provided in the document.


    8. The sample size for the training set

    Not applicable. This submission concerns a physical medical device (laser system), not an AI/machine learning model that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/machine learning model described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151890

    Validate with FDA (Live)

    Date Cleared
    2015-12-23

    (166 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "GBOX-15A/15B" are indicated for use in surgical applications requiring the hemostasis, ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), gastroenterology, general surgery, genitourinary surgery (urology), qynecology (GYN), neurosurgery, otolaryngology (ENT), ophthalmology, arthroscopy, podiatry, pulmonology, and thoracic surgery; and Laser Assisted Lipolysis (980 nm only).

    The "VELASII-30A/30B" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (UJrology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology Endovascular coagulation, Oral Surgery and Dental procedures, laser assisted lipolysis.

    The "VELASII -60A/60B/60C" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue. Such as: Dermatology, Aesthetics, Plastic Surgery, Vascular Surgery, Podiatry, Endovenous Occlusion of the greater saphenous vein of the thigh in patients with superficial vein reflux.

    Device Description

    Medical diode laser systems are a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.

    The medical diode laser systems for this unit is GaAlAs diode bar, and the wavelength are 810mm/980mm of GBOX-15A/15B, 810/980mm of VELAS II -30A/30B, 810/980/940nm of VELAS II -60A/60B/60C. It features impact structure, high efficiency and long lifetime. The beam divergence angle is too large lead to poor beam quality. In order to better meet the needs of Laser Diode practical applications, Laser Diode carried beam shaping is very necessary. With the GIGAA'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber.

    To start the laser unit, turn the main switch ON into " | " position. Immediately the power indicator will be green with the system fans working. At the same time, the LCD screen lights up as the picture shows below.

    After finishing setting the parameters, press the "Ready" button and then the system will remind you to wear the safety goggles (protect wavelength from 800nm to 1100nm). At this time when you press down the footswitch, the laser will emit.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Wuhan Gigaa Optronics Technology Company, Ltd. for their Medical Diode Laser Systems, including models GBOX-15A/15B, VELASII-30A/30B, and VELASII-60A/60B/60C. The purpose of a 510(k) submission is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device, meaning it is as safe and effective.

    The submission focuses on comparing the proposed devices to existing predicate devices rather than establishing novel acceptance criteria based on a specific clinical study of the proposed device's performance against a ground truth.

    Here's an analysis based on the provided text, addressing your questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not present "acceptance criteria" in the typical sense of quantitative performance metrics (e.g., sensitivity, specificity, accuracy) that a device must meet in a direct study against a ground truth. Instead, it demonstrates "substantial equivalence" to predicate devices by comparing their specifications and intended uses. The "reported device performance" is largely framed within the context of these comparative specifications and adherence to recognized standards.

    Here's a table summarizing the comparison, highlighting where the proposed devices align with or are similar to the predicate devices, which implicitly serves as the "acceptance criteria" for substantial equivalence.

    FeatureAcceptance Criteria (Predicate Device K082721 and K122567/K100558)Reported Device Performance (Wuhan Gigaa Optronics Technology Co., Ltd. Medical Diode Laser Systems)Comparison Conclusion (as stated in document)
    General
    Product CodeGEXGEXSame
    Device TypeDiode LaserDiode LaserSame
    ClassIIIISame
    Safety ClassificationClass I Type BClass I Type BSame
    Laser Class44Same
    Operation InterfaceColor LCD touch screenColor LCD touch screenSame
    Microprocessor ControlYesYesSame
    Indications for UseSurgical applications requiring ablation, hemostasis, vaporization, excision, incision, and coagulation of soft tissue in specified medical specialties (aesthetic, gastroenterology, general surgery, genitourinary surgery, gynecology, neurosurgery, otolaryngology, ophthalmology, arthroscopy, podiatry, pulmonology, thoracic surgery, and Laser Assisted Lipolysis for 980 nm). For some models, also includes use with endoscopic equipment and specific conditions like BPH, vascular lesions, hair removal, endovenous occlusion.Covers the same range of surgical applications and medical specialties: ablation, hemostasis, vaporization, excision, incision, and coagulation of soft tissue, including specific uses like Laser Assisted Lipolysis and endoscopic applications. Indications for GBOX-15A/B are identical to LaserPro; for VELAS II-30A/B and VELAS II-60A/B/C, they are "same" or "similar" to JOULE and QUANTA systems.Same or Similar. "The indications for use of 'VELAS II -30A/0B' are same to that of Quanta System QUANTA Diode Laser Family. The indications for use of 'VELAS II -60A/60B/60C' are same to that of JOULE 810/940/980 Multi-Platform System." Minor structural differences are deemed not to affect safety and effectiveness.
    Technical Specifications (Examples)
    Wavelength (GBOX-15A/B)LaserPro 810: 810nm±10%; LaserPro 980: 980nm±10%GBOX-15A: 810nm±10nm; GBOX-15B: 980nm±10nmSimilar. "Center wavelength is consistent, and GBOX wavelength accuracy is better than that of the predicate device."
    Wavelength (VELAS II)JOULE 810/940/980 (810nm, 940nm, 980nm); QUANTA 808/980 (808nm, 980nm)VELAS II-30A, VELAS II-60A: 810nm $\pm$ 10nm; VELAS II-30B, VELAS II-60B: 980nm $\pm$ 10nm; VELAS II-60C: 940nm $\pm$ 10nmSimilar. "Center wavelength is consistent. Reasonable accuracy range of VELAS II does not affect safety and effectiveness."
    Output Power (GBOX)Predicate: 1-15W, 1-20W, 1-25WGBOX: 1-15WSimilar. "Same as 1-15W. GBOX design output power has claimed to meet intended use."
    Output Power (VELAS II)JOULE: $\leq$ 100W; QUANTA: 30WVELAS II-30A/B: 1-30W; VELAS II-60A/B/C: 1-60WSimilar. "Output power of 'VELAS II -30A/30B' is Same to that for 'QUANTA 808, QUANTA 980'. Output power range of 'VELAS II -60A/60B/60C' is contained in that of 'JOULE 810, JOULE 940, JOULE 980', and it has been designed to meet the intended use requirement claimed."
    Pulse width (GBOX)Predicate: 0.05s-10sGBOX: $\le$ 2.5sSimilar. "Pulse width range of the device is contained in that of the predicate device, and it has been designed to meet the intended use requirement claimed."
    Pulse width (VELAS II)JOULE: $\le$ 2500ms (2.5s); QUANTA: 0.25-10000ms (0.25s-10s)VELAS II: 10ms-2.5sSimilar. "Pulse width range of 'VELAS II -30A/30B, VELAS II -60A/60B/60C' is contained in that of the two types predicate device, and it has been designed to meet the intended use requirement claimed."
    ComponentsLaser system, Color touch screen, Fiber and the handpiece, Foot switch, Power detector/calibratorLaser system, Color touch screen, Fiber, Foot switch, Power detector/calibrator. (GBOX lacks handpiece, VELAS II does not explicitly list handpiece as a separate component distinct from fiber)Similar. "Handle as the laser transmission system, should be used with optical fiber. So GBOX only uses fiber as the laser transmission system does not affect the safety and effectiveness." (Similar statement for VELAS II).
    Power SupplyPredicate K082721: 100-240 VAC, 540VA max, 50/60Hz; Predicate K122567: 230VAC, 50/60Hz; Predicate K100558: 100-240VAC, 50-60Hz, 350VAGBOX: 100-240VAC, 200VA, 50-60Hz; VELAS II-30A/B: 100-240VAC, 50/60Hz, 350VA; VELAS II-60A/B/C: 100-240VAC, 50/60Hz, 650VASimilar. "Only rated power is different, not affect safety and effectiveness."
    SterilityNon-sterile (main engine)Non-sterile (main engine), but fiber is sterile for single use.Similar. "Fiber steriled for single use of GBOX (and VELAS II) is more effective to ensure the safety of patients."
    Standards ComplianceIEC60601-1, IEC60601-1-2, IEC60601-2-22, IEC60825-1, ISO10993-1, ISO10993-4, ISO10993-5, ISO10993-10, ISO10993-11IEC60601-1, IEC60601-1-2, IEC60601-2-22, IEC60825-1, ISO10993-1, ISO10993-5, ISO10993-10 (For VELAS II, also ISO10993-4 and ISO10993-11 are listed in Section 9 Non-Clinical Tests)Same (referring to adherence to these specific standards for safety and performance testing).

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical tests performed on the "GBOX" and "VELAS II" Medical Diode Laser Systems for safety and performance in accordance with various IEC and ISO standards. It also mentions "software verification has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

    • Test Set: No explicit "test set" in the context of clinical data (e.g., patient cases, retrospective or prospective) is described for evaluating device performance against primary clinical endpoints. The testing described is primarily engineering and bench testing to demonstrate compliance with general safety and performance standards and comparison to predicate device specifications.
    • Sample Size: Not applicable in the context of a clinical test set. The non-clinical tests would have involved specific units of the device.
    • Data Provenance: The tests were performed by the manufacturer, Wuhan Gigaa Optronics Technology Co., Ltd. The document does not specify a country of origin for the non-clinical test data other than the manufacturer's location in China. The "tests" here refer to regulatory compliance and engineering evaluations, not clinical trials with human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided as the submission is not based on a clinical study requiring ground truth establishment by experts. The "ground truth" for this submission is effectively the established safety and performance profile of the predicate devices and compliance with recognized international standards for medical electrical equipment and laser products.

    4. Adjudication method for the test set

    Not applicable. There is no clinical test set requiring expert adjudication described in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical diode laser system, not an AI-powered diagnostic or interpretive tool that would involve "human readers" or "AI assistance" in the sense of image analysis or diagnostic support. Therefore, an MRMC study is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is not an algorithm or AI system for diagnostic or interpretive purposes. It's a surgical laser system where the performance is physical (e.g., power output, wavelength, tissue interaction) and its operation is directly controlled by a human operator.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this 510(k) submission is primarily:

    • Predicate Device Performance: The established safety and effectiveness of the legally marketed predicate devices to which the proposed devices are compared.
    • Compliance with Recognized Standards: Adherence to international standards for medical electrical equipment, laser safety, and biological evaluation (e.g., IEC 60601 series, IEC 60825-1, ISO 10993 series).
    • Engineering and Performance Specifications: The technical specifications (e.g., wavelength, output power, pulse width) are tested against internal design requirements and predicate device specifications.

    There is no mention of ground truth established via expert consensus, pathology, or outcomes data from a clinical trial for the proposed device itself. The claim of "Substantial Equivalence" relies on demonstrating that the proposed device is sufficiently similar to devices already found safe and effective.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1