Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K071040

    Validate with FDA (Live)

    Date Cleared
    2007-05-22

    (40 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wright 51 and 52 Sequential Compression System is for treatment of lymphedema, venous insufficiencies, and other edematous conditions.

    Device Description

    The Wright 51 and 52 Sequential Compression System consists of a sequential, segmental, intermittent, pneumatic compression pump and appropriate limb appliances. It is typically used to compress an upper or lower body extremity.

    The pump consists of an enclosure that contains a small air compressor, solenoid valves, printed circuit board including a programmable microprocessor, graphic display, and associated electronic circuits and pneumatic connections. Software controls the sequential opening and closing of the solenoid valves, which are connected to five (5) external pressure ports. The pressure ports are connected to the limb appliance through a hose assembly.

    The inflatable limb appliance is made of fabric and may comprise from one to five cells, with the length of the appliance dependent upon the physician's order or patients limb length. In operation, the pump fills the inflatable limb appliance with air, to prescribed, preset pressures -- each cell in sequence. This simple repetition of the sequence of compression and exhaust improves return blood circulation for the patient.

    AI/ML Overview

    The Wright 51 and 52 Sequential Compression System is a medical device intended for the treatment of lymphedema, venous insufficiencies, and other edematous conditions. The acceptance criteria and the study proving the device meets these criteria are detailed below, based on the provided document.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Bench Test)Reported Device Performance
    Safety: Device is as safe as predicate devices.Risk Analysis: Conducted to confirm the device's safety is comparable to predicate devices.
    Performance Specifications: Conformity to specified operational parameters for therapy delivery.Bench Performance Testing: Measurements verified the sequence, timing, and pressure of therapy delivery. The functionality of the operator interface was also verified.
    Electrical Safety and Electromagnetic Compatibility (EMC): Compliance with relevant consensus standards.Electrical Safety and EMC Testing: Conducted in accordance with consensus standards.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: Not explicitly stated. The document indicates "bench tests" were performed, implying a test set of the device itself (likely one or more units) rather than a human subject sample.
    • Data Provenance: The testing was conducted by or on behalf of the manufacturer, Wright Therapy Products, located in Oakdale, PA, USA. This suggests the data is prospective, generated specifically for this 510(k) submission.

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    • Experts: Not applicable for this type of submission. The evaluation was based on non-clinical testing (risk analysis and bench tests) against engineering specifications and consensus standards, not on expert interpretations of clinical data.

    4. Adjudication Method (Test Set)

    • Adjudication Method: Not applicable. The evaluation involved objective measurements and comparisons against predetermined performance specifications and safety criteria in a non-clinical setting.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No. The document explicitly states "Clinical Testing: None." The submission relies on substantial equivalence to legally marketed predicate devices through comparison of technical characteristics and non-clinical testing.

    6. Standalone Performance Study

    • Standalone Performance Study: Yes, in a non-clinical context. The "Nonclinical Testing" section describes a standalone evaluation of the device's performance through risk analysis, bench tests (sequence, timing, pressure, operator interface functionality), and electrical safety/EMC testing. This represents the algorithm/device-only performance in a simulated or controlled environment.

    7. Type of Ground Truth Used

    • Ground Truth: The ground truth for the non-clinical testing was based on:
      • Engineering Specifications: Predetermined parameters for sequence, timing, and pressure of therapy delivery, as well as operator interface functionality.
      • Consensus Standards: Standards for electrical safety and electromagnetic compatibility.
      • Predicate Device Characteristics: The safety profile of the legally marketed predicate devices served as a benchmark for the risk analysis.

    8. Sample Size for Training Set

    • Sample Size for Training Set: Not applicable. The document describes a medical device, not an AI/ML algorithm that employs a training set in the conventional sense. The device's control system uses a programmable microprocessor with software, which would have been developed and verified through standard software engineering practices, not through a 'training set' of data.

    9. How Ground Truth for Training Set Was Established

    • Ground Truth for Training Set: Not applicable for this device. The software controlling the pump would have been developed and verified against functional requirements and design specifications, which serve as the "ground truth" for its operation, established through engineering design processes rather than data labeling.
    Ask a Question

    Ask a specific question about this device

    K Number
    K961292

    Validate with FDA (Live)

    Date Cleared
    1996-11-06

    (216 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device is to be used to improve the return circulation of body fluid that has pooled in one or more of a patient's extremities. This treatment reduces debilitating pain, swelling, ulceration, and risk of dermatological impact, infection, and amputation.

    Device Description

    This is an electromechanical pneumatic device that employs a compressor and precision timer, regulator, and gage to sequentially inflate a sleeve appliance that fits over the patient's extremity. The device is approximately 8"W x 6.75"H x 5.5"D and weighs less than 15 lbs.

    AI/ML Overview

    The prompt asks to describe the acceptance criteria and the study that proves the device meets them, based on the provided text. However, the provided text describes a "Wright Linear Pump" which is a sequential compression pump, not an AI/ML powered medical device. Therefore, many of the requested fields are not applicable to the given text.

    Here's an analysis based on the provided information, noting where sections are not applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Explicit or Implied)Reported Device Performance
    Functional / Performance:
    Produces sequential compressionProduces sequential compression
    Max pressure < 101 mm HgFactory-preset for a maximum pressure less than 101 mm Hg
    Cell pressures reach prescribed setting 10+ seconds before next cell activatedPressures in each cell reach the prescribed setting 10 or more seconds before the next cell is activated.
    Maximum pressure in proximal cell sustained before end of cycleMaximum pressure in the proximal cell (cell C) is reached before the end of the cycle and, thus, sustained.
    Disengage cells in quick succession (proximal to distal) ensuring no reverse gradientDevice disengages cells in quick succession, from proximal to distal, to ensure a reverse gradient is never applied despite any slight irregularities inherent in mechanical pneumatic systems.
    Safety:
    Complies with physician-prescribed treatment pressuresComplies with physician-prescribed treatment pressures
    Addresses potential for "slight irregularities inherent in mechanical pneumatic systems"Designed to disengage cells in a specific sequence to avoid reverse gradient despite slight irregularities.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable in this context. The "test set" described relates to non-clinical technical testing of the prototype device's mechanical functions, not a dataset for an AI/ML model.
    • Data Provenance: Not applicable. The "data" refers to measurements from a prototype device's mechanical operation, not patient data or images.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. Ground truth, in the context of AI/ML, refers to labels or diagnoses. Here, the "ground truth" for the non-clinical tests would have been the objective physical measurements themselves (e.g., pressure readings from precision regulators), not subjective expert assessment.

    4. Adjudication method for the test set

    • Not applicable. Adjudication methods are relevant for resolving disagreements among multiple human experts establishing ground truth for AI/ML validation. This device's testing involved direct physical measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a mechanical medical device, not an AI-powered system designed to assist human readers (e.g., radiologists).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a mechanical device. The "standalone" performance here refers to the device operating according to its design specifications, without an AI algorithm.

    7. The type of ground truth used

    • Non-clinical test data: The ground truth was based on objective physical measurements from precision regulators and observation of the device's mechanical operation (e.g., timing of cell engagement/disengagement, pressure readings).

    8. The sample size for the training set

    • Not applicable. This is a mechanical device, not an AI/ML model that requires a training set. The "prototype device" was likely used for iterative design and testing, but not in the sense of an AI training set.

    9. How the ground truth for the training set was established

    • Not applicable. No AI training set.

    Summary regarding the device and its testing based on the provided text:

    The provided text details a mechanical sequential compression pump. The "study" described is a series of non-clinical tests conducted on a prototype device to verify its fundamental mechanical and pneumatic functions against design specifications. There are no mentions of clinical studies or any AI/ML components.

    The device's performance was evaluated by directly measuring parameters like:

    • Cell pressures using separate precision regulators.
    • The timing of pressure activation and deactivation in different cells.
    • The maximum pressure achieved.

    The acceptance criteria are implied by these measurements confirming the device operates as designed, particularly in ensuring sequential compression, preventing reverse gradients, and staying within pressure limits. The primary "proof" that the device meets these criteria comes from these direct, objective physical measurements on the prototype.

    Ask a Question

    Ask a specific question about this device

    K Number
    K961797

    Validate with FDA (Live)

    Date Cleared
    1996-11-06

    (181 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device is to be used to improve the return circulation of body fluid that has pooled in one or more of a patient's extremities. This treatment reduces debilitating pain, swelling, ulceration, and risk of dermatological impact, infection, and amputation. Compression pumps of various kinds are in wide use in hospitals, clinics, patient homes, and elsewhere for treating acute and chronic fluid accumulation disorders. Complies with physician-prescribed treatment pressures.

    Device Description

    This is an electromechanical pneumatic device that employs a compressor and precision timers, regulators, and gages to sequentially inflate a sleeve appliance that fits over the patient's extremity. This unit delivers a calibrated gradient pressure. The device is approximately 10"W x 6.75"H x 5.6"D and weighs less than 15 Ibs.

    AI/ML Overview

    The provided text describes a medical device, the WLP Pro Lite gradient sequential compression pump, and its comparison to an existing device (K830577B) for the purpose of a 510(k) submission. This type of document is for demonstrating substantial equivalence to a legally marketed predicate device, not typically a detailed study proving novel acceptance criteria for a new device's performance.

    Based on the provided information, I can extract the acceptance criteria related to its functional equivalence and performance characteristics from the "non-clinical test data" section. However, the document explicitly states "clinical test data: not applicable," which means a clinical study to prove efficacy or safety against defined acceptance criteria was not conducted or provided in this summary.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from Section 5.8.1)Reported Device Performance (from Section 5.8.1)
    Pressures in each individual cell reach the prescribed setting within ±5mm Hg from the panel gage reading.Pressures in each individual cell reach the prescribed setting within ±5mm Hg from the panel gage reading.
    Pressures in each cell reach the prescribed setting for that individual cell 10 or more seconds before the next cell is activated.Pressures in each cell reach the prescribed setting for that individual cell 10 or more seconds before the next cell is activated.
    Maximum pressure in the proximal cell (cell C) is reached before the end of the cycle and thus sustained.Maximum pressure in the proximal cell (cell C) is reached before the end of the cycle and thus sustained.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for the "prototype device" measurements. The measurements were taken on a "prototype device," implying a single or very limited number of units.
    • Data Provenance: Not specified, but likely from internal testing by Wright Linear Pump, Inc. It is a retrospective assessment against established design parameters, not a prospective study. There is no mention of country of origin for the data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. The "ground truth" for the non-clinical tests was established by comparison to a "calibrated Hg manometer," which is a standard measurement device, not expert consensus.
    • Qualifications of Experts: N/A.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. The test involved direct measurement against a calibrated standard, not a subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • MRMC Study: No. The document explicitly states "clinical test data: not applicable." This device is a sequential compression pump, not an imaging or diagnostic device that would typically involve human readers interpreting output.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: A form of standalone non-clinical performance was done, as described in Section 5.8.1. The device's pressure delivery and timing characteristics were measured and compared against specifications using a calibrated manometer. This is the device operating on its own, without a human in the loop for interpreting output (as might be the case for AI).

    7. The Type of Ground Truth Used

    • Ground Truth Type:
      • For pressure accuracy: Comparison to a calibrated Hg manometer (a physical gold standard measurement tool).
      • For timing (cell activation and sustained pressure): Intrinsic design specifications and observed performance.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This device is an electromechanical pump, not a machine learning or AI-driven system that requires a "training set" in the conventional sense. Its "design parameters" are likely derived from engineering principles and prior device models.

    9. How the Ground Truth for the Training Set was Established

    • Ground Truth for Training Set: Not applicable, as there is no "training set." The device's operational parameters and design would have been established through engineering design, previous product iterations (as it's similar to K830577B), and adherence to general medical device standards for mechanical and pneumatic systems.
    Ask a Question

    Ask a specific question about this device

    K Number
    K961891

    Validate with FDA (Live)

    Date Cleared
    1996-11-06

    (181 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device is to be used to improve the return circulation of body fluid that has pooled in one or more of a patient's extremities. This treatment reduces debilitating pain, swelling, ulceration, and risk of dermatological impact, infection, and amputation.

    Device Description

    This is an electromechanical pneumatic device that employs a compressor and precision timer, regulator, and gage to inflate a sleeve appliance that fits over the patient's extremity. The device is approximately 8"W x 6.75"H x 5.5"D and weighs less than 13 Ibs.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device (WLP Solo VII, a single-pressure pump for lymphedema). However, it does not contain information about acceptance criteria or a specific study proving the device meets those criteria, as typically understood in a regulatory submission for clinical performance.

    Here's an analysis based on the provided text, highlighting the absence of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Not Explicitly Stated (e.g., specific thresholds for pressure accuracy, cycle time variance, etc.)- Factory-preset for a maximum pressure less than 101 mm Hg.- Pressure in appliance reaches the prescribed setting 10 or more seconds before the end of the cycle.

    Explanation: The document mentions "similar pressure ranges, degrees of accuracy, cycle times," to predicate devices, and that the "Cell pressure of prototype device was measured comparing on-board gauge reading to calibrated Hg manometer." However, it does not define specific quantitative acceptance criteria for these parameters (e.g., "pressure accuracy must be within +/- 5 mmHg"). The reported performance elements are descriptive rather than criteria-meeting statements.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not applicable as a "test set" for clinical performance is not mentioned. The non-clinical test involves measuring "Cell pressure of prototype device" but the number of devices or measurements is not specified.
    • Data Provenance: Not applicable. The non-clinical tests are presumed to be conducted internally by the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not Applicable. The document explicitly states "clinical test data: not applicable." No ground truth was established by experts for clinical performance.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No clinical test set or related adjudication is mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This device is a compressed limb sleeve pump; it is not an AI-powered diagnostic or assistive technology for human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This device is a physical electromechanical pump, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not Applicable for clinical ground truth. For the non-clinical pressure measurement, the "calibrated Hg manometer" served as the reference for accuracy, acting as the ground truth for pressure readings.

    8. The sample size for the training set:

    • Not Applicable. This device uses electromechanical principles and does not involve machine learning or a "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set.

    Summary of Device Evaluation in the Provided Text:

    The document describes a 510(k) submission where the manufacturer is claiming substantial equivalence to predicate devices. The "study that proves the device meets the acceptance criteria" in this context is primarily focused on non-clinical performance and similarity to existing devices, rather than a clinical trial demonstrating efficacy against pre-defined acceptance criteria.

    The non-clinical tests included:

    • Measuring cell pressure of a prototype device comparing the on-board gauge reading to a calibrated Hg manometer (no specific acceptance criteria or results beyond "measured" are provided).
    • Verifying that pressure in the appliance reached the prescribed setting 10 or more seconds before the end of the cycle (again, no specific acceptance criteria or number of tests are provided).

    The submission explicitly states "clinical test data: not applicable," indicating that a clinical study with a test set, ground truth established by experts, or MRMC studies was not performed or deemed necessary for this 510(k) submission based on its classification and comparison to predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1