Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K173431

    Validate with FDA (Live)

    Date Cleared
    2018-05-18

    (197 days)

    Product Code
    Regulation Number
    884.6110
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Towako Transmyometrial Embryo Transfer Set is indicated for transfering IVF embryo(s) into the endometrial cavity through the myometrium for patients undergoing Assisted Reproductive Procedures.

    Device Description

    The Towako Transmyometrial Embryo Transfer Set is comprised of an echo tipped stainless steel 19 Gauge needle, matching stainless steel stylet and a 2 Fr polyethylene transfer catheter. The stylet fits into the needle with both bevels flush, and is inserted with the stylet in-situ to prevent blockage of the needle lumen and provide stiffness to the needle during insertion. When the needle is positioned correctly, the stylet is removed and the transfer catheter with the embryos loaded is inserted to allow embryo transfer. The device is used when difficulty is experienced in transferring embryos into the uterine cavity using a standard trans-cervical route. The Towako Transmyometrial Embryo Transfer Set is a sterile and single-use device.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for a medical device called the "Towako Transmyometrial Embryo Transfer Set." It describes the device, its intended use, comparison to a predicate device, and performance data.

    However, the document does not contain information about a study involving an AI/Machine Learning device and its performance against acceptance criteria. The listed performance data pertains to traditional medical device testing such as biocompatibility, mouse embryo assay, endotoxin testing, mechanical performance testing, and stability testing. There is no mention of a test set, ground truth, expert readers, or any other elements related to the evaluation of an AI algorithm as requested in your prompt.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria for an AI/ML device based on the provided text. The text describes the regulatory clearance process for a physical medical device, not a software or AI medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171611

    Validate with FDA (Live)

    Date Cleared
    2017-12-15

    (197 days)

    Product Code
    Regulation Number
    884.6100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Double Lumen Ovum Aspiration Need for laparoscopic or ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Device Description

    Double Lumen Ovum Pick-up Needles consist of a double lumen stainless steel needle with a manipulating handle. The manipulating handle has two ports: the central port through which oocytes are aspirated via the central lumen, and a secondary side port to allow flushing of follicles via a secondary lumen. The needle is intended to be used to aspirate oocytes from follicles of human ovaries via a laparoscopic or ultrasound guided transvaginal path.

    The Double Lumen Ovum Aspiration Needles are sterile and single-use devices.

    The Double Lumen Ovum Aspiration Needles are available in 16 or 17 gauge sizes and three lengths: 30, 33 and 35 cm. The needles are provided with an aspiration tube of 75 cm length and a vacuum tube with a length of 70 or 100 cm.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/Performance MetricAcceptance Criteria (Threshold/Standard)Reported Device Performance (Result)
    BiocompatibilityCytotoxicityConformance to ISO 10993-5:2009Conformed
    SensitizationConformance to ISO 10993-10:2010Conformed
    IrritationConformance to ISO 10993-10:2010Conformed
    Embryo CompatibilityMouse Embryo Assay (MEA)Not explicitly stated, but implied to show comparable development to control groupPercent of embryos developed to expanded blastocyst stage within 72 hours, assessed in comparison with the control group (implied positive result)
    Sterility/PurityEndotoxin TestingConformance to USP <85>Conformed
    Mechanical PerformanceNegative Pressure Leak TestNot explicitly stated, but implied to meet functional requirementsPerformed and implied to meet requirements
    Tensile Strength (Tubing to Cannula)Not explicitly stated, but implied to meet functional requirementsPerformed and implied to meet requirements
    Tensile Strength (Tubing to Bung)Not explicitly stated, but implied to meet functional requirementsPerformed and implied to meet requirements
    Durability/Shelf LifeNegative Pressure Leak Test (after aging)Not explicitly stated, but implied to maintain functional requirements after 3 yearsPerformed after three years of real time aging and implied to maintain requirements

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes design verification and validation activities for a medical device (Double Lumen Ovum Aspiration Needles), not a study involving human data, algorithms, or test sets in the context of medical imaging or AI.

    Therefore, the concepts of "sample size used for the test set," "data provenance," "country of origin of the data," and "retrospective or prospective" are not applicable to this type of device submission. The tests were performed on the device itself or its materials.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable as the document describes physical/material testing of a medical device, not a diagnostic or AI-driven system requiring expert ground truth establishment.

    4. Adjudication Method for the Test Set

    This information is not applicable for the reasons stated above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This submission is for a physical medical device, not an AI-based diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. This submission is for a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on established industry standards (e.g., ISO, USP) and internal functional requirements for biocompatibility, mechanical performance, and sterility. For the Mouse Embryo Assay (MEA), the "ground truth" would be the development of a control group of embryos.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this is not an algorithm or AI model.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. There is no "training set" or associated ground truth establishment for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160753

    Validate with FDA (Live)

    Device Name
    COOK Vacuum Pump
    Date Cleared
    2017-06-13

    (452 days)

    Product Code
    Regulation Number
    884.6120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The COOK Vacuum Pump is intended for the aspiration of eggs (ova), during assisted reproduction procedures using low flow, intermittent vacuum.

    Device Description

    The COOK Vacuum Pump (K-MAR-5200) is an electrically-powered, vacuum pump that is used for the collection of ova (eggs) from ovarian follicles for use in in vitro fertilization (IVF) procedures. The COOK Vacuum Pump provides vacuum levels ranging from -10 mmHg to -500 mmHg. . It is supplied non-sterile.

    The disposable Vacuum Line with Hydrophobic Filter (K-DVLF-240) is an accessory to the COOK Vacuum Pump and is used to connect the vacuum pump to an ovum aspiration needle. This component has been designed and tested to withstand the maximum vacuum pressures associated with the COOK Vacuum Pump. The Vacuum Line with Hydrophobic Filter (K-DVLF-240) is provided sterile (ethylene oxide sterilization) for single use only, and has a shelf-life of three years.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding the "COOK Vacuum Pump". This document primarily focuses on the device's substantial equivalence to a predicate device and does not contain information about clinical studies or performance data related to AI/algorithm-driven medical devices. Therefore, I cannot provide the requested information regarding acceptance criteria and study results for an AI device.

    The document discusses the following:

    • Device Name: COOK Vacuum Pump
    • Indications for Use: Aspiration of eggs (ova) during assisted reproduction procedures using low flow, intermittent vacuum.
    • Predicate Device: COOK Ultra Quiet Vacuum Pump & Regulator (K992070)
    • Technological Characteristics: Comparison of vacuum range, boost function, controller type, and presence of software.
    • Non-Clinical Performance Testing: Electrical Safety, Electromagnetic Compatibility, Software Verification and Validation, Sterilization Validation, Shelf-Life Testing.

    Specifically, regarding your request, this document does not include:

    • A table of acceptance criteria and reported AI device performance.
    • Sample sizes, data provenance, or details of a test set for an AI algorithm.
    • Information on experts used to establish ground truth for an AI test set.
    • Adjudication methods for an AI test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies for human readers with/without AI assistance.
    • Standalone AI algorithm performance.
    • Type of ground truth for an AI algorithm.
    • Sample size for an AI training set.
    • How ground truth for an AI training set was established.

    The "Software Verification and Validation Testing" mentioned refers to the general software development lifecycle and testing process of the embedded software controlling the vacuum pump, not to artificial intelligence or machine learning model validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170257

    Validate with FDA (Live)

    Date Cleared
    2017-02-24

    (28 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sydney IVF Culture Oil is intended for use as an oil overlay for culture of gametes, zygotes, or embryos in assisted reproduction technology (ART) and micro-manipulation procedures.

    Device Description

    Sydney IVF Culture Oil is designed to protect gametes and embryos during ART processing by providing a barrier between the embryo culture media and the air, thereby minimizing evaporation, and reducing fluctuations in osmolality. Sydney IVF Culture Oil consists of pharmaceutical grade mineral oil.

    Sydney IVF Culture Oil is provided in glass vials (50 ml). This product is a single use, aseptically-filtered device.

    AI/ML Overview

    The Sydney IVF Culture Oil is intended for use as an oil overlay for the culture of gametes, zygotes, or embryos in assisted reproduction technology (ART) and micromanipulation procedures. The study was conducted to expand the shelf-life of the device from 90 days to 20 weeks when stored at 2-8℃.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (After 20 weeks at 2-8°C)
    Mouse Embryo Assay (MEA): ≥80% 2-cell embryos expanded to blastocysts at 72 hoursMet (Specific percentage not provided, but stated as "met the following specifications")
    Endotoxin (LAL): <0.4 EU/ml in accordance with USP <85>Met (Specific value not provided, but stated as "met the following specifications")
    Sterility: No growth in accordance with USP <71>Met (Specific observation not provided, but stated as "met the following specifications")

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact number of units of Sydney IVF Culture Oil tested for the shelf-life study. It refers to a "real-time shelf-life study."
    • Data Provenance: The study was conducted by William A. Cook Australia Pty Ltd in Australia. It is a prospective, real-time shelf-life study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device is not an AI/imaging diagnostic device requiring expert interpretation for ground truth. The acceptance criteria are based on standardized laboratory assays (MEA, LAL, Sterility).

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the ground truth is established by objective laboratory assays, not expert consensus requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is a culture oil, not an AI diagnostic tool, and therefore human reader performance is not relevant.

    6. Standalone Performance

    Yes, a standalone performance study was done in the form of a real-time shelf-life study. This study evaluated the performance of the device (Sydney IVF Culture Oil) itself against pre-defined specifications without human intervention or interpretation as part of the primary outcome.

    7. Type of Ground Truth Used

    The ground truth used for the acceptance criteria are:

    • Mouse Embryo Assay (MEA): A biological assay, an established standard for assessing the suitability of materials for embryo culture, providing an objective measure of embryo development.
    • Endotoxin (LAL): A chemical assay (Limulus Amebocyte Lysate assay), a standard method for detecting bacterial endotoxins.
    • Sterility: A microbiological test, an established standard for confirming the absence of viable microorganisms.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning device that requires a training set. The study pertains to the physical and biological stability of a manufactured product over time.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153478

    Validate with FDA (Live)

    Date Cleared
    2016-08-15

    (257 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sydney IVF Gamete Buffer is used to physically wash and store gametes in preparation for the IVF process.

    Device Description

    Sydney IVF Gamete Buffer is a HEPES buffered physiologic solution used to physically wash gametes in preparation for the fertilization step in the IVF process. Sydney IVF Gamete Buffer can be used for both sperm and oocytes. This solution can be used to wash oocytes of excess blood following aspiration and also as a short term storage solution until transfer to a fertilisation medium. Sydney IVF Gamete Buffer is based upon the formulation of a 20 mM HEPES buffered media containing salts, non-essential amino acids, glucose, 10 mg/mL Human Serum Albumin (HSA) and 0.01 mg/mL Gentamicin. Sydney IVF Gamete Buffer is packaged in 20, 50 or 100 mL borosilicate Type I glass vials with Fluorotec coated rubber stoppers held in place with a tamper evident seal. These products are single use, sterile (aseptic filtration) devices. It has a shelf life of 20 weeks when stored at 2 - 8 °C.

    AI/ML Overview

    This document describes the Sydney IVF Gamete Buffer and its substantial equivalence to a predicate device. However, it does not contain the information you requested regarding the acceptance criteria and a study proving a new device's performance against those criteria in the context of an AI/human-in-the-loop system.

    This document is a 510(k) summary for a medical device (a buffer solution for IVF), not for a device that involves AI, image analysis, or human reader performance. The "performance data" section refers to bench testing for the buffer solution itself (pH, osmolality, endotoxin, etc.), and functional tests like the 2-cell Mouse Embryo Assay (MEA) and Human Sperm Survival Assay (HSSA) to confirm its suitability for its intended use. It's a comparison to ensure the new buffer is as safe and effective as an existing one, not a study of algorithm performance.

    Therefore, I cannot extract the requested information from the provided text.

    Here's why each point you asked for cannot be answered from this document:

    1. A table of acceptance criteria and the reported device performance: While the document lists performance specifications (pH, osmolality, endotoxin, MEA, HSSA) for the Sydney IVF Gamete Buffer and implies they are similar to the predicate, it doesn't present them in a formal acceptance criteria and reported performance table for a new device's diagnostic or AI performance.
    2. Sample size used for the test set and the data provenance: Not applicable. The "test set" here refers to samples of the buffer solution or biological assays, not a set of clinical data for an AI system.
    3. Number of experts used to establish the ground truth... and their qualifications: Not applicable. Ground truth for an AI system is established by human experts, not through chemical and biological assays of a buffer solution.
    4. Adjudication method: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. This type of study is for evaluating human performance with and without AI assistance, which is not relevant to a gamete buffer solution.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used: Not applicable in the context of AI. The "ground truth" for the buffer solution's performance is derived from standardized laboratory tests (e.g., pH meter readings, osmolality measurements, biological assay results).
    8. The sample size for the training set: Not applicable. There is no AI model being trained.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document describes a medical device of a completely different nature than what your questions target.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160814

    Validate with FDA (Live)

    Date Cleared
    2016-04-22

    (29 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sydney IVF Sperm Cryopreservation Buffer is intended for use as a buffer to prevent damage to sperm samples during cryopreservation and thawing.

    Device Description

    Sydney IVF Sperm Cryopreservation Buffer consists of an aqueous solution intended for use as a buffer to prevent damage to sperm samples during cryopreservation and thawing. The IVF technician will use the buffer to cryopreserve washed spermatozoa, including MESA (microsurgical epididymal sperm aspiration) and TESA (testicular sperm extraction) samples. Sydney IVF Sperm Cryopreservation Buffer contains glycerol as a cryoprotectant, and HEPES (4-(2-hydroxyethyl)-1-piperazineethanesulfonic acid) as a buffer. This product is used to cryopreserve washed sperm and store them for future use.

    Sydney IVF Sperm Cryopreservation Buffer contains Human Serum Albumin (HSA) (4 mg/mL) and Gentamicin (0.01 mg/mL). The device is available as a 20 mL fill only.

    The Sydney IVF Sperm Cryopreservation Buffer is provided in glass vials with Fluorotec coated rubber stoppers held in place with a tamper evident seal. These products are single use, sterile (aseptic filtration) devices.

    AI/ML Overview

    I am sorry, but the provided text does not contain the information required to describe the acceptance criteria and the study that proves a device meets the acceptance criteria in the context of an AI-powered medical device or diagnostic.

    The document is a 510(k) clearance letter from the FDA for a product called "Sydney IVF Sperm Cryopreservation Buffer." This product is a chemical solution, not an AI-powered device or a device that uses algorithms and data analysis in the way implied by the questions (e.g., test sets, training sets, ground truth from experts, MRMC studies).

    Therefore, I cannot populate the table or answer the specific questions about the study design because the document describes a chemical product's regulatory clearance based on "substantial equivalence" to a predicate device, focusing on its formulation, shelf-life, and biological performance (e.g., sperm survival, embryo assay), rather than the performance of an AI algorithm.

    To answer your prompt, I would need a document describing the validation of an AI-powered medical device or a related diagnostic algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151367

    Validate with FDA (Live)

    Date Cleared
    2016-01-27

    (251 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sydney IVF Hyaluronidase is intended for use to facilitate removal of the cumulus cells surrounding oocytes in assisted reproduction technology (ART) procedures.

    Device Description

    Sydney IVF Hyaluronidase is an enzymatic preparation containing 80 IU/mL Hyaluronidase (of ovine origin) in a Bicarbonate buffered media supplemented with 5 mg/mL Human Serum Albumin (HSA) and 10 ug/mL Gentamicin.

    Sydney IVF Hyaluronidase is provided sterile (aseptically filtered) and is packaged in glass vials with Fluorotec coated rubber stoppers held in place with a tamper evident seal. It is available in a 5 pack carton box with each vial containing 1 mL of Sydney IVF Hvaluronidase.

    AI/ML Overview

    The provided text describes the Sydney IVF Hyaluronidase device and its substantial equivalence to a predicate device, focusing on its formulation, shelf-life, and performance. However, it does not contain the kind of detailed information typically found in a study demonstrating the device meets specific acceptance criteria with reported device performance against those criteria in a tabular format, nor does it provide information about the methodology of such a study (sample sizes, expert involvement, adjudication, etc.).

    The document is a 510(k) summary for regulatory clearance, which primarily focuses on demonstrating substantial equivalence to a previously cleared device. It mentions "performance data" and "stability & shelf life" but doesn't present a formal study report with acceptance criteria and results in the requested format.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and the reported device performance.
    • Sample sizes used for the test set and data provenance.
    • Number of experts and their qualifications for ground truth.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study effect size.
    • Standalone performance.
    • Type of ground truth used (beyond "MEA data").
    • Sample size for the training set.
    • How ground truth for the training set was established.

    However, I can extract the relevant information known about the device's performance based on the document:

    The document states that the Sydney IVF Hyaluronidase (the proposed device) and the predicate device (K023353 - Sydney IVF Hyaluronidase) share similar performance specifications and that the proposed device has been validated in stability studies.

    Here's what can be pieced together from your request and the provided text:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategoryPredicate Device Specification (K023353)Proposed Device Performance (Sydney IVF Hyaluronidase)
    Osmolality285-295 mOsm/kgSame (285-295 mOsm/kg)
    Endotoxin< 0.40 EU/mLSame (< 0.40 EU/mL)
    SterilitySterile (aseptically filtered)Same (aseptically filtered)
    Embryo Toxicity (MEA)1-cell MEA (96hrs) with ≥75% of control that develop to blastocyst2-cell MEA (72hrs) with ≥80% of control that develop to blastocyst
    Shelf-Life6 months at -20°CValidated to 20 weeks at 2-8°C
    Stability Tests(Implied by predicate and proposed device sharing "performance specifications")pH, osmolality, endotoxin, MEA, sterility, concentrations of proline, pyruvate, ammonia, and enzymatic activity of Hyaluronidase

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified for any specific "test set." The document refers to "MEA data" and "stability studies" without providing sample sizes.
    • Data Provenance: The device is from William A. Cook Australia Pty Ltd. The "MEA data" and "stability studies" would presumably be conducted by them. The document mentions the ovine derived product has been "licensed and sold globally since 2007," and its performance is supported by "experience" during this time, suggesting real-world usage data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified. The "MEA data" itself serves as a performance metric, but the number or qualifications of experts involved in running or interpreting these assays are not mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This device is a reproductive media and supplement (Hyaluronidase Solution), not an AI or imaging device with "human readers." Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This is not an algorithm. The relevant "standalone performance" refers to the intrinsic performance of the chemical/biological solution, which is assessed through the MEA and stability studies.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The primary "ground truth" or performance indicator for embryo toxicity is the Mouse Embryo Assay (MEA), which assesses the development of mouse embryos to the blastocyst stage. This is a laboratory-based, biological assay. Stability tests also involve objective measurements of chemical and physical properties (pH, osmolality, concentrations, enzymatic activity).

    8. The sample size for the training set:

    • This concept is not applicable as this is not an AI or machine learning device that requires a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable (see point 8).

    In summary, the document addresses regulatory clearance based on substantial equivalence, and while it references performance data (like MEA and stability studies), it does not present a detailed study report with the specific methodological details requested for evaluating AI/diagnostic devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153290

    Validate with FDA (Live)

    Date Cleared
    2016-01-27

    (75 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sydney IVF Follicle Flush Buffer is intended for use during in vitro fertilization procedures for follicle flushing and oocyte collection.

    Sydney IVF Fertilization Medium is intended for use during in vitro procedures for insemination of oocytes.

    Sydney IVF Cleavage Medium is intended for use during in vitro fertilization procedures for culture and transfer of cleavage stage embryos.

    Sydney IVF Blastocyst Medium is intended for use during in vitro fertilization procedures for extended culture and transfer of embryos.

    Device Description

    The Sydney IVF Follicle Flush Buffer, Sydney IVF Fertilization Medium, Sydney IVF Cleavage Medium and Sydney IVF Blastocyst Medium are used for processing eggs and embryos during in vitro fertilization procedures.

    Sydney IVF Follicle Flush Buffer is used to flush the follicles during oocyte recovery. It contains Gentamicin (0.01mg/mL) and is available in 100 mL vials.

    Sydney IVF Fertilization Medium maintains oocytes in a 6% CO2 environment until insemination and fertilization is complete. Sydney IVF Fertilization Medium contains Humans Serum Albumin (HSA) (5 mg/mL) and Gentamicin (0.01 mg/mL). It is available in 20, 50 or 100 mL vials.

    Sydney IVF Cleavage Medium facilitates the first two days of embryonic growth post fertilization. Sydney IVF Cleavage Medium contains HSA (5 mg/mL) and Gentamicin (0.01 mg/mL). It is available in 20, 50 or 100 mL vials.

    Once an embryo has reached the Day 3 (eight-cell) stage, it is then transferred into Sydney IVF Blastocyst Medium. This medium has been metabolically balanced to maximize blastocyst development rates and is suitable for blastocyst transfer. Sydney IVF Blastocyst Medium contains HSA (5 mg/mL) and Gentamicin (0.01 mg/mL). It is available in 20 or 50 mL vials.

    The Sydney IVF Follicle Flush Buffer, Sydney IVF Fertilization Medium, Sydney IVF Cleavage Medium and Sydney IVF Blastocyst Medium are provided in glass vials with Fluorotec® coated rubber stoppers held in place with a tamper evident seal. These products are single use, sterile (aseptic filtration) devices.

    AI/ML Overview

    This document describes the FDA 510(k) clearance for Sydney IVF Follicle Flush Buffer, Sydney IVF Fertilization Medium, Sydney IVF Cleavage Medium, and Sydney IVF Blastocyst Medium. It is not a study proving a device meets acceptance criteria through clinical trials or performance metrics of an AI/software device. Instead, it demonstrates substantial equivalence to a predicate device based on similar technological characteristics and performance specifications for the storage and use of IVF media.

    Therefore, many of the requested elements are not applicable in this context. Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Specifications)Reported Device Performance (Validation/Testing)
    Endotoxin: < 0.40 EU/mLIncluded in stability tests
    MEA (Mouse Embryo Assay): Used to screen for embryo toxicityIncluded in stability tests
    pH: Similar to predicateIncluded in stability tests
    Osmolality: Similar to predicateIncluded in stability tests
    Sterility:Included in stability tests (aseptic filtration)
    Concentrations of constituents: (pyruvate, amino acids (proline and lysine), HSA by-product ammonia)Included in stability tests
    Shelf-life: 20 weeks at 2-8°CValidated in stability studies

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify exact sample sizes for the stability tests. It mentions "stability studies" and "stability tests included endotoxin, MEA, pH, osmolality, sterility and the concentrations of pyruvate, amino acids (proline and lysine), and the HSA by-product ammonia." This implies various batches of the manufactured media were tested over time.
    • Data Provenance: Not specified, but the manufacturing company is William A. Cook Australia Pty Ltd, so the testing data would likely originate from their facilities. The studies were likely internal validation studies conducted by the manufacturer. Whether retrospective or prospective is not explicitly stated, but stability studies are typically prospective, monitoring product characteristics over time.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not applicable. This is a submission for in vitro fertilization media (a consumable medical device), not a diagnostic device or AI algorithm that requires expert interpretation to establish ground truth for performance. The "ground truth" here is adherence to chemical, biological, and physical specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable. Adjudication methods are typically used in clinical studies or studies involving human readers/interpreters, which is not the case here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not applicable. This submission is for IVF media, not an AI/software device. No human readers or AI assistance are involved in performance evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No. This is not applicable. This is IVF media, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is based on established scientific and regulatory standards for reproductive media, including:
      • Chemical specifications (pH, osmolality, constituent concentrations)
      • Biological specifications (MEA for embryo toxicity, sterility)
      • Physical specifications (shelf-life stability)
    • These "ground truths" are not established by expert consensus in a clinical diagnostic sense, but rather by adherence to validated analytical methods and predefined acceptable ranges for critical quality attributes.

    8. The sample size for the training set

    • This information is not applicable. This is IVF media, not an AI/machine learning device that uses a "training set."

    9. How the ground truth for the training set was established

    • This information is not applicable. No training set exists for this type of device.

    In summary, this document is a regulatory submission for a traditional medical device (IVF media) seeking substantial equivalence to a predicate device. It demonstrates compliance through laboratory-based performance testing against predefined specifications, rather than clinical trials or AI algorithm validation studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152782

    Validate with FDA (Live)

    Date Cleared
    2015-12-21

    (87 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sydney IVF Sperm Medium is intended for use during in vitro fertilization procedures to process sperm.
    Sydney IVF Sperm Gradient Kit is intended for use during in vitro fertilization procedures to separate motile sperm from seminal plasma.
    Sydney IVF Spermient is intended for use during in vitro fertilization procedures to separate motile sperm from seminal plasma.

    Device Description

    The Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit and Sydney IVF Spermient are used for processing sperm during in vitro fertilization procedures.
    Sperm Medium is used to provide a "liquid" and nutritious environment for the sperm to maintain its motility for the "swim up" procedures and the following fertilization process. The medium contains 10 mg/mL Human Serum Albumin (HSA) and 0.01 mg/mL Gentamicin. It is available in 20, 50 or 100 mL vials.
    Sydney IVF Sperm Gradient Kit is used to separate sperm based on density, using density gradient solutions. Sperm Gradient Kit contains both 40% density solutions and is comprised of silane coated silica particles which are diluted in a HEPES buffer. The sperm gradient has the dual purpose of enriching motile sperm populations and removing sperm inhibitory compounds normally present in the eiaculate. It contains 10 mg/mL HSA and 0.01 mg/mL Gentamicin. Sydney IVF Sperm Gradient Kit is available in 20 or 50 mL vials, packaged in a carton box, consisting of one vial filled with the 40% sperm gradient solution, and a second vial filled with the 80% sperm gradient solution.
    Sydney IVF Spermient is intended to separate sperm based on density, using density gradient solutions. This is achieved by diluting the Spermient into various concentrations with a HEPES buffered solution. It contains colloidal silica coated with an inert polymer (silane) and has been formulated to be isotonic. Spermient contains 10 mg/mL HSA and 0.01 mg/mL Gentamicin. It is available in 20 or 100 mL vials.
    The Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit and Sydney IVF Spermient are provided in glass vials with Fluorotec® coated rubber stoppers held in place with a tamper evident seal.. These products are single use, sterile (aseptic filtration) devices.

    AI/ML Overview

    The provided document is a 510(k) summary for the Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit, and Sydney IVF Spermient. This document indicates that the device's main modification from its predicate was an extended shelf-life. The acceptance criteria and the study that proves the device meets them are described in the "PERFORMANCE DATA" section.

    Here's the breakdown:

    1. A table of acceptance criteria and the reported device performance:

    ParameterAcceptance CriteriaReported Device Performance
    Shelf-life20 weeks at 2-8°CValidated to 20 weeks at 2-8°C
    Endotoxin< 0.40 EU/mL (Implied from predicate comparison)Tested (Specific values not reported, but implied to meet criteria)
    Mouse Embryo Assay (MEA)Screen for embryo toxicity (Implied from predicate comparison)Tested (Implied to meet criteria as "product for embryo toxicity")
    Human Sperm Survival Assay (HSSA)Tested (Implicitly to ensure sperm viability)Tested (Implied to meet criteria)
    pHSimilar to predicate device (Implied from predicate comparison)Tested (Implied to meet criteria)
    Osmolality285-295 mOsm/kg (Implied from predicate comparison)Tested (Implied to meet criteria)
    SterilitySterile (Implied from manufacturing process)Tested (Implied to meet criteria)
    Concentrations of pyruvate and HSA by-product ammoniaSimilar to predicate device (Implied from predicate comparison)Tested (Implied to meet criteria)

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample size for the test set. It mentions "stability studies" and "testing" without providing numerical details for the number of batches or samples tested for each parameter.

    The provenance of the data is not explicitly stated as "country of origin." However, the applicant is William A. Cook Australia Pty Ltd, so the testing likely occurred in Australia or through their affiliated labs. The studies appear to be prospective as they were conducted to validate the extended shelf-life and other performance characteristics of the modified device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. For in vitro fertilization media, "ground truth" often refers to the validation results of specific assays (like MEA or HSSA) performed by trained lab personnel rather than experts establishing a "ground truth" in the diagnostic sense.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable and is not provided. The testing involves laboratory assays with established protocols and quantitative results, not subjective interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable as the device is a reproductive medium/supplement, not an AI-assisted diagnostic imaging device that involves human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a medical product (media/supplement) and not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device is based on laboratory analytical and biological assays (e.g., endotoxin limits, pH, osmolality, sterility, Mouse Embryo Assay, Human Sperm Survival Assay) that demonstrate the media's ability to maintain a suitable environment for sperm and not be toxic to embryos.

    8. The sample size for the training set:

    This is not applicable since the device is a medical product (media/supplement) and not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established:

    This is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1